Medical device recalls: May 2014
The FDA posted 146 medical device recall records in May 2014, from 87 recall events by 69 companies. 9 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Siemens Medical Solutions USA, Inc · 6 events, 6 records
- Hospira Inc. · 3 events, 8 records, 3 Class I
- Teleflex Medical · 2 events, 9 records
- Siemens Healthcare Diagnostics, Inc · 2 events, 8 records
- GE Healthcare · 2 events, 6 records
- Fresenius Medical Care Holdings, Inc. · 2 events, 3 records, 1 Class I
- Alere San Diego, Inc. · 2 events, 2 records, 1 Class I
- Baxter Corporation Englewood · 2 events, 2 records, 1 Class I
FDA root cause
- Nonconforming Material/Component · 38
- Device Design · 19
- Software design · 18
- Under Investigation by firm · 9
- Process design · 9
- Process control · 9
- Employee error · 8
- Labeling mix-ups · 5
Class I recalls (9)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Z-1639-2014 · Fresenius Medical Care Holdings, Inc. · posted May 22, 2014 · Open, Classified
Naturalyte Liquid Bicarbonate maybe contaminated
- Z-1523-2014 · Maquet Datascope Corp - Cardiac Assist Division · posted May 21, 2014 · Terminated
Potential mechanical failure of the fan assembly associated with the power supply. A fan assembly failure could result in the power supply overheating and cause the IABP to shut down without warning.
- Z-1539-2014 · Fisher & Paykel Healthcare, Ltd. · posted May 20, 2014 · Terminated
The Infant Nasal CPAP Prong has the potential to detach from the Nasal Tubing and therapy is likely to be interrupted.
- Z-1567-2014 · CareFusion 303, Inc. · posted May 14, 2014 · Terminated
CareFusion is recalling the Alaris Pump model 8100 version 9.1.18 because it may have a software issue that results in situation where the pump module will not properly delay an infusion when the "Delay Until" option or "Multidose" feature is used.
- Z-1566-2014 · Baxter Corporation Englewood · posted May 12, 2014 · Terminated
1. ABACUS v3.1 may calculate quantities of electrolytes that are double the expected values during the creation of TPN orders. 2. ABACUS v3.1 may automatically add additional sterile water to a formula equal to the volume of a premix resulting in an over…
- Z-1546-2014 · Alere San Diego, Inc. · posted May 8, 2014 · Terminated
Test Strips may report an inaccurately low INR result. Several patients had a therapeutic or near-therapeutic INR with the Alere INRatio2 PT/INR Professional Test Strip but a significantly higher INR (outside of therapeutic range) when performed by a…
- Z-1482-2014 · Hospira Inc. · posted May 1, 2014 · Terminated
Hospira has received customer reports of broken door assemblies on the Hospira/Abbott Acclaim Encore infusion pumps. If the door assembly breaks, it may prevent the door from closing properly and unrestricted flow may occur. If the door cannot be closed,…
- Z-1483-2014 · Hospira Inc. · posted May 1, 2014 · Terminated
Broken door assemblies on the Hospira/Abbott Acclaim Encore infusion pumps. If the door assembly breaks, it may prevent the door from closing properly and unrestricted flow may occur. If the door cannot be closed, the pump cannot be used which can result…
- Z-1512-2014 · Hospira Inc. · posted May 1, 2014 · Terminated
There are two situations that may occur when using the GemStar Docking Station, List Number 13075-XX-XX, in conjunction with the GemStar infusion pump: 1) when the Docking Station is used in conjunction with a GemStar Phase 3 pump (List 13000-XX, 13100-XX,…
Class II recalls (136)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Z-1673-2014 · Bayer Corp · posted May 30, 2014 · Terminated
There is a potential safety risk associated with potentially damaged locking pins within the bracket of non-wireless versions of Continuum systems shipped prior to March 2008.
- Z-1674-2014 · CryoLife, Inc. · posted May 30, 2014 · Terminated
Serum albumin component monomer failed to meet internally established end of shelf-life specification.
- Z-1675-2014 · Integra LifeSciences Corp. · posted May 30, 2014 · Terminated
Integra LifeSciences has identified through a small number of complaints that some lot numbers of Licox Test Kits (Catalog Number BC10R) may provide erroneous "calibration test failure" message when they are used with the Licox PtO2 Monitor (LCSO2, or may…
- Z-1669-2014 · Synthes, Inc. · posted May 29, 2014 · Terminated
Certain lots of the ProDisc-C 2.0mm Milling Bits have an increased potential to break during use.
