Artis Q.zen floor 10848353Recalls & reported problems
Siemens Healthcare GmbH · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
4 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 4
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04056869010014), received between Oct 7, 2020 and Oct 19, 2021 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Radiation Output Failure” (2 reports).
4 FDA recall records name this device's identifier, 4 still open.
86 other recalls cover devices cleared under the same FDA submission (K123529, K181407); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Mar 21, 2024Class IIOpen, Classified
Coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message TUBE HOT, have a break . the system cannot be operated normally, may result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an alternative system
Z-1598-2024 · Root cause: Device Design · Recall details · FDA record ↗
- Dec 16, 2022Class IIOpen, Classified
In rare cases, the system may only boot into backup mode after an abrupt shutdown and not reach full operating mode. This may result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an alternative system.
Z-1040-2023 · Root cause: Software design · Recall details · FDA record ↗
- Aug 10, 2022Class IIOpen, Classified
Siemens has become aware of a potential hardware issue with Artis Q, Artis Q.zen, Artis pheno and Artis icono systems with a specific lot of X-ray tubes. In rare cases, if the error detection mechanism fails, it will not be possible to release X-ray any longer until system shutdown.
Z-1645-2022 · Root cause: Device Design · Recall details · FDA record ↗
- Feb 23, 2022Class IIOpen, Classified
If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".
Z-0978-2022 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Radiation Output Failure2
- Circuit Failure1
- Loose or Intermittent Connection1
- Mechanical Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 19, 2021Radiation Output Failure; Radiation Output Failure
IT WAS REPORTED TO SIEMENS THAT A MALFUNCTION OCCURRED WHILE OPERATING THE ARTIS Q.ZEN FLOOR SYSTEM. DURING AN INTERVENTIONAL PROCEDURE, THE USER REPORTED THAT THE IMAGE WAS LOST. AT THIS TIME, WE ARE UNAWARE OF ANY IMPACT TO THE STATE OF HEALTH OF THE PATIENT INVOLVED. SIEMENS HAS REQUESTED ADDITIONAL INFORMATION IN ORDER TO CONDUCT AN INVESTIGATION OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12655371 ↗MalfunctionMay 26, 2021Radiation Output Failure; Radiation Output Failure
IT WAS REPORTED TO SIEMENS THAT A MALFUNCTION OCCURRED WHILE OPERATING THE ARTIS Q.ZEN FLOOR SYSTEM. DURING AN INTERVENTIONAL PROCEDURE, THE USER REPORTED THAT NO X-RAY WAS POSSIBLE. THE PROCEDURE WAS CONTINUED AND COMPLETED ON AN ALTERNATE SYSTEM. WE ARE UNAWARE OF ANY IMPACT TO THE STATE OF HEALTH OF THE PATIENT INVOLVED. SIEMENS HAS REQUESTED ADDITIONAL INFORMATION IN ORDER TO CONDUCT AN INVESTIGATION OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11884489 ↗MalfunctionNov 4, 2020Mechanical Problem; Mechanical Problem
IT WAS REPORTED TO SIEMENS THAT A MALFUNCTION OCCURRED WHILE OPERATING THE ARTIS Q.ZEN FLOOR SYSTEM. DURING AN INTERVENTIONAL PROCEDURE, THE USER REPORTED THAT NO STAND MOVEMENT WAS POSSIBLE. THE PROCEDURE WAS CONTINUED AND COMPLETED ON AN ALTERNATE SYSTEM. WE ARE UNAWARE OF ANY IMPACT TO THE STATE OF HEALTH OF THE PATIENT INVOLVED. SIEMENS HAS REQUESTED ADDITIONAL INFORMATION IN ORDER TO CONDUCT AN INVESTIGATION OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10784698 ↗MalfunctionOct 28, 2020Loose or Intermittent Connection; Loose or Intermittent Connection
IT WAS REPORTED TO SIEMENS THAT A MALFUNCTION OCCURRED WHILE OPERATING THE ARTIS Q.ZEN FLOOR SYSTEM. DURING AN INTERVENTIONAL PROCEDURE, THE USER REPORTED A PROBLEM WITH THE TABLE BECOMING STUCK. THE PROCEDURE WAS CONTINUED AND COMPLETED ON AN ALTERNATE SYSTEM. WE ARE UNAWARE OF ANY IMPACT TO THE STATE OF HEALTH OF THE PATIENT INVOLVED. SIEMENS HAS REQUESTED ADDITIONAL INFORMATION IN ORDER TO CONDUCT AN INVESTIGATION OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10747155 ↗MalfunctionOct 7, 2020Circuit Failure; Circuit Failure
IT WAS REPORTED TO SIEMENS THAT A MALFUNCTION OCCURRED WHILE OPERATING THE ARTIS Q.ZEN FLOOR SYSTEM. DURING AN INTERVENTIONAL PROCEDURE, THE USER REPORTED THAT THE SYSTEM LOCKED UP. THE PROCEDURE WAS CONTINUED AND COMPLETED ON AN ALTERNATE SYSTEM. WE ARE UNAWARE OF ANY IMPACT TO THE STATE OF HEALTH OF THE PATIENT INVOLVED. SIEMENS HAS REQUESTED ADDITIONAL INFORMATION IN ORDER TO CONDUCT AN INVESTIGATION OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10641735 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K123529, K181407. One submission often covers many models, so check the product and lot details in each record.
