Paradigm® REAL-Time Revel™ MMT-723RNAS (UDI 00643169513143)Recalls & reported problems
Medtronic Minimed, Inc. · PUMP MMT-723RNAS PRDGM 3.0 SK EN RC · Class 3
All models of Paradigm REAL-Time Revel (180)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This MiniMed Paradigm REAL-Time Revel insulin pump works with a glucose sensor. Many reports describe the pump showing incorrect messages or a blank display, difficulty programming, and mechanical or motor alarms, and some mention patients with high or low blood sugar. The records include 1,970 reports, 865 injuries and 18 deaths, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 1,970
- Deaths reported
- 18
- Injuries reported
- 865
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1,970 adverse event reports that list this exact device (DI 00643169513143), received between May 5, 2017 and Nov 18, 2025 — classified as 18 deaths, 865 injuries, 1,087 malfunctions.
The most frequently reported specific device problem is “Device Difficult to Program or Calibrate” (267 reports).
14 other recalls cover devices cleared under the same FDA submission (P980022); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem352
- Device Difficult to Program or Calibrate267
- Mechanical Problem236
- No Apparent Adverse Event223
- No Display/Image211
- Device Displays Incorrect Message208
- Patient Device Interaction Problem149
- Obstruction of Flow126
Patient problems
- No Consequences Or Impact To Patient946
- Hyperglycemia822
- Hypoglycemia229
- Elevated Ketones/ Diabetic Ketoacidosis35
- Death19
- No Known Impact Or Consequence To Patient9
- Loss of consciousness8
- Nausea8
Recent adverse event reports
The 25 most recent of 1,970 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 18, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER PASSED AWAY. THE EVENT INVOLVED PRODUCT(S) MMT-723RNAS. TROUBLESHOOTING WAS NOT PERFORMED. THE REASON FOR CUSTOMER PASSING WAS CANCER. IT WAS UNKNOWN IF THE CUSTOMER WAS USING THE INSULIN PUMP WITHIN 48 HOURS OF THE REPORTED EVENT. NO FURTHER COMPLICATIONS WERE REPORTED. MMT-723RNAS WAS RETURNED FOR ANALYSIS.
Patient problems: Cancer
MAUDE report 23578470 ↗DeathAug 14, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER PASSED AWAY ON (B)(6) 2020. CAUSE OF DEATH WAS HEART STROKE. OTHER RELEVANT HISTORY, INCLUDING PREEXISTING MEDICAL CONDITIONS: NO CONTRIBUTING FACTORS WERE IDENTIFIED AND EXPERIENCED HYPERGLYCEMIA WAS TREATED WITH HOSPITALIZATION. THE PUMP WAS NOT WORN AT THE TIME OF PASSING AND THE PUMP WAS NOT WORKING . IT WAS ALSO REPORTED CUSTOMER PASSED AWAY AT HOME. THE LAST KNOWN PRODUCT(S) FOR THIS CUSTOMER ARE AS FOLLOWS: MMT-332A, UNOMEDICAL AND MMT-723RNAS. NO PRODUCT RETURN IS REQUIRED FOR MMT-332A. NO PRODUCT RETURN IS REQUIRED FOR UNOMEDICAL AND NO PRODUCT RETURN IS REQUIRED FOR MMT-723RNAS.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Hyperglycemia
MAUDE report 19979782 ↗MalfunctionJul 19, 2021Mechanical Problem; Obstruction of Flow; Power Problem; Mechanical Problem; Obstruction of Flow; Power Problem
INFORMATION INDICATED BY MEDTRONIC THAT THE INSULIN PUMP HAD A MOTOR ERROR ALARM AND NO DELIVERY ALARM. CUSTOMER STATED THAT THE NEW BATTERY OF THE INSULIN PUMP WAS UNRESPONSIVE. CUSTOMER STATED THAT THE INSULIN PUMP TOOK LONGER TO REWIND. CUSTOMER DECLINED TROUBLESHOOTING FOR REWIND ANOMALY AS THE DEVICE WAS WORKING FINE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12193651 ↗MalfunctionNov 6, 2020Mechanical Problem; Mechanical Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT INSULIN PUMP HAD MOTOR ERROR ALARM. CUSTOMER REPORTED THAT THE DRIVE SUPPORT CAP WAS CORRECT. CUSTOMER REPORTED THAT THE INSULIN PUMP WAS NO DROPPED OR BUMPED. CUSTOMER WAS ABLE TO SUCCESSFULLY CLEAR THE ALARM. CUSTOMER WAS ABLE TO COMPLETE PUMP REWIND. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10797569 ↗MalfunctionSep 10, 2020Mechanical Problem; Mechanical Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD MOTOR ERROR. CUSTOMER STATED DRIVE SUPPORT CAP APPEARS NORMAL. CUSTOMER STATED THEY WERE ABLE TO CLEAR THE ALARM AND THEY WERE ABLE TO REWIND INSULIN PUMP. TROUBLESHOOTING WAS PERFORMED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10513378 ↗MalfunctionSep 2, 2020Loss of Data; Loss of Data
