Siemens ADVIA Centaur Systems HBc Total (HBcT), Reagent.… recallZ-0102-2015 · Siemens Healthcare Diagnostics, Inc · initiated Sep 15, 2014
Class II Terminated All recalls posted in October 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Increased imprecision and positive bias is observed with patient samples and negative controls and can result in an increase in initial reactive results.
FDA root cause category: Unknown/Undetermined by firm
What the company is doing
Siemens Healthcare issued An Urgent Medical Device Recall (UMDR) sent to all affected Siemens Healthcare Diagnostics customers in the United States on September 15, 2014. An Urgent Field Safety Notice (UFSN) was sent out for distribution to all customers outside the United States on September 15, 2014. These notices inform the customer of the imprecision and positive bias and a potential increase in initial reactive results associated with ADVIA Centaur Systems HBc Total Lot 100064. Customers were instructed to discontinue use of the affected lots and dispose of the product according to local regulations or procedures. The notices should be reviewed with the facility's laboratory or medical director. Each customer will be required to return a completed confirmation fax-back to Siemens.
Recall details
- Recall number
- Z-0102-2015
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Sep 15, 2014
- Posted by the FDA
- Oct 21, 2014
- Recalling firm
- Siemens Healthcare Diagnostics, Inc
- Quantity
- 5434 (200 tests)
- Product code
- LOM
- FDA submissions
- P040004
- Distribution
- Worldwide Distribution-USA (nationwide) and the countries of Austria Bahrain Belgium Croatia Cyprus Czech Republic Denmark Estonia Finland France Germany Great Britain Greece Hungary Iraq Ireland Italy Jordan Latvia Lebanon Lithuania Luxembourg Netherlands Norway Poland Portugal Romania Slovakia Spain Sweden Switzerland Turkey Unit.Arab Emir and Yemen.
Product description
Siemens ADVIA Centaur Systems HBc Total (HBcT), Reagent. For In Vitro Diagnostic Use Catalog Number: 07566733 Siemens Material Number: 10309508 The ADVIA Centaur HBc Total assay is an in vitro diagnostic test for the qualitative determination of total antibodies to the core antigen of the hepatitis B virus (HBc Total) in human serum or plasma (potassium EDTA, or lithium or sodium heparinized) using the ADVIA Centaur and ADVIA Centaur XP system.
Code information (models, lots, serial numbers)
Lot Number: 100064 Exp.Date:February 10, 2015
Other devices under the same FDA submission
2 devices cleared under P040004 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Atellica™ IM HBcT 10995597Siemens Healthcare Diagnostics Inc. · Qualitative and Quantitative Hepatitis B Virus Antibody assays8 reports
- ADVIA Centaur® HBcT 10309508Siemens Healthcare Diagnostics Inc. · Qualitative and Quantitative Hepatitis B Virus Antibody assays6 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 69292).
- Z-0103-2015 · TerminatedADVIA Centaur Systems HBc Total (HBcT) Kit. For In Vitro Diagnostic Use. Catalog Number: 07566733 Siemens Material Number: 10309508 The ADVIA Centaur HBc Total assay is an in vitr…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page