Medical device recalls: January 2015
The FDA posted 133 medical device recall records in January 2015, from 96 recall events by 69 companies. 6 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Siemens Medical Solutions USA, Inc · 7 events, 7 records
- Baxter Healthcare Corp. · 4 events, 4 records
- GE Healthcare · 4 events, 4 records
- DePuy Orthopaedics, Inc. · 3 events, 5 records
- Siemens Medical Solutions USA, Inc. · 3 events, 5 records
- Arrow International Inc · 3 events, 3 records
- Biomet Spine, LLC · 3 events, 3 records
- Synthes, Inc. · 3 events, 3 records
FDA root cause
- Device Design · 22
- Under Investigation by firm · 14
- Component change control · 13
- Software design · 12
- Component design/selection · 10
- Nonconforming Material/Component · 8
- Process control · 6
- Material/Component Contamination · 5
Class I recalls (6)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Lotus TAVR 23mm, ↗Class IZ-0915-2015 · Boston Scientific Corporation · posted Jan 13, 2015 · Terminated
Lotus valve became unlocked during release from the delivery system. This may lead to percutaneous insertion of another valve, resulting in a Valve in Valve (ViV) implantation. If this occurs, the first valve could embolize and it may be necessary to…
- Lotus TAVR 25mm, ↗Class IZ-0916-2015 · Boston Scientific Corporation · posted Jan 13, 2015 · Terminated
Lotus valve became unlocked during release from the delivery system. This may lead to percutaneous insertion of another valve, resulting in a Valve in Valve (ViV) implantation. If this occurs, the first valve could embolize and it may be necessary to…
- Lotus TAVR 27mm, ↗Class IZ-0917-2015 · Boston Scientific Corporation · posted Jan 13, 2015 · Terminated
Lotus valve became unlocked during release from the delivery system. This may lead to percutaneous insertion of another valve, resulting in a Valve in Valve (ViV) implantation. If this occurs, the first valve could embolize and it may be necessary to…
- Z-0880-2015 · Alere San Diego, Inc. · posted Jan 3, 2015 · Terminated
In certain cases, an INRatio PT/INR Monitor system may provide an INR result that is significantly lower than a result obtained using a reference INR system (laboratory method). This issue can arise if the patient has certain medical conditions. It can…
- INRatio Monitors ↗Class IZ-0881-2015 · Alere San Diego, Inc. · posted Jan 3, 2015 · Terminated
In certain cases, an INRatio PT/INR Monitor system may provide an INR result that is significantly lower than a result obtained using a reference INR system (laboratory method). This issue can arise if the patient has certain medical conditions. It can…
- INRatio 2 Monitors ↗Class IZ-0882-2015 · Alere San Diego, Inc. · posted Jan 3, 2015 · Terminated
In certain cases, an INRatio PT/INR Monitor system may provide an INR result that is significantly lower than a result obtained using a reference INR system (laboratory method). This issue can arise if the patient has certain medical conditions. It can…
Class II recalls (122)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- DrugCheck Ketamine Dip Test ↗Class IIZ-1033-2015 · Express Diagnostics Int'l., Inc. · posted Jan 30, 2015 · Terminated
This recall has been initiated due to potential false positives results when using the test to screen for Ketamine. Use of this product may give incorrect screening results.
- Z-1034-2015 · Animas Corporation · posted Jan 30, 2015 · Terminated
Calibration factors in the pump overwritten during a programming step. The force sensor could send a lower signal value to the pump processor, with loss of prime warnings, occlusion alarms and the pump unable to detect a cartridge during the prime…
- Z-1029-2015 · Orthosensor, Inc. · posted Jan 29, 2015 · Terminated
Presence of one inoperable (dead) sensor pad.
- Z-1031-2015 · Baxter Healthcare Corp. · posted Jan 29, 2015 · Terminated
The keypad buttons on HomeChoice devices may be activated without the operator pressing them.
- Z-1032-2015 · Fresenius Kabi USA, LLC · posted Jan 29, 2015 · Terminated
The lot failed pH specification.
- Z-1021-2015 · Synthes, Inc. · posted Jan 28, 2015 · Terminated
Certain lots have the potential for tip breakage while bending 2.0 mm thick plates. If device is used and breaks while in the Operating Room, surgical delay may occur while alternate bending or cutting instruments are located.
- Z-1022-2015 · Innokas Medical Oy · posted Jan 28, 2015 · Terminated
If the SpO2 or RRa value is violating the respective alarm limit at the time of switching to monitoring mode, the monitor does not issue visual and audible alarms.
