The Verigene Clostridium difficile Nucleic Acid Test (CDF)… recallZ-2682-2014 · Nanosphere, Inc. · initiated Oct 10, 2013

Class II Terminated All recalls posted in September 2014

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Nanosphere, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Nanosphere has received several reports relating to an increased rate of Processing Errors during procedures performed on the Processor SP caused by Tips in Tip Holder Assemblies for the following products and lots of Extraction Trays.

FDA root cause category: Nonconforming Material/Component

What the company is doing

Nanosphere sent a Field Notification Letter dated October 10, 2013, to all affected customers. The letter identified the product, the problem and the action to be taken by the customer. Instructions were included in the letter for customers to determine whether or not the extraction tray lots should be returned. If the extraction tray lots are to be returned, customers were to complete and return the attached form and replacement tip holder assemblies will be supplied at no cost. Customers could then dispose of the recalled unused tip holder assemblies. Customers with questions were instructed to contact their Nanosphere technical support representative. For questions regarding this recall call 1-888-837-4436, ext 3.

Recall details

Recall number
Z-2682-2014
Recall class
Class II
Status
Terminated
Initiated by the company
Oct 10, 2013
Posted by the FDA
Sep 23, 2014
Recalling firm
Nanosphere, Inc.
Quantity
80 units
Product code
OZN
FDA submissions
K123197
Distribution
Worldwide Distribution - USA including AZ, CA, FL, GA, IA, IL, IN, KY, MA, MD, MN, MO, MT, NC, NE, NJ, NY, OH, OR, PA, TX, UT, VA, WA, WI, OH and Internationally to Bulgaria, Germany, Italy, Japan, South Korea, and Spain.

Product description

The Verigene Clostridium difficile Nucleic Acid Test (CDF) is a qualitative, multiplexed in vitro diagnostic test for the rapid detection of toxin A (tcdA), toxin B (tcdB), and tcdC gene sequences of toxigenic strains Clostridium difficile and for presumptive identification of PCR ribotype 027 strains from unformed (liquid or soft) stool specimens collected from patients suspected of having C. difficile infection (CDI). Presumptive identification of the PCR ribotype 027 strain of C. difficile is by detection of the binary toxin (cdt) gene sequence and the single base pair deletion at nucleotide 117 in the tcdC gene. The tcdC gene encodes for a negative regulator in C. difficile toxin production. The test is performed on the Verigene System and utilizes automated specimen preparation and polymerase chain reaction (PCR) amplification, combined with a nanoparticle-based array hybridization assay to detect the toxin gene sequences associated with toxin-producing C. difficile.

Code information (models, lots, serial numbers)

Product Number: 20-009-022; Lot Number and Expiration Date: Lot: 100313022B, Expiration Date: 04/01/14

Other devices under the same FDA submission

1 device cleared under K123197 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.