Medical device recalls: November 2014
The FDA posted 93 medical device recall records in November 2014, from 56 recall events by 46 companies. 4 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Siemens Healthcare Diagnostics, Inc. · 3 events, 4 records, 2 Class I
- Nidek Inc · 3 events, 3 records
- Zimmer, Inc. · 2 events, 3 records
- Integra LifeSciences Corp. · 2 events, 2 records
- Medacta Usa · 2 events, 2 records
- Remel Inc · 2 events, 2 records
- Siemens Healthcare Diagnostics, Inc · 2 events, 2 records
- Siemens Medical Solutions USA, Inc · 2 events, 2 records
FDA root cause
- Labeling design · 24
- Process control · 10
- Nonconforming Material/Component · 9
- Device Design · 9
- Software design · 8
- Labeling Change Control · 5
- Component design/selection · 5
- Employee error · 4
Class I recalls (4)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Z-0226-2015 · Synthes (USA) Products LLC · posted Nov 26, 2014 · Terminated
Two nonconforming parts of the Titanium Polyaxial Reduction Head for Titanium Matrix Spine Screws may have been inadvertently packaged and shipped. It is believed that the inner diameter of the Reduction Head may not have been machined to the correct…
- Z-0170-2015 · Siemens Healthcare Diagnostics, Inc. · posted Nov 7, 2014 · Terminated
An increase in false positive susceptible results on Rapid Neg BP Combo Panel Type 3 and Rapid Neg Urine Combo Panel Type 1 used with MicroScan Microbiology Systems.
- Z-0171-2015 · Siemens Healthcare Diagnostics, Inc. · posted Nov 7, 2014 · Terminated
An increase in false positive susceptible results on Rapid Neg BP Combo Panel Type 3 and Rapid Neg Urine Combo Panel Type 1 used with MicroScan Microbiology Systems.
- Z-0125-2015 · Trivascular, Inc · posted Nov 3, 2014 · Terminated
TriVascular has identified select lots of Ovation Prime 29mm Aortic Body stent grafts that may be associated with rapid emptying of the fill polymer syringe when it is connected to the Autoinjector during the fill polymer injection step of the primary…
Class II recalls (79)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Z-0471-2015 · Toshiba American Medical Systems Inc · posted Nov 26, 2014 · Terminated
Potential problem with the cardiac function analysis software (CFA). Incorrect analysis results may be displayed in the Function Parameters for the Entire Heart displayed as analysis results of the CFA and in a Left-Ventricular Volume Curve generated…
- Z-0472-2015 · Toshiba American Medical Systems Inc · posted Nov 26, 2014 · Terminated
Potential problem with the cardiac function analysis software (CFA). Incorrect analysis results may be displayed in the Function Parameters for the Entire Heart displayed as analysis results of the CFA and in a Left-Ventricular Volume Curve generated…
- 8MM,POTTS SCISSORS ↗Class IIZ-0427-2015 · Intuitive Surgical, Inc. · posted Nov 25, 2014 · Terminated
Deviations in reprocessing steps from those stated in the reprocessing instructions, can cause surface degradation of the housing; and/or, accelerate mechanical wear of the instrument.
- 8MM,LARGE NEEDLE DRIVER ↗Class IIZ-0428-2015 · Intuitive Surgical, Inc. · posted Nov 25, 2014 · Terminated
Deviations in reprocessing steps from those stated in the reprocessing instructions, can cause surface degradation of the housing; and/or, accelerate mechanical wear of the instrument.
- Z-0429-2015 · Intuitive Surgical, Inc. · posted Nov 25, 2014 · Terminated
Deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface degradation of the housing and/or accelerate mechanical wear of the instrument.
- 8MM,LONG TIP FORCEPS ↗Class IIZ-0430-2015 · Intuitive Surgical, Inc. · posted Nov 25, 2014 · Terminated
Deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface degradation of the housing and/or accelerate mechanical wear of the instrument.
- 8MM,MICRO BIPOLAR FORCEPS ↗Class IIZ-0431-2015 · Intuitive Surgical, Inc. · posted Nov 25, 2014 · Terminated
Deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface degradation of the housing and/or accelerate mechanical wear of the instrument.
- 8MM,PROGRASP FORCEPS ↗Class IIZ-0432-2015 · Intuitive Surgical, Inc. · posted Nov 25, 2014 · Terminated
Deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface degradation of the housing and/or accelerate mechanical wear of the instrument.
