Skip to content

Arrow Select Kits (ASK) and Maximal Barrier Precautions… recallZ-0066-2015 · Arrow International Inc · initiated Aug 6, 2014

Class II Terminated All recalls posted in October 2014

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Arrow International Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The component supplier recalled their 0.9% Sodium Chloride Injection USP BD Posiflush SF Saline Flush Syringe 10mL because some unit packages may exhibit open seals which impact package integrity and potentially product sterility.

FDA root cause category: Packaging

What the company is doing

Arrow International sent an Urgent Medical Device Recall Notification letter, dated August 6, 2014, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately discontinue use and quarantine any affected product. Customers were instructed to remove the BD pre-filled saline syringe from the exterior of the packaging. To return the BD syringes, customers were instructed to complete the enclosed Recall Acknowledgement Form and fax it to 1-855-419-8507, Attn: Customer Service. A customer service representative will contact the customer with a Return Good Authorization (RGA) Number and will provide instructions for the return of producet to Arrow International. All customers were asked to complete the enclosed Recall Acknowledgement Form and fax it to 1-855-419-8507. Attn: Customer Service. Customers with questions were instructed to contact their local sales representative or Customer Service at 1-866-246-6990.

Recall details

Recall number
Z-0066-2015
Recall class
Class II
Status
Terminated
Initiated by the company
Aug 6, 2014
Posted by the FDA
Oct 14, 2014
Recalling firm
Arrow International Inc
Quantity
71255
Product code
DQY
FDA submissions
K011761, K073451, K780532, K810675, K862056
Distribution
Nationwide Distribution

Product description

Arrow Select Kits (ASK) and Maximal Barrier Precautions Kits (CDC) with BD Medical Posiflush 10mL Saline Flush Syringe.

Code information (models, lots, serial numbers)

Show the full code information (3k characters)

