Skip to content

Draeger Infinity Acute Care System Monitoring Solution recallZ-0714-2014 · Draeger Medical Systems, Inc. · initiated Dec 10, 2013

Class II Terminated All recalls posted in January 2014

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Draeger Medical Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

After a test in which a low SpO2 (yellow) medium grade alarm initiated in neo-natal mode and was subsequently paused, the high grade life threatening SpO2 alarm (red) did not activate when the SpO2 values were decreased to reach the red alarm limit.

FDA root cause category: Software design

What the company is doing

An Urgent Medical Device Recall Letter, dated December 2013, was sent to Users. Once available Drager will update components of the IACS Monitoring Solution System.

Recall details

Recall number
Z-0714-2014
Recall class
Class II
Status
Terminated
Initiated by the company
Dec 10, 2013
Posted by the FDA
Jan 10, 2014
Recalling firm
Draeger Medical Systems, Inc.
Quantity
439
Product code
MHX
FDA submissions
K113798
Distribution
US Distribution including the states of: MA, NH, IN, CA, TX, NY, OK, and IL.

Product description

Draeger Infinity Acute Care System Monitoring Solution

Code information (models, lots, serial numbers)

with software versions VG2.0.3 and higher.

Other devices under the same FDA submission

5 devices cleared under K113798 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.