- Z-1023-2015 · Beckman Coulter Inc. · posted Jan 28, 2015 · Terminated
N-Acetyl Cysteine (NAC), when administered in therapeutic concentrations for the treatment of acetaminophen overdose, interferes with the assays for these analytes: Cholesterol OSR6X16, Uric Acid OSR6X98, Lactate OSR6X93, and Lipase OSR6X30. NAC…
- Z-1024-2015 · Beckman Coulter Inc. · posted Jan 28, 2015 · Terminated
N-Acetyl Cysteine (NAC), when administered in therapeutic concentrations for the treatment of acetaminophen overdose, interferes with the assays for these analytes: Cholesterol OSR6X16, Uric Acid OSR6X98, Lactate OSR6X93, and Lipase OSR6X30. NAC…
- Z-1025-2015 · Beckman Coulter Inc. · posted Jan 28, 2015 · Terminated
N-Acetyl Cysteine (NAC), when administered in therapeutic concentrations for the treatment of acetaminophen overdose, interferes with the assays for these analytes: Cholesterol OSR6X16, Uric Acid OSR6X98, Lactate OSR6X93, and Lipase OSR6X30. NAC…
- Z-1026-2015 · Beckman Coulter Inc. · posted Jan 28, 2015 · Terminated
N-Acetyl Cysteine (NAC), when administered in therapeutic concentrations for the treatment of acetaminophen overdose, interferes with the assays for these analytes: Cholesterol OSR6X16, Uric Acid OSR6X98, Lactate OSR6X93, and Lipase OSR6X30. NAC…
- Z-1027-2015 · Siemens Medical Solutions USA, Inc · posted Jan 28, 2015 · Terminated
In case of a system crash, images may not be written to the hard disk and this may result in inconsistencies in the database. In case of a system crash (e.g. blue screen, power outage) images may not be written from cache to the hard disk and might get lost.
- Ebb Complete Tamponade System, BD-OTS Obstetric Catheter, REF CTS-1000, Rx Only, Sterile EO ↗Class IIZ-1014-2015 · Clinical Innovations, LLC · posted Jan 27, 2015 · Terminated
Potential for a balloon leak in all lots of (CTS-1000) manufactured since August 2014.
- Z-1015-2015 · Fresenius Medical Care Holdings, Inc. · posted Jan 27, 2015 · Terminated
Unqualified power assemblies shipped to customers.
- Z-1016-2015 · Mobius Imaging, LLC · posted Jan 27, 2015 · Terminated
There is a risk that during the transfer of an image and navigation data to the Brainlab Curve Image Guided Surgery Navigation System after a CT scan, an error may occur, causing either no navigation data or incorrect navigation data being transferred to…
- Z-1019-2015 · Sunrise Medical (US) LLC · posted Jan 27, 2015 · Terminated
There is the potential for the height adjustable handle to dislodge due to failing locking mechanisms on specific identified chairs.
- Z-1020-2015 · Sunrise Medical (US) LLC · posted Jan 27, 2015 · Terminated
There is the potential for the height adjustable handle to dislodge due to failing locking mechanisms on specific identified chairs.
- Z-0955-2015 · GE Healthcare · posted Jan 26, 2015 · Terminated
A required quality control test was not performed during installation associated with the software of the Revolution CT scanner.
- Z-1010-2015 · Baxter Healthcare Corp. · posted Jan 26, 2015 · Terminated
Potential for failure of the pouch packaging seal at high altitudes.
- Z-1011-2015 · Philips Medical Systems (Cleveland) Inc · posted Jan 26, 2015 · Terminated
Unintended detector and gantry movement due to software issues.
- Z-1012-2015 · Aesculap, Inc. · posted Jan 26, 2015 · Terminated
Some Miethke Shunt System accessories may have an incorrect MR safety symbol on the shelf box label. The label should have NO symbol but instead is marked with the MR conditional symbol. The product is MR unsafe. This information is only required to be in…
- Z-1013-2015 · Siemens Medical Solutions USA, Inc · posted Jan 26, 2015 · Terminated
Combination of CTVision with syngo RT Therapist / syngo RT Oncologist 4.3.SP1 automatic registration in Adaptive Targeting might result in wrong offset calculations. Applying this offset can lead to patient mistreatment. Cone Beam imaging is not affected…
- Z-1009-2015 · Synthes, Inc. · posted Jan 23, 2015 · Terminated
Under new testing protocols for MR Environment safety and compatibility, metal devices are no longer designated MR Safe. Some components in the CMF Mandible External Fixator Systems were labeled and/or etched with information indicating they are MR Safe.…
- Z-1005-2015 · Arrow International Inc · posted Jan 22, 2015 · Terminated
the Instructions for Use (IFU) were inadvertently removed from the kit.