- Z-1671-2014 · Abbott Laboratories · posted May 29, 2014 · Terminated
Abbott has confirmed that a performance shift in the ARCHITECT Intact PTH assay has the potential to generate falsely elevated results on patient samples. Results generated with impacted lots may demonstrate a positive shift relative to those generated…
- Z-1672-2014 · Abbott Laboratories · posted May 29, 2014 · Terminated
Abbott has confirmed that a performance shift in the ARCHITECT Intact PTH assay has the potential to generate falsely elevated results on patient samples. Results generated with impacted lots may demonstrate a positive shift relative to those generated…
- Z-1665-2014 · GE Healthcare · posted May 28, 2014 · Terminated
It has been identified that in a reject image workflow where multiple studies are simultaneously opened on the Centricity PACS RA1000 workstation and images are rejected from the active displayed study, images for the previously displayed study may also be…
- Z-1666-2014 · B. Braun Medical, Inc. · posted May 28, 2014 · Terminated
B.Braun Medical, Inc. (BBMI) has received reports of 24 gauge Sprotte Needles that are missing the accompanying introducer needles.
- Z-1667-2014 · Siemens Medical Solutions USA, Inc. · posted May 28, 2014 · Terminated
Software issues with ACUSON SC2000 ultrasound with software version 3/5 (VA35) - may incorrectly display the color flow data in the wrong position and when Doppler Frequency is adjusted, the system does not restart the pulsed wave Doppler or continuous…
- Z-1668-2014 · Siemens Medical Solutions USA, Inc. · posted May 28, 2014 · Terminated
Potential malfunction due to a software issue for ACUSON SC2000 ultrasound systems using software version 3.5 (VA35) when using the stress echo feature which could cause the need for the study to be repeated.
- Z-1662-2014 · AGFA Corp. · posted May 27, 2014 · Terminated
Failure to effectively install and test a hotfix to correct the known software defect within the IMPAX CV 7.8 SU3-OCR Service.
- Z-1663-2014 · Siemens Medical Solutions USA, Inc · posted May 27, 2014 · Terminated
Siemens discovered a processing error for the MAGNETOM Aera magnetic resonance system that may have resulted in some users not receiving Operator Manuals.
- Z-1664-2014 · Beaver-Visitec International Inc. · posted May 27, 2014 · Terminated
Contains a sharp tip needle when the product should contain a blunt tip cannula
- Z-1661-2014 · Medtronic Inc. Cardiac Rhythm Disease Management · posted May 24, 2014 · Terminated
Events related to a pacing rate outside of the intended setting, including events of sudden increased pacing rate up to the maximum setting of 180 pulses per minute (ppm), have been reported. This is caused by the development of high resistance on…
- Gluma Desensitizer Power Gel ↗Class IIZ-1651-2014 · Heraeus Kulzer, LLC. · posted May 23, 2014 · Terminated
The firm discovered that the Gluma Desensitizer Power Gel can be inadvertently extruded from the application syringe all at once when it is supposed to extrude on the desired area a small amount at a time. If extruded all at once, product could reach the…
- Z-1652-2014 · Siemens Medical Solutions USA, Inc · posted May 23, 2014 · Terminated
Siemens Radiation Oncology became aware that customers may be using the Siemens Linear Accelerator in combination with stereotactic accessories which have not been validated as being compatible with Siemens LINAC models.
- Z-1653-2014 · Siemens Healthcare Diagnostics, Inc · posted May 23, 2014 · Terminated
Samples containing fluorescein may show interference with the ADVIA Centaur Systems TSH3 Ultra, Vitamin D and the ADVIA Centaur BRAHMS Procalcitonin assays.
- Z-1654-2014 · Siemens Healthcare Diagnostics, Inc · posted May 23, 2014 · Terminated
Samples containing fluorescein may show interference with the ADVIA Centaur Systems TSH3 Ultra, Vitamin D and the ADVIA Centaur BRAHMS Procalcitonin assays.
- Z-1655-2014 · Siemens Healthcare Diagnostics, Inc · posted May 23, 2014 · Terminated
Samples containing fluorescein may show interference with the ADVIA Centaur Systems TSH3 Ultra, Vitamin D and the ADVIA Centaur BRAHMS Procalcitonin assays.