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD PUMP ERROR ALARM MULTIPLE TIMES. CUSTOMER WAS UNABLE TO CLEAR THE ALARM. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10481208 ↗DeathAug 27, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED VIA PHONE CALL THAT THE CUSTOMER PASSED AWAY IN HOSPITAL. THE CUSTOMER WAS HOSPITALIZED ON (B)(6) 2020 FOR HIGH BLOOD GLUCOSE. THE CAUSE OF DEATH WAS SEPSIS. THE CALLER STATED THAT THE CUSTOMER HAD MANY HIGH BLOOD GLUCOSE INCIDENTS THAT MAY HAVE LED TO THE CUSTOMER'S PASSING. THE CUSTOMER¿S BLOOD GLUCOSE WAS UNKNOWN AT THE TIME OF HOSPITALIZATION AND PASSING. THE CUSTOMER WAS WEARING THE INSULIN PUMP AT THE TIME OF HOSPITALIZATION. THE INSULIN PUMP HAD BEEN DISCONNECTED MORE THAN 48 HOURS PRIOR TO PASSING, DUE TO OTHER MEDICAL COMPLICATIONS. THE CUSTOMER WAS NOT USING SENSORS. THE CALLER AGREED TO RETURN THE INSULIN PUMP FOR ANALYSIS.
Patient problems: Death; Hyperglycemia
MAUDE report 10460769 ↗MalfunctionAug 24, 2020Battery Problem; Battery Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD LOW BATTERY ALERT AND BATTERY KEPT DRAINING. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WAS RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10441872 ↗MalfunctionAug 13, 2020No Display/Image; No Display/Image
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP SCREEN HARD TIME SEEING THE DISPLAY ON THE IT, AS PER CUSTOMER INSULIN PUMP LACKING IT'S BRIGHTNESS. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10400406 ↗MalfunctionAug 12, 2020Device Alarm System; Moisture or Humidity Problem; Device Alarm System; Moisture or Humidity Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP WAS VIBRATING WITHOUT AN ALARM AND ALSO HAD WATER IN IT. TROUBLESHOOTING WAS PERFORMED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10395381 ↗MalfunctionAug 12, 2020Mechanical Problem; Mechanical Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD OVER REPEATED MOTOR ERRORS. THE CUSTOMER STATED THEY WERE NOT ABLE TO CLEAR THE ALARM AND NOT ABLE TO COMPLETE REWIND PROCESS. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10394888 ↗MalfunctionAug 12, 2020Obstruction of Flow; Power Problem; Obstruction of Flow; Power Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP QUIT WORKING. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10394855 ↗MalfunctionAug 12, 2020Mechanical Problem; Mechanical Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD MOTOR ERROR. CUSTOMER REPORTED THE DRIVE SUPPORT CAP APPEARS NORMAL. CUSTOMER WAS ABLE TO SUCCESSFULLY CLEAR THE ALARM. CUSTOMER WAS ABLE TO COMPLETE PUMP REWIND. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: Hyperglycemia
MAUDE report 10394826 ↗MalfunctionAug 11, 2020Device Difficult to Program or Calibrate; Obstruction of Flow; Device Difficult to Program or Calibrate; Obstruction of Flow
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD BUTTON ERROR ALARM. THE CUSTOMER ALSO STATED THAT, THE INSULIN PUMP HAD RANDOM OR UNEXPLAINED NO DELIVERY ALARMS. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10389827 ↗MalfunctionAug 7, 2020Device Difficult to Program or Calibrate; Device Difficult to Program or Calibrate
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE UP ARROW ON THE INSULIN PUMP WAS NOT RESPONDING AND HE USED FINGERNAIL TO PRESS IT. THE TIME ON THE INSULIN PUMP ADVANCED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10377690 ↗InjuryAug 6, 2020No Display/Image; No Display/Image
CUSTOMER'S DAUGHTER REPORTED VIA PHONE CALL THAT CUSTOMER WAS HOSPITALIZED DUE TO UNKNOWN REASON FOR UNKNOWN DATE. CUSTOMER STATED THAT THE INSULIN PUMP WAS DIED AS A RESULT OF WHICH THEY WERE HOSPITALIZED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: Hyperglycemia
MAUDE report 10371734 ↗InjuryAug 6, 2020Electrical /Electronic Property Problem; Power Problem; Electrical /Electronic Property Problem; Power Problem
THE CUSTOMER REPORTED VIA PHONE CALL THAT THEY WERE EXPERIENCING HIGH BLOOD GLUCOSE LEVEL. BLOOD GLUCOSE LEVEL AT THE TIME OF THE INCIDENT WAS 450 MG/DL. CUSTOMER STATED INSULIN PUMP WAS NOT WORKING AND ALSO HAD PUMP ERROR ALARM. CUSTOMER STATED THEY WERE UNABLE TO CLEAR THE ALARM. CUSTOMER DECLINED TROUBLESHOOTING FOR HIGH BLOOD GLUCOSE LEVEL. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: Hyperglycemia
