cobas e 411 Immunoassay Analyzer 04775201001Recalls & reported problems
Roche Diagnostics GmbH · Class 1
All models of cobas e 411 Immunoassay Analyzer (2)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Roche cobas immunoassay analyzer has 683 adverse event reports, with no deaths and no injuries. Most describe lab results that could not be reproduced or were too high or too low, such as hormone, pregnancy and hepatitis tests where repeat testing gave different answers. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 680
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 680 adverse event reports that list this exact device (DI 04015630924721), received between Jul 12, 2018 and Sep 30, 2026 — classified as 680 malfunctions.
The most frequently reported specific device problem is “Non Reproducible Results” (378 reports).
6 other recalls cover devices cleared under the same FDA submission (K961481); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Non Reproducible Results378
- Low Test Results331
- High Test Results168
- Incorrect, Inadequate or Imprecise Result or Readings81
- Incorrect Measurement28
- Output Problem16
- Mechanical Problem14
- Insufficient Device Problem Information7
Patient problems
- No Clinical Signs, Symptoms or Conditions584
- No Consequences Or Impact To Patient66
- No Known Impact Or Consequence To Patient21
- Test Result4
- Insufficient Information3
- Cardiac Enzyme Elevation1
- Myocardial Infarction1
- No Patient Involvement1
Recent adverse event reports
The 25 most recent of 680 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings; Output Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50947527 ↗MalfunctionSep 30, 2026False Positive Result
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50946531 ↗MalfunctionJun 3, 2026High Test Results; Non Reproducible Results
THERE WAS AN ALLEGATION OF A QUESTIONABLE ELECSYS TOTAL PSA RESULT FOR 1 PATIENT SAMPLE ON A COBAS E 411 ANALYZER (RACK SYSTEM) MODULE. THE INITIAL RESULT WAS 0.695 NG/ML. THE REPEAT RESULTS ON (B)(6) 2026 WERE BOTH 0.006 NG/ML, ACCOMPANIED BY DATA FLAGS. THE PATIENT COMPLAINED ABOUT THE INITIAL RESULT WHEN THEY GOT A DIFFERENT RESULT FROM ANOTHER LABORATORY. THE REPEAT RESULTS WERE DEEMED TO BE CORRECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25377970 ↗MalfunctionMay 22, 2026Low Test Results; Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE VITAMIN B12 RESULTS FOR 1 PATIENT SAMPLE ON A COBAS E 411 ANALYZER (RACK SYSTEM). THE INITIAL RESULT WAS >37 PMOL/L. THE REPEAT RESULT WAS 457.9 PMOL/L.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25257069 ↗MalfunctionMay 19, 2026High Test Results; Low Test Results; Non Reproducible Results
WE RECEIVED AN ALLEGATION ABOUT DISCREPANT RESULTS FOR 1 PATIENT SAMPLE TESTED WITH ELECSYS HBSAG II ASSAY AND 1 PATIENT SAMPLE TESTED WITH ELECSYS ANTI-HBS ASSAY ON A COBAS E 411 ANALYZER (RACK SYSTEM). SAMPLE 1 WAS TESTED FOR HBSAG II: INITIAL RESULT: 1.67 (REACTIVE). ON (B)(6) 2026: REPEAT RESULT: 0.641 (NON-REACTIVE). SAMPLE 2 WAS TESTED FOR ANTI-HBS: INITIAL RESULT: <2.00 IU/L (NON-REACTIVE). THE PHYSICIAN QUESTIONED THE INITIAL RESULT, AND THE SAMPLE WAS REPEATED. ON (B)(6) 2026: REPEAT RESULT: >1000 IU/L (REACTIVE).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25222863 ↗MalfunctionMay 15, 2026High Test Results; Low Test Results; Non Reproducible Results
