AGFA Digital Radiography X-Ray system DX-D 100 recallZ-2081-2014 · AGFA Corp. · initiated Jun 6, 2014
Class II Terminated All recalls posted in July 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from AGFA Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
When liquid comes in contact with the DX-D 100 touch screen, the device may incorrectly recognize this as user input altering device settings.
FDA root cause category: Use error
What the company is doing
Agfa Healthcare sent an "URGENT FIELD SAFETY NOTICE" dated June 6, 2014, to the US and Canadian customers. The letter described the safety alert and mitigation. Acknowledgment, via FAX-Back or email, that the information was received and understood was requested from the customers. Although medical staff are required to disinfect hands on a frequent basis, users must ensure that their hands are dry before using the touch screen of the DX-0100 because liquid residue may activate the action buttons on the touch screen. Customers were instructed: "Do not operate touch screen monitor with wet hands! "Do not let liquids come in contact with the touch screen while the DXD 100 is powered on! "Always double check your parameter settings prior to exposing the patient. We would like to remind you of this specific warning statement in the DX-0100 User Manual. Please complete the feedback form as soon as possible and return it to us by June 16, 2014. Should the above information not apply to your facility or should the device have been transferred to another organization, please be so kind as to indicate this on the attached feedback form and pass this Urgent Field Safety Notice to the organization where the device has been transferred. We thank you for your careful attention to this issue and your continued support. If you have any questions about this matter, please contact your local Agfa Healthcare organization: 1-877-777-2432, prompt 1 and reference PR1406020002
Recall details
- Recall number
- Z-2081-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Jun 6, 2014
- Posted by the FDA
- Jul 21, 2014
- Recalling firm
- AGFA Corp.
- Quantity
- 216 units
- Product code
- IZL
- FDA submissions
- K103597
- Distribution
- USA (nationwide) including the states of AL, CA, FL, GA, IL, IN, IA, KY, LA, MA, MI, MN, MS, MO, NE, NJ, NY, NC, OH, SC, TN, TX, VA, WV, WI, WY and Internationally to Canada.
Product description
AGFA Digital Radiography X-Ray system DX-D 100 Agfa's DX-D 100 is indicated for use in providing diagnostic quality images to aid the physician with diagnosis. Systems can be used with MUSICA2 image processing to create radiographic images of the skeleton including skull, spinal column and extremities: chest, abdomen, and other body parts. Agfa's DX-D 100 is not indicated for use in mammography.
Code information (models, lots, serial numbers)
Show the full code information (3k characters)
Serial # A5410000053, A5410000054, A5410000067, A5410000068, A5410000069, A5410000070, A5410000072, A5410000073, A5411000315, A5411000331, A5411000332, A5411000141, A5411000144, A5411000038, A5411000069, A5411000133, A5411000175, A5411000119, A5411000123, A5411000333, A5411000334, A5410000038, A5411000019, A5411000047, A5411000048, A5411000053, A5410000041, A5410000042, A5410000046, A5410000016, A5410000039, A5410000021, A5410000023, A5411000074, A5411000080, A5411000024, A5411000184, A5411000193, A5411000196, A5410000100, A5410000114, A5410000115, A5410000116, A5410000117, A5411000342, A5411000343, A5411000344, A5411000366, A5411000368, A5411000237, A5411000027, A5411000020, A5411000021, A5411000132, A5411000070, A5411000078, A5411000082, A5411000154, A5411000156, A54110000107, A5410000082, A5410000066, A5411000236, A5411000155, A5411000151, A5411000148, A5411000149, A5411000134, A5411000120, A5411000125, A5411000147, A5410000079, A5411000185, A5411000191, A5411000192, A5411000194, A5411000239, A5411000052, A5411000030, A5411000314, A5411000348, A5411000268, A5411000347, A5411000349, A5411000140, A5411000142, A5411000097, A5410000022, A5411000017, A5410000027, A5411000153, A5411000233, A5411000234, A5411000235, A5411000318, A5410000044, A5411000037, A5410000043, A5411000181, A5411000158, A5411000152, A5411000263, A5411000309, A5411000199, A5411000209, A5411000210, A5411000232, A5411000346, A5411000124, A5411000174, A5411000177, A5411000230, A5411000173, A5411000316, A5411000145, A5411000188, A5411000197, A5411000202, A5411000203, A5411000206, A5411000208, A5411000335, A5411000336, A5411000159, A5411000041, A5411000050, A5411000088, A5411000247, A5411000079, A5411000040, A5411000240, A5411000085, A5411000277, A5411000278, A5411000279, A5411000280, A5411000281, A5411000282, A5411000245, A5411000064, A5411000065, A5411000187, A5411000198, A5411000207, A5411000211, A5411000212, A5411000231, A5411000039, A5411000032, A5411000083, A5411000122, A5411000108, A5411000110, A5411000241, A5411000246, A5411000114, A5411000033, A5411000139, A5411000312, A5411000182, A5411000190, A5411000218, A5411000249, A5411000084, A5411000098, A5411000238, A5411000116, A5410000042, A5411000023, A5411000337, A5411000338, A5411000340, A5411000314, A5411000117, A5411000121, A5411000262, A5411000270, A5411000272, A5411000273, A5411000293, A5411000260, A5411000264, A5411000269, A5411000271, A5411000291, A5411000292, A5411000261, A5411000265, A5411000266, A5411000267, A5411000363, A5411000034, A5411000025, A5411000081, A5411000146, A5411000092, A5411000093, A5411000026, A5411000068, A5411000131, A5411000060, A5411000018, A5411000100, A5411000313, A5411000284, A5411000076, A5411000135, A5411000180, A5411000285, A5411000311, A5411000286, A5411000062, A5411000136, A5411000178, A5411000179, A5411000283
Other devices under the same FDA submission
1 device cleared under K103597 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- DX-D 100 5411/050Agfa · System, x-ray, mobile2 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page