- Z-1656-2014 · Siemens Healthcare Diagnostics, Inc · posted May 23, 2014 · Terminated
Samples containing fluorescein may show interference with the ADVIA Centaur Systems TSH3 Ultra, Vitamin D and the ADVIA Centaur BRAHMS Procalcitonin assays.
- Z-1657-2014 · Siemens Healthcare Diagnostics, Inc · posted May 23, 2014 · Terminated
Samples containing fluorescein may show interference with the ADVIA Centaur Systems TSH3 Ultra, Vitamin D and the ADVIA Centaur BRAHMS Procalcitonin assays.
- Z-1658-2014 · Siemens Healthcare Diagnostics, Inc · posted May 23, 2014 · Terminated
Samples containing fluorescein may show interference with the ADVIA Centaur Systems TSH3 Ultra, Vitamin D and the ADVIA Centaur BRAHMS Procalcitonin assays.
- Z-1659-2014 · Siemens Healthcare Diagnostics, Inc · posted May 23, 2014 · Terminated
Samples containing fluorescein may show interference with the ADVIA Centaur Systems TSH3 Ultra, Vitamin D and the ADVIA Centaur BRAHMS Procalcitonin assays.
- Z-1660-2014 · Physio-Control, Inc. · posted May 23, 2014 · Terminated
Physio-Control has become aware of incidents where customers have attempted to use their LIFEPAK 1000 defibrillator and the device has shut down unexpectedly due to a very low battery. A defibrillator in this scenario has the potential to fail to deliver a…
- Z-1650-2014 · Smiths Medical ASD, Inc. · posted May 22, 2014 · Terminated
The hose could separate from the blanket.
- Z-1640-2014 · GE Healthcare, LLC · posted May 21, 2014 · Terminated
When the E-NMT-01 module is used in conjunction with the ElectroSensor, the Neuromuscular Transmission (NMT) values may indicate a deeper level of muscle relaxation than the actual level of muscle relaxation. In the clinical situation visual movements of…
- Z-1642-2014 · Fujifilm Medical System USA, Inc. · posted May 21, 2014 · Terminated
The MXR-35 X-ray Generator's labeling did not include the certification statement as required by 21 CFR Part 10 10.2.
- Z-1643-2014 · Gambro Renal Products, Incorporated · posted May 21, 2014 · Terminated
The firm received several complaints for internal blood leaks involving various lot numbers produced in late 2013.
- Z-1644-2014 · Gambro Renal Products, Incorporated · posted May 21, 2014 · Terminated
The firm received several complaints for internal blood leaks involving various lot numbers produced in late 2013.
- Z-1645-2014 · Siemens Medical Solutions USA, Inc · posted May 21, 2014 · Terminated
Table may lose calibration during patient treatment, which may cause the possibility of mistreatment.
- Z-1646-2014 · Philips Healthcare Informatics, Inc. · posted May 21, 2014 · Terminated
When using XIRIS Rio version (8.1.50.0) in conjuncture with iSite PACS a synchronization error can occur and if not noticed, information for the incorrect patient may be displayed
- Dimension Vista CSA and CSAE ↗Class IIZ-1647-2014 · Siemens Healthcare Diagnostics, Inc. · posted May 21, 2014 · Terminated
A complaint was received of Dimension Vista CSA and CSAE under-recovery of cyclosporine when patient samples are run from the Small Sample Container. The under-recovery that has been observed is 18 to 44%. In March 2014, Siemens recalled due to the…
- VITROS TT4 Calibrator ↗Class IIZ-1648-2014 · Ortho-Clinical Diagnostics · posted May 21, 2014 · Terminated
Total T4 Calibrators and Reagent Packs may have calibration failures or low Quality Control results.
- Z-1649-2014 · Medline Industries Inc · posted May 21, 2014 · Terminated
This lot did not go through the correct sterilization procedures. This product may potentially be non-sterile.
- Z-1635-2014 · Microline Surgical · posted May 20, 2014 · Terminated
Grasper jaw may break when force is applied to the jaw
- TRIGEN INTERTAN NAIL:Class IIZ-1636-2014 · Smith & Nephew Inc · posted May 20, 2014 · Terminated
A manufacturing error resulted in the set screw being inserted too deep in the affected nails.
- TRIGEN INTERTAN 1.5 NAIL:Class IIZ-1637-2014 · Smith & Nephew Inc · posted May 20, 2014 · Terminated
A manufacturing error resulted in the set screw being inserted too deep in the affected nails.
- MAMMOMAT Inspiration.Class IIZ-1638-2014 · Siemens Medical Solutions USA, Inc · posted May 20, 2014 · Terminated
There is a potential and possible hazard to the user when using the MAMMOMAT Inspiration PC monitor at the control desk, in that the holder of the PC monitor can break causing an unstable monitor to fall causing possible serious injury.
- Z-1630-2014 · SCM True Air Technologies LLC · posted May 19, 2014 · Terminated
The firm is recalling the products due to lack of design controls, lack of a Device Master Record, and failure to approve specific vendors.
- Z-1631-2014 · SCM True Air Technologies LLC · posted May 19, 2014 · Terminated
The firm is recalling the products due to lack of design controls, lack of a Device Master Record, and failure to approve specific vendors.
- Z-1632-2014 · SCM True Air Technologies LLC · posted May 19, 2014 · Terminated
The firm is recalling the products due to lack of design controls, lack of a Device Master Record, and failure to approve specific vendors.
- Z-1633-2014 · Arrow International Inc · posted May 19, 2014 · Terminated
The labeling states that the kit contains Introducer Needle: Echogenic 18 Ga. x 2.5 (6.35cm) XTW and 5mL ArroW Raulerson Spring-Wire Introduction Syringe; however, the kit contains, and should state, Introducer Needle: 18 Ga. x 2.5 (6.35cm) XTW and 5mL…
- Z-1634-2014 · Terumo Cardiovascular Systems Corporation · posted May 19, 2014 · Terminated
There is a potential for inaccuracies in blood gas readings/measurements. There is a potential risk of respiratory alkalosis or cerebral vasoconstriction with low paCO2. If the Hemoglobin and Hematocrit measurements are not accurate, there is a potential…
- syngo Imaging XS ↗Class IIZ-1622-2014 · Siemens Medical Solutions USA, Inc · posted May 16, 2014 · Terminated
A complaint reported that the archiving method at one customer site was changed from implicit to explicit, and the cleanup procedure enabled by the explicit archivingmethod may have deleted data from archive itself.
- LATEX-FREE ↗Class IIZ-1623-2014 · Hospira Inc. · posted May 16, 2014 · Terminated
The Tyvek cover may not be completely sealed to the rest of the package allowing openings at the edge of the sterile package for the following products: List No. 04693-01, Latex Free Extension Set; List No. 06543-01, Nonvented Y-Set for Transurethral…
- LATEX-FREE ↗Class IIZ-1624-2014 · Hospira Inc. · posted May 16, 2014 · Terminated
The Tyvek cover may not be completely sealed to the rest of the package allowing openings at the edge of the sterile package for the following products: List No. 04693-01, Latex Free Extension Set; List No. 06543-01, Nonvented Y-Set for Transurethral…
- LATEX-FREE ↗Class IIZ-1625-2014 · Hospira Inc. · posted May 16, 2014 · Terminated
The Tyvek cover may not be completely sealed to the rest of the package allowing openings at the edge of the sterile package for the following products: List No. 04693-01, Latex Free Extension Set; List No. 06543-01, Nonvented Y-Set for Transurethral…
- LATEX-FREE ↗Class IIZ-1626-2014 · Hospira Inc. · posted May 16, 2014 · Terminated
The Tyvek cover may not be completely sealed to the rest of the package allowing openings at the edge of the sterile package for the following products: List No. 04693-01, Latex Free Extension Set; List No. 06543-01, Nonvented Y-Set for Transurethral…
- LATEX-FREE ↗Class IIZ-1627-2014 · Hospira Inc. · posted May 16, 2014 · Terminated
The Tyvek cover may not be completely sealed to the rest of the package allowing openings at the edge of the sterile package for the following products: List No. 04693-01, Latex Free Extension Set; List No. 06543-01, Nonvented Y-Set for Transurethral…
- Z-1628-2014 · Philips Medical Systems, Inc. · posted May 16, 2014 · Terminated
The risk of battery failure increases with age, when a battery remains in use longer than 3 years after date of manufacture or 500 charge-discharge cycles. Such failure can result in overheating that in rare cases can cause the battery to ignite or explode.
- Anspach Irrigation Clip 40 ↗Class IIZ-1629-2014 · The Anspach Effort, Inc. · posted May 16, 2014 · Terminated
The Irrigation Clips were assembled with the incorrect (smaller) clips.
- Z-1621-2014 · InterValve Inc · posted May 15, 2014 · Terminated
Leak issues were discovered through accelerated age testing.