- Z-1006-2015 · Aptalis Pharmatech Inc. · posted Jan 22, 2015 · Terminated
Customer notification that the device may be difficult to open or close.
- Z-1007-2015 · Arrow International Inc · posted Jan 22, 2015 · Terminated
The tunneling sheath packaged within the kits can crack during placement. If the tunneling sheath cracks, it can cause a delay in treatment for the time it takes to retrieve a replacement kit.
- Z-1000-2015 · BD Biosciences, Systems & Reagents · posted Jan 20, 2015 · Terminated
One lot of CD8 PE is contaminated with CD4 FITC and may cause an unexpected staining pattern and incorrect patient data.
- Z-1001-2015 · Genetic Testing Institute,inc · posted Jan 20, 2015 · Terminated
Some Lots of PF4 Concentrated Wash (10X) are a slightly more dilute concentration. This may lead to slightly increased Optical Density (OD) values. Samples with a higher background reactivity may demonstrate false positive result (borderline reactivity) at…
- Z-1002-2015 · BD Biosciences, Systems & Reagents · posted Jan 20, 2015 · Terminated
Two lots of the Anti-Lambda APC-H7 antibody are contaminated with CD38 antibody.
- Z-1003-2015 · Dick's Sporting Goods · posted Jan 20, 2015 · Terminated
Reports of customers falling from the Inversion Table and sustaining injuries. The Inversion Table is a fitness product, not a medical device. DSG removed medical claims from product labels and advertising.
- Z-0993-2015 · Maquet Medical Systems USA · posted Jan 16, 2015 · Terminated
The wrong movement results on the MAGNUS OR table. The button commands on the TEGRIS touchscreen for Lower Leg Up and Upper Leg Down are switched in the software for the integration with the MAGNUS operating table system.
- Ysio Max, Luminos dRF Max and Agile Max systems with software version VE10 and SmartOrtho license. ↗Class IIZ-0994-2015 · Siemens Medical Solutions USA, Inc · posted Jan 16, 2015 · Terminated
Potential for composed images to be flipped before being sent to PACS on systems with software version VE10 and SmartOrtho license. Flipped images may be reversed so the associated annotations, e.g. labels (R/L), may be displayed incorrectly, potentially…
- Z-0995-2015 · Ferno-Washington Inc · posted Jan 16, 2015 · Terminated
The wheel castor assemblies may be loosening on the stretchers.
- Z-0996-2015 · TransMotion Medical Inc · posted Jan 16, 2015 · Terminated
The firm is finding that the stretcher chair controllers are moving the chairs involuntarily.
- Philips IntelliVue TcG10 ↗Class IIZ-0997-2015 · Philips Medical Systems, Inc. · posted Jan 16, 2015 · Terminated
Philips has discovered that the Instructions for Use (IFU) for the IntelliVue TcG10 Transcutaneous Gas Measurement Module is missing the contraindication that the device is not to be used on patients under gas anesthesia.
- Z-0998-2015 · Cordis Corporation · posted Jan 16, 2015 · Terminated
One unit from a demo lot of POWERFLEX PRO PTA Dilatation Catheter was left at a hospital and was not sterilized.
- Z-0956-2015 · Brainlab AG · posted Jan 15, 2015 · Terminated
iPlan RT Radiation Treatment Planning Software: Potentially incorrect patient positioning when using multiple localized CT image data sets.
- Z-0957-2015 · Brainlab AG · posted Jan 15, 2015 · Terminated
iPlan RT Radiation Treatment Planning Software: Potentially incorrect patient positioning when using multiple localized CT image data sets.
- Z-0979-2015 · Philips Medical Systems (Cleveland) Inc · posted Jan 15, 2015 · Terminated
Improper assembly of the vertical motor/brake system may lead to uncommanded vertical motion.
- Z-0980-2015 · Siemens Medical Solutions USA, Inc · posted Jan 15, 2015 · Terminated
Use of any unauthorized third-party components on Siemens LINACs may lead to patient or operator injury, damage to the LINAC, and can potentially compromise customer warranty. The dark current dose must be considered in the overall dose calculation, or an…
- Z-0981-2015 · Straumann USA, LLC · posted Jan 15, 2015 · Terminated
Drill set may contain incorrect drill.
