DX-D 100 5411/050Recalls & reported problems
Agfa · DX-D 100 MOB WLS DR 50KW (300KHU)/110V · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05414904094440), received between Jan 3, 2025 and Apr 22, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
6 other recalls cover devices cleared under the same FDA submission (K103597); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component1
- Unintended System Motion1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 22, 2025Unintended System Motion; Unintended System Motion
A CUSTOMER IN THE UNITED KINGDOM REPORTED UNINTENDED MOVEMENT OF A DX-D 100 SYSTEM. THE USER STATED THE SYSTEM WAS BEING MOVED WHEN IT ACCELERATED INTO A WALL AND BEGAN BEEPING. THERE WERE NO PATIENTS INVOLVED. THE LOCAL AGFA FIELD SERVICE ENGINEER RESPONDED TO INVESTIGATE THE ISSUE AND COMPLETED SEVERAL SYSTEM CHECKS. DURING THE INVESTIGATION, AGFA SERVICE REPLACED THE DMC BOARD AND THE SYSTEM WORKED AS INTENDED. THERE HAVE BEEN NO ADDITIONAL REPORTED EVENTS. AS PART OF THE DEVICE INVESTIGATION BY AGFA AND THE SUPPLIER, THE REMOVED BOARD WAS TESTED AND WORKED AS INTENDED. THE ROOT CAUSE FOR THE REPORTED EVENT IS UNKNOWN. THERE HAS BEEN NO REPORTED HARM TO PATIENT OR USERS FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21880089 ↗MalfunctionJan 3, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
A CUSTOMER IN AUSTRALIA REPORTED THE COLLIMATOR FELL OFF A DX-D 100 SYSTEM. THE LOCAL AGFA FIELD SERVICE ENGINEER RESPONDED AND CONFIRMED SCREWS ATTACHED TO THE COLLIMATOR ADJUSTMENT WERE LOOSE. THE LOCAL ENGINEER REATTACHED THE SYSTEM COLLIMATOR AND TIGHTENDED THE SCREWS. THERE WERE NO REPORTED INJURIES TO USER OR PATIENTS. THE SYSTEM IS NOW WORKING AS INTENDED AND THERE ARE NO FUTURE ACTIONS RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21069717 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K103597. One submission often covers many models, so check the product and lot details in each record.
- Z-1462-2024 · Feb 23, 2024 · Open, ClassifiedAgfa DX-D 100 system-Digital Radiography mobile X-ray System. A mobile X-Ray modality, consisting of X-Ray console, generator, tube, digital detector, and workstation. This modalit
- Z-1698-2017 · Jan 20, 2016 · TerminatedAGFA Digital Radiography X-Ray System DX-D100 DX-D100 is indicated for use in providing diagnostic quality images to aid the physician with diagnosis
- Z-1745-2015 · May 29, 2015 · TerminatedAGFA DX-D 100 is a mobile digital radiography x-ray system
- Z-2081-2014 · Jun 6, 2014 · TerminatedAGFA Digital Radiography X-Ray system DX-D 100 Agfa's DX-D 100 is indicated for use in providing diagnostic quality images to aid the physician with diagnosis. Systems can be use
- Z-2176-2014 · May 13, 2014 · TerminatedAgfa Digital Radiography X-Ray System DX-D 100 Product Usage: Is indicated for use in providing diagnostic quality images to aid the physician with diagnoses.
- Z-1487-2013 · May 15, 2013 · TerminatedAGFA Digital Radiography X-Ray System DX-D 100 Agfa's DX-D 100 is indicated for use in providing diagnostic quality images to aid the physician with diagnosis. Systems can be used
Other System, x-ray, mobile devices
- Philips MobileDiagnost wDR 2.2Sedecal · System, X-Ray, Mobile3 recalls7 reports
- DRX-Revolution Mobile X-ray SystemCarestream Health, Inc. · System, x-ray, mobile2 recalls12 reports
- Optima XR200amxGE Medical Systems, L.L.C. · System, x-ray, mobile2 recalls1 report
- Optima XR240amxGE Medical Systems, L.L.C. · System, x-ray, mobile2 recalls0 reports
- Optima XR220amxGE Medical Systems, L.L.C. · System, x-ray, mobile2 recalls0 reports
- DRX-Revolution Mobile X-ray SystemCarestream Health, Inc. · System, x-ray, mobile1 recall16 reports
- SEDECAL Soltus 500Sedecal · System, X-Ray, Mobile1 recall4 reports
- Optima XR240amx HD UpgradeGE Medical Systems, L.L.C. · System, x-ray, mobile1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.