ARROW IPN036821Recalls & reported problems
Teleflex Incorporated · Arrowg+ard Blue(R) MAC(TM) Two-Lumen Central Venous Access for use with 7 - 8 Fr. Catheters · Class 2
All models of ARROW (1494)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 30801902122248), received between Oct 1, 2021 and Nov 20, 2025 — classified as 1 injury, 7 malfunctions.
The most frequently reported specific device problem is “Unraveled Material” (4 reports).
2 FDA recall records name this device's identifier, 2 still open.
9 other recalls cover devices cleared under the same FDA submission (K011761); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Apr 21, 2026Open, Classified
Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.
Z-2385-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
- Apr 21, 2026Open, Classified
Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.
Z-2377-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Most reported problems
Device problems
- Unraveled Material4
- Deformation Due to Compressive Stress2
- Malposition of Device1
- Material Separation1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Foreign Body In Patient1
- Insufficient Information1
- Perforation of Vessels1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 20, 2025Unraveled Material; Unraveled Material
IT WAS REPORTED THAT: "DR. WAS PLACING THE GUIDE WIRE AND IT CAUGHT ON SOMETHING. WHEN IT WAS REMOVED IT WAS FRAYED. HAD TO PULL THE WHOLE THING OUT AND OPEN NEW KIT. THERE WAS NO REPORTED PATIENT HARM OR INJURY."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23604166 ↗MalfunctionOct 8, 2024Unraveled Material; Unraveled Material
IT WAS REPORTED "THE GUIDEWIRE IN THE KIT UNCOILED WHILE WITHDRAWING FROM THE PATIENT." THERE WAS NO PATIENT HARM OR INJURY. NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20399637 ↗MalfunctionOct 8, 2024Unraveled Material; Unraveled Material
IT WAS REPORTED "THE GUIDEWIRE IN THE KIT UNCOILED WHILE WITHDRAWING FROM THE PATIENT." THERE WAS NO PATIENT HARM OR INJURY. NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20399052 ↗MalfunctionOct 1, 2024Unraveled Material; Unraveled Material
IT WAS REPORTED THAT: "THE INTERNAL WIRE HAD COME OUT OF THE COILED EXTERNAL WIRE WITH COIL STILL IN TACT. THERE WAS NO REPORTED PATIENT HARM OR CONSEQUENCE.".
Patient problems: Insufficient Information
MAUDE report 20342141 ↗InjuryOct 12, 2021Malposition of Device; Malposition of Device
PHYSICIAN USED ULTRASOUND TO IDENTIFY THE RIGHT INTERNAL JUGULAR VEIN. HE GOT FLASHBACK WITH THE INTRODUCER. PUT THE GUIDEWIRE DOWN. ADVANCED A 2 LUMEN PSI IN THE USUAL FASHION. IT WAS DISCOVERED LATER WHEN THE PATIENT HAD A THORACOTOMY, THAT THE SHEATH WENT THROUGH THE VESSEL. IT WAS REPORTED THE PATIENT HAD A GUNSHOT WOUND TO THE RIGHT UPPER QUADRANT. A CHEST TUBE WAS PLACED IN THE EMERGENCY ROOM. AFTER THE SHEATH WAS PLACED THE PATIENT EXPERIENCED A HEMOTHORAX WITH SIGNIFICANT PNEUMOTHORAX, ESTIMATED 3 TO 5 LITERS OF BLOOD. A THORACOTOMY WAS PERFORMED FOR THE INCREASED OUTPUT FROM THE CHEST TUBE. ON EXPLORATION OF THE GUNSHOT WOUND THE SURGEON SAW THE SHEATH IN THE THORAX. IT WAS REPORTED THE INCREASED OUTPUT FROM THE CHEST TUBE WAS RELATED TO THE INJURY CAUSED BY THE SHEATH. THE VESSEL WAS REPAIRED AT THE SAME TIME AS THE GUNSHOT WOUND INJURY. THE PATIENT SURVIVED AND WAS DISCHARGED 2-3 DAYS LATER.