- Z-1607-2014 · HeartWare Inc · posted May 14, 2014 · Terminated
Patients and caregivers are instructed that if a battery does not provide two hours of support or behaves unusually, that battery should no longer be used and should be replaced.
- Z-1620-2014 · Baxter Corporation Englewood · posted May 14, 2014 · Terminated
Possible breach in sterile barrier for inner and outer bags of the product.
- Z-1553-2014 · Siemens Healthcare Diagnostics · posted May 13, 2014 · Terminated
Instrument Cover Gas Spring failures. Over time, the gas spring may lose its effectiveness and fail to support the cover in its full or partially open position. This may lead to the cover falling.
- Z-1610-2014 · Ortho Development Corporation · posted May 13, 2014 · Terminated
A femoral component containing pegs was found in a box for the pegless version.
- Z-1611-2014 · Mindray DS USA, Inc. dba Mindray North America · posted May 13, 2014 · Terminated
The MPM included a cardiac ST/Arrhythmia analysis feature which is not cleared for marketing.
- Z-1612-2014 · Medline Industries Inc · posted May 13, 2014 · Terminated
There is a potential that the specified mucus specimen traps are not sterile. Use of the affected products may result in the contamination of patient specimens.
- Z-1613-2014 · Siemens Medical Solutions USA, Inc · posted May 13, 2014 · Terminated
Systems with the affected serial number are experiencing a software error which may lead to incorrect patient demographics display.
- Coaxial Micro-Introducer Kit. ↗Class IIZ-1614-2014 · Greatbatch Medical · posted May 13, 2014 · Terminated
Greatbatch identified that the documentation does not support the five-year shelf life of the Bard Access Division Coaxial Micro-Introducer Set.
- ExploR 7x26mm Modular Radial Stem, Catalog number 11-210062. Elbow prosthesis orthopedic implant.Class IIZ-1615-2014 · Biomet, Inc. · posted May 13, 2014 · Terminated
This lot of the ExploR Modular Radial Stem may be missing the wedge ramp threads. If the wedge ramp threads are missing from the stem, the likelihood of the set screw loosening from the radial stem intraoperatively is increased. If the set screw loosens,…
- Z-1616-2014 · Roche Diagnostics Operations, Inc. · posted May 13, 2014 · Terminated
Roche Diabetes Care has become aware the ACCU-CHEK Compact Plus test strips may produce erroneously low blood glucose readings in patients undergoing Ceftriaxone therapy. The interference is not described in the product labeling.
- McKesson Radiology-PACS. ↗Class IIZ-1617-2014 · Mckesson Medical Immaging · posted May 13, 2014 · Terminated
High priority alert icon was not displayed in the Study List when there are no Stat studies present. However, the High priority alert icon is visible when studies marked as Stat priority exist in the Study List.
- Z-1618-2014 · Fresenius Medical Care Holdings, Inc. · posted May 13, 2014 · Terminated
2008-Series Hemodialysis Machines with Revision K Actuator Test Boards may fail to detect a disconnected Ultrafiltration (UF) Pump
- Z-1619-2014 · Fresenius Medical Care Holdings, Inc. · posted May 13, 2014 · Terminated
2008-Series Hemodialysis Machines with Revision K Actuator Test Boards may fail to detect a disconnected Ultrafiltration (UF) Pump
- Z-1608-2014 · RM Temena GmbH · posted May 12, 2014 · Terminated
Ultrasound needles may be defective and leak at the needle hub.
- Z-1609-2014 · RM Temena GmbH · posted May 12, 2014 · Terminated
Ultrasound needles may be defective and leak at the needle hub.
- Z-1603-2014 · Zimmer, Inc. · posted May 10, 2014 · Terminated
During a transfer of products from a recently shutdown facility, the firm discovered raw material labeled as latex free actually contained latex.
- Z-1604-2014 · Zimmer, Inc. · posted May 10, 2014 · Terminated
During a transfer of products from a recently shutdown facility, the firm discovered raw material labeled as latex free actually contained latex.
- Z-1605-2014 · Zimmer, Inc. · posted May 10, 2014 · Terminated
During a transfer of products from a recently shutdown facility, the firm discovered raw material labeled as latex free actually contained latex.
- Z-1606-2014 · Zimmer, Inc. · posted May 10, 2014 · Terminated
During a transfer of products from a recently shutdown facility, the firm discovered raw material labeled as latex free actually contained latex.