THE INITIAL REPORTER RECEIVED QUESTIONABLE ELECSYS HCG+BETA AND ELECSYS HIV COMBI PT RESULTS FOR TWO PATIENT SAMPLES TESTED ON THE COBAS E 411 ANALYZER (RACK SYSTEM). PATIENT SAMPLE 1 ELECSYS HIV COMBI PT ON (B)(6) 2026, THE INITIAL RESULT WAS 13.32 COI (REACTIVE). ON (B)(6) 2026, THE REPEAT RESULT WAS 0.346 COI (NON-REACTIVE). PATIENT SAMPLE 2 ELECSYS HCG+BETA ON (B)(6) 2026, THE INITIAL RESULT WAS 10.88 MIU/ML. THE REPEAT RESULT WAS 0.100 MIU/ML WITH A DATA FLAG.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25183505 ↗MalfunctionMay 5, 2026Incorrect, Inadequate or Imprecise Result or Readings; High Test Results; Low Test Results; Non Reproducible Results
THE INITIAL REPORTER RECEIVED QUESTIONABLE TESTOSTERONE II ASSAY (TESTOSTERONE II) RESULTS FROM ONE PATIENT SAMPLE TESTED ON THE COBAS E 411 ANALYZER (DISK SYSTEM). THE INITIAL RESULT OF THE INITIAL PATIENT SAMPLE IS 539.5 NG/DL (18.72 NMOL/L). THE REPEAT RESULT IS 372.9 NG/DL (12.939 NMOL/L) WITH A DATA FLAG. THE RESULT OF A NEW PATIENT SAMPLE IS 38.61 NG/DL (1.34 NMOL/L).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25080657 ↗MalfunctionApr 29, 2026High Test Results; Non Reproducible Results
THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR ONE PATIENT SAMPLE TESTED WITH THE ELECSYS TESTOSTERONE II ASSAY ON A COBAS E 411 ANALYZER. NO QUESTIONABLE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. THE REPEAT VALUE WAS DEEMED CORRECT. THE SAMPLE INITIALLY RESULTED IN A TESTOSTERONE VALUE OF 8.97 NG/ML. THE SAMPLE WAS REPEATED ON (B)(6) 2026, RESULTING IN A VALUE OF 4.29 NG/ML.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25027217 ↗MalfunctionApr 28, 2026Low Test Results; Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE ELECSYS PROBNP II RESULTS FOR 1 PATIENT SAMPLE ON A COBAS E 411 ANALYZER (RACK SYSTEM). THE INITIAL RESULT WAS 10.00 PG/ML WITH A DATA FLAG. THE SAMPLE WAS RERUN TWICE, AND THE RESULTS WERE 84.47 PG/ML AND 84.01 PG/ML.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25019419 ↗MalfunctionApr 27, 2026Low Test Results; Non Reproducible Results
1. THERE WAS AN ALLEGATION OF A QUESTIONABLE TG G2 ELECSYS RESULT FOR ONE PATIENT FROM THE COBAS E 411 ANALYZER (RACK SYSTEM). 2. PATIENT AGE RANGE: ASKU. 3. PATIENT WEIGHT RANGE: ASKU. 4. PATIENT SEX BREAKDOWN: ASKU. 5. PATIENT RACE BREAKDOWN: ASKU. 6. PATIENT ETHNICITY BREAKDOWN: ASKU.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25002395 ↗MalfunctionApr 27, 2026Non Reproducible Results
1. DESCRIBE THE EVENT: THERE WAS AN ALLEGATION OF QUESTIONABLE RESULTS FOR 1 PATIENT SAMPLE TESTED FOR ELECSYS T4 ON A COBAS E 411 ANALYZER (RACK SYSTEM). 2. PATIENT AGE RANGE: 74 YEARS. 3. PATIENT WEIGHT RANGE: 75 KG. 4. PATIENT SEX BREAKDOWN: MALE. 5. PATIENT RACE BREAKDOWN: ASKU. 6. PATIENT ETHNICITY BREAKDOWN: ASKU.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25002177 ↗MalfunctionApr 27, 2026Low Test Results; Non Reproducible Results