- 8MM,MARYLAND BIPOLAR FORCEPS ↗Class IIZ-0433-2015 · Intuitive Surgical, Inc. · posted Nov 25, 2014 · Terminated
Deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface degradation of the housing and/or accelerate mechanical wear of the instrument.
- 8MM MONOPOLAR CURVED SCISSORS ↗Class IIZ-0434-2015 · Intuitive Surgical, Inc. · posted Nov 25, 2014 · Terminated
Deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface degradation of the housing and/or accelerate mechanical wear of the instrument.
- 8MM RESANO FORCEPS, ↗Class IIZ-0435-2015 · Intuitive Surgical, Inc. · posted Nov 25, 2014 · Terminated
Deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface degradation of the housing and/or accelerate mechanical wear of the instrument.
- 8MM PERMANENT CAUTERY HOOK, ↗Class IIZ-0436-2015 · Intuitive Surgical, Inc. · posted Nov 25, 2014 · Terminated
Deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface degradation of the housing and/or accelerate mechanical wear of the instrument.
- 8MM PERMANENT CAUTERY SPATULA, ↗Class IIZ-0437-2015 · Intuitive Surgical, Inc. · posted Nov 25, 2014 · Terminated
Deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface degradation of the housing and/or accelerate mechanical wear of the instrument.
- Z-0438-2015 · Intuitive Surgical, Inc. · posted Nov 25, 2014 · Terminated
Deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface degradation of the housing and/or accelerate mechanical wear of the instrument.
- 8MM, TENACULUM FORCEPS ↗Class IIZ-0439-2015 · Intuitive Surgical, Inc. · posted Nov 25, 2014 · Terminated
Deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface degradation of the housing and/or accelerate mechanical wear of the instrument.
- 8MM, CARDIAC PROBE GRASPER ↗Class IIZ-0440-2015 · Intuitive Surgical, Inc. · posted Nov 25, 2014 · Terminated
Deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface degradation of the housing and/or accelerate mechanical wear of the instrument.
- Z-0441-2015 · Intuitive Surgical, Inc. · posted Nov 25, 2014 · Terminated
- Z-0442-2015 · Intuitive Surgical, Inc. · posted Nov 25, 2014 · Terminated
Deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface degradation of the housing and/or accelerate mechanical wear of the instrument.
- 8MM, DUAL BLADE RETRACTOR ↗Class IIZ-0443-2015 · Intuitive Surgical, Inc. · posted Nov 25, 2014 · Terminated
Deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface degradation of the housing and/or accelerate mechanical wear of the instrument.
- 8MM, MEGA SUTURECUT ND ↗Class IIZ-0444-2015 · Intuitive Surgical, Inc. · posted Nov 25, 2014 · Terminated
Deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface degradation of the housing and/or accelerate mechanical wear of the instrument.
- 8MM, SMALL GRASPING RETRACTOR ↗Class IIZ-0445-2015 · Intuitive Surgical, Inc. · posted Nov 25, 2014 · Terminated
Deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface degradation of the housing and/or accelerate mechanical wear of the instrument.
- 8MM,MEDIUM-LARGE CLIP APPLIER ↗Class IIZ-0446-2015 · Intuitive Surgical, Inc. · posted Nov 25, 2014 · Terminated
Deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface degradation of the housing and/or accelerate mechanical wear of the instrument.
- 8MM, CURVED BIPOLAR DISSECTOR ↗Class IIZ-0447-2015 · Intuitive Surgical, Inc. · posted Nov 25, 2014 · Terminated
Deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface degradation of the housing and/or accelerate mechanical wear of the instrument.
- Z-0448-2015 · Intuitive Surgical, Inc. · posted Nov 25, 2014 · Terminated
Deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface degradation of the housing and/or accelerate mechanical wear of the instrument.
- Z-0449-2015 · Remel Inc · posted Nov 25, 2014 · Terminated
The product may appear cloudy or discolored and may not perform correctly in presumptive diagnosis of mycobacterial disease.
