GE Healthcare, Proteus XR/a with wall stand model number… recallZ-0954-2015 · GE Healthcare · initiated Dec 8, 2014
Class II Terminated All recalls posted in January 2015
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from GE Healthcare or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
GE Healthcare has recently become aware of a potential safety issue involving the breakage of the two steel cables which support the bucky device in the wall stand of Proteus XR/a X-ray imaging systems. No injuries reported.
FDA root cause category: Nonconforming Material/Component
What the company is doing
Consignees were sent on 12/8/2014 a GE Healthcare "Urgent Medical Device Correction" letter GEHC Ref# 17107, dated December 8, 2014. The letter was addressed to Hospital Administrators /Risk Managers, Managers of Radiology / Cardiology and Radiologists / Cardiologists. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product correction and Contact information (For questions in USA call 800-437-1171, For other countries contact your local GE Healthcare Service Representative.
Recall details
- Recall number
- Z-0954-2015
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Dec 8, 2014
- Posted by the FDA
- Jan 12, 2015
- Recalling firm
- GE Healthcare
- Quantity
- 3096 (2486 US, 610 OUS).
- Product code
- KPR
- FDA submissions
- K993090
- Distribution
- USA: All states in continental US including, Guam, DC. OUS; ARAB EMIRATES, AUSTRIA, BOSNIA AND HERZEGOVINA, BRAZIL, BULGARIA, CAMBODIA, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, CZECH REPUBLIC, ECUADOR, EGYPT, FINLAND, FRANCE, GERMANY, GREECE, HONDURAS, HONG KONG, HUNGARY, INDONESIA, IRELAND, ISRAEL, ITALY, JAPAN, KAZAKHSTAN, KUWAIT, LATVIA, LITHUANIA, MALAYSIA, MEXICO, NETHERLANDS, NEW ZEALAND, NORWAY, PAKISTAN, PHILIPPINES, POLAND, POLYNESIA, PORTUGAL, RUSSIAN FEDERATION, SAUDI ARABIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, UKRAINE, UNITED KINGDOM, URUGUAY, VENEZUELA, VIETNAM.
Product description
GE Healthcare, Proteus XR/a with wall stand model number 600-0301.
Code information (models, lots, serial numbers)
Show the full code information (80k characters)
FMI 17107 - Proteus Wall Stand Counterpoise Cables Break Serial # System ID 00000044028HL9 928656ER 00000063303HL2 719595ORTHO 00000045940HL4 803751PR2 00000041478HL9 812238FPCRAD 00000057994HL6 208359PROT 00000046527HL8 310732TIPRO 00000047064HL1 318255PRO 00000036130HL3 319352PROT 00000053610HL2 336765WFBIRAD 00000021190HL4 360457ORTHO 00000037481HL9 405377PRO 00000031767HL7 406638CATOMO 00000034159HL4 407628PRO 00000031902HL0 409938TC1 00000074625HL5 419578RAD2 00000096491HL6 419636RAD 00000096684HL6 435563RAD 00000041757HL6 435586PROT 00000057200HL8 435658RAD 00000088306HL6 505983ORTHPROT 00000020347HL1 508698NEPATS 00000049345HL2 573882NSCRAD1 00000064320HL5 601788RAD1 00000048917HL9 601968MAERM1 00000016587HL8 602285PROT2 00000106226HL4 608366PRO 00000047063HL3 616752NWPROTEUS 00000033643HL8 623327XR3 00000059646HL0 630312PROT1 00000082840HL0 630856SCPRO 00000057436HL8 702396XR1 00000102867HL9 703771PROT 00000049772HL7 703784RAD1 00000067466HL3 706641CTC 00000024904HL5 708226PROT 00000041647HL9 708226RM2 00000062869HL3 757606PRO1 00000063690HL2 770252GBJ3 00000049128HL2 801IMCXR 00000036909HL0 804594RM4 00000112434HL6 805681RAD4 00000107242HL0 805681RAD5 00000027096HL7 814833PROT 00000046048HL5 818409RAD3 00000081590HL2 870245PROT 00000041480HL5 904264PRO 00000050595HL8 904381RAD 00000068296HL3 912279CCGA 00000062303HL3 912748PROT 00000035901HL8 918352DRAD1 00000061157HL4 MCCPROT1 00000061158HL2 MCCPROT2 103733HL2 918494XRTRAUMA2 00000020349HL7 210297NCRAD 00000044647HL6 252466CPRAD1 00000060369HL6 303269RM2 00000089696HL9 402294EBPROT 00000038564HL1 520202PRO1 00000038565HL8 520202PRO2 00000073629HL8 609878PRO1 00000075649HL4 609878PRO2 00000045941HL2 816282LSPRO1 00000078007HL2 817250RAD 00000061302HL6 856341PROT 00000059336HL8 203932PROT2 00000106552HL3 412232PROT1 00000019948HL9 812256RT1 00000045545HL1 662434RM2 00000046670HL6 808KWAJPROT 00000084213HL8 412315PROT1 00000050315HL1 318368PRO 00000111660HL7 276694PRO 00000068951HL3 256486XR1 00000058725HL3 931766ER1 00000031770HL1 662779RAD1 000000000000NA 706544MA2 00000104753HL9 0004177304A 00000018751HL8 201358ER1 00000042692HL4 201418ER 00000043018HL1 201418ER 00000023401HL3 201418RAD1 00000021897HL4 201418RAD3 00000060463HL7 201894PRO1 00000060864HL6 201894PRO2 00000112815HL6 203319PROT 00000065033HL3 203466PRO 00000016999HL5 203679RAD 00000058726HL1 203688C262 00000065367HL5 203688PROT2 00000074013HL4 203688PROT2 00000063510HL2 203739PROT 00000022750HL4 203797PROT2 00000028477HL8 203852RM5 00000057596HL9 203932PROT1 00000038056HL8 217333MXR1 00000028501HL5 217337CXR3 00000081384HL0 217366CWXR2 00000021397HL5 217366CXR1 00000034089HL3 217383CXR20 00000033499HL5 217383CXR24 00000044381HL2 217431CFAXR1 00000026066HL1 217431CXR1 00000062685HL3 217443UXR2 00000024374HL1 217732LPR1 00000017462HL3 217788PRT1 00000023841HL0 217824PRT 00000065364HL2 217854RAD1 00000052078HL3 217893CRA2 00000052238HL3 218229PROT 00000016191HL9 218759RAD 00000084370HL6 218773RAD 00000080250HL4 218828RAD1 00000061264HL8 218983RAD 00000017830HL1 219284ER 00000016991HL2 219347RAD 00000017822HL8 219492RAD1 00000018607HL2 219492RAD2 00000022756HL1 219757ER 00000028879HL5 219757WP1 00000111658HL1 219836PRO2 00000028783HL9 225761R5 00000033602HL4 228436XR1 00000057198HL4 228875PRXRA 00000041103HL3 229243RAD1 00000023267HL8 229249SSC 00000074884HL8 229257MR1 00000027844HL0 229259OUT 00000083631HL2 229259PR2 00000036119HL6 229276PROT 00000021034HL4 229386TER 00000057595HL1 229434PER 00000019691HL5 229890CR4 00000018991HL0 231547PROTEUS 00000065484HL8 231726PROTEUS2 00000064530HL9 231726PROTEUS3 00000028630HL2 231924ER 00000016391HL5 231924RM3 00000028785HL4 239254RAD 00000033660HL2 239368XR1 00000049934HL3 239403RAD 00000025617HL2 239432XR1 00000022389HL1 240379RAD 00000035861HL4 248652ER 00000025565HL3 248652RAD4 00000030083HL0 248858ER1 00000017618HL0 717733R1B 00000041407HL8 717761PRO1 00000017640HL4 717859CR1 00000016345HL1 718262RAD1 00000047004HL7 718470XRA 00000064049HL0 718960PROT 00000086110HL4 718960PROT 00000092087HL6 718960PROT1 00000021894HL1 719560GD1 00000063302HL4 719767PROTEUS 00000048109HL3 719846XR 00000084573HL5 719LVRAD1 00000074623HL0 724347PROT 00000102140HL1 724357RM1 00000019535HL4 724537PROT1 00000104749HL7 724588PROT 00000050598HL2 724626PROT1 00000079885HL0 724935UCC 00000016985HL4 724983PX 00000025564HL6 724MTVXR 00000023293HL4 727341ER 00000037716HL8 727341PPER1 00000037377HL9 727341RAD1 00000017622HL2 727521ER 00000042380HL6 727528RAD1 00000041883HL0 727588XRA5 00000047372HL8 731422RM2 00000023840HL2 731425RS 00000031156HL3 731425TRAUMA2 00000028102HL2 731425TRAUMA3 00000030902HL1 731584RAD 00000021837HL0 732222ER2 00000019787HL1 732294ER 00000076135HL3 732295RAD 00000035183HL3 732321ERP2 00000021657HL2 732321ERPRO 00000034453HL1 732499PROT5 00000032882HL3 732499PROT7 00000028975HL1 732530ERP1 00000028881HL1 732530ERP2 00000098078HL9 732840PRO5OP 00000015944HL2 732970RAD 00000016594HL4 734475ORTHO 00000021038HL5 734712PROT1 00000027091HL8 734712T2 00000027946HL3 734930UMED 00000018812HL8 740356PROT 00000023540HL8 740368PR2 00000021838HL8 740368PR4 00000080937HL6 740383PROT 00000106549HL9 740393PRO5 00000041829HL3 740393PROT1 00000050599HL0 740393PROT4 00000047232HL4 740420PRO 00000037887HL7 740454BHER 00000031439HL3 740454BHRM6 00000042861HL5 740454SZUC 00000036692HL2 740474PROT 00000028298HL8 740474R2 00000032399HL8 740592PROD 00000044743HL3 740779PRO1 00000043746HL7 740779PRO3 00000080251HL2 740845PRO 00000049771HL9 740941WRM1 00000030011HL1 757398PRO1 00000023173HL8 757827RM3 00000017631HL3 757934RM1 00000017636HL2 757934RM5 00000020776HL1 760241R2 00000093930HL6 760241RAD1 00000089699HL3 760876RAD 00000033672HL7 760934OP1 00000018990HL2 763236RAD 00000028177HL4 763520N2 00000062561HL6 763581PRO 00000037408HL2 763744PROT 00000029622HL8 763780R2 00000025415HL1 765455IUPROT 00000028290HL5 765455PROT 00000100702HL0 770228RM2 00000037888HL5 770267RM4 00000015868HL3 770389HRAD 00000061066HL7 770434PROT 00000040835HL1 770507EL1 00000075398HL8 770533LR1 00000020592HL2 770748PER 00000050596HL6 770918ER1 00000028505HL6 770918ER2 00000021901HL4 770918PRO1 00000020354HL7 772468ER1 00000016983HL9 773525TPR 00000028181HL6 773564WPRO1 00000028504HL9 773IMPRO11 00000066676HL8 775445PROTEUS 00000016599HL3 775738RAD 00000042078HL6 775982FERNLEY 00000016347HL7 775982XR1 00000029493HL4 775SMPROT 00000050602HL2 781581UR3 00000033000HL1 785239PRO 00000043331HL8 785239PRO2 00000046434HL7 785239PRO3 00000078011HL4 785354PROT 00000073846HL8 785354PROT2 00000085135HL2 785354PROT4 00000063305HL7 785357PROT 00000071486HL5 785452PRO 00000034921HL7 785632PRO 00000061519HL5 801256RAD 00000038331HL5 801268CHT 00000017824HL4 801268ER 00000040770HL0 801314RM1 00000063180HL4 801321OSG2 00000045155HL9 801387PRO1 00000062098HL9 801387PRO2 00000035370HL6 801461PROT 00000064816HL2 801461RM2 00000021658HL0 801501RAD 00000049455HL9 801535RM21 00000037748HL1 801535RM23 00000041581HL0 801562RM2 00000035180HL9 801634RM3 00000057038HL2 801639MI 00000059448HL1 801714XR 00000023176HL1 801774ER 00000051941HL3 801786XR 00000060982HL6 801840RM1 00000061156HL6 801840RM2 00000031152HL2 801855RM4 00000081379HL0 802775PRO 00000068442HL3 803502XR 00000013301HL5 803641ER 00000080462HL5 803641XR4 00000021498HL1 803699PR1 00000055683HL7 803713R1 00000057439HL2 803713R3 00000015955HL8 803791LR1 00000070369HL4 804254RMA 00000021898HL2 804541ER 00000024467HL3 804747DHPER 00000078170HL8 804784RAD1 00000024735HL3 805383XR1 00000089900HL5 805418XR1 00000032165HL3 805525OPXR3 00000069561HL9 805563RAD1 00000030961HL7 805648OPXR2 00000019784HL8 805648RM1 00000024370HL9 805681ORTHO 00000059447HL3 806250RAD 00000050592HL5 806647RAD 00000051286HL3 806647RAD 00000047230HL8 808257PROTRM1 00000053316HL6 808473NMCPRTRM1 00000053415HL6 808473NMCPRTRM3 00000024471HL5 808973PROT 00000051940HL5 810648PRO 00000045758HL0 810844AVANT 00000029205HL2 810985ER 00000019150HL2 812237TH2 00000048620HL9 812254ER 00000020772HL0 864459AR1 00000024660HL3 865218RAD1 00000023296HL7 865218RAD2 00000025118HL1 865218RAD3 00000105863HL5 870208RF 00000103416HL4 870208XR1 00000020861HL1 870423CR2 00000065590HL2 870541PROT6 00000031860HL0 901722RAD1 00000029158HL3 903232LRRT 00000042196HL6 903525RAD1 00000036747HL4 903614PROTEUS 00000062686HL1 903654RAD 00000034663HL5 904202CSRAD1 00000102864HL6 904202ER1 00000095427HL1 904202ER2 00000036321HL8 904276RM1 00000024555HL5 904391RAD1 00000026637HL9 904399ER1 00000061301HL8 904399RADWEST 00000068783HL0 904819WGRAD 00000034025HL7 904821CHEST1 00000111678HL9 906337PROT 00000031158HL9 906341RAD1 00000109254HL3 906632PRO 00000048621HL7 906932PROT 00000060865HL3 907580RM2 00000032396HL4 907762SXR 00000044312HL7 907796PROT 00000072092HL0 907874PROT 00000029230HL0 909336RAD1 00000018494HL5 909788R1 00000017637HL0 909788R2 00000067634HL6 910259PMRAD1 00000045157HL5 910343NHR5 00000062560HL8 910609HSPROT 00000025559HL6 910667BFRAD1 00000049454HL2 910907WARM2PROT 00000049456HL7 910907WARM2PROT 00000053614HL4 910907WARM8PRO 00000086589HL9 910938CHPROT 00000025485HL4 912275RM4 00000052232HL6 912283R209 00000104278HL7 912369PROT 00000038209HL3 912466ER05 00000024375HL8 912564PROT 00000056338HL7 912576PROTEUS 00000019534HL7 912691C02 00000040537HL3 912754PROT 00000078010HL6 912754PROT2 00000030012HL9 913588KER 00000028976HL9 913588KR1 00000039752HL1 913588KUIVP 00000025125HL6 913684CPRO 00000042560HL3 913684RM3 00000062199HL5 913ATCHPROT 00000029057HL7 914378PROTEUS 00000041885HL5 914666PRO 00000034685HL8 915521ERRAD 00000075268HL3 915629RAD 00000021840HL4 915735RAD1 00000023036HL7 915853RAD 00000024584HL5 916784PRO 00000028877HL9 916984PRO 00000048519HL3 918307SFSP1 00000077763HL1 918323PROT1 00000060066HL8 918453PROT1 00000084368HL0 918458PROT1 00000087774HL6 918505PROT 00000017004HL3 918540MRAD 00000076133HL8 918579PROTEUS 00000096318HL1 918689PROT1 00000028571HL8 918744TOMO1 00000020207HL7 918825PROTEUS 00000017827HL7 919350NWRAD1 00000019152HL8 919497PROT 00000032494HL7 919662RAD1 00000035099HL1 919663CRAD 00000030644HL9 919781PROT2 00000016408HL7 919781WRAD1 00000108819HL4 919787RAD3 00000035367HL2 919934JA1 00000035369HL8 919934JA2 00000080060HL7 919934JMSER 00000016815HL3 919934JRAD1 00000016737HL9 919934JRAD2 00000018292HL3 920262RAD1 00000018502HL5 920262RAD1 00000080939HL2 920457PROT 00000048357HL8 920622PROT 00000019228HL6 920682PR1 00000034709HL6 920787PROT 00000029881HL0 920926RAD1 00000024737HL9 925734RAD1 00000019881HL2 928348PRO 00000030185HL3 928639RM2 00000024746HL0 928645RM1 00000072091HL2 931388CPXR 00000086104HL7 931752RAD 00000028786HL2 931967RM2 00000089695HL1 936348MVPRO 00000041812HL9 936756RAD1 00000048919HL5 937237PRO1 00000027843HL2 937278ER1 00000027850HL7 937278ER2 00000056090HL4 937298PRO1 00000063512HL8 937298PRO1 00000034780HL7 937352RM1 00000084367HL2 937382PROT 00000061395HL0 937384PRO1 00000053411HL5 937458PRO1 00000032400HL4 937498PRO1 00000062562HL4 940297PROT 00000036469HL5 940384PRO 00000030271HL1 940384RAD 00000004255HL6 940676RM3 00000042455HL6 940676RM3 00000036912HL4 940676RM4 00000043071HL0 941473RM3 00000046761HL3 941629RM5 00000017140HL5 952403RAD 00000088065HL8 952758PRO 00000028631HL0 954201RAD1 00000028637HL7 954201RAD2 00000099705HL6 954355PROTEUS 00000060064HL3 954385RAD 00000028246HL7 954476RAD4 00000087139HL2 956632XREC 00000036352HL3 970242RMORAD 00000030177HL0 970479XR2 00000056596HL0 970624XR1 00000056735HL4 970624XR2 00000041736HL0 970668XRA 00000088856HL0 970675XRA 00000082844HL2 970945ORTHO 00000035366HL4 970945RAD 00000039521HL0 970BASRAD 00000019955HL4 970SILTRAD 00000083637HL9 972230PRO 00000025487HL0 972420LR4 00000028679HL9 972420LR5 00000095074HL1 972566PEDIPRO 00000032207HL3 972566PROER1 00000016343HL6 972566PTOMO 00000032206HL5 972566RAD8 00000017152HL0 972596OPRAD1 00000017466HL4 972596OPRAD1 00000018295HL6 972596PER3 00000046525HL2 972875RAD 00000029127HL8 973365PROTEUS 00000023178HL7 973908RM8 00000083036HL1 973971ARI 00000083636HL1 973971ARI 00000023106HL8 973971OP 00000019157HL7 978249ER1 00000045523HL8 978355PRD 00000042477HL0 978521MVRAD2 00000034841HL7 979764PROT 00000045525HL3 979830BCRAD1 00000057996HL1 979865PRO2 00000032299HL0 985230DCRAD1 00000108376HL5 985230SCRAD1 00000027941HL4 985646RAD2 00000054532HL7 985646RAD3 00000060863HL8 985652ER 00000106224HL9 985867RAD1 00000077526HL2 989224PROTEUS 00000047008HL8 989723RAD 00000028378HL8 989839RAD 00000059952HL2 ALLIANCEXR113 00000114260HL3 FOU719390RAD 00000022588HL8 FWR4112 00000023839HL4 FWR4113 00000023872HL5 FWR4113 00000060273HL0 HUGHESPRO1 00000087775HL3 LMICPRAD1 00000087393HL5 MH589RAD1 00000029686HL3 UVRMCRM2 115529HL0 817BSFWPRO 116076HL1 618436R1 113347HL9 817FCARLPRO 89699HL3 760876RAD 117111HL5 850623NWFXR 1-165721296 519735PRO 114991HL3 303341XR 115530HL8 605842RAD MISSING_SN_GON4219475 706335PRO1 00000102144HL3 870226PROT 00000103729HL0 0004034785 00000018689HL0 201358ER2 00000075125HL5 203453PROT 00000061517HL9 203688PROTLW 00000060398HL5 208882QCX1 00000044410HL9 210590OWRAD 00000041617HL2 210616PRO 00000045704HL4 210617NEI1 00000086106HL2 210703RAD1 00000019768HL1 212570RAD 00000033899HL6 217443UXR7 00000107243HL8 219836PRO1 00000037153HL4 248615PROT 00000015948HL3 248858ER 00000018690HL8 248937RM3 00000017463HL1 251435I02 00000052564HL2 251847RAD 00000023295HL9 252482CR2 00000035579HL2 252757ORTHO1 00000036131HL1 252757ORTHO3 00000034924HL1 253597XRM1 00000097810HL6 253680PROTEUS 00000035973HL7 253735MS1 00000075650HL2 253815PROT 00000018495HL2 256705XR1 00000028873HL8 256974PR1 00000050245HL0 256990FPR1 00000039750HL5 260373RM1 00000041261HL9 260426RAD 00000027944HL8 262249PROT 00000015888HL1 262268RAD1 00000017464HL9 262387RAD1 00000036992HL6 262642PROT 00000092360HL7 262646PRO 00000018603HL1 262741PR1 00000025421HL9 262741PR2 00000031794HL1 262928ER2 00000054166HL4 262948PRO1 00000061520HL3 269983PRO 00000025486HL2 270745RAD1 00000027947HL1 270745TOMO 00000017823HL6 276546PRORAD 00000019161HL9 276546PRORAD 00000020947HL8 276546PRORAD 00000071636HL5 276628PRO 00000019538HL8 281338RAD 00000106554HL9 281379FCYOPT 00000087392HL7 281395KATYPRO 00000093645HL0 281450FCP 00000064326HL2 281461PROT 00000057195HL0 281566ALCPRO 00000070005HL4 281576PROT 00000029233HL4 281599PRO 00000032395HL6 281837BPROD1 00000075126HL3 281923PWPRO1 00000047001HL3 281929PRORAD 00000076313HL6 281955RAD 00000100471HL2 281FCDPPROT 00000106547HL3 281FCERAPRO 00000113342HL0 281FCERSC220 00000104279HL5 281FCLPPROT 00000060464HL5 281OBMCPRO 00000077760HL7 281SLUTPRO 00000081383HL2 281UTSWPRO 00000054662HL2 281WHSLXRD 00000051586HL6 301209PRO 00000021505HL3 301290RAD1 00000079708HL4 301609RM2 00000022466HL7 301652RAD 00000062198HL7 301744RAD 00000049931HL9 301809RAD 00000036260HL8 301868R 00000103730HL8 301877RAD3 00000017454HL0 301896RAD1 00000017810HL3 302392PRO 00000021700HL0 302651PRO3 00000056259HL5 303426RM1A 00000023022HL7 303426RM2 00000030960HL9 303629RM05 00000023035HL9 303629RM06 00000031549HL9 303629RM08 00000028874HL6 303661RAD 00000022394HL1 303673TXR1 00000023034HL2 303673TXR2 00000105204HL2 303678FCFSXR 00000020953HL6 303699ER 00000070587HL1 303699RM3 00000103413HL1 303AFCXR 00000107786HL6 303FCBMOPTIMA 00000114262HL9 303NWMCXR 00000058405HL2 304253PROT 00000018385HL5 304424PRO 00000039696HL0 304424RM6 00000018323HL6 304424RM8 00000095422HL2 304598PROT 00000020067HL5 304599PRODA 00000019957HL0 304599PRODB 00000034541HL3 305345STRAD 00000034253HL5 307527PROTEUS 00000077287HL1 308425FPROT 00000041989HL5 308534NPOXR 00000099465HL7 308865KRRAD 00000030184HL6 308928RAD 00000036613HL8 309346PRAD1 00000033603HL2 309582PROT 00000016993HL8 309655FXR6 00000019597HL4 309655MORTON2 00000029530HL3 309663MHXR1 00000026059HL6 309664CBN2 00000068297HL1 313369PROT1 00000029288HL8 313640PROT1 00000045760HL6 314251SJBL 00000045792HL9 314251SJBR 00000037249HL0 314251SJMXR1 00000056595HL2 314739P1 00000081135HL6 314919JRAD1 00000062870HL1 315719PRO1 00000055252HL1 316685PRO1 00000049933HL5 316721WWRAD2 00000016409HL5 316722WWRAD 00000029086HL6 317888PRO1 00000039279HL5 317888PRO2 00000073628HL0 318259RM2 00000064174HL6 318443XRM1 00000029335HL7 318445LCXR1 00000070368HL6 318495WDR1 00000034712HL0 318878PRO 00000032393HL1 319235PROT 00000037749HL9 319642PROT 00000017467HL2 320654R1 00000024911HL0 320843RAD 00000086115HL3 321FCR1 00000016816HL1 3271539RAD 00000077530HL4 330297PX 00000031276HL9 330682ER 00000017468HL0 330740PX 00000080463HL3 330841PXD 00000111842HL1 331551DMGLORTHO 00000109667HL6 331551DMGLPROT1 00000107530HL8 331551DMGLPROT3 00000046882HL7 334244RAD2 00000018500HL9 334273RAD1 00000022473HL3 334636TRM1 00000028676HL5 334670RAD 00000050101HL5 336625R5203 00000053612HL8 336625R5207 00000098073HL0 337289SMRAD1 00000089288HL5 337363VR2 00000053199HL6 337531FRM2 00000038831HL4 337531FRM3 00000072363HL5 337531PRM5 00000044039HL6 352597RAD1 00000035182HL5 352689PRO 00000067465HL5 360397BGP 00000096320HL7 360397D8K 00000032496HL2 360412XR1 00000063175HL4 360417JTRAD1 00000063179HL6 360452RAD1 00000053542HL7 360476BANGOR 00000051288HL9 360476PSNS 00000016998HL7 360566RM1 00000028500HL7 360651XR1 00000042690HL8 360992CLK1 00000042384HL8 361698DPR1 00000022501HL1 361980RAP 00000101179HL0 386326ER 00000041259HL3 386FHIRAD 00000043171HL8 386HPORAD 00000031446HL8 386PORAD 00000031225HL6 386TWLRAD 00000029121HL1 401431PR1 00000020211HL9 401727P1 00000035302HL9 401766ARP1 00000064047HL4 402354MRAD 00000087142HL6 402393MERAD 00000088602HL8 402484AMIRAD 00000053200HL2 402745PROT 00000039280HL3 402768RAD1 00000086875HL2 402837CTCRAD 00000030412HL1 402947WRAD 00000037818HL2 404265ER1 00000035240HL1 404321PRO1 00000032761HL9 404466ED 00000016182HL8 404501P1 00000015898HL0 404501P2 00000049343HL7 404785PRO4 00000048257HL0 404785PROT 00000101400HL0 405271UPRO2 00000098076HL3 405271UPRO3 00000031854HL3 405360PRO 00000026060HL4 405379XR1 00000059950HL6 405636PRO 00000059577HL7 405682FTCPRO 00000074270HL0 405734TPR 00000072365HL0 406466TMC1 00000056597HL8 406653WXR 00000023538HL2 406748CXR 00000049347HL8 406768PXR 00000106550HL7 409267RAD1 00000036061HL0 409724RAD3 00000027849HL9 409727MJRAD2 00000063516HL9 409729PROT 00000040637HL1 409883RAD1 00000044038HL8 409899DCRAD1 00000086105HL4 410362ER 00000030726HL4 410601ER1 00000049346HL0 410604KIRAD 00000094526HL1 410787DRS8 00000029494HL2 410871RAD2 00000068786HL3 4109681 00000085826HL6 412372R1WDR 00000020590HL6 412624PROT1 00000087604HL5 412664ED1 00000045882HL8 412782PROT1 00000073181HL0 412784ORTHO2 00000019229HL4 412784PROT1 00000016344HL4 412784PROT3 00000042691HL6 412784PROT4 00000063694HL4 412788PROT 00000022748HL8 413499PROT1 00000023373HL4 413528R1 00000062564HL0 413738RAD1 00000019949HL7 414274PR1 00000022390HL9 414647RAD1 00000017825HL1 414805PRO2 00000070717HL4 414805PRO3 00000018611HL4 417269NX1 00000048625HL8 417354CMX 00000046431HL3 417461MRC 00000064169HL6 417625MSSU 00000090335HL1 417678ERXR 00000033829HL3 417820JX1 00000033716HL2 417820JX2 00000033958HL0 417820JX3 00000029628HL5 417820X1 00000029337HL3 417820X2 00000076948HL9 417847XR2 00000029126HL0 417887SNI2 00000033552HL1 419383RAD1 00000033387HL2 419407GE3 00000079709HL2 419479RAD 00000099140HL6 419547RAD1 00000078967HL7 419627RAD1 00000032885HL6 419843RAD 00000032813HL8 423431SDS 00000032294HL1 423431XR1 00000031273HL6 423431XR2 00000060102HL1 423586RAD 00000030014HL5 423778ERM1 00000030009HL5 423778ERM2 00000028289HL7 423778EXR2 00000025553HL9 423778EXR3 00000017617HL2 425228RC 00000019160HL1 425251VMA1 00000045630HL1 425374RAD1 00000037003HL1 425672RMB 00000074785HL7 425741RAD1 00000071150HL7 425885RAD1 00000022160HL6 432268RAD 00000104758HL8 432362RAD 00000062992HL3 434793RAD 00000087391HL9 435251NSHPC 00000062761HL2 435251OPPRM1 00000096689HL5 435528RAD 00000048801HL5 435635HCPRO1 00000053417HL2 435652PR1 00000086872HL9 435755PRO 00000050998HL4 435792RM1 00000083634HL6 435792RM3 00000100469HL6 435893RM3 00000088304HL1 435FCMCPRO1 00000086590HL7 440748PROTEUS 00000082523HL2 440967RAD1 00000037247HL4 440997PROT 00000031863HL4 443849ER2 00000086877HL8 469764MBPRAD 00000053413HL1 478275RAD2 00000035574HL3 478757CGTC 00000093303HL6 478PCHRAD 00000074786HL5 480345XR1 00000037380HL3 480598PRO 00000031269HL4 480603XR1 00000046336HL4 480730PROT 00000038830HL6 480984XR1 00000111898HL3 480MAYOXR 00000076947HL1 501202BNLER 00000080461HL7 501202BNLER 00000018694HL0 501268PROT 00000070589HL7 501268PROT2 00000059949HL8 501622PROT 00000047374HL4 502484PROT 00000041257HL7 502955RAD1 00000057949HL0 503216PRO 00000034865HL6 503224COPROT 00000074271HL8 503391SKY 00000024731HL2 503994RM2 00000103415HL6 504678PRO 00000018615HL5 504897SBER 000004-1188HL4 505727PROT 00000039602HL8 505763RAD1 00000086874HL5 505786PROT 00000045405HL8 505887RAD1 00000043961HL2 505887RAD2 00000048800HL7 505983PROT1 00000096227HL4 505DZXR 00000065131HL5 505LOBOXR 00000045880HL2 507212PRO 00000047010HL4 507212PRO 00000047428HL8 507274PRO 00000048111HL9 507274PRO 00000036696HL3 507283PROTEUS 00000090333HL6 507372PROTEUS 00000070887HL5 507407PROTEUS 00000020213HL5 507474PR1 00000087776HL1 507537PROTEUS 00000041101HL7 507723RAD 00000032814HL6 507836PROTEUS 00000050094HL2 508430FMC1 00000046763HL9 508548BRAD 00000025614HL9 508676PCR1 00000028374HL7 508676PCR2 00000022558HL1 508862CCH2 00000091771HL6 508996HRAD 00000113757HL9 508SANBR1 00000058225HL4 509247FAR1 00000055874HL2 509247FAR2 00000033900HL2 509276PROT 00000022601HL6 509332PRAD 00000033601HL6 509332PRAD 00000049932HL7 509525MCX1 00000052563HL4 509525MPROTEUS 00000049769HL3 509525VPROTEUS 00000105548HL2 509697SX1 00000063692HL8 509829VVX1 00000106220HL7 509966SX1 00000043859HL8 512887PR01 00000038408HL1 512887PR03 00000107528HL2 512FCBRAD 00000093644HL3 512FCJRAD 00000082841HL8 512LWMOBRAD 00000111894HL2 512SDERCPRAD 00000076310HL2 513233PRO 00000077529HL6 513389PRO1 00000023542HL4 513529XR 00000079182HL2 513853PRO1 00000064534HL1 513862WRPRO 00000040674HL4 513872PRO1 00000040892HL2 513872PRO2 00000063306HL5 513872PROER 00000110790HL3 515282XRA1 00000046673HL0 516674RM1 00000031862HL6 516676ER 00000023302HL3 517788PROTEUS 00000017472HL2 518243RAD1 00000017471HL4 518243RAD2 00000034240HL2 518254RAD3 00000034023HL2 518371RAD 00000113337HL0 518523PROT 00000043862HL2 518618RAD 00000033717HL0 520498PR1 00000033830HL1 520498PR2 00000016393HL1 520669XR 00000093305HL1 530528PROT 00000073848HL4 530BGPRO 00000056334HL6 540283RAD 00000080061HL5 540536UCARE 00000107531HL6 540591RAD 00000042903HL5 540741PRER 00000061396HL8 540741XR 00000041826HL9 540774VO 00000037688HL9 540776ROCLG 00000062868HL5 541276YX1 00000033426HL8 541322RM2 00000033553HL9 541322RM3 00000104755HL4 541342XR 00000080941HL8 541523PROTEUS 00000081136HL4 541548RM2 00000086873HL7 541549RM1 00000047231HL6 541OIT119A 00000047226HL6 541OIT119B 00000047067HL4 541OIT119C 00000047068HL2 541OIT119D 00000046766HL2 541OIT119E 00000047002HL1 541OIT134A 00000046883HL5 541OIT134B 00000046881HL9 541OIT134D 00000047005HL4 541OIT134E 00000046884HL3 541OIT134F 00000062993HL1 559459XR3 00000051287HL1 559683XR1 00000028380HL4 561207PBCC1 00000028638HL5 561207PBCC2 00000041102HL5 561627RAD 00000049590HL3 561650RAD 00000062866HL9 561881RAD 00000070367HL8 563252PR2 00000049131HL6 563387PRT2 00000021706HL7 563422PR 00000078166HL6 563568PROT 00000017821HL0 570409RAD 00000080460HL9 570455PRO 00000021185HL4 570455RAD1 00000104602HL5 570853PRO 00000023905HL3 571223RAD 00000056809HL7 573358R1 00000025061HL3 573442RAD 00000032264HL4 573556JCR4 00000064819HL6 573635JCOR1 00000022500HL3 573686ER 00000036060HL2 573729SRM1 00000028174HL1 574258PRO 00000048362HL8 574753PRT3 00000080732HL1 574753PRT4 00000096526HL9 575624RM1 00000082842HL6 580213VPRO 00000033424HL3 580220ROOM1 00000041993HL7 580371PRO 00000082527HL3 580481XRAY 00000063301HL6 580548BCPRO 00000056044HL1 580928PRO 00000079886HL8 585343RMC 00000028880HL3 586263MO2 00000043499HL3 586263RM1 00000061063HL4 586323PRO 00000066822HL8 586803PRO 00000031790HL9 601484PRO1 00000037589HL9 601583PROT1 00000042860HL7 601765PROT1 00000046429HL7 601783RAD1 00000061399HL2 601792PROT1 00000085130HL3 601876PROT1 00000020217HL6 601968COLRAD1 00000089290HL1 601968ERFP 00000029227HL6 601968PROT1 00000022914HL6 602220RM11 00000028296HL2 602246RM4 00000015953HL3 602249ERRAD 00000067235HL2 602251XR1 00000068441HL5 602251XR1 00000016188HL5 602285PROT1 00000017000HL1 602344RM1 00000018109HL9 602562PROT 00000086108HL8 602CIGNAPROT1 00000020955HL1 603788PROT 00000029229HL2 603882SJ4 00000015878HL2 603882SJ5 00000071925HL2 605328PROTEUS 00000046671HL4 605332PROTEUS 00000032295HL8 605342RAD 00000059948HL0 605347RAD2 00000061064HL2 605348RAD 00000044516HL3 605384PROTEUS 00000040980HL5 605428RAD 00000041210HL6 605462RAD 00000045108HL8 605532PROTEUS 00000035580HL0 605583PROTEUS 00000066385HL6 605665PROTEUSLC 00000038464HL4 605697PROTEUS 00000042902HL7 605697PROTEUS2 00000044646HL8 605698RAD 00000112431HL2 605755RAD 00000056339HL5 605835PROTEUS 00000057438HL4 605867RAD 00000031901HL2 605874RAD 00000029954HL5 605882PROTEUS 00000071487HL3 605996PROTEUS 00000020872HL8 606285PROT 00000047427HL0 606598FCI 00000080252HL0 606598XR1 00000042128HL9 606723PROT 00000050601HL4 606928PROT 00000031897HL2 607274EAST2 00000040769HL2 608258PROT 00000086588HL1 608326RAD 00000021191HL2 608364PR1 00000020588HL0 608796PR1 00000099139HL8 608WTCPRO 00000036792HL0 609426PROT 00000022749HL6 609465PROT 00000037717HL6 609625PROT 00000041135HL5 609646PROT 00000030410HL5 609652PROT 00000088599HL6 610565JOIPRO 00000031545HL7 610861PRO 00000083397HL0 610861PROFP 00000066675HL0 610JOIPROT 00000084009HL0 612313PRO 00000022155HL6 614234WESTPRO 00000045759HL8 614478PRO 00000023598HL6 614868PRO4 00000019232HL8 614898PROER 00000104750HL5 615435WRADOP 00000059848HL2 615446HER 00000104751HL3 615745PROT 00000023298HL3 616292PROTEUS 00000048493HL1 616365PROTEUS 00000023595HL2 616391PROTEUS1 00000030006HL1 616391PROTEUS3 00000042563HL7 616391WPROTEUS 00000050393HL8 616494PROTEUS 00000104281HL1 616685PROTEUS 00000042900HL1 616739PROTEUS 00000016189HL3 616752ER 00000043498HL5 616752M6PROTEUS 00000045408HL2 616752PROTEUS 00000034539HL7 617243PR1 00000080731HL3 617499PRAD 00000040588HL6 617MCOP1 00000069103HL0 618282RBRM2 00000103732HL4 618436R1 00000096155HL7 618664FCRAD 00000057777HL5 620257PRO 00000032300HL6 620285RAD 00000023260HL3 620327PRO 00000090859HL0 620332RAD 00000034686HL6 620724CX2 00000076043HL9 620786PRO 00000076509HL9 620872PRO 00000078551HL9 623876PROT 00000074784HL0 623876XR1 00000028175HL8 623882XR1 00000069843HL1 623925XR1 00000049765HL1 630323PROT 00000089291HL9 630351DOPROT1 00000046531HL0 630545PROT 00000050591HL7 630545PROT2 00000048492HL3 630545PROT3 00000048916HL1 630545PROT4 00000099909HL4 630551PROT2 00000111895HL9 630646PROT 00000049038HL3 630790PROT 00000041373HL2 630856PRO1 00000107790HL8 630871DMGWPROT1 00000075267HL5 630941PROT 00000021036HL9 630953LOBRAD 00000048518HL5 630968PRO 00000071146HL5 630968PROT 00000082186HL8 636239JR2 00000085136HL0 636239MOBR1 00000106555HL6 636390MSDR 00000025488HL8 636498MPR1 00000064171HL2 636530P1 00000098081HL3 636695MWR1 00000091293HL1 636789PDR 00000086817HL4 636866XR1 00000023172HL0 641295RAD1 00000052231HL8 641394PROT 00000064168HL8 641394PROT2 00000035374HL8 641424PROT 00000085133HL7 641446XRA 00000051825HL8 641456PROT 00000062991HL5 641484PROT 00000036911HL6 641664PROT 00000034344HL2 641754ERRAD 00000096319HL9 651254PRO 00000096154HL0 651265PRO 00000086586HL5 651293PRO 00000083218HL8 651439PRO 00000086114HL6 651789EPPA 00000064701HL6 651789PRO 00000098876HL6 651789PROVH 00000060368HL8 660530SFR2 00000088601HL0 660885WDR1 00000041733HL7 660890GVPRO 00000047006HL2 660890PROT3 00000017621HL4 661327RAD1 00000070714HL1 662285PRO 00000023904HL6 662377R1 00000062304HL1 662473R1 00000071147HL3 662726PRO1 00000053611HL0 671USNHPR 00000087607HL8 678289PRO 00000042327HL7 678318PRO1 00000086871HL1 678686PRO 00000045705HL1 701742RAD 00000059648HL6 701854RAD 00000099910HL2 702255XR2 00000038260HL6 702382XR1 00000023261HL1 702696XR 00000019149HL4 702731ORTH3 00000019148HL6 702731ORTHO1 00000018753HL4 702731ORTHO2 00000018814HL4 702731ORTHO4 00000080249HL6 702732MPXR1 00000056594HL5 702869XR1 00000057039HL0 702869XR1 00000056336HL1 702869XR2 00000057948HL2 702869XR2 00000043525HL5 703339RAD 00000083219KL6 703441RAD 00000017151HL2 703481RAD 00000019956HL2 703591RAD 00000034973HL8 703670ER1 00000103731HL6 703697PROT1 00000019159HL3 703751RAD 00000083058HL8 703849RAD1 00000057041HL6 704323PRO1 00000070888HL3 704323UPRO 00000065132HL3 704660PRO1 00000032662HL9 704660RAD1 00000033661HL0 704783PRO1 00000033600HL8 704783PRO2 00000023601HL8 704895RAD1 00000052076HL7 706245TR1 00000022156HL4 706554MSR1 00000053616HL9 706756RAD2 00000085824HL1 708237PROT1 00000107527HL4 708246PROT 00000057043HL2 708478PROT 00000036978HL5 708756ER1 00000083398HL8 708799PROT 00000022237HL2 708873XRA 00000058126HL4 708WSBCPROT1 00000065037HL4 712225XRA 00000086113HL8 712251PROTEUS 00000039491HL6 712737PROTEUS 00000041814HL5 712737PROTEUS 00000036861HL3 713442KSRAD 00000038443HL8 713463TCRAD 00000063805HL6 713547HCPROT 00000065483HL0 713566RAD 00000062473HL4 713566RAD2 00000073179HL4 713636HCRAD 00000106548HL1 713636SITIPRO 00000039753HL9 713662UTPROT 00000042079HL4 713770RAD 00000047488HL6 713948PMCRAD1 00000059773HL2 713982SCPROT 00000071634HL0 713HCALDPROT 00000084575HL0 713UTNWPRO 00000074012HL6 713UTPROTEUS 00000054531HL9 715241PR2 00000029689HL7 715241PRO 00000067055HL4 715284HRAD 00000050249HL2 715294PROT 00000051828HL2 715362PR1 00000042780HL7 715393PR5 00000065369HL1 715424PR1 00000061622HL7 715478PR1 00000063300HL8 715478PR1 00000037592HL3 715539PRO1 00000059576HL9 715588PR1 00000028577HL5 715644RAD 00000035153HL6 715672RAD 00000082524HL0 715793PR 00000057992HL0 715822RAD 00000017155HL3 715833PROD 00000029952HL9 715847PROTEUS 00000023268HL6 716633RAD 00000023405HL4 716633RAD1 00000075127HL1 716636PRAD1 00000077284HL8 716636PRAD1 00000053416HL4 716712RM1 00000054164HL9 716839RAD1 00000107787HL4 717270RAD1 00000086587HL3 717270RAD2 00000039370HL2 717464R1 00000088068HL2 717531MJR1 00000102865HL3 717561PRO1 00000032296HL6 717627RAD1 00000100280HL7 717761PRO2 00000058403HL7 717761PROT 00000029283HL9 717761RAD1 00000023168HL8 718267ER 00000059951HL4 718267PROTEUS 00000023368HL4 718567ER 00000055346HL1 718579RAD1 00000029449HL6 718604ER1 00000029022HL1 718604PROTEUS1 00000046672HL2 718604PROTEUS2 00000098080HL5 718840RAD2 00000107788HL2 719264RAD 00000046762HL1 719346PRO1 00000111469HL3 719487RAD 00000063806HL4 719584XR1 00000090332HL8 719638RAD 00000054935HL2 719667RM1 00000059844HL1 719686RAD 00000057435HL0 719SLVXR1 00000058603HL2 719SLVXR2 00000066674HL3 724224UCC 00000031610HL9 724230PROT1 00000022467HL5 724627PROT2 00000040988HL8 724779PROT 00000049588HL7 724779PROT2 00000080624HL0 724864PROT1 00000040779HL1 731659RF1 00000029724HL2 731784RM1 00000038055HL0 732223PRO 00000017455HL7 732360ER 00000034923HL3 732360PAT 00000021164HL9 732505RAD 00000094363HL9 732530URGW 00000015943HL4 734434PROT 00000043526HL3 734446PROT 00000077286HL3 734712AABJ 00000094199HL7 734712AABJ2 00000079512HL0 740351PROT 00000085134HL5 740506PROT 00000063515HL1 757275PRO1 00000056739HL6 757314RAD1 00000059649HL4 757314RAD2 00000045524HL6 757673ER 00000031898HL0 757673PRO1 00000050996HL8 757673PRO2 00000042562HL9 757673RM2 00000076700HL4 757822TCC118 00000077070HL1 757822TCC122 00000040432HL7 757889MM2 00000017818HL6 757889RM3 00000043858HL0 757889RM5 00000089698HL5 760379RAD 00000045546HL9 760830PRO 00000094656HL6 763587PRO 00000111843HL9 763795PRO 00000057199HL2 765254PROT 00000071424HL6 765472RAD1 00000089292HL7 765472VA1 00000050390HL4 765674PROT 00000052237HL5 765674PROT1 00000032881HL5 770229GT 00000070209HL2 770297NDR2 00000023300HL7 770751PRO 00000081880HL7 770838TDVAPRO 00000045614HL5 770965BR2 00000080253HL8 770968PRO 00000034669HL2 770971PROT 00000015893HL1 773265PROT 00000032298HL2 773265PROT2 00000020866HL0 773282TOMO8 00000075399HL6 775SLYONXR 00000071923HL7 785263PRO 00000033377HL3 785434PRO 00000059717HL9 785475PRO 00000015887HL3 785562PRO 00000055684HL5 801273XRAY 00000054412HL2 801294XRAY 00000060396HL9 801302MOBRM1 00000037747HL3 801302RM1 00000045761HL4 801321OSG 00000060061HL9 801341SS1 00000062301HL7 801387RM1 00000041886HL3 801489XR 00000043589HL1 801501XR 00000057774HL2 801525PRO 00000114259HL5 801525PRTS 00000015867HL5 801581RM6 00000070208HL4 801627PRO 00000020584HL9 801734RM1 00000064531HL7 801777RM1 00000068443HL1 801777RM2 00000049039HL1 801779RAD 00000087140HL0 801779RAD2 00000090857HL4 801786XR2 00000085132HL9 801871XR1 00000044042HL0 801972WKMD 00000099141HL4 801AFP1 00000087143HL4 801RCXR1 00000031226HL4 803358UCR1 00000031270HL2 803536R1 00000028519HL7 803541BRAD 00000037576HL6 803581RM2 00000024732HL0 803749R1 00000035917HL4 803895RAD 00000042199HL0 804217RAD1 00000043173HL4 804287PRO1 00000096690HL3 804325RSPRO 00000078167HL4 804595XR1 00000078168HL2 804595XR1 00000041580HL2 804639RAD 00000106221HL5 804893RAD 00000076311HL0 804893RAD1 00000076698HL0 804893RAD1 00000054534HL3 804HDMCRM2 00000059774HL0 805737RAD1 00000036120HL4 805PIXRA 00000020206HL9 806212RAD1 00000079888HL4 806355RAD1 00000095426HL3 806663RAD 00000062019HL5 808448XRA 00000030415HL4 808696RM1 00000098075HL5 808873PROA 00000066963HL0 808876PRO 00000050391HL2 810342ER1 00000025121HL5 810342RMB 00000023870HL9 810762ER1 00000050594HL1 810985PRO1 00000022327HL1 812242APN1 00000022240HL6 812242APN2 00000036987HL6 812242APN3 00000075817HL7 812242PROT 00000086593HL1 812242UAP1 00000089904HL7 812242UAP2 00000053540HL1 812343PROT 00000026638HL7 812450ER1 00000041449HL0 813253PRO 00000031769HL3 813964PRO 00000032888HL0 814255R2P 00000084773HL1 815300PVPROT 00000083635HL3 815395OSFER1 00000107789HL0 815395OSFXRF 00000060366HL2 815462PROT1 00000017456HL5 815842JR1 00000048113HL5 816348PROT 00000087777HL9 816348PROT2 00000033957HL2 816561DICPRO 00000049589HL5 816HILLYARD 00000086816HL6 817336PRO 00000035025HL6 817346RAD 00000034971HL2 817346RAD1 00000025562HL0 817347HPROER 00000040838HL5 817543ACC 00000038258HL0 817605CCHRAD 00000084771HL5 817644FCRAD 00000037346HL4 817685ER2 00000022555HL7 817685HEBTOMO1 00000022753HL8 817685HEBTOMO2 00000046427HL1 817837PRO 00000102866HL1 817995PRODR 00000089902HL1 818360XR1 00000056335HL3 828433JIP 00000044379HL6 830816RAD 00000051829HL0 830896VAPRO 00000057950HL8 843228MCASRAD 00000043812HL7 843228PR1 00000033642HL0 843784RAD 00000107534HL0 843867RAD1 