Roche MODULAR ANALYTICS EVO Serum Work Area (SWA)… recallZ-1790-2014 · Roche Diagnostics Operations, Inc. · initiated May 21, 2014
Class II Terminated All recalls posted in June 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Roche Diagnostics Operations, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Under certain conditions loose ProCell/CleanCell aspiration tube filters may cause: consecutive discrepant results; sipper alarm; foam on reservoirs and system reagent alarms.
FDA root cause category: Labeling design
What the company is doing
Roche sent an Urgent Medical Device Correction (UMDC) letter dated May 20, 2014 via UPS Grount (receipt required) to affected customers. The letter identified the affected product, problem and actions to be taken. The letter instructed customers to: 1. Follow the instructions in the Precautions Regarding Handling the Aspiration Tube Filter section in this UMDC; 2. Discontinue performing ProCell/Clean Cell aspiration tube filter check as this maintenance item will be performed by your Roche FSR as a part of service maintenance; 3. If you have distributed the products listed within this UMDC to other sites or facilities, please provide them with a copy of the UMDC; and 4. Complete the faxback form enclosed with this UMDC and fax it to 1-888-410-6620. For questions contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions about the information contained in this UMDC.
Recall details
- Recall number
- Z-1790-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- May 21, 2014
- Posted by the FDA
- Jun 14, 2014
- Recalling firm
- Roche Diagnostics Operations, Inc.
- Quantity
- 462
- Product code
- JJE
- FDA submissions
- K961481
- Distribution
- US Nationwide Distribution
Product description
Roche MODULAR ANALYTICS EVO Serum Work Area (SWA), Operator's Manual Software Version 08-02; Filter label: Filter (N) Mat. No 03149773001, Contents 1, 768-320 Filter (N). Product Usage: Fully automated immunoassay analyzer intended for the in-vitro quantitative/qualitative determination of analytes in body fluids.
Code information (models, lots, serial numbers)
Show the full code information (4k characters)
1305-31, 1305-21, 1305-22, 1308-35, 1308-36, 130833, 130834, 1308-44, 1310-20, 1418-12, 1307-33, 1421-11, 1417-20, 1423-16, 143115, 1428-07, 1430-16, 1430-17, 1423-15, 1429-15, 1305-40, 1541-05, 1539-15, 1548-15, 1544-11, 1431-14, 1548-17, 1545-03, 1547-20, 1557-12, 1544-06, 1548-08, 1429-13, 1303-31, 1303-32, 1556-16, 1549-04, 1548-16, 1540-20, 1538-10, 1563-06, 1556-17, 1539-13, 1539-14, 1571-12, 1572-19, 1684-05, 1570-14, 1571-13, 1679-13, 1684-06, 1683-17, 1570-17, 1684-07, 1684-08, 1685-16, 1572-20, 1673-01, 1691-17, 1691-19, 1691-18, 1698-05, 1698-06, 1691-14, 1728-10, 1694-19, 1711-17, 1708-20, 1709-03, 1601-01, 1698-03, 1601-03, 1705-20, 1714-18, 1714-19, 1847-02, 1686-01, 1733-17, 1733-19, 1733-16, 1736-06, 1843-02, 1570-15, 1849-01, 1855-15, 1857-22, 1860-10, 1691-20, 1696-05, 1567-01, 1860-24, 1853-02, 1855-10, 1855-11, 1858-02, 1858-03, 1860-12, 1860-13, 1860-14, 1736-04, 1849-02, 1853-01, 1860-11, 1857-23, 1857-24, 1739-19, 1857-25, 1857-26, 1861-18, 1861-19, 1861-23, 1853-03, 1860-16, 1860-23, 1860-15, 1864-16, 1860-19, 1985-13, 1989-19, 1996-02, 1861-20, 1865-19, 1307-34, 1992-19, 1999-16, 1999-17, 1864-19, 1980-15, 1535-09, 1999-20, 1993-01, 1996-01, 1999-18, 1905-05, 1905-03, 1921-03, 1925-03, 1860-22, 1860-20, 1905-01, 1921-01, 1858-04, 1860-17, 1905-02, 1921-02, 1418-03, 1418-04, 1925-04, 1905-06, 1999-19, 2050-04, 2050-03, 2050-06, 2050-07, 2050-05, 1537-15, 2052-01, 2052-02, 2050-08, 1725-18, 1694-20, 