POWEREASE Instruments Set Screw Breakoff Instrument, REF… recallZ-0952-2015 · Medtronic Sofamor Danek USA Inc · initiated Dec 3, 2014
Class II Terminated All recalls posted in January 2015
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Sofamor Danek USA Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The retaining tabs component of the Set Screw Retaining Drive may be oversized, which will not allow proper mating between the Set Screw Retaining Drive and the set screw.
FDA root cause category: Process design
What the company is doing
Medtronic sent an Urgent - Voluntary Market Removal letter dated December 3, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Instructions to Customers: 1.) Please remove the product from normal storage locations. Do not use this product. 2.) As applicable, contact your Medtronic Sales Representative to coordinate the return of affected product. Alternatively, you may call Medtronic's Global Quality Department at (800) 876 -3133, extension 3197 for assistance If you need assistance or have any related questions or concerns, please do not hesitate to contact your Medtronic Sales Representative or the Global Quality Department at (800) 876 - 3133, extension 3197. We sincerely apologize for this inconvenience, but this action is necessary to assure that our high standard of quality is maintained.
Recall details
- Recall number
- Z-0952-2015
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Dec 3, 2014
- Posted by the FDA
- Jan 12, 2015
- Recalling firm
- Medtronic Sofamor Danek USA Inc
- Quantity
- 24 units
- Product code
- HWE
- FDA submissions
- K111520
- Distribution
- US Distribution to the states of : CA, KS, KY, NC, TX, WA, MI, OH, IA, PA, OR, MS, FL, NY, GA and AR.
Product description
POWEREASE Instruments Set Screw Breakoff Instrument, REF 2346010, Medtronic Sofamor Danek, USA, Inc. Spinal surgical instrumrent.
Code information (models, lots, serial numbers)
Lot # - CA14C107
Other devices under the same FDA submission
9 devices cleared under K111520 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- POWEREASE™ Instruments 2300000Medtronic Sofamor Danek, Inc. · STIMULATOR, ELECTRICAL, EVOKED RESPONSE1 recall203 reports
- POWEREASE™ Instruments 2342306MMedtronic Sofamor Danek, Inc. · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT80 reports
- POWEREASE™ Instruments 2342281MMedtronic Sofamor Danek, Inc. · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT19 reports
- POWEREASE™ Instruments 2342306SMedtronic Sofamor Danek, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)8 reports
- POWEREASE™ Instruments 2342301MMedtronic Sofamor Danek, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)8 reports
- POWEREASE™ Instruments 2342305MMedtronic Sofamor Danek, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)8 reports
- POWEREASE™ Instruments 2342301SMedtronic Sofamor Danek, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)4 reports
- POWEREASE™ Instruments 2343431MMedtronic Sofamor Danek, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)1 report
- Indigo™ 31132080Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page