WE RECEIVED AN ALLEGATION ABOUT DISCREPANT RESULTS FOR 1 PATIENT SAMPLE TESTED WITH ELECSYS FT4 IV ASSAY ON A COBAS E 411 ANALYZER (RACK SYSTEM). INITIAL RESULT: 0.039 NG/DL (ACCOMPANIED BY A DATA FLAG). REPEAT RESULT: 2.36 NG/DL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25001608 ↗MalfunctionApr 22, 2026Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE HEPATITIS B SURFACE ANTIGEN (HBSAG) RESULTS FOR MULTIPLE PATIENT SAMPLES ON A COBAS E 411 ANALYZER (RACK SYSTEM). MULTIPLE PREVIOUS PATIENT SAMPLES WERE REPEATED AND THE RESULTS WERE POSITIVE. THE EXACT RESULTS AND NUMBER OF PATIENT SAMPLES WAS NOT PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24953465 ↗MalfunctionApr 14, 2026Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE IGE AND INSULIN RESULTS FOR AN UNSPECIFIED AMOUNT OF PATIENT SAMPLES ON A COBAS E 411 ANALYZER (RACK SYSTEM). ONLY ONE EXAMPLE WAS PROVIDED. IGE: THE INITIAL RESULT WAS 180 IU/ML. THE REPEAT RESULT WAS 520 IU/ML. INSULIN: THE INITIAL RESULT WAS 260 UU/ML. THE REPEAT RESULT WAS 130 UU/ML.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24876602 ↗MalfunctionApr 14, 2026Low Test Results; Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE ELECSYS VITAMIN B12 II RESULTS FOR 3 PATIENT SAMPLES ON A COBAS E 411 ANALYZER (RACK SYSTEM). ON (B)(6) 2026, SAMPLE 1 HAD AN INITIAL RESULT OF <36.90 PMOL/L. THE REPEAT RESULT WAS 389.50 PMOL/L. ON (B)(6) 2026, SAMPLE 2 HAD AN INITIAL RESULT OF<36.90 PMOL/L. THE REPEAT RESULT WAS 206.8 PMOL/L. ANOTHER SAMPLE TUBE COLLECTED AT THE SAME TIME WAS TESTED, AND THE RESULT WAS 238 PMOL/L. THE TWO REPEAT RESULTS WERE DEEMED CORRECT. ON (B)(6) 2026, SAMPLE 3 HAD AN INITIAL RESULT OF <36.90 PMOL/L. THE REPEAT RESULT WAS 220.44 PMOL/L. NO QUESTIONABLE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24876379 ↗MalfunctionApr 13, 2026Incorrect, Inadequate or Imprecise Result or Readings; High Test Results; Output Problem
THE INITIAL REPORTER RECEIVED A QUESTIONABLE ELECSYS HCG STAT RESULT FROM ONE PATIENT SAMPLE TESTED ON THE COBAS E 411 ANALYZER (RACK SYSTEM). THE INITIAL RESULT WAS 5.05 MIU/ML (POSITIVE PREGNANCY RESULT). THE FIRST REPEAT RESULT WAS <1.0 MIU/ML WITH A DATA FLAG (NEGATIVE PREGNANCY RESULT). THE SECOND REPEAT RESULT WAS <1.0 MIU/ML WITH A DATA FLAG (NEGATIVE PREGNANCY RESULT). THE QUESTIONABLE POSITIVE PREGNANCY RESULT WAS NOT REPORTED OUTSIDE THE LABORATORY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24862898 ↗MalfunctionApr 7, 2026Non Reproducible Results
THERE WAS AN ALLEGATION OF A DISCREPANT NEGATIVE RESULT FOR 1 PATIENT SAMPLE TESTED FOR ELECSYS RUBELLA IGG IMMUNOASSAY (RUBELLA IGG) ON A COBAS E 411 ANALYZER (RACK SYSTEM). THE INITIAL RESULT WAS 0.212 IU/ML (NEGATIVE). THE DOCTOR QUESTIONED THIS RESULT. THE SAMPLE WAS REPEATED WITH A RESULT OF 119.7 IU/ML (POSITIVE). THIS RESULT WAS BELIEVED TO BE CORRECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24797864 ↗MalfunctionApr 6, 2026High Test Results; Low Test Results; Non Reproducible Results
THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR FOUR PATIENT SAMPLES TESTED WITH ELECSYS HCG+BETA ON A COBAS E411 RACK. THE FIRST SAMPLE INITIALLY RESULTED IN AN HCG+BETA VALUE OF 340.3 MIU/ML WITH A DATA FLAG. THE SAMPLE WAS REPEATED ON (B)(6) 2026, RESULTING IN A VALUE OF < 0.100 MIU/ML WITH A DATA FLAG. THE SECOND SAMPLE INITIALLY RESULTED IN AN HCG+BETA VALUE OF 4.36 MIU/ML ON (B)(6) 2026. THE SAMPLE WAS REPEATED, RESULTING IN A VALUE OF < 0.1 MIU/ML WITH A DATA FLAG. THE THIRD SAMPLE INITIALLY RESULTED IN AN HCG+BETA VALUE OF 5.14 MIU/ML WITH A DATA FLAG ON 18-(B)(6) 2026. THE SAMPLE WAS REPEATED ON (B)(6) 2026, RESULTING IN A VALUE OF < 0.100 MIU/ML WITH A DATA FLAG. THE SAMPLE WAS REPEATED ON (B)(6) 2026, RESULTING IN A VALUE OF < 0.100 MIU/ML WITH A DATA FLAG. THE FOURTH SAMPLE INITIALLY RESULTED IN AN HCG+BETA VALUE OF 9.60 MIU/ML WITH A DATA FLAG ON (B)(6) 2026. THE SAMPLE WAS REPEATED ON (B)(6) 2026, RESULTING IN A VALUE OF < 0.100 MIU/ML WITH A DATA FLAG. THE SAMPLE WAS REPEATED ON (B)(6) 2026, RESULTING IN A VALUE OF < 0.100 MIU/ML WITH A DATA FLAG.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24788858 ↗MalfunctionApr 1, 2026Non Reproducible Results
THE INITIAL REPORTER RECEIVED A QUESTIONABLE VITAMIN B12 ASSAY RESULT FROM ONE PATIENT SAMPLE TESTED ON THE COBAS E 411 ANALYZER (RACK SYSTEM). THE INITIAL RESULT WAS >2000 PG/ML WITH A DATA FLAG. THE REPEAT RESULT WAS 80.5 PG/ML. THE INITIAL RESULT WAS NOT REPORTED OUTSIDE OF THE LABORATORY. THE REPEAT RESULT WAS DEEMED CORRECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24755021 ↗MalfunctionApr 1, 2026Insufficient Device Problem Information
WE RECEIVED AN ALLEGATION ABOUT DISCREPANT RESULTS FOR 1 PATIENT SAMPLE TESTED WITH ELECSYS TSH ASSAY AND ELECSYS FT4 G4 ASSAY ON A COBAS E 411 ANALYZER (RACK SYSTEM). TSH: INITIAL RESULT: "NO VALUE" WAS GENERATED AND (ACCOMPANIED BY A "<SIGNAL" DATA FLAG). 1ST REPEAT RESULT: 0.005 UIU/ML (ACCOMPANIED BY A DATA FLAG). 2ND REPEAT RESULT: "NO VALUE" WAS GENERATED AND (ACCOMPANIED BY A "<SIGNAL" DATA FLAG). 3RD REPEAT RESULT: "NO VALUE" WAS GENERATED AND (ACCOMPANIED BY A "<SIGNAL" DATA FLAG). 4TH REPEAT RESULT: 0.005 UIU/ML (ACCOMPANIED BY A DATA FLAG). 5TH REPEAT RESULT: "NO VALUE" WAS GENERATED AND (ACCOMPANIED BY A "<SIGNAL" DATA FLAG). 6TH REPEAT RESULT: 1.6 UIU/ML (TESTED IN A DIFFERENT LABORATORY). FT4: INITIAL RESULT: "NO VALUE" WAS GENERATED AND (ACCOMPANIED BY A "<SIGNAL" DATA FLAG). 1ST REPEAT RESULT: "NO VALUE" WAS GENERATED AND (ACCOMPANIED BY A "<SIGNAL" DATA FLAG). 2ND REPEAT RESULT: "NO VALUE" WAS GENERATED AND (ACCOMPANIED BY A "<SIGNAL" DATA FLAG). 3RD REPEAT RESULT: "NO VALUE" WAS GENERATED AND (ACCOMPANIED BY A "<SIGNAL" DATA FLAG). 4TH REPEAT RESULT: 16 PMOL/L (TESTED IN A DIFFERENT LABORATORY).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24753443 ↗MalfunctionMar 30, 2026High Test Results; Low Test Results; Non Reproducible Results
THE INITIAL REPORTER COMPLAINED OF DISCREPANT RESULTS FOR 1 PATIENT SAMPLE TESTED FOR ELECSYS HCG+BETA (HCG + B) ON A COBAS E 411 ANALYZER (RACK SYSTEM). THE INITIAL RESULT WAS 0.167 MIU/ML WITH A DATA FLAG. THIS RESULT WAS QUESTIONED BY LABORATORY STAFF. THE SAMPLE WAS RECENTRIFUGED, AND THE REPEAT RESULT WAS 184.2 MIU/ML WITH A DATA FLAG. THE SAMPLE WAS ALIQUOTED INTO A SECONDARY TUBE, AND THE RESULT WAS < 0.100 MIU/ML WITH A DATA FLAG. THE SAMPLE WAS CENTRIFUGED AGAIN AND ALIQUOTED INTO A SECONDARY TUBE, WITH A RESULT OF 190.7 MIU/ML WITH A DATA FLAG. THE SAMPLE WAS REPEATED AGAIN WITH A RESULT OF 46.89 MIU/ML WITH A DATA FLAG.