- Z-1571-2014 · Codman & Shurtleff, Inc. · posted May 9, 2014 · Terminated
A small number of specific pumps may trigger a low battery alarm at the beginning of the implanted life.
- Z-1572-2014 · Codman & Shurtleff, Inc. · posted May 9, 2014 · Terminated
A small number of specific pumps may trigger a low battery alarm at the beginning of the implanted life.
- Z-1573-2014 · SCC Soft Computer · posted May 9, 2014 · Terminated
When an isolate is resulted without a SNOMED code, isolate information in the downstream system may be incomplete or missing. As a result, there is the potential for the delay or omission of patient treatment updates.
- Z-1574-2014 · The Anspach Effort, Inc. · posted May 9, 2014 · Terminated
Anspach product code L-3SD contained product code S-3SD which is shorter in length than the labeled product.
- NavioPFSTM System, Orthopedic Sterotaxic Instrument, model #'s NPFS-02000, NPFS-02010, NPFS 02020. ↗Class IIZ-1575-2014 · Blue Belt Technologies MN · posted May 9, 2014 · Terminated
Blue Belt Technologies is conducting a voluntary recall for their Navio PFS system which is used during orthopedic knee surgery. One of the electronic modules inside the computer cart assembly may have a defect in one of its components that could result in…
- Z-1576-2014 · William C. Domb, D.M.D., A Professional Corporation · posted May 9, 2014 · Terminated
William Domb is recalling the Enaly 1000 BT-12 Ozone Generator because it is not approved or cleared by the FDA for medical use.
- Z-1577-2014 · Nellcor Puritan Bennett Inc. (dba Covidien LP) · posted May 9, 2014 · Terminated
Reports of the audible alarm not sounding when the ventilator is in an alarm state. The ventilator, as well as the visual and remote alarms, continue to operate normally.
- BAG 9732315 MOUSE STERILE 15 PKClass IIZ-1578-2014 · Medtronic Navigation, Inc. · posted May 9, 2014 · Terminated
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
- Z-1579-2014 · Medtronic Navigation, Inc. · posted May 9, 2014 · Terminated
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
- BIT 9733516 2.9mm DRILL STERILEClass IIZ-1580-2014 · Medtronic Navigation, Inc. · posted May 9, 2014 · Terminated
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
- Z-1581-2014 · Medtronic Navigation, Inc. · posted May 9, 2014 · Terminated
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
- Z-1582-2014 · Medtronic Navigation, Inc. · posted May 9, 2014 · Terminated
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
- Z-1583-2014 · Medtronic Navigation, Inc. · posted May 9, 2014 · Terminated
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
- Z-1584-2014 · Medtronic Navigation, Inc. · posted May 9, 2014 · Terminated
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
- Z-1585-2014 · Medtronic Navigation, Inc. · posted May 9, 2014 · Terminated
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
- Z-1586-2014 · Medtronic Navigation, Inc. · posted May 9, 2014 · Terminated
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
- Z-1587-2014 · Medtronic Navigation, Inc. · posted May 9, 2014 · Terminated
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
- MOUSE 9732721 STERILE O-ARM 10PKClass IIZ-1588-2014 · Medtronic Navigation, Inc. · posted May 9, 2014 · Terminated
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
- Z-1589-2014 · Medtronic Navigation, Inc. · posted May 9, 2014 · Terminated
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
- Z-1590-2014 · Medtronic Navigation, Inc. · posted May 9, 2014 · Terminated
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
- Z-1591-2014 · Medtronic Navigation, Inc. · posted May 9, 2014 · Terminated
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
- Z-1592-2014 · Medtronic Navigation, Inc. · posted May 9, 2014 · Terminated
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
- Z-1593-2014 · Medtronic Navigation, Inc. · posted May 9, 2014 · Terminated
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
- SHUNT KIT 9733605 NON-INVASIVEClass IIZ-1594-2014 · Medtronic Navigation, Inc. · posted May 9, 2014 · Terminated
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
- SPHERES 8801071 1/TRAY 12PKClass IIZ-1595-2014 · Medtronic Navigation, Inc. · posted May 9, 2014 · Terminated
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
- SPHERES 8801075 5/TRAY 12PKClass IIZ-1596-2014 · Medtronic Navigation, Inc. · posted May 9, 2014 · Terminated
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
- Z-1597-2014 · Medtronic Navigation, Inc. · posted May 9, 2014 · Terminated
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
- Z-1598-2014 · Medtronic Navigation, Inc. · posted May 9, 2014 · Terminated
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
- Z-1599-2014 · Medtronic Navigation, Inc. · posted May 9, 2014 · Terminated
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
- Z-1600-2014 · Medtronic Navigation, Inc. · posted May 9, 2014 · Terminated
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
- Z-1601-2014 · Medtronic Navigation, Inc. · posted May 9, 2014 · Terminated
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
- Z-1602-2014 · Medtronic Navigation, Inc. · posted May 9, 2014 · Terminated
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
- Z-1568-2014 · Medtronic Neuromodulation · posted May 8, 2014 · Terminated
Some refill kits in the affected lots may contain an extension tubing set that is occluded and healthcare professionals may be unable to aspirate or fill the SynchroMed pumps. There is no problem with the Lioresal Intrathecal (baclofen injection)…
- Z-1569-2014 · Shimadzu Medical Systems · posted May 8, 2014 · Terminated
Shimadzu Corporation is recalling the Shimadzu Digital X-Ray System because they have received a report of a lost image under certain circumstances.
