Philips/Children's Medical Ventures Smart Monitor 2… recallZ-1789-2014 · Childrens Medical Ventures · initiated Apr 29, 2014

Class II Terminated All recalls posted in June 2014

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Childrens Medical Ventures or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The battery pack wire harness is improperly assembled rendering the device inoperable. The wiring issue causes an error code to be displayed and results in the device sounding an alarm repeatedly at power up. The presence of the error code and the devices alarm prevents the device from functioning until the issue is addressed.

FDA root cause category: Component design/selection

What the company is doing

Philips sent an Urgent Medical Device Correction letter dated April 29, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. This Customer Information is intended to inform you about: 1.what the problem is and under what circumstances it can occur 2.the actions that you as a customer can take to minimize the effect of the problem 3.the actions planned by Philips to correct the problem. If you need any further information or support concerning this issue, please contact your local Philips representative: For further questions please call (412) 380-8881. Philips apologizes for any inconveniences caused by this problem.

Recall details

Recall number
Z-1789-2014
Recall class
Class II
Status
Terminated
Initiated by the company
Apr 29, 2014
Posted by the FDA
Jun 13, 2014
Recalling firm
Childrens Medical Ventures
Quantity
112
Product code
NPF
FDA submissions
K061256
Distribution
Worldwide Distribution - USA including the states of MI, WI, SC, FL, TX, KS, MA, AZ, UT, WV, IN, MN, NV and VA., and the countries of Australia, Belgium and Canada.

Product description

Philips/Children's Medical Ventures Smart Monitor 2 Professional Series infant apnea monitor.

Code information (models, lots, serial numbers)

989805603611 / 1014557, 989805603631 / 1014559, 989805604511 / 1028969, 989805604521 / 1028970, 989805604531 / 1028971, 989805604631 / 1030271, 989805605051 / 1038140, 989805605061 / 1038141, 989805608801 / U1014557, 989805608811 / U1014559, 989805608831 / U1028969, 989805608841 / U1028970, 989805608851 / U1038140, 989805604231 / 1023384

Other devices under the same FDA submission

4 devices cleared under K061256 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.