ARIATELE TELEMETRY TRANSMITTER, Model 96281, with SpO2… recallZ-0871-2014 · Spacelabs Healthcare, Llc · initiated Dec 19, 2013
Class II Terminated All recalls posted in January 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Spacelabs Healthcare, Llc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The AriaTele Model 96281 with SpO2 monitoring Option C is recalled because the AriaTele (transmitter) display and the Telemetry Central Station display may show a SpO2 value when the sensor is not connected to the patient or in certain sensor failed conditions.
FDA root cause category: Software design
What the company is doing
On January 6, 2014, Spacelabs sent an "URGENT - MEDICAL DEVICE CORRECTION Spacelabs Healthcare AriaTele Model 96281 with SpO2 Option C" letter, dated 03 January 2014 to US consignees. On January 9, 2014, Spacelabs e-mailed the "URGENT - MEDICAL DEVICE CORRECTION Spacelabs Healthcare AriaTele Model 96281 with SpO2 Option C" letter (translated as necessary), dated 03 January 2014, to foreign consignees. The letter described product, problem and actions to be taken. The Consignees where instructed to follow the "Advise on what action the clinician takes" on the letter. Spacelabs Healthcare will contact you at the earliest possible date to schedule a convenient tie for Spacelabs to correct all of your facility's affected transmitters (software update) at no cost. If you have any questions about this corrective action program, please contact Spacelabs Healthcare at 1-800-522-7025 and select 2 for Technical Support.
Recall details
- Recall number
- Z-0871-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Dec 19, 2013
- Posted by the FDA
- Jan 30, 2014
- Recalling firm
- Spacelabs Healthcare, Llc
- Quantity
- 1057 units (988 in the US and 69 outside of the US).
- Product code
- MHX
- FDA submissions
- K121480
- Distribution
- Worldwide Distribution: US (nationwide) including states of: AZ, CA, FL, GA, IA, IL, LA, MA, ME, MI, MO, MS, NY, PA, SC, TN, TX, VA, and Puerto Rico; and Internationally to: Canada, Chile, and Panama.
Product description
ARIATELE TELEMETRY TRANSMITTER, Model 96281, with SpO2 Option C. Model 96281 is intended for use with either adult or neonatal patient populations in a hospital environment.
Code information (models, lots, serial numbers)
Show the full code information (15k characters)
Serial Numbers IN the US: 6281- 002002, 6281- 002003, 6281- 002004, 6281- 002005, 6281- 002006, 6281- 002007, 6281- 002008, 6281- 002009, 6281- 002010, 6281- 002011, 6281- 002012, 6281- 002013, 6281-000295, 6281-000401, 6281-000402, 6281-000403, 6281-000404, 6281-000405, 6281-000406, 6281-000407, 6281-000408, 6281-000503, 6281-000504, 6281-000505, 6281-000506, 6281-000507, 6281-000508, 6281-000512, 6281-000513, 6281-000514, 6281-000515, 6281-000516, 6281-000517, 6281-000518, 6281-000519, 6281-000520, 6281-000521, 6281-000522, 6281-000523, 6281-000524, 6281-000526, 6281-000527, 6281-000528, 6281-000529, 6281-000530, 6281-000531, 6281-000532, 6281-000533, 6281-000534, 6281-000535, 6281-000536, 6281-000537, 6281-000538, 6281-000539, 6281-000540, 6281-000541, 6281-000542, 6281-000543, 6281-000544, 6281-000545, 6281-000546, 6281-000547, 6281-000549, 6281-000550, 6281-000551, 6281-000552, 6281-000553, 6281-000554, 6281-000555, 6281-000556, 6281-000557, 6281-000558, 