- Abbott m2000sp, an automated system for performing sample preparation for nucleic acid testing.Class IIZ-0463-2015 · Abbott Molecular · posted Nov 25, 2014 · Terminated
Some versions of Application Specifications (App Spec) are incompatible with m2000sp system software version 6.0 and 7.0. This issue will cause Error Code 9000 (An unexpected error has occurred: Software error) to be generated when the Available Reagent…
- Z-0421-2015 · Hospira Inc. · posted Nov 24, 2014 · Terminated
Some patients are obtaining access to the dosage reset codes required to change the settings on GemStar infusion pumps and are subsequently changing the infusion rate set by their doctors. Patients are obtaining access to these codes via website sponsors…
- Z-0422-2015 · Hospira Inc. · posted Nov 24, 2014 · Terminated
Some patients are obtaining access to the dosage reset codes required to change the settings on GemStar infusion pumps and are subsequently changing the infusion rate set by their doctors. Patients are obtaining access to these codes via website sponsors…
- Z-0423-2015 · Zimmer, Inc. · posted Nov 24, 2014 · Terminated
To provide clarifying instructions relating to the assembly/disassembly of the Trabecular Metal Tibial Impactor with/from TM tibial Impactor Pads due to complaints regarding Tibal Impactor pad fracture.
- Z-0424-2015 · Zimmer, Inc. · posted Nov 24, 2014 · Terminated
To provide clarifying instructions relating to the assembly/disassembly of the Trabecular Metal Tibial Impactor with/from TM tibial Impactor Pads due to complaints regarding Tibal Impactor pad fracture.
- Z-0425-2015 · Discus Dental LLC · posted Nov 24, 2014 · Terminated
Product sold after April 1, 2014 were inadvertently filled with hydrogen peroxide teeth whitening gel (6% concentration as dispensed) on one side of the dual barrel syringe instead of the intended component (phosphate gel). Consequently, use of this…
- Z-0418-2015 · Materialise USA LLC · posted Nov 21, 2014 · Terminated
The box of guides for a specific case arrived intact but contained two (2) femoral guides instead of one (1) femoral guide and one (1) tibial guide. The device did not meet specification: One tibia guide missing and one femur guide contained an edge which…
- Z-0420-2015 · Siemens Healthcare Diagnostics, Inc. · posted Nov 21, 2014 · Terminated
Certain lots of the reagent cartridges show confirmed positive bias averaging 0.4% Hemoglobin A1c units and occasionally up to 1.0% HbA1c units for patient samples when compared to the National Glycohemoglobin Standardization Program (NGSP). QC samples…
- Z-0235-2015 · Siemens Medical Solutions USA, Inc · posted Nov 20, 2014 · Terminated
Siemens became aware of a very rare occurrence where an intended movement of the SimView NT simulation system components could not be stopped with usual hand control commands or by use of the collision detection functionality. It is possible that…
- Z-0416-2015 · Owens & Minor Distribution, Inc. · posted Nov 20, 2014 · Terminated
Product wheels may be too loose, too tight, or fall off.
- Z-0227-2015 · Monteris Medical Corp · posted Nov 19, 2014 · Terminated
Monteris Medical received a complaint of a separation of the distal sapphire tip from the laser delivery probe.
- Z-0228-2015 · Monteris Medical Corp · posted Nov 19, 2014 · Terminated
Monteris Medical received a complaint of a separation of the distal sapphire tip from the laser delivery probe.
- Z-0225-2015 · Medacta Usa · posted Nov 18, 2014 · Terminated
The MectaLIF Oblique Handle has the laser marking "MEDIAL" on the incorrect side and could result in incorrect insertion of the device during surgery.
- Z-0222-2015 · Nidek Inc · posted Nov 17, 2014 · Terminated
Retrospective review found that a Engineering Change Order initiated in May 2013 for a memory function problem was not reported as required.
- Z-0223-2015 · Nidek Inc · posted Nov 17, 2014 · Terminated
Contact failure in F1 and F4 fuse holders was detected on the EC-5000 CXIII Excimer Laser Corneal Surgery System that could cause a serious adverse event for the patient.
- Z-0224-2015 · Nidek Inc · posted Nov 17, 2014 · Terminated
When the MC-500 is turned on and coagulation is performed without changing the laser color, the delivery unit occasionally fails to emit the laser, or the power output is lower than the set program.
- MOSAIQ ↗Class IIZ-0221-2015 · Elekta, Inc. · posted Nov 15, 2014 · Terminated
A problem can exist in MOSAIQ resulting in the display of incorrect numeric data due to a dose rounding error on printed reports.
- Z-0216-2015 · Silver Star Brands · posted Nov 14, 2014 · Terminated
Silver Star Brands is recalling the Transport Chair and Rollator in One because they have received several reports of the wheel assembly failing resulting in damaged wheels..
