Align Radial Stems, Prosthesis, Elbow, Hemi recallZ-0582-2015 · Skeletal Dynamics · initiated Oct 16, 2014
Class II Terminated All recalls posted in December 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Skeletal Dynamics or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Report received where the Align Radial Stem fractured.
FDA root cause category: Device Design
What the company is doing
Skeletal Dynamic sent an Urgent Medical Device Safety Alert dated October 16, 2014, to all affected customers via certified mail. The Safety Alert included the warning "that radial head prostheses cannot be expected to withstand the activity levels and loads of normal healthy bone and joint tissue. Failure of the component can occur as a result of loss of fixation, strenuous activity, malalignment, trauma, non-union or excessive loads (estimated body weight equivalent of 350 lbs or greater)" with instructions to notify any surgeons that have used the product. A response form has been included with instructions to return the signed form to acknowledge the receipt and understanding of the new warning. Product labeling has been revised to include these labels in future shipments. Customers with questions were instructed to contact Customer Service at 877-753-5396 for assistance. For questions regarding this recall call 305-596-7585, ext 7026.
Recall details
- Recall number
- Z-0582-2015
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Oct 16, 2014
- Posted by the FDA
- Dec 12, 2014
- Recalling firm
- Skeletal Dynamics
- Quantity
- 1700
- Product code
- KWI
- FDA submissions
- K092721
- Distribution
- Nationwide Distribution including GA, NJ, NM, IL, CA, OK, CO, NY, LA, OH, FL, VA, MD, MS, TX, IA, AL, WA, TN, ME, CT, WI, AZ, NE, SC, NC and Puerto Rico.
Product description
Align Radial Stems, Prosthesis, Elbow, Hemi Replacement of the radial head for degenerative or post-traumatic disabilities presenting pain, crepitation, and decreased motion at the radio-humeral and/or proximal radio-ulnar joint with: Joint destruction and/or subluxation and resistance to conservative treatment. Intended use is also for primary replacement after fracture of the radial head and symptomatic sequelae after radial head resection.
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Cat #s: ALN-RST-0700, ALN-RST-0702, ALN-RST-0704, ALN-RST-0706, ALN-RST-0708, ALN-RST-0800, ALN-RST-0802, ALN-RST-0802, ALN-RST-0804, ALN-RST-0806, ALN-RST-0808, ALN-RST-0900, ALN-RST-0902, ALN-RST-0904, ALN-RST-0906, ALN-RST-0908, ALN-RST-1000, ALN-RST-1002, ALN-RST-1004, ALN-RST-1006, ALN-RST-1008, ALN-RST-1100, ALN-RST-1102, ALN-RST-1104, ALN-RST-1106, ALN-RST-1108; Lot #s: AN0610023, AN0212065, AN0913033, AN0913034, AN0610024, AN0913035, AN0913036, AN0610025, AN0913037, AN0913038, AN0610026, AN0913039, AN0913040, AN0610027, AN0913041, AN0913042, AN0610028, AN0212066, AN0913043, AN0913044, AN0610029, AN0913045, AN0913046, AN0610030, AN0913047, AN0913048, AN0610031, AN0913049, AN0913050, AN0610032, AN0913051, AN0913052, AN0610033, AN0212067, AN0913053, AN0913054, AN0610034, AN0913055, AN0913056, AN0610035, AN0913057, AN0913058, AN0610036, AN0913059, AN0913060, AN0610037, AN0913061, AN0913062, AN0610038, AN0212068, AN0913063, AN0913064, AN0610039, AN0212069, AN0913065, AN0913066, AN0610040, AN0913067, AN0913068, AN0610041, AN0913069, AN0913070, AN0410008, AN0610042, AN0410009, AN0913071, AN0913072, AN0610043, AN0212070, AN0913073, AN0913074, AN0610044, AN0913075, AN0913076, AN0610045, AN0913077, AN0913078, AN0610046, AN0913079, AN0913080, AN0610047, AN0913081, AN0913082.
Other devices under the same FDA submission
6 devices cleared under K092721 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Align ALIGN Radial Head & Lock Screw, 24mmSkeletal Dynamics L.L.C. · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER9 reports
- Align ALIGN Radial Head & Lock Screw, 26mmSkeletal Dynamics L.L.C. · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER3 reports
- Align ALIGN Radial Stem, 9mm x 0mmSkeletal Dynamics L.L.C. · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER2 reports
- Align ALIGN Radial Stem, 7mm x 0mmSkeletal Dynamics L.L.C. · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 report
- Align ALIGN Radial Stem, 8mm x 0mmSkeletal Dynamics L.L.C. · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 report
- Align ALIGN Radial Stem, 10mm x 0mmSkeletal Dynamics L.L.C. · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.