ACCU-CHEK Compact Plus Test Strips recallZ-1616-2014 · Roche Diagnostics Operations, Inc. · initiated Apr 14, 2014

Class II Terminated All recalls posted in May 2014

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Roche Diagnostics Operations, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Roche Diabetes Care has become aware the ACCU-CHEK Compact Plus test strips may produce erroneously low blood glucose readings in patients undergoing Ceftriaxone therapy. The interference is not described in the product labeling.

FDA root cause category: Device Design

What the company is doing

On 4/14/14, Roche Diagnostics Operations issued URGENT MEDICAL DEVICE CORRECTION notification to consignees which included description of the affected product and issue, clinical significance, and actions required. Consignees are directed to call the ACCU-CHEK Customer Care Service Center at 1-800-858-8072 for any questions about the information contained in the notification.

Recall details

Recall number
Z-1616-2014
Recall class
Class II
Status
Terminated
Initiated by the company
Apr 14, 2014
Posted by the FDA
May 13, 2014
Recalling firm
Roche Diagnostics Operations, Inc.
Quantity
590,136 boxes
Product code
NBW
FDA submissions
K113614
Distribution
Nationwide Distribution.

Product description

ACCU-CHEK Compact Plus Test Strips For use with: ACCU-CHEK Compact Plus Meters; 17 count sample strip 51 count retail strip; 102 count retail strip; 51 count MedBen strip 51 count Mail Order strip; Sample Kit 16 per case. The ACCU-CHEK Compact Plus Test Strips are for use with the ACCU-CHEK Compact Plus Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips or palm.

Code information (models, lots, serial numbers)

5599423160, 5599415160, 5919967160, 5919959160, 5919983160, 5907675001

Other devices under the same FDA submission

1 device cleared under K113614 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.