- Z-0217-2015 · Edwards Lifesciences, LLC · posted Nov 14, 2014 · Terminated
Edwards Lifesciences is conducting a field action on the Vent Catheters due to the potential for disruption of the white plastic coating on the stylet.
- Z-0218-2015 · Accuray Incorporated · posted Nov 14, 2014 · Terminated
Software upgrade to correct potential safety issue related to CyberKnife System that occurs when upgrading the Treatment Delivery Software for the first generation Iris Variable Aperture Collimator.
- Z-0219-2015 · Synthes, Inc. · posted Nov 14, 2014 · Terminated
The 10 mm/135 degree Titanium Cannulated Trochanteric Fixation Nail 170 MM (Sterile) was assembled incorrectly. If the non conformance goes undetected, it may impact the locking function post-operatively and compromise the bone reduction and construct…
- Z-0205-2015 · Medtronic Vascular · posted Nov 13, 2014 · Terminated
Two reports have been received on the 6F launcher Guide Catheter lot (Model Number LA6JR40K; lot No 0007244836} for incorrect usable length of the Guide Catheter. It was reported that the catheters had a usable length of 90 cm instead of 55 cm, which was…
- AMS 700 Series Inflatable Penile Prosthesis Accessory Kit (AMS 700 Line Accessory Kit), 72401850. ↗Class IIZ-0209-2015 · American Medical Systems, Inc. · posted Nov 13, 2014 · Terminated
AMS' supplier has issued a recall of certain lots of Lightning Bolt Keith needles supplied in AMS 700 Accessory Kits. Based on this, AMS is recalling AMS 700 Accessory Kits which contain these lots of needles. The supplier made the decision to issue this…
- Z-0210-2015 · Fisher & Paykel Healthcare, Ltd. · posted Nov 13, 2014 · Terminated
Breakage of the nut that secures the heater head in place, which can cause the heater head to become partially detached and swing towards the bassinet. There is the potential that a heater head could contact a patient situated on the bassinet and cause a…
- Cardiohelp-I SystemClass IIZ-0211-2015 · Maquet Cardiovascular Us Sales, Llc · posted Nov 13, 2014 · Terminated
Cardiohelp- i System may have a software issue that can potentially result in an erroneous display of a "Battery Needs Service" message after startup of the units when using either AC or DC power
- Masimo SET uSpO2 Cable ↗Class IIZ-0213-2015 · Masimo Corporation · posted Nov 13, 2014 · Terminated
Firm has identified a small number of Oximetry Cables with crossed-internal wires. The effect on performance when these wires are crossed is that when a low SpO2 value is measured, a high SpO2 value would be displayed on the monitoring device.…
- Z-0202-2015 · Medacta Usa · posted Nov 12, 2014 · Terminated
Medacta has registered in its database 3 cases of breakage of the GMK Intramedullary Extension Rod 100 mm, code 02.07.10.0162.
- Z-0204-2015 · Roscoe Medical Inc · posted Nov 12, 2014 · Terminated
The caster may separate completely from the device frame causing injury to user.
- Z-0199-2015 · DePuy Orthopaedics, Inc. · posted Nov 10, 2014 · Terminated
Issuing a device correction because if the pin bushing is over loaded or inverted, there is the potential for partial or full displacement and fracture of the plastic around the pin bushing.
- Z-0200-2015 · Remel Inc · posted Nov 10, 2014 · Terminated
Recalled product may not provide adequate recovery of Legionella pneumophila when used as directed and may result in false negative identification of Legionella pneumophila.
- Z-0201-2015 · Integra LifeSciences Corp. · posted Nov 10, 2014 · Terminated
Integra LifeSciences Corporation received 2 complaints that the USB port on the Licox Pt02 monitor does not consistently provide the user the ability to extract the Pt02 trend data according to the User's Manual for the device.
- Integra Titan Reverse Shoulder System right and left head cutting templates, 142.5 degrees. ↗Class IIZ-0197-2015 · Integra LifeSciences Corp. · posted Nov 8, 2014 · Terminated
A single lot of left and a single lot of right Reverse Shoulder System cutting templates were manufactured incorrectly. Specifically, the threaded handle has been welded backwards on the template resulting in the cutting angle being the reverse of what it…
- Z-0198-2015 · Zimmer, Inc. · posted Nov 8, 2014 · Terminated
Cleaning process validation failure.