- Z-1343-2022 · May 24, 2022 · TerminatedArtis zeego (Model no. 10280959) and Artis Q.zeego (Model no. 10848283), Interventional Fluoroscopic X-Ray System
- Z-0967-2022 · Feb 23, 2022 · Open, ClassifiedArtis zee floor, Model Number 10094135
- Z-0971-2022 · Feb 23, 2022 · Open, ClassifiedArtis zee floor MN, Model Number 10094142
- Z-0974-2022 · Feb 23, 2022 · Open, ClassifiedArtis Q floor, Model Number 10848280
- Z-0979-2022 · Feb 23, 2022 · Open, ClassifiedArtis Q.zen ceiling, Model Number 10848354
- Z-0980-2022 · Feb 23, 2022 · Open, ClassifiedArtis Q.zen biplane, Model Number 10848355
- Z-0968-2022 · Feb 23, 2022 · Open, ClassifiedArtis zee ceiling, Model Number 10094137
- Z-0977-2022 · Feb 23, 2022 · Open, ClassifiedArtis Q zeego, Model Number 10848283
- Z-0973-2022 · Feb 23, 2022 · Open, ClassifiedArtis zeego, Model Number 10280959
- Z-0972-2022 · Feb 23, 2022 · Open, ClassifiedArtis zee biplane MN, Model Number 10094143
- Z-0969-2022 · Feb 23, 2022 · Open, ClassifiedArtis zee multi-purpose, Model Number 10094139
- Z-0976-2022 · Feb 23, 2022 · Open, ClassifiedArtis Q biplane, Model Number 10848282
- Z-0975-2022 · Feb 23, 2022 · Open, ClassifiedArtis Q ceiling, Model Number 10848281
- Z-0970-2022 · Feb 23, 2022 · Open, ClassifiedArtis zee biplane, Model Number 10094141
- Z-0300-2022 · Oct 20, 2021 · TerminatedArtis Q floor with software VD12- angiography systems developed for single and biplane diagnostic imaging and interventional procedure Model Number: 10848280
…and 71 more.
Other System, x-ray, angiographic devices
- ARTIS icono ceiling 11328100Siemens Healthcare GmbH · System, x-ray, angiographic9 recalls1 report
- Artis Q ceiling 10848281Siemens Healthcare GmbH · System, x-ray, angiographic4 recalls1 death49 reports
- Artis Q floor 10848280Siemens Healthcare GmbH · System, x-ray, angiographic4 recalls18 reports
- Artis zee biplane 10094141Siemens Healthcare GmbH · System, x-ray, angiographic3 recalls63 reports
- Artis Q.zen ceiling 10848354Siemens Healthcare GmbH · System, x-ray, angiographic4 recalls10 reports
- Artis zee floor MN 10094142Siemens Healthcare GmbH · System, x-ray, angiographic2 recalls0 reports
- iMRX System iMRX System AR-DR MR30RSIMRIS Imaging, Inc. · System, X-Ray, Angiographic1 recall0 reports
- Allia MoveoGe Medical Systems · System, X-Ray, Angiographic1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page