1733-18, 2056-22, 2056-23, 1419-12, 1422-17, 1307-41, 1536-04, 2065-12, 1905-07, 1905-08, 1864-17, 1864-18, 2061-20, 2065-06, 1736-03, 2065-13, 2076-18, 2068-24, 2080-02, 2071-24, 2071-23, 2068-23, 2081-05, 2081-06, 2070-01, 2079-19, 2080-01, 2081-04, 2070-02, 2086-14, 2086-20, 2084-16, 2084-17, 2084-18, 2085-09, 2086-17, 2086-18, 2082-39, 2084-05, 2084-12, 2084-13, 2084-14, 2084-15, 2085-05, 2085-07, 2085-08, 2188-05, 2188-06, 2188-07, 2086-15, 2086-16, 2086-19, 2188-08, 2188-09, 2188-10, 2084-06, 2084-07, 2071-22, 2084-08, 2084-09, 2084-10, 2084-11, 2085-04, 2085-10, 2188-04, 2084-19, 2084-20, 2084-21, 2076-19, 2076-20, 2190-10, 2190-19, 2190-20, 2190-06, 2190-07, 2190-08, 2190-09, 2190-11, 2190-12, 2081-07, 2081-08, 2081-09, 2081-10, 2081-11, 2083-04, 2083-05, 2084-22, 2084-23, 2076-21, 2190-13, 2083-06, 2083-07, 2083-08, 2079-18, 2081-12, 2190-14, 2080-03, 2188-36, 2188-37, 2188-38, 2188-31, 2188-34, 2188-35, 2187-33, 2188-39, 2188-40, 2190-17, 2190-18, 2190-21, 2082-40, 2083-01, 2188-32, 2190-03, 2190-04, 2190-05, 2190-15, 2190-16, 1549-02, 2065-04, 2071-20, 2071-21, 2189-40, 2190-01, 2190-02, 2083-03, 2085-18, 2189-37, 2189-38, 2189-39, 2085-16, 2189-36, 2192-24, 1561-12, 1544-10, 2083-02, 2085-17, 1537-17, 2187-36, 2187-37, 2187-38, 2102-06, 2102-07, 2102-08, 2085-12, 2192-21, 2192-22, 2086-12, 2106-28, 2106-29, 2110-09, 2110-10, 2189-01, 2085-13, 2192-28, 2192-29, 2085-14, 2192-26, 2192-27, 2192-31, 2086-23, 2086-24, 2102-03, 2195-23, 2085-11, 2085-15, 2192-25, 2192-32, 1711-16, 2086-25, 2086-26, 2101-39, 2188-03, 2189-04, 2188-01, 2188-02, 2119-13, 2189-05, 2189-06, 2086-13, 2086-21, 2189-02, 2189-03, 2102-04, 2102-05, 2119-14, 2192-18, 2192-19, 2126-07, 2132-14, 2192-20, 2126-08, 2189-07, 2140-02, 1545-05, 1686-03, 2146-17, 2146-18, 2146-19, 2251-16, 2251-17, 2251-18, 2252-02, 2252-04, 2253-15, 2147-05, 2147-06, 2251-19, 2251-20, 2252-03, 2253-16, 2253-17, 2253-18, 2260-11, 2260-12, 2260-13, 2257-08, 2257-09, 2257-10, 2257-11, 2252-05, 2253-19, 2253-20, 2254-01, 2255-20, 2257-04, 2056-21, 2147-04, 2251-15, 2253-12, 2253-13, 2253-14, 2257-05, 2257-06, 2257-07, 1725-19, 1857-27, 2076-22, 2260-17, 2260-18, 2260-19, 2260-20, 2262-11, 2262-12, 1725-20, 1691-15, 2260-14, 2260-15, 2260-16, 2256-01, 1684-02, 2256-02, 2262-16, 2262-17, 2263-12, 2262-19, 2262-20, 2267-20, 2263-13, 2263-14, 1847-01, 2071-25, 2075-20, 2262-18, 2275-11, 2275-12, 2267-17, 2267-18, 2263-16, 2263-17, 2275-14, 2290-05, 2267-19, 2275-09, 2275-10, 2275-13, 1556-12, 2275-15, 2275-16, 2278-07, 2290-06, 2278-08, 2283-16, 2296-20, 1563-04, 2278-09, 2263-15, 2278-10, 2278-11, 2283-17, 1534-12, 1601-02, 2330-19, 2331-01, 1544-13, 2330-20, 2331-02, 2330-16, 2330-17, 2330-18, 1559-13, 1307-35.
Other devices under the same FDA submission
2 devices cleared under K961481 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- cobas e 411 Immunoassay Analyzer 04775201001Roche Diagnostics GmbH · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE685 reports
- cobas e 411 Immunoassay Analyzer 04775279001Roche Diagnostics GmbH · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE533 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 68294).
- Z-1791-2014 · Terminatedcobas 6000 analyzer series Operators Manual Roche cobas; Filter label: Filter (N) Mat. No 03149773001, Contents 1, 768-320 Filter (N) Product Usage: Fully automated immunoassay…
- Z-1792-2014 · Terminatedcobas 8000 analyzer series Operators Manual Roche cobas Filter label: Filter (N) Mat. No 03149773001, Contents 1, 768-320 Filter (N) Product Usage: Fully automated immunoassay…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page