- Philips Medical Systems Allura Xper FD20C Radiological Imaging with software version R8.2.O ↗Class IIZ-0181-2015 · Philips Medical Systems, Inc. · posted Nov 7, 2014 · Terminated
Faulty Automatic Motion Controller (AMC), a problem in the Power On Self Test (POST) error handling was detected, can result in a hazardous movement of the C-arc. system.
- Z-0182-2015 · Philips Medical Systems, Inc. · posted Nov 7, 2014 · Terminated
Faulty Automatic Motion Controller (AMC), a problem in the Power On Self Test (POST) error handling was detected, can result in a hazardous movement of the C-arc. system.
- Z-0183-2015 · Siemens Medical Solutions USA, Inc · posted Nov 7, 2014 · Terminated
A cable inside the C-arm systems may have been routed in an inappropriate manner. Improper routing of the cable may result in increased wear over time. Without additional measures, the potential exists for a cable inside the C-arm to break which could…
- Bard PerFix Light Plug ↗Class IIZ-0191-2015 · Davol, Inc., Subs. C. R. Bard, Inc. · posted Nov 7, 2014 · Terminated
The patient record peel-off label product code and size did not match the other labeling and actual product in the packaging for the Bard Small PerFix Light Plug.
- OEC UroView 2800. ↗Class IIZ-0192-2015 · GE OEC Medical Systems, Inc · posted Nov 7, 2014 · Terminated
GE Healthcare initiated a field correction because prior revision of components were potentially installed on certain OEC 9800, FlexiView 8800, OEC MiniView 6800 and OEC UroView 2800 systems.
- OEC MiniView 6800. ↗Class IIZ-0193-2015 · GE OEC Medical Systems, Inc · posted Nov 7, 2014 · Terminated
GE Healthcare initiated a field correction because prior revision of components were potentially installed on certain OEC 9800, FlexiView 8800, OEC MiniView 6800 and OEC UroView 2800 systems.
- Z-0194-2015 · GE OEC Medical Systems, Inc · posted Nov 7, 2014 · Terminated
GE Healthcare initiated a field correction because prior revision of components were potentially installed on certain OEC 9800, FlexiView 8800, OEC MiniView 6800 and OEC UroView 2800 systems.
- Series OEC 9800. ↗Class IIZ-0195-2015 · GE OEC Medical Systems, Inc · posted Nov 7, 2014 · Terminated
GE Healthcare initiated a field correction because prior revision of components were potentially installed on certain OEC 9800, FlexiView 8800, OEC MiniView 6800 and OEC UroView 2800 systems.
- Z-0196-2015 · Steris Corporation · posted Nov 7, 2014 · Terminated
The 5085 and 5085SRT stainless steel column shrouds may become damaged as a result of misalignment during extreme table articulation, excessive side force on the table column, or pressure on the base of the table. Severe damage to the shrouds may prevent…
- Z-0037-2015 · Richard Wolf Medical Instruments Corp. · posted Nov 6, 2014 · Terminated
Richard Wolf Medical Instrument Corporation is recalling Hulka Clip Tubal Occluding Device lot 51000232. The devices are being recalled as a part of an investigation into a biocompatibility test discrepancy where the recalled lot (lot 51000232) passed…
- Z-0174-2015 · Straumann USA, LLC · posted Nov 6, 2014 · Terminated
Product transfer piece was fitted with a Narrow Connect (NC) instead of a Regular Connect (RC)
- Access 2 Immunoassay Systems, Catalog No. 81600N, 386220, Instructions for Use Part No. B14253A. ↗Class IIZ-0175-2015 · Beckman Coulter Inc. · posted Nov 6, 2014 · Terminated
Beckman Coulter is recalling the Small Form Factor Console PC (SFF PC) (B23083) for use with the Access 2 Immunoassay systems because it may experience a "MFC Exception" error during normal operation of the Access 2 Immunoassay Systems.
- Z-0176-2015 · Beckman Coulter Inc. · posted Nov 6, 2014 · Terminated
Beckman Coulter is recalling the Small Form Factor Console PC (SFF PC) (B23083) for use with the DxC 600i analyzer because it may experience a "MFC Exception" error during normal operation of the analyzer system.
- Z-0177-2015 · St Jude Medical · posted Nov 6, 2014 · Terminated
St. Jude Medical is performing a voluntary recall of batch 4671318, model number 401353 of the Response Fixed Curve DECA 6F 65cm CSL 2/8 Catheter. Specifically, the firm has identified that every device packaged within batch 4671318 contains the incorrect…
- Z-0178-2015 · Dokitz NovoSci · posted Nov 6, 2014 · Terminated
The single leg of the PAY connector has been observed to become detached.