Product Numbers: ASK-04001-VCUH, ASK-04510-MM3, ASK-05001-TSC, ASK-05052-QV, ASK-05052-SFL, ASK-05502-UWMC, ASK-05541-MS2, ASK-09804-SPH, ASK-09810-PCMH1, ASK-09903-ECL, ASK-09903-PCMH1, ASK-11142-MGH3, ASK-12703-MGH2, ASK-12703-SPH, ASK-15703-MH, ASK-15854-FAM ASK-17702-PJHH1, ASK-21142-HF2, ASK-21242-PCMH1, ASK-21242-SFM1,  ASK-21242-UVA3, ASK-29803-HF3, ASK-29803-MM, ASK-29803-MS, ASK-29903-UK, ASK-29903-UK, ASK-34041-PTG, ASK-34052-PTG, ASK-35052-PTG, ASK-42703-AMC, ASK-42703-OHU1, ASK-42703-PCCH1, ASK-42703-PCSJ, ASK-42703-PPMC, ASK-42703-PSBM, ASK-42703-PSCH, ASK-42703-PSFM, ASK-42703-PTRH, ASK-42703-SJMT, ASK-42703-UHC, ASK-42802-ECL1, ASK-42802-PMCG, ASK-42802-PUCS, ASK-42854-CMC2, ASK-45703-MM2, ASK-45703-NYU1, ASK-45703-PCCH1, ASK-45703-PCHS1, ASK-45703-PCSJ, ASK-45703-PDCH, ASK-45703-PGBM1, ASK-45703-PHCH, ASK-45703-PHF, ASK-45703-PHF1, ASK-45703-PHPA2, ASK-45703-PIA, ASK-45703-PLGH1, ASK-45703-PMHG, ASK-45703-PSA, ASK-45703-PTMC, ASK-45703-PUAB, ASK-45703-PUC ASK-45703-PUHC, ASK-45703-PUR, ASK-45703-PVCU, ASK-45703-UHC, ASK-45703-UKM, ASK-45802-PCMH1, ASK-45854-NG, ASK-45854-PCHS1, ASK-45854-PECL, CDC-02041-MK1A, CDC-05041-HPK1A, CDC-05052-HPK1A,  CDC-21242-X1A, CDC-24703-X1A, CDC-25123-X1A, CDC-25402-X1A, CDC-25502-X1A, CDC-25553-X1A, CDC-26553-X1A, CDC-29803-X1A, CDC-29903-X1A CDC-34052-HPK1A, CDC-35041-VPS, CDC-35052-HPK1A, CDC-35052-VPS,  CDC-35063-VPS, CDC-35563-VPS, CDC-42703-XP1A, CDC-42802-XP1A, CDC-42802-XP1A, CDC-44041-HPK1A, CDC-44052-HPK1A, CDC-45041-HPK1A, CDC-45041-VPS, CDC-45541-HPK1A, CDC-45541-VPS, CDC-45552-HPK1A, CDC-45703-XP1A, CDC-45703-XPB1A, CDC-45802-XP1A, CDC-45854-XP1A,  CDC-46702-XP1A, with lot numbers: 23F13H1030, 23F13H0046,RF3053077, 23F13J0406, 23F13H0230, 23F13J0435, 23F13J0440, 23F13H1031, 23F13J0181, 23F13G0409, 23F13H0060, 23F13H0231, RF3053088, 23F13H0232, 23F13H0062,  23F13H0233, RF3053092, RF3053096, 23F13J0072, 23F13H0520, 23F13H1032, 23F13H0064, 23F13J0446, 23F13H0804, 23F13J0187, 23F13H0069, 23F13J0193, 23F13H1033, 23F13G0724, 23F13H0898, RF3053035, 23F13J0202, 23F13H0297, RF3053111, 23F13H0521, 23F13G0402, 23F13G0404, 23F13H1034, 23F13G0495, 23F13J0399, 23F13H0072, 23F13H0808, 23F13H0810, 23F13H0313, RF3053655, 23F13J0208, 23F13H0315, 23F13H0768, 23F13J0254, RF3052973, 23F13J0204, 23F13J0037, 23F13J0352, 23F13J0207, 23F13H0687, 23F13H0512, 23F13J0349, 23F13J0452, RF3040503, 23F13G0472, RF3052873, 23F13J0400, 23F13J0401, 23F13H0816, 23F13H0817, 23F13H0818, 23F13J0241, 23F13H0451, 23F13H0861, 23F13H0858, 23F13H0819, 23F13G0830, 23F13H0519, 23F13H0821, 23F13J0351, 23F13G0473, 23F13J0217, 23F13J0073, RF3065022, 23F13H0045, 23F13H0916, 23F13H0796, 23F13J0403, 23F13J0542, 23F13H0127, 23F13J0454, 23F13J0057, 23F13J0291, 23F13J0080, 23F13J0076, 23F13H0994, 23F13H0013, 23F13H0993,  23F13H0022, RF3064650, 23F13H0919, 23F13J0160, 23F13H1018, 23F13J0162, 23F13J0166, 23F13H0911, RF3052890, 23F13J0167, RF3052884, RF3052536, RF3041482, 23F13J0178, RF3054066, 23F13J0135, 23F13J0014, 23F13H0349, 23F13H0351, 23F13H0266, 23F13H0036, 23F13H0265, 23F13H0707, 23F13H0709, 23F13H0941, 23F13J0319, 23F13J0133, 23F13J0390, 23F13J0137, 23F13G0578, 23F13J0331, 23F13H0750, 23F13H0267, 23F13H0752, 23F13H0753, 23F13J0304, 23F13J0298, 23F13J0301, 23F13H0751, 23F13J0303, 23F13G0217, 23F13H0268, 23F13H0269, 23F13H0270, RF3052159, RF3054123, RF3054125, 23F13H0759, 23F13J0249, 23F13H0668, 23F13H0664, 23F13H0706.

Other devices under the same FDA submission

355 devices cleared under K011761, K073451, K780532, K810675, K862056 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates