ARIATELE TELEMETRY TRANSMITTER, Model 96281, with SpO2… recallZ-2586-2014 · Spacelabs Healthcare Inc · initiated Jul 2, 2014

Class II Terminated All recalls posted in September 2014

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Spacelabs Healthcare Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The transmitter display and the Telemetry Central Station display may show a SpO2 value when the sensor is not connected to the patient or in certain sensor failed conditions. This is an expansion of recall Z-0871-2014 because additional serial numbers were identified as being part of this recall action as part of the firm's retrospective analysis of all recalls since 2011.

FDA root cause category: Software design

What the company is doing

Spacelabs notified both consignees via email of the expanded recall July 2, 2014. A f/u letter was sent by overnight delivery to the single new consignee the same day. Spacelabs sent an announcement of the expanded recall to both consignees identified in the expanded recall August 25, 2014. A field service technician from Spacelabs will visit each consignee and upgrade the software as necessary. If you have any questions about this corrective action program, please contact SpaceLabs Healthcare at 1-800- 522-7025, and select 2 for Technical Support.

Recall details

Recall number
Z-2586-2014
Recall class
Class II
Status
Terminated
Initiated by the company
Jul 2, 2014
Posted by the FDA
Sep 4, 2014
Recalling firm
Spacelabs Healthcare Inc
Quantity
8 units
Product code
MHX
FDA submissions
K121480
Distribution
US Distribution to TX.

Product description

ARIATELE TELEMETRY TRANSMITTER, Model 96281, with SpO2 Option C. Physiological patient monitor. Used in a hospital environment.

Code information (models, lots, serial numbers)

Serial numbers: 6281-002500, 6281-002501, 6281-002659, 6281-002661, 6281-002662, 6281-002663, 6281-002664, and 6281-003056.  Model 96281, with SpO2 Option C.

Other devices under the same FDA submission

1 device cleared under K121480 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.