- Z-1570-2014 · Medtronic Neuromodulation · posted May 8, 2014 · Terminated
This recall provides important new information regarding overinfusion associated with the Medtronic SynchroMed II Implantable Pump. Overinfusion can result in a life-threatening overdose and can also result in drug withdrawal due to premature emptying of…
- Z-1555-2014 · Teleflex Medical · posted May 7, 2014 · Terminated
Product does not meet minimum needle attachment strength requirements; thus, the needle can become detached from the suture during use.
- Z-1556-2014 · Teleflex Medical · posted May 7, 2014 · Terminated
Product does not meet minimum needle attachment strength requirements; thus, the needle can become detached from the suture during use.
- Z-1557-2014 · Teleflex Medical · posted May 7, 2014 · Terminated
Product does not meet minimum needle attachment strength requirements; thus, the needle can become detached from the suture during use.
- Z-1558-2014 · Teleflex Medical · posted May 7, 2014 · Terminated
Product does not meet minimum needle attachment strength requirements; thus, the needle can become detached from the suture during use.
- Z-1559-2014 · Teleflex Medical · posted May 7, 2014 · Terminated
Product does not meet minimum needle attachment strength requirements; thus, the needle can become detached from the suture during use.
- Z-1560-2014 · Teleflex Medical · posted May 7, 2014 · Terminated
Product does not meet minimum needle attachment strength requirements; thus, the needle can become detached from the suture during use.
- Deknatel Teleflex Medical Silk Black Braided Silk Suture Nonabsorbable Surgical Suture U.S.P. ↗Class IIZ-1561-2014 · Teleflex Medical · posted May 7, 2014 · Terminated
Product does not meet minimum needle attachment strength requirements; thus, the needle can become detached from the suture during use.
- Siemens ADVIA Centaur FolateClass IIZ-1564-2014 · Siemens Healthcare Diagnostics, Inc · posted May 7, 2014 · Terminated
Serum controls may go out of range low with the ADVIA Centaur Systems Folate assay and a shift in whole blood patient may result with kit lots ending in 219 and 223
- Z-1565-2014 · SpineFrontier, Inc. · posted May 7, 2014 · Terminated
It has been verified through field use that there are two possible breakages that may occur when using SI50001 Revision D Lock Shaft Inserter, possibly resulting in instrument fragments entering the sterile field: Impact Cap: The impact cap on the Lock…
- Z-1547-2014 · GE Healthcare · posted May 6, 2014 · Terminated
Medical device software may be defective and may cause user-set alarm limits to result in false positive or false negative Oxygen saturation and pulse rate alarm notifications to the caregiver.
- Z-1548-2014 · GE Healthcare · posted May 6, 2014 · Terminated
Medical device software may be defective and may cause user-set alarm limits to result in false positive or false negative Oxygen saturation and pulse rate alarm notifications to the caregiver.
- Z-1549-2014 · GE Healthcare · posted May 6, 2014 · Terminated
Medical device software may be defective and may cause user-set alarm limits to result in false positive or false negative Oxygen saturation and pulse rate alarm notifications to the caregiver.
- Z-1550-2014 · GE Healthcare · posted May 6, 2014 · Terminated
Medical device software may be defective and may cause user-set alarm limits to result in false positive or false negative Oxygen saturation and pulse rate alarm notifications to the caregiver.