6281-000559, 6281-000560, 6281-000561, 6281-000562, 6281-000563, 6281-000564, 6281-000565, 6281-000566, 6281-000567, 6281-000570, 6281-000574, 6281-000575, 6281-000576, 6281-000577, 6281-000578, 6281-000579, 6281-000580, 6281-000581, 6281-000582, 6281-000583, 6281-000584, 6281-000585, 6281-000586, 6281-000587, 6281-000588, 6281-000589, 6281-000590, 6281-000591, 6281-000592, 6281-000593, 6281-000594, 6281-000595, 6281-000596, 6281-000597, 6281-000598, 6281-000599, 6281-000600, 6281-000601, 6281-000602, 6281-000603, 6281-000604, 6281-000605, 6281-000606, 6281-000607, 6281-000608, 6281-000609, 6281-000610, 6281-000611, 6281-000612, 6281-000613, 6281-000614, 6281-000615, 6281-000616, 6281-000617, 6281-000618, 6281-000619, 6281-000620, 6281-000621, 6281-000622, 6281-000623, 6281-000624, 6281-000625, 6281-000626, 6281-000627, 6281-000628, 6281-000629, 6281-000630, 6281-000631, 6281-000632, 6281-000633, 6281-000634, 6281-000635, 6281-000636, 6281-000637, 6281-000638, 6281-000639, 6281-000640, 6281-000641, 6281-000642, 6281-000643, 6281-000644, 6281-000645, 6281-000646, 6281-000647, 6281-000648, 6281-000649, 6281-000650, 6281-000651, 6281-000652, 6281-000653, 6281-000654, 6281-000655, 6281-000656, 6281-000657, 6281-000658, 6281-000659, 6281-000660, 6281-000661, 6281-000662, 6281-000663, 6281-000664, 6281-000665, 6281-000666, 6281-000667, 6281-000668, 6281-000669, 6281-000670, 6281-000671, 6281-000672, 6281-000673, 6281-000674, 6281-000675, 6281-000676, 6281-000677, 6281-000678, 6281-000679, 6281-000680, 6281-000681, 6281-000682, 6281-000683, 6281-000684, 6281-000685, 6281-000686, 6281-000687, 6281-000688, 6281-000689, 6281-000690, 6281-000691, 6281-000692, 6281-000693, 6281-000694, 6281-000695, 6281-000696, 6281-000697, 6281-000698, 6281-000699, 6281-000700, 6281-000701, 6281-000702, 6281-000703, 6281-000704, 6281-000705, 6281-000706, 6281-000707, 6281-000708, 6281-000709, 6281-000710, 6281-000711, 6281-000712, 6281-000713, 6281-000714, 6281-000715, 6281-000716, 6281-000717, 6281-000718, 6281-000719, 6281-000720, 6281-000721, 6281-000722, 6281-000723, 6281-000724, 6281-000725, 6281-000853, 6281-000854, 6281-000855, 6281-000856, 6281-000857, 6281-000858, 6281-000859, 6281-000860, 6281-000861, 6281-000862, 6281-000863, 6281-000864, 6281-000865, 6281-000866, 6281-000867, 6281-000868, 6281-000869, 6281-000870, 6281-000871, 6281-000872, 6281-000938, 6281-000939, 6281-000940, 6281-000941, 6281-000942, 6281-000943, 6281-000944, 6281-000945, 6281-000946, 6281-000947, 6281-000948, 6281-000949, 6281-000950, 6281-000951, 6281-000952, 6281-000953, 6281-000954, 6281-000955, 6281-000956, 6281-000957, 6281-000958, 6281-000959, 6281-000960, 6281-000961, 6281-000962, 6281-000963, 6281-000964, 6281-000965, 6281-000966, 6281-000967, 6281-000968, 6281-000969, 6281-000970, 6281-000971, 6281-000972, 6281-000973, 6281-000974, 6281-000975, 6281-000976, 6281-000977, 6281-000978, 6281-000979, 6281-000981, 6281-000982, 6281-000983, 6281-000984, 6281-000985, 6281-000986, 6281-000987, 6281-000988, 6281-000989, 6281-000990, 6281-000991, 6281-000992, 6281-001005, 6281-001006, 6281-001007, 6281-001008, 6281-001009, 6281-001010, 6281-001011, 6281-001012, 6281-001013, 6281-001014, 6281-001015, 6281-001016, 6281-001017, 6281-001018, 6281-001019, 6281-001020, 6281-001021, 6281-001022, 6281-001023, 6281-001024, 6281-001025, 6281-001026, 6281-001027, 6281-001028, 6281-001029, 6281-001030, 6281-001031, 6281-001036, 6281-001037, 6281-001080, 6281-001081, 6281-001082, 6281-001083, 6281-001084, 6281-001085, 6281-001086, 6281-001087, 6281-001088, 6281-001089, 6281-001090, 6281-001091, 6281-001092, 6281-001093, 6281-001094, 6281-001095, 6281-001096, 6281-001097, 6281-001098, 6281-001099, 6281-001100, 6281-001101, 6281-001102, 6281-001103, 6281-001104, 6281-001105, 6281-001106, 6281-001107, 6281-001108, 6281-001109, 6281-001110, 6281-001111, 6281-001112, 6281-001113, 6281-001114, 6281-001115, 6281-001116, 6281-001117, 6281-001118, 6281-001119, 6281-001120, 6281-001121, 6281-001122, 6281-001123, 6281-001124, 6281-001125, 6281-001126, 6281-001127, 6281-001128, 6281-001129, 6281-001130, 6281-001131, 6281-001132, 6281-001133, 6281-001134, 6281-001135, 6281-001136, 6281-001137, 6281-001138, 6281-001139, 6281-001140, 6281-001141, 6281-001142, 6281-001143, 6281-001144, 6281-001145, 6281-001146, 6281-001147, 6281-001148, 6281-001149, 6281-001150, 6281-001151, 6281-001152, 6281-001153, 6281-001154, 6281-001155, 6281-001156, 6281-001157, 6281-001158, 6281-001159, 6281-001160, 6281-001161, 6281-001162, 6281-001163, 6281-001164, 6281-001165, 6281-001166, 6281-001167, 6281-001168, 6281-001169, 6281-001170, 6281-001171, 6281-001178, 6281-001179, 6281-001180, 6281-001181, 6281-001182, 6281-001183, 6281-001184, 6281-001185, 6281-001186, 6281-001187, 6281-001188, 6281-001189, 6281-001190, 6281-001191, 6281-001192, 6281-001193, 6281-001194, 6281-001195, 6281-001196, 6281-001197, 6281-001222, 6281-001223, 6281-001224, 6281-001225, 6281-001226, 6281-001227, 6281-001228, 6281-001229, 6281-001230, 6281-001231, 6281-001232, 6281-001233, 6281-001234, 6281-001235, 6281-001236, 6281-001237, 6281-001238, 6281-001239, 6281-001240, 6281-001241, 6281-001242, 6281-001243, 6281-001244, 6281-001245, 6281-001246, 6281-001247, 6281-001248, 6281-001249, 6281-001250, 6281-001251, 6281-001252, 6281-001253, 6281-001254, 6281-001255, 6281-001256, 6281-001257, 6281-001258, 6281-001259, 6281-001260, 6281-001261, 6281-001262, 6281-001263, 6281-001264, 6281-001265, 6281-001266, 6281-001267, 6281-001268, 6281-001269, 6281-001270, 6281-001271, 6281-001272, 6281-001273, 6281-001274, 6281-001275, 6281-001276, 6281-001277, 6281-001278, 6281-001279, 6281-001280, 6281-001281, 6281-001282, 6281-001283, 6281-001284, 6281-001285, 6281-001286, 6281-001287, 6281-001288, 6281-001289, 6281-001290, 6281-001291, 6281-001292, 6281-001293, 6281-001294, 6281-001295, 6281-001296, 6281-001297, 6281-001298, 6281-001299, 6281-001300, 6281-001301, 6281-001302, 6281-001303, 6281-001304, 6281-001305, 6281-001306, 6281-001307, 6281-001308, 6281-001309, 6281-001310, 6281-001311, 6281-001312, 6281-001313, 6281-001314, 6281-001315, 6281-001316, 6281-001317, 6281-001318, 6281-001319, 6281-001320, 6281-001321, 6281-001322, 6281-001323, 6281-001331, 6281-001332, 6281-001337, 6281-001338, 6281-001390, 6281-001391, 6281-001392, 6281-001393, 6281-001394, 6281-001395, 6281-001396, 6281-001397, 6281-001398, 6281-001399, 6281-001400, 6281-001401, 6281-001402, 6281-001403, 6281-001404, 6281-001405, 6281-001406, 6281-001407, 6281-001408, 6281-001409, 6281-001410, 6281-001412, 6281-001413, 6281-001414, 6281-001415, 6281-001416, 6281-001417, 6281-001418, 6281-001419, 6281-001420, 6281-001421, 