- Z-0982-2015 · Biomet Spine, LLC · posted Jan 15, 2015 · Terminated
The outer diameter of the Drill shaft is oversized; thereby, resulting in interference fit when inserting the Drill into the Fixed or Variable Sleeve Assembly.
- Z-0983-2015 · Biomet Spine, LLC · posted Jan 15, 2015 · Terminated
The inner diameter of the Sleeve shaft is undersized, resulting in interference fit when inserting the Anchored Lateral Straight Retractable Awl or the Anchored Lateral Straight Retractable Drill into the Sleeve.
- Z-0985-2015 · GE Healthcare · posted Jan 15, 2015 · Terminated
Potential safety issue due to a required torque check on three bolted joints of the Revolution CT scanner. No injuries or illnesses reported.
- Z-0986-2015 · Stryker Instruments Div. of Stryker Corporation · posted Jan 15, 2015 · Terminated
Faulty SurgiCounter Charging Cradle can prevent the scanner from charging or downloading information after a procedure.
- Z-0987-2015 · Siemens Healthcare Diagnostics, Inc. · posted Jan 15, 2015 · Terminated
Enabling sending of preliminary/initial results on the ADVIA Chemistry XPT creates a risk that a critical result could be overwritten by the same result rather than showing the repeated result on CentraLink. A result could be erroneous and critical and…
- Femto LDV Z-Generation Femtosecond Surgical Lasers are used to perform ophthalmic surgeries. ↗Class IIZ-0988-2015 · Ziemer Usa Inc · posted Jan 15, 2015 · Terminated
The devices were not shipped with the Starter Kits.
- Z-0989-2015 · Med Tec Inc · posted Jan 15, 2015 · Terminated
The product (wool wrist loop) was manufactured 7cm shorter than product specifications.
- Z-0990-2015 · Med Tec Inc · posted Jan 15, 2015 · Terminated
The product (wool wrist loop) was manufactured 7cm shorter than product specifications.
- Z-0854-2015 · Siemens Medical Solutions USA, Inc · posted Jan 14, 2015 · Terminated
The Small Focus of a 3 focal spot x-ray tube may fail and further release of radiation will not be possible. The Large Focus and Micro Focus focal spots will still be operational without any restriction. Manual switchover is needed, resulting in a short…
- Z-0966-2015 · GE Healthcare · posted Jan 14, 2015 · Terminated
Potential unintentional system motion or rotation due to a positioning user interface issue with Discovery IGS 730 and 740 products.
- VITROS 3600 Immunodiagnostic System, Catalog Number 6802783, IVD --- Ortho Clinical Diagnostics. ↗Class IIZ-0967-2015 · Ortho-Clinical Diagnostics · posted Jan 14, 2015 · Terminated
Software Anomaly: the firm has identified an anomaly with VITROS System Software Version 3.1 and below, and determined that the software may not properly identify an expired calibration.
- Z-0968-2015 · Ortho-Clinical Diagnostics · posted Jan 14, 2015 · Terminated
Software Anomaly: the firm has identified an anomaly with VITROS System Software Version 3.1 and below, and determined that the software may not properly identify an expired calibration.
- VITROS 5600 Integrated System, Catalog Number 6802413, IVD --- Ortho Clinical Diagnostics. ↗Class IIZ-0969-2015 · Ortho-Clinical Diagnostics · posted Jan 14, 2015 · Terminated
Software Anomaly: the firm has identified an anomaly with VITROS System Software Version 3.1 and below, and determined that the software may not properly identify an expired calibration.
- EXp Tibial Insert ↗Class IIZ-0970-2015 · Stelkast Co · posted Jan 14, 2015 · Terminated
Difficulty locking EXp tibial inserts into tibial trays, caused by a manufacturing condition where a lead in radius on the anterior locking tab is not present on certain insert lots. The absence of this lead in may prevent or make assembly of the insert…
- SOMATOM ForceClass IIZ-0971-2015 · Siemens Medical Solutions USA, Inc · posted Jan 14, 2015 · Terminated
Artifacts are found in acquired imaging on the SOMATOM Force.
- Z-0972-2015 · Siemens Medical Solutions USA, Inc · posted Jan 14, 2015 · Terminated
Suboptimal routing of the cable may result in increased wear over time. Without additional measures, damage of a cable may result in limited functionality and in rare cases, failure of a system. An ongoing procedure could be terminated in rare cases.