MAUDE report 10371704 ↗MalfunctionAug 5, 2020Device Alarm System; Display or Visual Feedback Problem; Mechanical Problem; Device Difficult to Program or Calibrate; Moisture or Humidity Problem; Device Alarm System; Display or Visual Feedback Problem; Mechanical Problem; Device Difficult to Program or Calibrate; Moisture or Humidity Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP'S ACT BUTTON WAS STICKY. CUSTOMER REPORTED THAT THE INSULIN PUMP HAD CONDENSATION INSIDE THE SCREEN AND ALARMS WERE FAINT. CUSTOMER HAD RESTART THE REWIND TO COMPLETE AT TIMES AND BACKLIGHT WAS DIM. CUSTOMER DECLINED TO TROUBLESHOOTING WITH 24 HOURS HELPLINE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10366805 ↗MalfunctionAug 5, 2020Mechanical Problem; Mechanical Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD MOTOR ERROR ALARM. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10366244 ↗MalfunctionAug 5, 2020Device Alarm System; Device Alarm System
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP WAS NOT GIVING ANY WARNING AND NOT BEEPING. CUSTOMER STATED THAT THE INSULIN PUMP DID NOT BEEP DURING TONE TEST.NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10366025 ↗MalfunctionAug 3, 2020Mechanical Problem; Mechanical Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THEIR INSULIN PUMP HAD MOTOR POSITION ENCODER ALARM. CUSTOMER REPORTED THAT THEIR INSULIN PUMP HAD MOTION SENSOR TEST FAILURE ALARM AND MOTOR ERROR. CUSTOMER REPORTED THE DRIVE SUPPORT CAP WAS NORMAL. CUSTOMER STATED THE INSULIN PUMP WAS NOT EXPOSED TO A HIGH MAGNETIC FIELD. CUSTOMER WERE TO CLEAR THE ALARMS AND ABLE TO COMPLETE REWIND THE INSULIN PUMP. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10360353 ↗InjuryAug 3, 2020Insufficient Flow or Under Infusion; Power Problem; Insufficient Flow or Under Infusion; Power Problem
THE CUSTOMER REPORTED VIA PHONE CALL THAT THEY WERE HOSPITALIZED FOR DIABETIC KETOACIDOSIS ON (B)(6) 2019. BLOOD GLUCOSE READING WAS 500 MG/DL. THE CUSTOMER STATED THAT DOCTOR TOOK OFF INSULIN PUMP BECAUSE IT WAS NOT WORKING. THE CUSTOMER WAS TREATED WITH INTRAVENOUS DRIP AT THE HOSPITAL. CUSTOMER HAD SYMPTOMS SUCH AS WEAKNESS AND NO BALANCE. TROUBLESHOOTING FOR THE CUSTOMER¿S HIGH BLOOD GLUCOSE WAS PERFORMED. THE CUSTOMER ALLEGED THAT INSULIN PUMP WAS UNDER DELIVERING INSULIN. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: Elevated Ketones/ Diabetic Ketoacidosis; Hyperglycemia; Weakness
MAUDE report 10355716 ↗MalfunctionJul 31, 2020Patient Device Interaction Problem; Patient Device Interaction Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP DID NOT WORKING DUE TO HIGH BLOOD GLUCOSE. CUSTOMER HAD USING INSULIN PUMP SYSTEM WITHIN 48 HOURS OF REPORTED HIGH BLOOD GLUCOSE EVENT. THE CUSTOMER TREATED HIGH BLOOD GLUCOSE WITH SHOT. CUSTOMER WAS NEITHER IN EMERGENCY ROOM, NOR ADMITTED INTO HOSPITAL. DRIVE SUPPORT CAP WAS NORMAL. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: Hyperglycemia
MAUDE report 10348737 ↗InjuryJul 30, 2020Obstruction of Flow; Obstruction of Flow
THE CUSTOMER'S SON REPORTED VIA PHONE CALL THAT THE CUSTOMER EXPERIENCED HIGH BLOOD GLUCOSE LEVELS. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 561 MG/DL. THEY ALSO REPORTED THAT THE INSULIN PUMP RECEIVED INSULIN FLOW BLOCK ALARM DURING BOLUS. TROUBLESHOOTING WAS PERFORMED AND THE INSULIN EXITED DURING FILL TUBING. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: Hyperglycemia
MAUDE report 10343091 ↗MalfunctionJul 30, 2020Device Difficult to Program or Calibrate; Device Difficult to Program or Calibrate
THE CUSTOMER CALLED AND REPORTED THAT THEIR INSULIN PUMP HAD AN UNRESPONSIVE KEYBOARD. THE CUSTOMER'S BLOOD GLUCOSE LEVEL WAS UNKNOWN AT THE TIME OF THE INCIDENT. THE CUSTOMER STATED THE ACT BUTTON AND THE UP AND DOWN ARROW BUTTONS ARE STICKY AND HARD TO GET TO WORK. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND REVERT TO THE BACKUP PLAN. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10342995 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P980022. One submission often covers many models, so check the product and lot details in each record.
