Baxter***Buretrol Solution Set***105" (2.7 m)***150 mL… recallZ-2613-2014 · Baxter Healthcare Corp · initiated Jul 14, 2014
Class II Terminated All recalls posted in September 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Baxter Healthcare Corp or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Baxter Healthcare Corporation is voluntarily issuing a recall for specific lots of Buretrol Solution set product codes due to complaints for separation between the burette chamber and the drip chamber. Affected product codes are: 2C7564, 2C8864 and 2H8864.
FDA root cause category: Nonconforming Material/Component
What the company is doing
Action to be taken if product was purchased directly from Baxter: Baxter is requesting that you take the following actions: 1. Locate and remove all affected product from your facility (the product code and lot number can be found on the individual product package or shipping carton). 2. Contact Baxter Healthcare Center for Service to arrange for return and credit. The Center for Service can be reached at 888-229-0001 between the hours of 7:00 am and 6:00 pm Central Time. Please have your Baxter eight-digit ship-to account number ready when calling. 3. Complete the enclosed customer reply form, and return it to Baxter by either fax Or scanned e-mail. 4. If you are a dealer, wholesaler, or distributor/reseller that distributed any product to other facilities, please conduct a recall with your end-user customers in accordance with your customary procedures. Baxter distributed this product to customers between September 27, 2013, and April 25, 2014. Action to be taken if the products were purshased from a distributor or re-seller. 1. Locate and remove all affected product from your facility (the product code and lot number can be found on the individual product package or shipping carton). 2. Contact Baxter Healthcare Center for Service to arrange for return and credit. The Center for Service can be reached at 888-229-0001 between the hours of 7:00 am and 6:00 pm Central Time. Please have your ship-to account number ready when calling. 3. Follow your supplier's reply and recall process. Please do not return the customer reply form to Baxter.
Recall details
- Recall number
- Z-2613-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Jul 14, 2014
- Posted by the FDA
- Sep 9, 2014
- Recalling firm
- Baxter Healthcare Corp
- Quantity
- 234196
- Product code
- FPA
- FDA submissions
- K123874
- Distribution
- Nationwide Distribution.
Product description
***Baxter***Buretrol Solution Set***105" (2.7 m)***150 mL Burette***Drip Chamber Filter Valve***3 Injection Sites Male Luer Lock Adapter with***Retractable Collar***Fluid path is Sterile, Nonpyrogenic***Indications for Use: For the administration of fluids from a container into the patient's vascular system through a vascular access device.***
Code information (models, lots, serial numbers)
2C7564- Buretrol Solution Set with 150mL INTERLINK Burette, lots: DR13E06019 DR13E07017 DR13E07025 DR13I14023 DR13I16010 DR13I16028 DR13I23016 DR13I23024 DR13J03033 DR13J04015 DR13J04023 DR13J08024 DR13J09014 DR13J09022 DR13J24021 DR13J24039 DR13K07024 DR13K08022 DR13K12024 DR13K13022 DR13K18021 DR13K19029 DR13K25026 DR13K26024 DR13L02023 DR13L03039 DR13L10018 DR13L11016 DR13L18011 DR13L18029 DR13L19035 DR14A09052 DR14A10035 DR14A23012 DR14A23020 DR14A29027 DR14A30017 DR14B06049 DR14B07013 DR14B07047 DR14B11023 DR14B13037 DR14B13045 DR14B14019 DR14B20016 DR14B20032 DR14B21014 DR14B26013 DR14B26039 DR14C05015 DR14C05031
Other devices under the same FDA submission
9 devices cleared under K123874 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- INTERLINK/BURETROL 2C7564Baxter Healthcare Corporation · Set, administration, intravascular19 reports
- CLEARLINK/BURETROL 2C8864Baxter Healthcare Corporation · Set, administration, intravascular18 reports
- CLEARLINK/BURETROL 2H8864Baxter Healthcare Corporation · Set, administration, intravascular10 reports
- CLEARLINK/BURETROL 2C8862Baxter International Inc. · Set, administration, intravascular8 reports
- CLEARLINK/BURETROL 2H8865Baxter Healthcare Corporation · Set, administration, intravascular4 reports
- CLEARLINK/BURETROL 1C8712Baxter International Inc. · Set, administration, intravascular3 reports
- CLEARLINK/BURETROL 3C0157Baxter International Inc. · Set, administration, intravascular3 reports
- CLEARLINK/BURETROL 2R8864Baxter Healthcare Corporation · Set, administration, intravascular2 reports
- INTERLINK/BURETROL 2H7565Baxter Healthcare Corporation · Set, administration, intravascular1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 68843).
- Z-2614-2014 · Terminated***Baxter***BURETROL Solution Set***105" (2.7 m) 150 mL Burette***Drip Chamber Filter Valve***3 Luer Activated Valves***Male Luer Lock Adapter***with Retractable Collar***Fluid pat…
- Z-2615-2014 · Terminated***Baxter***Non-DEHP***BURETROL Solution Set 105" (2.7 m), 150 mL Burette***Drip Chamber Filter Valve 3 Luer Activated Valves***Male Luer Lock Adapter***with Retractable Collar***F…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page