- Z-0973-2015 · Synthes, Inc. · posted Jan 14, 2015 · Terminated
Potential failure and/or corrosion of the internal spring of the retractor, which may have been manufactured from an incorrect material. Surgical delay may occur. Spring particles could transfer from the instrument to the patient, and may lead to an…
- Z-0974-2015 · Becton Dickinson & Co. · posted Jan 14, 2015 · Terminated
A low level of surface and subsurface contamination of Listeria monocytogenes on 9 lots of non-sterile prepared plated media containing sheep blood, typically visible immediately upon removal from the packaging. The contamination is typically observed as…
- Z-0975-2015 · Becton Dickinson & Co. · posted Jan 14, 2015 · Terminated
A low level of surface and subsurface contamination of Listeria monocytogenes on 9 lots of non-sterile prepared plated media containing sheep blood, typically visible immediately upon removal from the packaging. The contamination is typically observed as…
- Z-0976-2015 · Becton Dickinson & Co. · posted Jan 14, 2015 · Terminated
A low level of surface and subsurface contamination of Listeria monocytogenes on 9 lots of non-sterile prepared plated media containing sheep blood, typically visible immediately upon removal from the packaging. The contamination is typically observed as…
- Z-0977-2015 · SpineFrontier, Inc. · posted Jan 14, 2015 · Terminated
A crack may develop in the handle to the Inline Persuader, which may progress if the user continues to use the affected instrument.
- Timberline Anchored Lateral Variable Retractable Sleeve LITPCR8600-2101-07, Model No. 86300204 ↗Class IIZ-0978-2015 · Biomet Spine, LLC · posted Jan 14, 2015 · Terminated
The inner diameter of the Sleeve shaft is undersized, resulting in interference fit when inserting the Anchored Lateral Straight Retractable Awl or the Anchored Lateral Straight Retractable Drill into the Sleeve shaft.
- Z-0840-2015 · GE OEC Medical Systems, Inc · posted Jan 13, 2015 · Terminated
mA Accuracy may exceed the design and labeling specification of+ 10%. Per 1020.32(f): Deviation of x-ray tube potential and current from the indicated values shall not exceed the maximum deviation as stated by the manufacturer in accordance with 1020.30(h)(3).
- Z-0841-2015 · GE OEC Medical Systems, Inc · posted Jan 13, 2015 · Terminated
mA Accuracy may exceed the design and labeling specification of+ 10%. Per 1020.32(f): Deviation of x-ray tube potential and current from the indicated values shall not exceed the maximum deviation as stated by the manufacturer in accordance with 1020.30(h)(3).
- Z-0958-2015 · Invivo Corporation · posted Jan 13, 2015 · Terminated
The MR 200 devices failed the initial power up test due to capacitor short and hi-pot testing; possible loss of Patient Monitoring or User/Service Electrical Shock.
- IC PROFILER-MR - Model 1123. ↗Class IIZ-0959-2015 · Sun Nuclear Corporation · posted Jan 13, 2015 · Terminated
Has a Potential Pinch Hazard to user performing Machine QA in an MRI environment.
- Z-0960-2015 · Siemens Medical Solutions USA, Inc. · posted Jan 13, 2015 · Terminated
There are confirmed reports of a component coming loose inside the CT enclosure and damaging the system. components.
- Z-0963-2015 · Philips Medical Systems, Inc. · posted Jan 13, 2015 · Terminated
Device may malfunction, which could cause therapy to be delayed, disabled, or delivered inadvertently.
- MENTOR MemoryGel Resterilizable Gel Sizer Moderate Plus Profile STERILE. Breast prosthesis sizer.Class IIZ-0964-2015 · Mentor Texas, LP. · posted Jan 13, 2015 · Terminated
The IFU provided in the affected units of the MENTOR MemoryGel Resterilizable Sizer is incorrect. Those units include the IFU for the MENTOR MemoryShape Resterilizable Sizer.
- Z-0965-2015 · Biocare Medical Llc · posted Jan 13, 2015 · Terminated
Three lots of MUM-1 may not meet their labeled shelf life, as a drop in staining intensity over time has been observed.
- Z-0947-2015 · Phadia US Inc · posted Jan 12, 2015 · Terminated
During an investigation of instrument logs it was determined that In specific circumstances involving multiple steps, a rack sequencing error may occur. This will result in a mismatch between the sample ID and the test result reported for all subsequent…
- Z-0950-2015 · Intertrade Imports Inc. · posted Jan 12, 2015 · Terminated
A surveillance sample collected by FDA confirmed the presence of Bacillus Cereus and Bacillus Lentus in the contact lenses. I Color Color Complete brand color contact lenses are unapproved medical devices.
