Ovation Prime Abdominal Stent Graft System - 29mm Aortic… recallZ-0125-2015 · Trivascular, Inc · initiated Sep 20, 2014
Class I Terminated All recalls posted in November 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Trivascular, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
TriVascular has identified select lots of Ovation Prime 29mm Aortic Body stent grafts that may be associated with rapid emptying of the fill polymer syringe when it is connected to the Autoinjector during the fill polymer injection step of the primary implant procedure.
FDA root cause category: Under Investigation by firm
What the company is doing
Urgent Field Safety Notice letter was sent to all consignees on 9/19/14 via trackable mail system. Users are notified that TriVascular has identified select lots of Ovation Prime 29mm Aortic Body stent grafts that may be associated with rapid emptying of the fill polymer syringe when it is connected to the Autoinjector during the fill polymer injection step of the primary implant procedure.
Recall details
- Recall number
- Z-0125-2015
- Recall class
- Class I (most serious)
- Status
- Terminated
- Initiated by the company
- Sep 20, 2014
- Posted by the FDA
- Nov 3, 2014
- Recalling firm
- Trivascular, Inc
- Quantity
- 0 Devices
- Product code
- MIH
- FDA submissions
- P120006
- Distribution
- International Distribution Only -- Canada, Turkey, Austria, Ireland, Belgium, Italy, Switzerland, Spain, France, and Cyprus.
Product description
Ovation Prime Abdominal Stent Graft System - 29mm Aortic Body stent grafts. The TriVascular Ovation Prime Abdominal Stent Graft System is a low-profile endovascular device delivered via catheter to treat descending abdominal aortic aneurysms (AAAs).
Code information (models, lots, serial numbers)
The devices subject to this action are labeled for distribution outside of the United States; therefore, no UDI is on the device label or package. Model Number: TV-AB2980-D Lot Numbers: FS031814-11 FS052214-22 FS052814-36 FS040414-21 FS050614-56 FS040714-53 FS040514-02 FS042614-05 FS050614-53 FS032614-46 FS050214-27 FS050314-12 FS052214-48 FS052214-53 FS031814-44 FS031814-49 FS032714-04 FS032714-20 FS040414-30 FS040514-08 FS050114-01 FS050114-08 FS050514-18 FS050514-27 FS050614-54 FS050614-55 FS051914-32 FS051914-34 Expiry dates ranging "May-17" through "Jun-17"
Other devices under the same FDA submission
42 devices cleared under P120006 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Ovation iX Abdominal Stent Graft System Aortic BodyTrivascular, Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment11 deaths105 reports
- Ovation iX Abdominal Stent Graft System Aortic BodyTrivascular, Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment4 deaths87 reports
- Ovation iX Abdominal Stent Graft System Aortic BodyTrivascular, Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths72 reports
- Alto™ Abdominal Stent Graft System TV-AB2980-NEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment1 death65 reports
- Alto™ Abdominal Stent Graft System TV-AB2680-NEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment5 deaths56 reports
- Alto™ Abdominal Stent Graft System TV-AB3480-NEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment6 deaths56 reports
- Ovation iX Abdominal Stent Graft System Aortic BodyTrivascular, Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths54 reports
- Alto™ Abdominal Stent Graft System TV-AB2380-NEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment4 deaths48 reports
- Alto™ Abdominal Stent Graft System TV-AB2080-NEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment20 reports
- Ovation iX Abdominal Stent Graft System Aortic BodyTrivascular, Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment20 reports
- Alto™ Abdominal Stent Graft System TV-AB2680-NEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment8 reports
- Alto™ Abdominal Stent Graft System TV-AB3480-NEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment1 death8 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.