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24717287 ↗MalfunctionMar 25, 2026Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE ELECSYS HCG+BETA RESULTS FROM THE COBAS E411 RACK ANALYZER. SAMPLE 1: THE INITIAL RESULT WAS 7.18 MIU/ML. THIS RESULT DID NOT FIT WITH THE CLINICAL PICTURE OF THE PATIENT, AS THE CUSTOMER IS A FERTILITY CLINIC AND THEY WERE EXPECTING A NEGATIVE RESULT. THE PHYSICIAN REQUESTED REPEAT TESTING, AND THE RESULT WAS <0.100 MIU/ML WITH A DATA FLAG. THIS WAS THE EXPECTED RESULT. SAMPLE 2: ON (B)(6) 2026, THE INITIAL RESULT WAS 3.88 MIU/ML WITH A DATA FLAG. THIS RESULT WAS QUESTIONED BY THE FERTILITY CLINIC PHYSICIAN SINCE THE PATIENT WAS NOT PREGNANT. THE REPEAT RESULT WAS 0.739 MIU/ML, WHICH WAS MORE IN LINE WITH THE CLINICAL PICTURE. THE QUESTIONABLE RESULTS WERE NOT REPORTED OUTSIDE OF THE LABORATORY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24689966 ↗MalfunctionMar 25, 2026Incorrect, Inadequate or Imprecise Result or Readings; Output Problem
THE INITIAL REPORTER RECEIVED QUESTIONABLE ELECSYS HCG+B ASSAY RESULTS FROM SEVERAL PATIENT SAMPLES TESTED ON THE COBAS E 411 ANALYZER (RACK SYSTEM). ONE EXAMPLE OF DISCREPANT RESULTS WAS PROVIDED: THE INITIAL RESULT WAS 0.100 IU/ML. THE REPEAT RESULT WAS 247.7 IU/ML. THE INITIAL RESULT WAS QUESTIONED AS IT DID NOT MATCH THE PATIENT'S HISTORY, PROMPTING THE RERUN. THE REPEAT RESULT WAS DEEMED CORRECT
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24689371 ↗MalfunctionMar 19, 2026Incorrect Measurement
THE INITIAL REPORTER COMPLAINED THAT THE ELECSYS FREE PSA (FREE PSA) WAS GREATER THAN THE ELECSYS TOTAL PSA (TOTAL PSA) FOR 1 PATIENT SAMPLE TESTED ON A COBAS E 411 ANALYZER (RACK SYSTEM). THIS MEDWATCH WILL COVER TOTAL PSA. REFER TO MEDWATCH WITH A1 PATIENT IDENTIFIER (B)(6) FOR INFORMATION ON THE FREE PSA RESULTS. THE FREE PSA RESULT WAS 0.010 NG/ML WITH A DATA FLAG. THE REPEAT RESULT WAS 0.010 NG/ML. THE TOTAL PSA RESULT WAS 0.006 NG/ML WITH A DATA FLAG. THE REPEAT RESULT WAS 0.006 NG/ML.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24637010 ↗MalfunctionMar 19, 2026Incorrect Measurement
THE INITIAL REPORTER COMPLAINED THAT THE ELECSYS FREE PSA (FREE PSA) WAS GREATER THAN THE ELECSYS TOTAL PSA (TOTAL PSA) FOR 1 PATIENT SAMPLE TESTED ON A COBAS E 411 ANALYZER (RACK SYSTEM). THIS MEDWATCH WILL COVER FREE PSA. REFER TO MEDWATCH WITH A1 PATIENT IDENTIFIER (B)(6) FOR INFORMATION ON THE TOTAL PSA RESULTS. THE FREE PSA RESULT WAS 0.010 NG/ML WITH A DATA FLAG. THE REPEAT RESULT WAS 0.010 NG/ML. THE TOTAL PSA RESULT WAS 0.006 NG/ML WITH A DATA FLAG. THE REPEAT RESULT WAS 0.006 NG/ML.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24637009 ↗
FDA clearance and classification
Cleared through 510(k) K961481 on Jun 14, 1996 (decision: Substantially Equivalent). Submitted by Boehringer Mannheim Corp. as “Elecsys 2010 Analyzer”. The FDA review took 57 days. FDA record ↗
The FDA regulates product code JJE (analyzer, chemistry (photometric, discrete), for clinical use) under 21 CFR 862.2160 as a Class I device. 21 CFR 862.2160 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K961481. One submission often covers many models, so check the product and lot details in each record.
