10 mm/135 degree Titanium Cannulated Trochanteric Fixation… recallZ-0219-2015 · Synthes, Inc. · initiated Oct 29, 2014

Class II Terminated All recalls posted in November 2014

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Synthes, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The 10 mm/135 degree Titanium Cannulated Trochanteric Fixation Nail 170 MM (Sterile) was assembled incorrectly. If the non conformance goes undetected, it may impact the locking function post-operatively and compromise the bone reduction and construct stability which may lead to non-union/malunion.

FDA root cause category: Employee error

What the company is doing

An urgent medical device recall notice, dated October 29, 2014, was sent to end users that identified the product, problem and action to be taken. Customers were instructed to return the device to Synthes along with completing the verification section of the notice.

Recall details

Recall number
Z-0219-2015
Recall class
Class II
Status
Terminated
Initiated by the company
Oct 29, 2014
Posted by the FDA
Nov 14, 2014
Recalling firm
Synthes, Inc.
Quantity
6
Product code
HSB
FDA submissions
K011857
Distribution
US Distribution in the states of: AK, FL, OH, OK, and TX.

Product description

10 mm/135 degree Titanium Cannulated Trochanteric Fixation Nail 170 MM - Sterile: intended to treat stable and unstable fractures of the proximal femur

Code information (models, lots, serial numbers)

part number: 456.316S, lot number: 7782247