- Z-0951-2015 · Intertrade Imports Inc. · posted Jan 12, 2015 · Terminated
A surveillance sample collected by FDA confirmed the presence of Bacillus Cereus and Bacillus Lentus in the contact lenses, product is not sterile. I Color Color Complete brand color contact lenses are unapproved medical devices.
- Z-0952-2015 · Medtronic Sofamor Danek USA Inc · posted Jan 12, 2015 · Terminated
The retaining tabs component of the Set Screw Retaining Drive may be oversized, which will not allow proper mating between the Set Screw Retaining Drive and the set screw.
- Z-0953-2015 · DePuy Orthopaedics, Inc. · posted Jan 12, 2015 · Terminated
Complaints have indicated that the ATTUNE INTUITION Impaction Handle lever has fractured during impaction. If the lever fractures, there is the potential for fractured pieces of the device, or the inner spring, if released, to be left in the patient if not…
- Z-0954-2015 · GE Healthcare · posted Jan 12, 2015 · Terminated
GE Healthcare has recently become aware of a potential safety issue involving the breakage of the two steel cables which support the bucky device in the wall stand of Proteus XR/a X-ray imaging systems. No injuries reported.
- Ethicon PERMA-HAND Silk siliconized black braided 10 strands per packet Ethicon LLC- Made in USA.Class IIZ-0492-2015 · Ethicon, Inc. · posted Jan 9, 2015 · Terminated
Ethicon is unable to ensure that PERMA-HAND silk suture (siliconized black braid) meets the expected sterility assurance level for sterile products.
- Z-0948-2015 · Abbott Molecular · posted Jan 9, 2015 · Terminated
The x-drive assembly may fail prematurely. This could cause the manipulator or liquid handling arm to contact an object, shake and drip, or improperly pipet.
- Merlin@home RF Remote Monitoring Transmitter, Model EX1150, with software versions 6.1M or 7.0. ↗Class IIZ-0949-2015 · St Jude Medical Cardiac Rhythm Management Division · posted Jan 9, 2015 · Terminated
Transmitter may initiate a software reset resulting in backup operation in some implanted St. Jude Medical Radio Frequency (RF) enabled Implantable Cardioverter Assura, Unify Assura, and Quadra Assura ICDs and Assurity and Allure Pacemakers.
- Z-0942-2015 · Centurion Medical Products Corporation · posted Jan 8, 2015 · Terminated
According to the recall notice received from Hospira, the recall was initiated due to a confirmed customer report of particulate in a single unit. Hospira identified the particulate as human hair, embedded in and attached to a pinched area of the stopper.
- Z-0943-2015 · Centurion Medical Products Corporation · posted Jan 8, 2015 · Terminated
According to the recall notice received from Hospira, the recall was initiated due to a confirmed customer report of particulate in a single unit. Hospira identified the particulate as human hair, embedded in and attached to a pinched area of the stopper.
- Z-0944-2015 · Centurion Medical Products Corporation · posted Jan 8, 2015 · Terminated
According to the recall notice received from Hospira, the recall was initiated due to a confirmed customer report of particulate in a single unit. Hospira identified the particulate as human hair, embedded in and attached to a pinched area of the stopper.
- Z-0945-2015 · Centurion Medical Products Corporation · posted Jan 8, 2015 · Terminated
According to the recall notice received from Hospira, the recall was initiated due to a confirmed customer report of particulate in a single unit. Hospira identified the particulate as human hair, embedded in and attached to a pinched area of the stopper.
- Z-0946-2015 · Siemens Healthcare Diagnostics, Inc. · posted Jan 8, 2015 · Terminated
Dimension Vista Myoglobin Calibrator (MYO CAL) may produce a positive shift in MYO QC and Patient test results that exceeds acceptance criteria for this product. A positive shift observed up to 12% at MYO concentrations within and above the reference range…
- Z-0930-2015 · Steris Corporation · posted Jan 7, 2015 · Terminated
STERIS has learned that groups of LED lights (modules) contained within the vLED light head may illuminate inconsistently.
- Z-0931-2015 · Synthes (USA) Products LLC · posted Jan 7, 2015 · Terminated
The device may operate solely in reverse mode, not operate in reverse mode when intended, not function, or unintentionally start. Unintended start of the hand piece or wrong mode/direction could cause bone/soft tissue, peripheral nerve, and /or vascular…
- Z-0932-2015 · Baxter Healthcare Corp. · posted Jan 7, 2015 · Terminated
System error 01779 is produced when the battery cannot be charged due to specific voltage differences between the battery and the charger.
