The Abbott m2000sp is an automated system for performing… recallZ-0948-2015 · Abbott Molecular · initiated Nov 23, 2013
Class II Terminated All recalls posted in January 2015
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Abbott Molecular or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The x-drive assembly may fail prematurely. This could cause the manipulator or liquid handling arm to contact an object, shake and drip, or improperly pipet.
FDA root cause category: Nonconforming Material/Component
What the company is doing
Abbott Molecular sent a different Field Correction Recall / Urgent Field Safety Notice letters dated October 24, 2014 to each of three sets of customers. Customers whose instruments' X-Drive Assembly belts were perviously inspected by a Field Service Representative and found to not require replacement were advised to contact their local Abbott Molecular Representative or Abbott Molecular Technical Support at 800-553-7042 x 2 should they hear any unusual noises associated with the LiHa or RoMa movement or misalignment of the LiHa or RoMa. Customers whose instruments' X-Drive Assembly belts were previously inspected and replaced by a Field Service Representative were notifieied that this maintenance activity took place. Customers with questions were instructed to contact Abbott Molelcular. For questions regarding this recall call 224-361-7000. Customers whose instruments' X-Drive Assembly belts have not been inspected were notified that a Field Service Representative is scheduled to visit their facility to inspect and if necessary, replace the X-Drive Assembly belt(s). Customers were advised to contact their local Abbott Molecular Representative should they hear any unusual noises associated with the LiHa or RoMa movement or misalignment of the LiHa or RoMa.
Recall details
- Recall number
- Z-0948-2015
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Nov 23, 2013
- Posted by the FDA
- Jan 9, 2015
- Recalling firm
- Abbott Molecular
- Quantity
- 176 devices
- Product code
- JJH
- FDA submissions
- K092705
- Distribution
- Worldwide Distribution - USA including CA, FL, GA, IL, MI, NC, NJ, NY, OH, OK, OR, and TX and Internationally to Australia, Austria, Belgium, Bosnia And Herzegovina, Brazil, Canada, China, Colombia, Costa Rica, Denmark, Dominican Republic, Egypt, Finland, France, Germany, Indonesia, Ireland, Israel, Italy, Kazakhstan, Kenya, Malawi, Mali, Mexico, Morocco, Mozambique, New Zealand, Nigeria, Norway, Pakistan, Poland, Russia, Saudi Arabia, Senegal, South Africa, South Korea, Taiwan, Thailand, Turkey, Ukraine, and United Kingdom.
Product description
The Abbott m2000sp is an automated system for performing sample preparation for nucleic acid testing.
Code information (models, lots, serial numbers)
List Number: 09K14-02; Model: E-series; Serial Numbers: 10664 - 10833, 1004, 1005, 10142, 10154, 10171, 10416
Other devices under the same FDA submission
2 devices cleared under K092705 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- m2000 09K14-02Abbott Molecular Inc. · CLINICAL SAMPLE CONCENTRATOR23 reports
- m2000 9K15-01Abbott Molecular Inc. · real time Nucleic acid amplification system6 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page