- Z-0179-2015 · Leica Microsystems, Inc. · posted Nov 6, 2014 · Terminated
Bond Ready to Use Primary Antibody Thyroglobulin does not function as intended up to the expiry date on the product labelling. This was detected through an on-going stability program. There is a link between the age of the product and staining intensity.
- Z-0172-2015 · Stryker Instruments Div. of Stryker Corporation · posted Nov 5, 2014 · Terminated
The SmartLife Aseptic Housings are being recalled in that the top section of the housing may separate from the bottom section due to insufficient welding.
- Z-0173-2015 · Stryker Instruments Div. of Stryker Corporation · posted Nov 5, 2014 · Terminated
The SmartLife Aseptic Housings are being recalled in that the top section of the housing may separate from the bottom section due to insufficient welding.
- Z-0168-2015 · Iba Dosimetry Gmbh · posted Nov 4, 2014 · Terminated
Error in the software. A deviation between reconstructed and planned dose distribution may not be detected prior to treatment and this can result in an over or under-estimation of the pretreatment delivered dose.
- Z-0169-2015 · Cynosure, Inc. · posted Nov 4, 2014 · Terminated
If the laser system is turned on and the user adjusts, installs or removes the XPL pulsed light handpiece, there is a potential risk of an electrical shock
- Z-0166-2015 · Smiths Medical ASD, Inc. · posted Nov 1, 2014 · Terminated
Smiths Medical has identified an issue with a single batch (Lot Number 2752712) of CADD"-Solis Medication Safety Software Administrator 3.1 CDs. CADD"-Solis Medication Safety Software Administrator 3.1 CDs were sent with the CADD"-Solis Medication…
Class III recalls (10)
Using the device is not likely to cause health problems, but it violates FDA rules.
- Z-0229-2015 · Siemens Healthcare Diagnostics, Inc · posted Nov 19, 2014 · Terminated
Siemens Healthcare Diagnostics has confirmed that approximately 2000 ReadyPacks from the ADVIA Centaur Systems BNP reagent contains an ADVIA Centaur Systems FT4 human-readable side label.
- Etest Ceftaroline (CPT32) WW B30. ↗Class IIIZ-0230-2015 · Biomerieux · posted Nov 19, 2014 · Terminated
The products have a wrong expiration date on their labeling, 5 years instead of 2 years.
- Etest Ceftaroline (CPT32) WWB100. ↗Class IIIZ-0231-2015 · Biomerieux · posted Nov 19, 2014 · Terminated
The products have a wrong expiration date on their labeling, 5 years instead of 2 years.
- Etest Ceftaroline (CPT32) US B30. ↗Class IIIZ-0232-2015 · Biomerieux · posted Nov 19, 2014 · Terminated
The products have a wrong expiration date on their labeling, 5 years instead of 2 years.
- Etest Ceftaroline (CPT32) US B100. ↗Class IIIZ-0233-2015 · Biomerieux · posted Nov 19, 2014 · Terminated
The products have a wrong expiration date on their labeling, 5 years instead of 2 years.
- Etest Ceftaroline (CPT32) US B30. ↗Class IIIZ-0234-2015 · Biomerieux · posted Nov 19, 2014 · Terminated
The products have a wrong expiration date on their labeling, 5 years instead of 2 years.
- Z-0214-2015 · Siemens Healthcare Diagnostics, Inc · posted Nov 14, 2014 · Terminated
Stability of the ADVIA Centaur Cleaning Solution is not assured
- Z-0207-2015 · CooperSurgical, Inc. · posted Nov 13, 2014 · Terminated
Incorrect expiration date printed on the MILEX PESSARY KIT packaging
- CooperSurgical MILEX PESSARY KITS-Class IIIZ-0208-2015 · CooperSurgical, Inc. · posted Nov 13, 2014 · Terminated
Incorrect expiration date printed on the MILEX PESSARY KIT packaging
- IMMULITE Calcitonin Control Module (IMMULITE, IMMULITE 1000, IMMULITE 2000, IMMULITE 2000 XPi)Class IIIZ-0212-2015 · Siemens Healthcare Diagnostics, Inc. · posted Nov 13, 2014 · Terminated
A typographical error regarding the expiration date was identified in the control module instructions for use.