Patient problems: Perforation of Vessels
MAUDE report 12618529 ↗MalfunctionOct 6, 2021Deformation Due to Compressive Stress; Deformation Due to Compressive Stress
THE COMPLAINT IS REPORTED AS: TIP OF DILATORS BENT AND SEVERE KINKING WITH STAINLESS STEEL WIRE IN KIT. IT WAS REPORTED TWO DILATORS USED ON THE SAME PATIENT FOR A RIGHT SUBCLAVIAN VEIN CENTRAL LINE ATTEMPT. AFTER THE USER PULLED THE SECOND DILATOR OUT THAT WAS BENT, THE USER DECIDED TO ABANDON THE PCI AND INSTEAD PUT AN ARROW 18GA LONG CATHETER OVER IT 10.5CM IN LENGTH. NO PATIENT INJURY REPORTED. NO MEDICAL INTERVENTION REQUIRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12585260 ↗MalfunctionOct 6, 2021Deformation Due to Compressive Stress; Deformation Due to Compressive Stress
THE COMPLAINT IS REPORTED AS: TIP OF DILATORS BENT AND SEVERE KINKING WITH STAINLESS STEEL WIRE IN KIT. IT WAS REPORTED TWO DILATORS USED ON THE SAME PATIENT FOR A RIGHT SUBCLAVIAN VEIN CENTRAL LINE ATTEMPT. AFTER THE USER PULLED THE SECOND DILATOR OUT THAT WAS BENT, THE USER DECIDED TO ABANDON THE PCI AND INSTEAD PUT AN ARROW 18GA LONG CATHETER OVER IT 10.5CM IN LENGTH. NO PATIENT INJURY REPORTED. NO MEDICAL INTERVENTION REQUIRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12585145 ↗MalfunctionOct 1, 2021Material Separation; Material Separation
CUSTOMER REPORTED THE GUIDEWIRE WAS SHEERED DURING PROCEDURE AND WAS LODGED IN THE PATIENT. IT WAS REPORTED THE GUIDEWIRE BROKE OFF COMPLETELY AND 8 CM WAS LEFT INSIDE THE PATIENT. THE ICU NURSE NOTICED THE GUIDEWIRE POKING OUT OF THE PATIENT'S GROIN AND REMOVED IT AT THE BEDSIDE. NO MEDICAL INTERVENTION WAS REQUIRED. NO PATIENT HARM OR COMPLICATION REPORTED.
Patient problems: Foreign Body In Patient
MAUDE report 12565027 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K011761. One submission often covers many models, so check the product and lot details in each record.
- Z-2555-2019 · Aug 6, 2019 · Terminated8.5 Fr. X 10 cm ARROW+gard Blue PSI Kit Cath-Gard for use with 7.5 - 8 Fr.Catheters Product Code: CDC-29803-1A - Product Usage: The MACTM Two-Lumen Central Venous Access Device wit
- Z-1983-2018 · Apr 11, 2018 · TerminatedMAC(TM) Two-Lumen Central Venous Access Kit with ARROWg+ard Blue(R) Access Device, Integral Hemostasis Valve and Sharps Safety Features for use with 7.5 - 8 Fr. Catheters; Product
- Z-1324-2018 · Dec 22, 2017 · Terminated(1) ARROWg+ard Blue(R) MAC(TM) Two-Lumen Central Venous Access Kit for use with 7.5 - 8 Fr. Catheters, Product Number: ASK-21142-RH; (2) Arrowg+ard Blue(R) MAC(TM) Two-Lumen Centra
- Z-1002-2018 · Nov 27, 2017 · TerminatedARROWg+ard Blue Product Usage: The Arrow CVC is indicated to provide short-term (< 30 days) central venous access for treatment of diseases or conditions requiring central venous
- Z-1000-2018 · Nov 27, 2017 · TerminatedARROWg+ard Blue MAC" Two-Lumen Central Venous Access Kit Product Usage: The Arrow CVC is indicated to provide short-term (< 30 days) central venous access for treatment of disease
- Z-2333-2017 · May 9, 2017 · Terminated1) ARROWg+ard Blue MAC Two-Lumen Central Venous Access Kit 2) ARROWg+ard Blue PSI Kit 3) ARROWg+ard Blue MAC Two-Lumen Central Venous Access Kit 4) ARROWg+ard Blue PSI Kit The MACT
- Z-2825-2015 · Aug 27, 2015 · TerminatedArrow Kits with Medtronic Covidien 0.9% Sodium Chloride Flush Syringes
- Z-0066-2015 · Aug 6, 2014 · TerminatedArrow Select Kits (ASK) and Maximal Barrier Precautions Kits (CDC) with BD Medical Posiflush 10mL Saline Flush Syringe.
- Z-0783-2010 · Nov 24, 2009 · TerminatedMAC (TM) Two Lumen Central Venous Access Kit with ARROWg+ard Blue (R) Access Device, Integral Hemostasis Valve, Sharps Safety Features and Maximal Barrier Precautions for use with
Other INTRODUCER, CATHETER devices
- VIZIGO D138502Biosense Webster Inc · INTRODUCER, CATHETER2 recalls4 deaths2,839 reports
- VIZIGO D138501Biosense Webster Inc · INTRODUCER, CATHETER2 recalls2 deaths2,005 reports
- Swartz™ 407451St. Jude Medical, Inc. · INTRODUCER, CATHETER5 deaths728 reports
- Edwards eSheath+ Introducer set 914ESPEdwards Lifesciences LLC · Introducer, Catheter22 deaths532 reports
- Fast-Cath™ 406849St. Jude Medical, Inc. · INTRODUCER, CATHETER420 reports
- VIZIGO D138503Biosense Webster Inc · INTRODUCER, CATHETER362 reports
- Agilis™ G408318St. Jude Medical, Inc. · INTRODUCER, CATHETER333 reports
- Agilis™ NxT Steerable Introducer Dual-Reach™ SI-1371-MSt. Jude Medical, Inc. · Introducer, catheter1 death301 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page