Siemens Dimension HB1C Flex reagent cartridge recallZ-0420-2015 · Siemens Healthcare Diagnostics, Inc. · initiated Sep 30, 2014

Class II Terminated All recalls posted in November 2014

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Certain lots of the reagent cartridges show confirmed positive bias averaging 0.4% Hemoglobin A1c units and occasionally up to 1.0% HbA1c units for patient samples when compared to the National Glycohemoglobin Standardization Program (NGSP). QC samples may exhibit a similar bias.

FDA root cause category: Nonconforming Material/Component

What the company is doing

Siemens sent an Urgent Medical Device Correction letter dated September 2014 was provided to all Dimension Vista HbA1c and Dimension HB1C customers who received the affected lots to notify them of the issue, the potential risk to health, and actions to be taken by customers. Customers were instructed to discontinue use of and discard the affected devices. Customers need to review the letter with their Medical Director. Siemens will replace any unused inventoryof the affected lots at no charge. Customers were asked to complete the attached form and fax it to 302-631-7795. Customers should retain a copy of the letter with their laboratory records, and forward the letter to those who may have received this product. Customers with questions should contact their Siemens Customer Care Center - Technical Solutions or their local Siemens technical support representative. For questions regarding this recall call 800-441-9250.

Recall details

Recall number
Z-0420-2015
Recall class
Class II
Status
Terminated
Initiated by the company
Sep 30, 2014
Posted by the FDA
Nov 21, 2014
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
53, 046
Product code
LCP
FDA submissions
K102045, K102510
Distribution
Worldwide Distribution: US (nationwide) and internationally to: Canada.

Product description

Siemens Dimension HB1C Flex reagent cartridge; Dimension Hemoglobin A1c (HB1C) Flex reagent cartridge on the Dimension clinical chemistry system is an in vitro diagnostic assay for the quantitative determination of hemoglobin A1c (HbA1c) in human anticoagulated whole blood. Measurements of hemoglobin A1c are effective in monitoring long-term glucose control in individuals with diabetes mellitus.

Code information (models, lots, serial numbers)

Siemens Dimension Vista HbA1c Flex reagent cartridge: Catalog Number K3105A, Siemens Material Number (SMN) 10470481, Lot Numbers 13294BA 13308BA 13329AA 13350AA 14006AA 14041BA 14069BB 14083BA   Siemens Dimension HB1C Flex reagent cartridge: Catalog Number DF105A, Siemens Material Number (SMN) 10483822, Lot Numbers GA4266 BA4273 BA4280 BA4287 GA4301 GA4315 GC4322 GA4343 GA4350 GA4357 GA5013 GA5020

Other devices under the same FDA submission

1 device cleared under K102045, K102510 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.