- Z-0499-2018 · Jul 5, 2017 · TerminatedRoche / Hitachi MODULAR Analyzer Systems-Immunoassay Analyzer-Model/Catalog/Part Number: 11568248001/11568248692 - Elecsys 2010 analyzer (disk system) and 11804014001/11804014692 -
- Z-0095-2017 · Jul 15, 2016 · Terminatedcobas e411 Immunoassay Analyzer e411: Immunoassay Analyzer Product Usage: The cobas p501/p701 instrument stores primary and secondary sample tubes. The tubes are stored in a contr
- Z-1790-2014 · May 21, 2014 · TerminatedRoche MODULAR ANALYTICS EVO Serum Work Area (SWA), Operator's Manual Software Version 08-02; Filter label: Filter (N) Mat. No 03149773001, Contents 1, 768-320 Filter (N). Produc
- Z-1218-2012 · Jan 31, 2012 · TerminatedElecsys 2010 Analyzer Software version 06-07 The Elecsys 2010 analyzer is intended for the in vitro quantitative and qualitative analysis of analytes in body fluids.
- Z-0165-2009 · Jun 24, 2008 · TerminatedRoche/Hitachi Modular E Module immunoassay analyzer, Roche Diagnostics, Ind., IN; GMMI Nos. 04998642001 and 03617505001.
- Z-0868-2008 · Mar 21, 2008 · TerminatedRoche/Hitachi Modular Analytics System, Modular E Module GMMI immunoassay analyzer; Catalog numbers 04998642001 and 03617505001. Intended for qualitative and quantitative in vitr
Common questions about the cobas e 411 Immunoassay Analyzer 04775201001
Has the cobas e 411 Immunoassay Analyzer 04775201001 been recalled?
No FDA recall record in our data names this exact device identifier (DI 04015630924721), based on the FDA recall file dated Nov 27, 2026. However, 6 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.
What problems have been reported with the cobas e 411 Immunoassay Analyzer 04775201001?
The FDA's MAUDE database lists 680 adverse event reports for this device, received between Jul 12, 2018 and Sep 30, 2026. In those reports, the most common specific device problems are “Non Reproducible Results” (378), “Low Test Results” (331) and “High Test Results” (168); the most common patient problems are “Test Result” (4), “Cardiac Enzyme Elevation” (1) and “Myocardial Infarction” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the cobas e 411 Immunoassay Analyzer 04775201001?
None of the reports are classified as deaths, 0 as injuries and 680 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the cobas e 411 Immunoassay Analyzer 04775201001 FDA approved?
It was cleared through 510(k) K961481 on Jun 14, 1996. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the cobas e 411 Immunoassay Analyzer 04775201001, and is it still sold?
The FDA's device database lists Roche Diagnostics GmbH as the company responsible for it (brand cobas e 411 Immunoassay Analyzer, model 04775201001). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE devices
- Alinity 03R6701Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall1 death1,184 reports
- Alinity 03R6501Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall903 reports
- Atellica CH 930 Analyzer 11067000Siemens Healthcare Diagnostics Inc. · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE5 recalls677 reports
- DxI 9000 Access Immunoassay Analyzer C11137Beckman Coulter, Inc. · Analyzer, Chemistry (Photometric, Discrete), For Clinical Use17 recalls1 death3 reports
- ARCHITECT 3L77-01Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE2 recalls1 death692 reports
- ARCHITECT 3M74-02Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall650 reports
- ARCHITECT 2P24-40Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall1 death645 reports
- VITROS 6802413Ortho-Clinical Diagnostics, Inc. · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE12 recalls68 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page