The Verigene Gram Negative Blood Culture Nucleic Acid Test… recallZ-2681-2014 · Nanosphere, Inc. · initiated Oct 10, 2013
Class II Terminated All recalls posted in September 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Nanosphere, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Nanosphere has received several reports relating to an increased rate of Processing Errors during procedures performed on the Processor SP caused by Tips in Tip Holder Assemblies for the following products and lots of Extraction Trays.
FDA root cause category: Nonconforming Material/Component
What the company is doing
Nanosphere sent a Field Notification Letter dated October 10, 2013, to all affected customers. The letter identified the product, the problem and the action to be taken by the customer. Instructions were included in the letter for customers to determine whether or not the extraction tray lots should be returned. If the extraction tray lots are to be returned, customers were to complete and return the attached form and replacement tip holder assemblies will be supplied at no cost. Customers could then dispose of the recalled unused tip holder assemblies. Customers with questions were instructed to contact their Nanosphere technical support representative. For questions regarding this recall call 1-888-837-4436, ext 3.
Recall details
- Recall number
- Z-2681-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Oct 10, 2013
- Posted by the FDA
- Sep 23, 2014
- Recalling firm
- Nanosphere, Inc.
- Quantity
- 400 units
- Product code
- PEN
- FDA submissions
- K132843
- Distribution
- Worldwide Distribution - USA including AZ, CA, FL, GA, IA, IL, IN, KY, MA, MD, MN, MO, MT, NC, NE, NJ, NY, OH, OR, PA, TX, UT, VA, WA, WI, OH and Internationally to Bulgaria, Germany, Italy, Japan, South Korea, and Spain.
Product description
The Verigene Gram Negative Blood Culture Nucleic Acid Test (BC-GN), performed using the sample-to-result Verigene System, is a qualitative multiplexed in vitro diagnostic test for the simultaneous detection and identification of selected gram-negative bacteria and resistance markers. BC-GN is performed directly on blood culture media using blood culture bottles identified as positive by a continuous monitoring blood culture system and which contain gram negative bacteria as determined by gram stain.
Code information (models, lots, serial numbers)
Product Number: 20-009-021; Lot Number and Expiration Date: Lot: 100113021C, Expiration Date: 03/30/14
Other devices under the same FDA submission
3 devices cleared under K132843 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Gram-Negative Blood Culture (BC-GN) Nucleic Acid Test Cartridge 20-006-021Luminex Corporation · Gram-Negative bacteria and associated resistance markers1 recall1 death15 reports
- Gram-Negative Blood Culture (BC-GN) Nucleic Acid Test Cartridge 20-006-021Nanosphere, Inc. · Gram-Negative Bacteria And Associated Resistance Markers2 deaths2 reports
- Gram-Negative Blood Culture (BC-GN) Utility Tray 20-011-021Luminex Corporation · Gram-Negative bacteria and associated resistance markers1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 69034).
- Z-2682-2014 · TerminatedThe Verigene Clostridium difficile Nucleic Acid Test (CDF) is a qualitative, multiplexed in vitro diagnostic test for the rapid detection of toxin A (tcdA), toxin B (tcdB), and tcd…
- Z-2685-2014 · TerminatedThe Verigene Enteric Pathogens Nucleic Acid Test (EP) is a multiplexed, qualitative test for simultaneous detection and identification of common pathogenic enteric bacteria and gen…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page