- Z-1749-2026 · Feb 13, 2026 · Open, ClassifiedParadigm REAL-Time Insulin Pump (MMT-522, MMT-722)
- Z-1750-2026 · Feb 13, 2026 · Open, ClassifiedParadigm REAL-Time Revel Insulin Pump (MMT-523, MMT-723)
- Z-1256-2025 · Jan 31, 2025 · Open, ClassifiedParadigm insulin pump, REF: MMT-522, MMT-523, MMT-551, MMT-722, MMT-723, MMT-751
- Z-1589-2020 · Jun 27, 2019 · Open, ClassifiedMedtronic MiniMed Paradigm Insulin pump, model number MMT-522K
- Z-1593-2020 · Jun 27, 2019 · Open, ClassifiedMedtronic MiniMed Paradigm Insulin pump, model number MMT-723K
- Z-1591-2020 · Jun 27, 2019 · Open, ClassifiedMedtronic MiniMed Paradigm Insulin pump, model number MMT-523
- Z-1588-2020 · Jun 27, 2019 · Open, ClassifiedMedtronic MiniMed Paradigm Insulin pump, model number MMT-722
- Z-1594-2020 · Jun 27, 2019 · Open, ClassifiedMedtronic MiniMed Paradigm Insulin pump, model number MMT-723
- Z-1590-2020 · Jun 27, 2019 · Open, ClassifiedMedtronic MiniMed Paradigm Insulin pump, model number MMT-722K
- Z-1592-2020 · Jun 27, 2019 · Open, ClassifiedMedtronic MiniMed Paradigm Insulin pump, model number MMT-523K
- Z-1587-2020 · Jun 27, 2019 · Open, ClassifiedMedtronic MiniMed Paradigm Insulin pump, model number MMT-522
- Z-2442-2015 · Jul 22, 2015 · TerminatedGuardian Real-Time Monitor, Model No. CSS7100. Indicated for continuous or periodic monitoring of glucose levels in the fluid under the skin, in adults (18 years and older) with d
- Z-2179-2014 · Jul 21, 2014 · TerminatedMedtronic MiniMed Guardian Monitor, Model No. CSS7100, CSS7100K Product Usage: The Medtronic MiniMed model CSS7100 and CSS7100K Guardian REAL-Time monitors are used in combinatio
- Z-2544-2014 · Mar 13, 2014 · TerminatedMedtronic MiniMed Paradigm Insulin Infusion Pumps, Paradigm models: MMT-511, MMT-512, MMT-712, MMT-712E, MMT-515, MMT-715, MMT-522, MMT-522K, MMT-722, MMT-722K, MMT-523, MMT-523K
Other Sensor, glucose, invasive devices
- Dexcom G4 Platinum Sensor 9500-27Dexcom, Inc. · Sensor, Glucose, Invasive2 recalls2 deaths99,426 reports
- Dexcom G5 Transmitter Kit STT-GF-001Dexcom, Inc. · Sensor, Glucose, Invasive1 death58,932 reports
- Dexcom G5 Mobile IOS App SW10611Dexcom, Inc. · Sensor, Glucose, Invasive54,544 reports
- Dexcom G5 Transmitter 9438-06Dexcom, Inc. · Sensor, Glucose, Invasive15,360 reports
- Dexcom G5 Mobile Receiver MT22719Dexcom, Inc. · Sensor, Glucose, Invasive1 death12,594 reports
- Dexcom G4 SBT Transmitter 9438-05Dexcom, Inc. · Sensor, Glucose, Invasive1 death9,041 reports
- Dexcom G5 Mobile Receiver MT22719-BLUDexcom, Inc. · Sensor, Glucose, Invasive7,685 reports
- Dexcom G5 Mobile Receiver Kit STK-GF-001Dexcom, Inc. · Sensor, Glucose, Invasive1 recall6,345 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page