- Z-0933-2015 · Kci USA, Inc. · posted Jan 7, 2015 · Terminated
KCI has received reports that, in a small number of cases, the CelluTome Harvester (a component of the CelluTome Epidermal Harvesting System), blades have the potential to drift during shipping and could prematurely protrude into openings of harvester top…
- Z-0934-2015 · Vision Rt Inc · posted Jan 7, 2015 · Terminated
Potential use error when the external Gate Controller is turned ON after the Vision RT software has been opened.
- Lumbar Extension Machine ↗Class IIZ-0935-2015 · MedX Holdings, Inc. · posted Jan 7, 2015 · Terminated
Review of DHRs revealed that HiPot, Continuity, load and friction re-testing was not completed on these devices during manufacturing.
- Cervical Extension Machine ↗Class IIZ-0936-2015 · MedX Holdings, Inc. · posted Jan 7, 2015 · Terminated
Review of DHRs revealed that HiPot, Continuity, load and friction re-testing was not completed on these devices during manufacturing.
- Z-0937-2015 · Orthovita, Inc., dBA Stryker Orthobiologics. · posted Jan 7, 2015 · Terminated
There is the potential for a breach in the inner or outer packaging pouches of all lots of the Stryker Orthobiologics Imbibe Aliquot Needle Bone Cement Needle.
- Z-0938-2015 · Orthovita, Inc., dBA Stryker Orthobiologics. · posted Jan 7, 2015 · Terminated
There is the potential for a breach in the inner or outer packaging pouches of all lots of the Imbibe Bone Marrow Aspiration Needle.
- Z-0939-2015 · Baxter Healthcare Corp. · posted Jan 7, 2015 · Terminated
There are additional and updated warning and cautions that are not in the Patient At-Home Guide for HomeChoice Automated PD Systems and HomeChoice PRO Automated PD Systems.
- IS3000 Vision Side Cart (VSC) used in conjunction with the da Vinci Si Surgical System IS3000. ↗Class IIZ-0940-2015 · Intuitive Surgical, Inc. · posted Jan 7, 2015 · Terminated
Intuitive Surgical is initiating a voluntary correction related to da Vinci Si System (IS3000) Vision Side Carts, with threaded-in caster wheels, which can result in the caster wheel bolts gradually unthreading and the wheels loosening.
- CSA Medical truFreeze System ↗Class IIZ-0941-2015 · CSA Medical · posted Jan 7, 2015 · Terminated
An increase in complaint trending for a loss of audible sound associated with the timer has been observed.
- Z-0911-2015 · Siemens Medical Solutions USA, Inc. · posted Jan 6, 2015 · Terminated
Siemens Medical Solutions USA, Inc. has received reports of unintended radial motion of the detectors. This unintended motion, should it occur, may cause the patient compression related injuries.
- Z-0912-2015 · Siemens Medical Solutions USA, Inc. · posted Jan 6, 2015 · Terminated
Siemens Medical Solutions USA, Inc. has received reports of unintended radial motion of the detectors. This unintended motion, should it occur, may cause the patient compression related injuries.
- Z-0913-2015 · Siemens Medical Solutions USA, Inc. · posted Jan 6, 2015 · Terminated
Siemens Medical Solutions USA, Inc. has received reports of unintended radial motion of the detectors. This unintended motion, should it occur, may cause the patient compression related injuries.
- Z-0914-2015 · Siemens Medical Solutions USA, Inc. · posted Jan 6, 2015 · Terminated
There is a potential for patient finger injury on the e.cam systems.
