Vanguard 360 Femoral Boss Reamer recallZ-2242-2014 · Biomet, Inc. · initiated Jul 1, 2014

Class II Terminated All recalls posted in August 2014

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The Vanguard 360 Femoral Boss Reamers were manufactured oversized to the print. The instrument may not fit into the mating component.

FDA root cause category: Process change control

What the company is doing

Urgent Medical Device Recall Notices dated July 15, 2014 were mailed with verification of delivery via FedEx. Response forms were provided to all consignees in order for Biomet to track returns. The letters informed the consignees to return the devices to Biomet and conduct a sub-recall sending the recall letter to all customers.

Recall details

Recall number
Z-2242-2014
Recall class
Class II
Status
Terminated
Initiated by the company
Jul 1, 2014
Posted by the FDA
Aug 15, 2014
Recalling firm
Biomet, Inc.
Quantity
461
Product code
JWH
FDA submissions
K093293
Distribution
California, IA, FL, MD, UT, SC, AL, IL, CO, MA, KS, GA, WA, NV, AR, NH, OK,Pennsylvania, New York, Texas, Virginia, New Jersey, North Carolina, Kentucky, Michigan, Ohio, Wisconsin, Louisiana, Arizona, Indiana, South Dakota, Missouri & Montana. In addition to Korea, B.V., Japan, Puerto Rico, UK, Chile, New Zealand, Australia, & Canada

Product description

Vanguard 360 Femoral Boss Reamer, Label: REF 32-488474 VANGUARD 360 1NSTRUMENT FEMORAL BOSS REAMER NON-STERILE, BIOMET ORTHOPEDICS 56 EAST BELL DRIVE P.O. BOX 587 WARSAW, IN 46581 USA

Code information (models, lots, serial numbers)

Catalog Number: 32-48847 4 Lot Numbers: 022120, 022180, 162530, 232460, 475500 and 475510. 699240, 793350, 793570 Additional Scope: 065030, 095780, 164290, 235870, 253670, 273630, 308620, 418970, 420030, 421210, 570460, 585260, 619690, 708240, 711430, 726490, 816160, 896810, 929690, 953540, 968950

Other devices under the same FDA submission

32 devices cleared under K093293 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.