6281-001422, 6281-001423, 6281-001424, 6281-001425, 6281-001426, 6281-001427, 6281-001428, 6281-001429, 6281-001430, 6281-001431, 6281-001432, 6281-001433, 6281-001434, 6281-001435, 6281-001436, 6281-001437, 6281-001438, 6281-001439, 6281-001440, 6281-001441, 6281-001449, 6281-001450, 6281-001451, 6281-001452, 6281-001453, 6281-001454, 6281-001459, 6281-001468, 6281-001469, 6281-001470, 6281-001471, 6281-001472, 6281-001595, 6281-001596, 6281-001597, 6281-001598, 6281-001610, 6281-001611, 6281-001615, 6281-001616, 6281-001648, 6281-001802, 6281-001803, 6281-001804, 6281-001805, 6281-001806, 6281-001807, 6281-001808, 6281-001809, 6281-001810, 6281-001811, 6281-001812, 6281-001813, 6281-001814, 6281-001815, 6281-001816, 6281-001817, 6281-001818, 6281-001819, 6281-001820, 6281-001914, 6281-001915, 6281-001916, 6281-001917, 6281-001918, 6281-001919, 6281-001920, 6281-001921, 6281-001922, 6281-001923, 6281-001924, 6281-001925, 6281-001926, 6281-001927, 6281-001928, 6281-001929, 6281-001930, 6281-001931, 6281-001932, 6281-001933, 6281-001934, 6281-001935, 6281-001936, 6281-001937, 6281-001938, 6281-001939, 6281-001940, 6281-001941, 6281-001942, 6281-001943, 6281-001944, 6281-001945, 6281-001946, 6281-001947, 6281-001948, 6281-001949, 6281-001950, 6281-001951, 6281-001952, 6281-001953, 6281-001954, 6281-001955, 6281-001956, 6281-001957, 6281-001958, 6281-001959, 6281-001960, 6281-001961, 6281-001962, 6281-001963, 6281-001964, 6281-001965, 6281-001966, 6281-001967, 6281-001968, 6281-001969, 6281-001970, 6281-001971, 6281-001972, 6281-001973, 6281-001974, 6281-001975, 6281-001976, 6281-001977, 6281-001978, 6281-001979, 6281-001980, 6281-001998, 6281-002014, 6281-002015, 6281-002016, 6281-002017, 6281-002018, 6281-002019, 6281-002020, 6281-002021, 6281-002022, 6281-002023, 6281-002024, 6281-002025, 6281-002026, 6281-002027, 6281-002028, 6281-002029, 6281-002030, 6281-002031, 6281-002032, 6281-002033, 6281-002034, 6281-002035, 6281-002036, 6281-002037, 6281-002038, 6281-002039, 6281-002040, 6281-002041, 6281-002042, 6281-002043, 6281-002044, 6281-002045, 6281-002046, 6281-002047, 6281-002048, 6281-002049, 6281-002050, 6281-002051, 6281-002052, 6281-002053, 6281-002054, 6281-002055, 6281-002056, 6281-002057, 6281-002058, 6281-002059, 6281-002060, 6281-002061, 6281-002062, 6281-002063, 6281-002064, 6281-002065, 6281-002066, 6281-002067, 6281-002068, 6281-002069, 6281-002070, 6281-002071, 6281-002072, 6281-002073, 6281-002074, 6281-002075, 6281-002076, 6281-002077, 6281-002078, 6281-002079, 6281-002080, 6281-002081, 6281-002082, 6281-002083, 6281-002084, 6281-002085, 6281-002086, 6281-002087, 6281-002088, 6281-002089, 6281-00209 , 6281-002090, 6281-002091, 6281-002092, 6281-002093, 6281-002094, 6281-002095, 6281-002096, 6281-002100, 6281-002101, 6281-002102, 6281-002103, 6281-002104, 6281-002105, 6281-002106, 6281-002107, 6281-002108, 6281-002109, 6281-002110, 6281-002111, 6281-002112, 6281-002113, 6281-002114, 6281-002115, 6281-002116, 6281-002117, 6281-002118, 6281-002119, 6281-002120, 6281-002121, 6281-002122, 6281-002123, 6281-002124, 6281-002125, 6281-002126, 6281-002127, 6281-002128, 6281-002129, 6281-002130, 6281-002131, 6281-002132, 6281-002133, 6281-002134, 6281-002135, 6281-002136, 6281-002137, 6281-002138, 6281-002139, 6281-002140, 6281-002141, 6281-002142, 6281-002143, 6281-002144, 