- Z-0918-2015 · Trumpf Medical Systems, Inc. · posted Jan 6, 2015 · Terminated
Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes
- Cable remote control JUPITER / TS7500, Material no.1601926, Serial number: to (incl.) 101465919. ↗Class IIZ-0919-2015 · Trumpf Medical Systems, Inc. · posted Jan 6, 2015 · Terminated
Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes
- Z-0920-2015 · Trumpf Medical Systems, Inc. · posted Jan 6, 2015 · Terminated
Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes
- Z-0921-2015 · Trumpf Medical Systems, Inc. · posted Jan 6, 2015 · Terminated
Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes
- IR remote control JUPITER / TS7500, Material no.1601927, Serial number: to (incl.) 102222333. ↗Class IIZ-0922-2015 · Trumpf Medical Systems, Inc. · posted Jan 6, 2015 · Terminated
Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes
- Z-0923-2015 · Trumpf Medical Systems, Inc. · posted Jan 6, 2015 · Terminated
Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes
- Z-0924-2015 · Trumpf Medical Systems, Inc. · posted Jan 6, 2015 · Terminated
Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes
- Z-0925-2015 · Trumpf Medical Systems, Inc. · posted Jan 6, 2015 · Terminated
Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes
- Z-0926-2015 · Trumpf Medical Systems, Inc. · posted Jan 6, 2015 · Terminated
Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes
- Z-0927-2015 · Trumpf Medical Systems, Inc. · posted Jan 6, 2015 · Terminated
Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes
- Cable remote control JUC, BUTTON, Material no.1239823, Serial number: to (incl.) 100143048. ↗Class IIZ-0928-2015 · Trumpf Medical Systems, Inc. · posted Jan 6, 2015 · Terminated
Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes
- Z-0929-2015 · Trumpf Medical Systems, Inc. · posted Jan 6, 2015 · Terminated
Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes
- Z-0903-2015 · DePuy Orthopaedics, Inc. · posted Jan 5, 2015 · Terminated
ATTUNE INTUITION Impactors (cat. nos. 254401003, 254401004, and 254401006) have broken and produced small pieces. Should a fracture occur and not be observed during surgery, there is the potential for these small fractured pieces of the instrument to be…
- Z-0904-2015 · DePuy Orthopaedics, Inc. · posted Jan 5, 2015 · Terminated
ATTUNE INTUITION Impactors (cat. nos. 254401003, 254401004, and 254401006) have broken and produced small pieces. Should a fracture occur and not be observed during surgery, there is the potential for these small fractured pieces of the instrument to be…
- Z-0905-2015 · DePuy Orthopaedics, Inc. · posted Jan 5, 2015 · Terminated
ATTUNE INTUITION Impactors (cat. nos. 254401003, 254401004, and 254401006) have broken and produced small pieces. Should a fracture occur and not be observed during surgery, there is the potential for these small fractured pieces of the instrument to be…
- Z-0906-2015 · Exactech, Inc. · posted Jan 5, 2015 · Terminated
Mislabeled
- Z-0907-2015 · Spinal Elements, Inc · posted Jan 5, 2015 · Terminated
Spinal Elements is recalling the Ti-Bond related brochures that includes LT-10000 and MP-14000 (all revisions) the content regarding the Ti-Bond coating within the literature is unsubstantiated in context and/or misleading for the coating's use in spine…
- Z-0908-2015 · DePuy Orthopaedics, Inc. · posted Jan 5, 2015 · Terminated
Use of excessive force when impacting may lead to intra-operative femoral fracture.
- Z-0909-2015 · Ebi, Llc · posted Jan 5, 2015 · Terminated
Inoperative screw head seat splay and/or fracture as a result of positioning of the rod which is a part of the Cypher MIS Screw System.
- Z-0910-2015 · McKesson Israel Ltd. · posted Jan 5, 2015 · Terminated
An ECG procedure time stamp is incorrect, due to a software error in the McKesson Cardiology ECG Management with software versions 13.1 and 13.1.1.
- ABX PENTRA Reagent Container, Model No. B1037307 (15 mL), B1034626 (10 mL), and B1034634 (4 mL)Class IIZ-0902-2015 · Horiba Instruments, Inc dba Horiba Medical · posted Jan 2, 2015 · Terminated
HORIBA Medical is recalling the ABX PENTRA 400 Reagent Container because they do not always sit properly into the reagent rack. In low reagent level scenarios, this could lead to insufficient reagent volume being pipetted and could lead to incorrect…
Class III recalls (5)
Using the device is not likely to cause health problems, but it violates FDA rules.
- Z-1028-2015 · Smiths Medical ASD, Inc. · posted Jan 29, 2015 · Terminated
The outer box label is marked with the correct expiration date of 08/2016. The inner Tyvek Unit Label is marked with 08-2017 which exceeds the actual expiration date by one year.
- Z-1030-2015 · Arrow International Inc · posted Jan 29, 2015 · Terminated
Certain product codes and lot numbers of Kits with Biopatch Dressing were packaged with the incorrect size Biopatch Dressing. No risk to patients is anticipated.
- Z-1004-2015 · Ortho-Clinical Diagnostics · posted Jan 20, 2015 · Terminated
Calibration may not occur when using calibrator barcode labels supplied with VITROS Chemisty Products Calibrator Kit 2.
- Infinite 200 ↗Class IIIZ-0991-2015 · Tecan US, Inc. · posted Jan 15, 2015 · Terminated
Incorrect lumi firmware version installed (E.027 instead of V2.00)
- Infinite 200PRO ↗Class IIIZ-0992-2015 · Tecan US, Inc. · posted Jan 15, 2015 · Terminated
Incorrect lumi firmware version installed (E.027 instead of V2.00)