Medtronic MiniMed Guardian Monitor, Model No. CSS7100… recallZ-2179-2014 · Medtronic MiniMed Inc. · initiated Jul 21, 2014
Class II Terminated All recalls posted in August 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic MiniMed Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Medtronic MiniMed is recalling the Guardian REAL-Time monitors because of a manufacturing issue that has the potential to compromise the water resistance of the affected Guardian monitors.
FDA root cause category: Device Design
What the company is doing
Medical Device Recall letters dated July 21, 2014 andJuly 22, 2014 were sent to distributors, patients, and healthcare providers who purchased the Guardian REAL time monitors. The letters inform the customers of the problems identified and the actions to be taken. Distributors with questions are instructed to email [email protected]. If healthcare professionals want to speak to Medtronic's medical officers then they are instructed to call (818) 576-4211, ext. 64511. Customers and healthcare professionals' patients are instructed to call Guardian Recall Hotline at (800) 646-4633, ext. 21090, 8am-6pm, CST.
Recall details
- Recall number
- Z-2179-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Jul 21, 2014
- Posted by the FDA
- Aug 10, 2014
- Recalling firm
- Medtronic MiniMed Inc.
- Quantity
- 86 units
- Product code
- MDS
- FDA submissions
- P980022
- Distribution
- Worldwide:Distribution - US Nationwide and the countries of Croatia, Denmark, Finland, Germany, Netherlands, Norway, Sweden, Switzerland.
Product description
Medtronic MiniMed Guardian Monitor, Model No. CSS7100, CSS7100K Product Usage: The Medtronic MiniMed model CSS7100 and CSS7100K Guardian REAL-Time monitors are used in combination with the Minilink transmitter and Sof-sensor glucose sensor for continuous monitoring of subcutaneous glucose concentration. The information provided by these devices is intended for use in identifying potential episodes of high or low blood glucose. The information provided is not intended to be used directly for making therapy adjustment but rather to assist the user in determining that an additional fingerstick glucose measurement may be appropriate. Per the product labeling, all therapy adjustments should be based on fingerstick glucose measurements obtained using standard home glucose meters and not on the sensor glucose values provided by the Guardian REAL-Time monitor. The CSS7100 and CSS7100K monitors are identical expect for the fact that the CSS7100K monitor is a pediatric version that has a minimum programmable low alert setting of 90 mg/dL. The minimum programmable low glucose alert setting for the adult model (CSS7100) is 40 mg/dL.
Code information (models, lots, serial numbers)
CSS7100 and CSS7100K
Other devices under the same FDA submission
277 devices cleared under P980022 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Paradigm® REAL-Time Revel™ MMT-723NASMedtronic Minimed, Inc. · Sensor, glucose, invasive15 deaths3,413 reports
- Paradigm® REAL-Time Revel™ MMT-723LNASMedtronic Minimed, Inc. · Sensor, glucose, invasive1 recall12 deaths2,995 reports
- Paradigm® REAL-Time Revel™ MMT-723LNASMedtronic Minimed, Inc. · Sensor, glucose, invasive1 recall18 deaths2,909 reports
- Paradigm® REAL-Time Revel™ MMT-723LNASMedtronic Minimed, Inc. · Sensor, glucose, invasive1 recall23 deaths2,570 reports
- Paradigm® REAL-Time Revel™ MMT-723RNASMedtronic Minimed, Inc. · Sensor, glucose, invasive18 deaths1,970 reports
- Paradigm® REAL-Time Revel™ MMT-723NAPMedtronic Minimed, Inc. · Sensor, glucose, invasive7 deaths1,564 reports
- Paradigm® REAL-Time Revel™ MMT-723NALMedtronic Minimed, Inc. · Sensor, glucose, invasive9 deaths1,504 reports
- Paradigm® REAL-Time Revel™ MMT-723NABMedtronic Minimed, Inc. · Sensor, glucose, invasive8 deaths1,492 reports
- Paradigm® REAL-Time Revel™ MMT-723LNALMedtronic Minimed, Inc. · Sensor, glucose, invasive9 deaths1,271 reports
- Paradigm® REAL-Time Revel™ MMT-523NASMedtronic Minimed, Inc. · Sensor, glucose, invasive9 deaths1,259 reports
- Paradigm® REAL-Time Revel™ MMT-723LNAPMedtronic Minimed, Inc. · Sensor, glucose, invasive7 deaths1,256 reports
- Paradigm® REAL-Time Revel™ MMT-723LNABMedtronic Minimed, Inc. · Sensor, glucose, invasive10 deaths1,243 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.