RAPIDPoint 500 Measurement Clinical Chemistry System recallZ-2079-2014 · Siemens Healthcare Diagnostics Inc · initiated May 23, 2014
Class II Terminated All recalls posted in July 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Healthcare Diagnostics Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
RAPIDPoint 500 Measurement Cartridges may have an error code one or more electrolyte parameters (e.g., Na+, Ca++, K+, Cl-)
FDA root cause category: Process design
What the company is doing
Siemens Healthcare Diagnostics issued an Urgent Field Safety Notice by e-mail to all affected Siemens Healthcare Diagnostics Regional offices on May 27, 2014 both in the United States and elsewhere for communication with affected customers. This notice informs customers of the issue with the RAPIDPoint 500 Measurement cartridges. Questions contact: Siemens Customer Care Center or your local Siemens technical support representative at 877-229-3711. What steps can be taken to mitigate the issue: " Install a new RAPIDPoint 500 Measurement Cartridge. or " Install a 405 Measurement Cartridge - The RAPIDPoint 405 cartridge is available in four sizes: Since the Lactate parameter is not available on the RAPIDPoint 405 cartridge, however, you will need to arrange to get your lactate results from other instruments in your central laboratory. or " Use an alternative system to obtain your electrolyte values.
Recall details
- Recall number
- Z-2079-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- May 23, 2014
- Posted by the FDA
- Jul 18, 2014
- Recalling firm
- Siemens Healthcare Diagnostics Inc
- Quantity
- 20,507 cartridges
- Product code
- KHP
- FDA submissions
- K113216
- Distribution
- Nationwide Foreign: U.A.E. Albania Austria Australia Bosnia Herzeg. Belgium Bulgaria Bahrain Brazil Canada Switzerland P.R. China Colombia Czech Republic Germany Denmark Algeria Estonia Egypt Spain Finland France United Kingdom Georgia Greece Hong Kong Croatia Hungary Indonesia Ireland Israel India Italy Japan Republic Korea Kuwait Lithuania Luxembourg Latvia Malaysia Mexico Netherlands Norway New Zealand Peru Poland Portugal Qatar Romania Serbia Russian Fed. Saudi Arabia Sweden Singapore Slovenia Turkey Taiwan
Product description
RAPIDPoint 500 Measurement Clinical Chemistry System RAPIDPoint 500 Measurement Cartridge Lactate (750 Tests) Part number: 10491449; RAPIDPoint 500 Measurement Cartridge Lactate (400 Tests) Part Number: 10491448 RAPIDPoint 500 Measurement Cartridge Lactate (250 Tests) Part Number: 10491447 RAPIDPoint 500 Measurement Cartridge Lactate (100 Tests) Part Number:10844813
Code information (models, lots, serial numbers)
Serial Numbers: 2334500011 through 2412705915
Other devices under the same FDA submission
1 device cleared under K113216 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- RAPIDPoint® 500 System 10492730Siemens Healthcare Diagnostics Inc. · ACID, LACTIC, ENZYMATIC METHOD1 recall1 death32 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.