- Z-1551-2014 · GE Healthcare · posted May 6, 2014 · Terminated
Medical device software may be defective and may cause user-set alarm limits to result in false positive or false negative Oxygen saturation and pulse rate alarm notifications to the caregiver.
- Z-1552-2014 · Tenex Health Inc · posted May 6, 2014 · Terminated
The sterile barrier in the packaging may be compromised due to cracks in the thermoform tray that may occur during transport of the product in extreme cold weather conditions, in conjunction with mechanical impact.
- Z-1554-2014 · Bausch & Lomb Inc- Greenville Solutions Plant · posted May 6, 2014 · Terminated
Out of specification results for one of the two disinfectants.
- Z-1562-2014 · AlloSource, Inc. · posted May 6, 2014 · Terminated
The donor was hemodiluted.
- Z-1563-2014 · Mar Cor Purification · posted May 6, 2014 · Terminated
Mar Cor Purification manufactured Central Water Plant (CWP) reverse osmosis system, models 102, 104 and 106. This correction has been initiated due to a potential situation with a specific lot of the water inlet solenoid valve. The potential exists that…
- Z-1528-2014 · INO Therapeutics (dba Ikaria) · posted May 2, 2014 · Terminated
An anomaly impacts the monitoring system of INOmax DSIR Plus (software 3.0.0). When the integrated pneumatic backup switch is activated with low or no gas flow through the injector module, an intended NO cell failure alarm will occur. In addition, an…
- Z-1537-2014 · Bio-Rad Laboratories, Inc. · posted May 2, 2014 · Terminated
There is an issue with the buffers which impacts the quantification of the Hemoglobin F peak. Calibration and Quality control are likely to fail if the affected lots are used.
- Z-1538-2014 · Biorep Technologies · posted May 2, 2014 · Terminated
The support arm attached with an adhesive to the body of the blade is detaching while attempting to engage the visor component.
- Z-1540-2014 · Alere San Diego, Inc. · posted May 2, 2014 · Terminated
These lots may not recover within range when tested using certain commercially available controls.
- Z-1541-2014 · Med-Mizer, Inc. · posted May 2, 2014 · Terminated
All configurations of the SS Retractabed, Clinical Contour, made prior to June 7 2008 may have had insufficient welding on the foot cross tube. This tube is responsible for holding the two support rails together when using the foot side rails and keeps…
- Z-1542-2014 · Med-Mizer, Inc. · posted May 2, 2014 · Terminated
All configurations of the SS Retractabed, Clinical Contour, made prior to June 7 2008 may have had insufficient welding on the foot cross tube. This tube is responsible for holding the two support rails together when using the foot side rails and keeps…
- Z-1543-2014 · Med-Mizer, Inc. · posted May 2, 2014 · Terminated
All configurations of the SS Retractabed, Clinical Contour, made prior to June 7, 2008 may have had insufficient welding on the foot cross tube. This tube is responsible for holding the two support rails together when using the foot side rails and keeps…
- Z-1544-2014 · Stryker Endoscopy · posted May 2, 2014 · Terminated
Gravity sterilization parameters provided in the 12 Instrument Sterilization reprocessing guide did not fully consider all worst-case scenarios during validation testing.
- Z-1533-2014 · GE Healthcare, LLC · posted May 1, 2014 · Terminated
GE Healthcare has recently become aware of a potential safety issue with certain GE MRI Products. Patient egress from the system can be delayed or impeded if the cradle release handle and/or cradle release block mechanism malfunctions or becomes…
- Z-1534-2014 · Teleflex Medical · posted May 1, 2014 · Terminated
The product did not meet minimum and/or average minimum Teleflex resorption strength requirements.
- Z-1535-2014 · Teleflex Medical · posted May 1, 2014 · Terminated
The product did not meet minimum and/or average minimum Teleflex resorption strength requirements.
- Z-1536-2014 · BD Biosciences, Systems & Reagents · posted May 1, 2014 · Terminated
Two lots of Anti EpCAM PE-Analyte Specific Reagent are mislabeled - the actual antibody present in the vial is CD9 PE.
Class III recalls (1)
Using the device is not likely to cause health problems, but it violates FDA rules.
- Z-1545-2014 · Siemens Healthcare Diagnostics Inc · posted May 4, 2014 · Terminated
Device with Sample Interference Notes (SIN) enabled not cleared for US marketing.
Source: FDA Medical Device Recalls and FDA enforcement reports (recall class), via openFDA. Records are grouped by the date the FDA posted them. Methodology · Not medical advice.