Skip to content

MooreMedical Supporting Health & Care...MooreBrand Blood… recallZ-0909-2014 · Nipro Diagnostics, Inc. · initiated Oct 2, 2013

Class II Terminated All recalls posted in February 2014

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Nipro Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Manufactured with an incorrect factory set unit of measure.

FDA root cause category: Software design (manufacturing process)

What the company is doing

Nipro Diagnostics sent an Urgent Medical Device Voluntary Recall letter dated December 31, 2013 to all affected customers. The customers were instructed to do the following: 1) Screen their inventory immediately for any of the serial numbers listed on the back of the products. 2) If they have any of the affected products, return the product by contacting Stericycle at 1-866-236-4518. 3) Complete and return the enclosed response form via fax at 1-888-358-7294. Customers with questions were instructed to call Stericycle at 1-866-236-4518. For questions regarding this recall call 954-677-9201.

Recall details

Recall number
Z-0909-2014
Recall class
Class II
Status
Terminated
Initiated by the company
Oct 2, 2013
Posted by the FDA
Feb 3, 2014
Recalling firm
Nipro Diagnostics, Inc.
Quantity
501
Product code
NBW
FDA submissions
K090495
Distribution
Worldwide Distribution - USA (nationwide, HI and Puerto Rico and Internationally to Canada, European Union, Jamaica, Suriname, Trinidad and Tobago, Kazakhastan, Russian, Turkey, Australia, Bangladesh, Pakistan, Cambodia, Laos, Myanmar, Thailand, Vietnam, and India.

Product description

MooreMedical Supporting Health & Care...MooreBrand Blood Glucose Monitor for Self-Testing & Multi-Patient Use TRUEbalance. Manufactured by: NIPRO DIAGNOSTICS, Fort Lauderdale, FL 33309 U.S.A. Is intended for the quantitative determination of glucose in human whole blood taken from finger or forearm.

Code information (models, lots, serial numbers)

Finished Good: KK0468; Lot Numbers: KL01079, KM0337TI, KN1834TI, KN1973TI, KN2074TI, KN 2089TI, KN2090TI, KN2102TI, KP0047TI, KP0076TI, KP0128TI, KP0925TI.

Other devices under the same FDA submission

3 devices cleared under K090495 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 67152).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page