6281-002145, 6281-002146, 6281-002147, 6281-002148, 6281-002149, 6281-002150, 6281-002151, 6281-002152, 6281-002153, 6281-002154, 6281-002155, 6281-002156, 6281-002157, 6281-002158, 6281-002159, 6281-002160, 6281-002161, 6281-002162, 6281-002163, 6281-002164, 6281-002165, 6281-002166, 6281-002167, 6281-002168, 6281-002169, 6281-002170, 6281-002171, 6281-002172, 6281-002173, 6281-002174, 6281-002175, 6281-002176, 6281-002177, 6281-002178, 6281-002179, 6281-002180, 6281-002181, 6281-002182, 6281-002183, 6281-002184, 6281-002185, 6281-002186, 6281-002187, 6281-002188, 6281-002189, 6281-002190, 6281-002191, 6281-002375, 6281-002376, 6281-002377, 6281-002378, 6281-002379, 6281-002380, 6281-002381, 6281-002382, 6281-002383, 6281-002384, 6281-002507, 6281-002545, 6281-002546, 6281-002547, 6281-002548, 6281-002549, 6281-002550, 6281-002551, 6281-002552, 6281-002553, 6281-002554, 6281-002555, 6281-002556, 6281-002557, 6281-002558, 6281-002559, 6281-002560, 6281-002561, 6281-002562, 6281-002563, 6281-002564, 6281-002565, 6281-002566, 6281-002567, 6281-002568, 6281-002569, 6281-002570, 6281-002571, 6281-002572, 6281-002573, 6281-002574, 6281-002575, 6281-002576, 6281-002577, 6281-002578, 6281-002579, 6281-002580, 6281-002581, 6281-002582, 6281-002583, 6281-002584, 6281-002585, 6281-002586, 6281-002587, 6281-002588, 6281-002589, 6281-002590, 6281-002591, 6281-002592, 6281-002593, 6281-002594, 6281-002595, 6281-002596, 6281-002597, 6281-002598, 6281-002599, 6281-002600, 6281-002601, 6281-002602, 6281-002603, 6281-002604, 6281-002605, 6281-002606, 6281-002607, 6281-002608, 6281-002609, 6281-002610, 6281-002611, 6281-002612, 6281-002613, 6281-002614, 6281-002615, 6281-002616, 6281-002617, 6281-002618, 6281-002619, 6281-002620, 6281-002621, 6281-002622, 6281-002623, 6281-002624, 6281-002625, 6281-002626, 6281-002627, 6281-002681, 6281-002682, 6281-002683, and 6281-002684. SERIAL NUMBERS OUTSIDE THE US: 6281-000451, 6281-000771, 6281-000772, 6281-000773, 6281-000774, 6281-000775, 6281-000776, 6281-000993, 6281-000994, 6281-000995, 6281-000996, 6281-000997, 6281-000998, 6281-000999, 6281-001000, 6281-001001, 6281-001002, 6281-001003, 6281-001004, 6281-001038, 6281-001039, 6281-001040, 6281-001041, 6281-001042, 6281-001043, 6281-001044, 6281-001045, 6281-001046, 6281-001047, 6281-001048, 6281-001049, 6281-001050, 6281-001051, 6281-001052, 6281-001053, 6281-001054, 6281-001055, 6281-001056, 6281-001057, 6281-001058, 6281-001059, 6281-001060, 6281-001061, 6281-001062, 6281-001063, 6281-001064, 6281-001065, 6281-001066, 6281-001067, 6281-001068, 6281-001069, 6281-001070, 6281-001071, 6281-001172, 6281-001173, 6281-001174, 6281-001175, 6281-001475, 6281-001476, 6281-001514, 6281-001515, 6281-001516, 6281-001517, 6281-001705, 6281-001706, 6281-001994, 6281-001995, 6281-001996, and 6281-001997. Spacelabs is currently conducting a internal audit and discovered that there are additional lot numbers for this recall. 6681-003056, 6281-002500, 6281-002501, 6281-002659, 6281-002661, 6281-002662, 6281-002663, and 6281-002664.
Other devices under the same FDA submission
1 device cleared under K121480 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- AriaTele 96281Spacelabs Healthcare (Washington), Inc · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)3 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page