00000109666HL8 843PPCELMSR1 00000034239HL4 844336PROT 00000035578HL4 844336PROT2 00000019876HL2 845703RAD 00000027848HL1 845703RAD2 00000037929HL7 845938RM2 00000078966HL9 847292PROT 00000060062HL7 847543CLCA 00000059335HL0 847543CLCB 00000059574HL4 847543CLCF 00000024744HL5 850431ER1 00000024734HL6 850431ER2 00000104280HL3 850623NWFXR 00000081139HL8 850654PRO 00000062017HL9 850696WXR1 00000101180HL8 850862PRO 00000088855HL2 856234PROT 00000053346HL3 856363PROT 00000070004HL7 856451PROT 00000020353HL9 859226ER 00000043494HL4 859987PROT 00000068620HL4 860667RAD1 00000020768HL8 863419ER 00000113336HL2 864299XRA1 00000037117HL9 864560XRA1 00000039278HL7 864898RM1 00000062100HL3 865525RAD 00000064323HL9 865573XR1 00000036610HL4 870972PRO1 00000037119HL5 870972PRO2 00000106223HL1 901NAVY 00000072093HL8 903886PRO1 00000065129HL9 903TOPRO 00000086107HL0 904230RM1 00000032492HL1 904858MRAD 00000051823HL3 904928RAD 00000111892HL6 904998RM1 00000039954HL3 907234PROT 00000023837HL8 907796ORTHO 00000022331HL3 907835PROTEUS 00000081877HL3 907861VER 00000034093HL5 908522PRO 00000044744HL1 910450CHBAY 00000045629HL3 910450NHCL3 00000044740HL9 910450NHCL6 00000090334HL4 910904CFHPROT 00000046337HL2 910907WA5 00000062990HL7 910907WACC1 00000059341HL8 910907WAJOEL1 00000038444HL6 910907WAPROT 00000059845HL8 910907WARM6 00000052230HL0 910907WARM7 00000065130HL7 912287OTC 00000101038HL8 913397PROT 00000068622HL0 913785FDR1 00000036468HL7 915533RAD1 00000062302HL5 915577ERRAD 00000087605HL2 915869RAD 00000028983HL5 916278PROT1 00000039281HL1 916351PRO1 00000043330HL0 916575PROMN 00000023110HL0 916781PRO 00000032559HL7 916781PROT1 00000041618HL0 916781PROT2 00000028034HL7 916874PRO1 00000028040HL4 916874PRO3 00000028037HL0 916874PRO4 00000045111HL2 916LAGPRO1 00000077285HL5 916MCCPRO 00000019947HL1 918287PRAD 00000043959HL6 918294MPPRO 00000046339HL8 918294MPPROT 00000054533HL5 918392PROTEUS 00000041477HL1 918486PROTEUS 00000094652HL5 918567PROT1 00000048798HL3 918599XR112 00000046626HL6 918599XRER 00000100278HL1 918647PROT1 00000028521HL3 918787GRAD1 00000028980HL1 919572SWSM1 00000020963HL5 919781RAD1 00000016986HL2 919787WRAD2 00000023114HL2 919954PROT1 00000054165HL6 919954RM10PROT 00000079711HL8 919BCPROT 00000017826HL9 920288RAD1 00000018386HL3 920288RAD2 00000018388HL9 920288RAD2 00000028639HL3 920303ER 00000028780HL5 920303PROT 00000022585HL4 920327RAD 00000028787HL0 920451RAD1 00000036790HL4 920720PRT 00000020215HL0 920738RAD1 00000020685HL4 920738RAD2 00000020721HL7 920738RAD3 00000039055HL9 920793PRO1 00000038562HL5 920793PRO2 00000034432HL5 920866PR1 00000016592HL8 920952ER1 00000016736HL1 920952RAD2 00000031900HL4 925210RAD1 00000015952HL5 925210RAD2 00000015949HL1 925210RAD3 00000022058HL2 925370ORTHO1 00000022063HL2 925370ORTHO2 00000087389HL3 928453UCXR1 00000038832HL2 928725RM1 00000105864HL3 928737RAD 00000038463HL6 928758RAD 00000020076HL6 931403RAD1 00000020068HL3 931520PROT1 00000021035HL1 931520PROT2 00000074014HL2 936825GSJRAD1 00000042284HL0 937268R113 00000059269HL1 937440PROH 00000088307HL4 937440PROH2 00000039367HL8 937440PRON 00000038465HL1 937440PROS 00000041136HL3 937440RM3 00000069101HL4 937440RM4 00000084369HL8 937440SPRO1 00000041645HL3 937558PRO 00000050246HL8 937558SCPRO 00000086591HL5 937OVXR 00000081380HL8 940320PROTEUS 00000047519HL4 940898PROD 00000064172HL0 949517RM4 00000086818HL2 952541PRO 00000114263HL7 952738PRO 00000033586HL9 952944RAD 00000046428HL9 956366RAD1 00000028476HL0 956632MCRAD1 00000038035HL2 970453XR1 00000032884HL9 970565RAD1 00000022757HL9 970668XR 00000041648HL7 970DOAXR1 00000035154HL4 972331RAD1 00000055347HL9 972487BGDX2 00000036466HL1 972487BGOP1 00000044035HL4 972487BGPRO 00000041172HL8 972487BGRAD 00000034842HL5 972566DMSRAD1 00000033875HL6 972665BPPR1 00000033898HL8 972665BPPR2 00000034946HL4 972665BPSC1 00000034866HL4 972665BPSC2 00000033671HL9 972665DDA1 00000033714HL7 972665DDA2 00000084774HL9 972744PRO1 00000087608HL6 972899FMPRO 00000043021HL5 972935ERPRO 00000041828HL5 972981PRO10 00000037343HL1 972981PROTEUS1 00000042325HL1 972981TOMO9 00000045282HL1 973331PRO 00000041579HL4 973439PRO 00000038257HL2 973503PRO 00000029129HL4 973625RM2 00000040431HL9 973746PRO 00000029538HL6 973989RM2 00000058404HL5 978372WRAD 00000088303HL3 979245PRO 00000072528HL3 980487PRO1 00000035023HL1 985732RAD1 00000072529HL1 989246RAD 00000045793HL7 989246RM2 00000088067HL4 989343ER 00000018815HL1 989352RMB 00000035981HL0 989802PRO1 00000035371HL4 989839RM5 00000065035HL8 CVSPRAD1 00000109668HL4 FCERPL200 00000111891HL8 FCFRISCOPROT 00000102139HL3 FCGARX-RAY 00000103417HL2 FCMESQPROT 00000106553HL1 FCRICHPROT 00000053315HL8 FHAPKPRO 00000044029HL7 FHEOPROTEUS3 00000022557HL3 FHORLER1 00000023170HL4 FHORLER2 00000051589HL0 FHORLPRO4 00000048624HL1 FHWPPRO3 00000046526HL0 GAFBXRA 00000048005HL3 INSIGHTXR107 00000086592HL3 MC532RAD1 00000112429HL6 PMUC7B 00000081589HL4 WC231RM1 00000077528HL8 WC231RM4 00000022388HL3 WH231RAD1 114269HL4 605322CLIFFRAD 00000053412HL3 205481PRO 00000053538HL5 205592DCPRO 00000045210HL2 205750RAD 00000086876HL0 205795PRO 00000037575HL8 206248RAD1 00000113339HL6 206248RAD3 00000082322HL9 206288XR1 00000035237HL7 206302XRM1 00000104282HL9 207255RAD1 00000045795HL2 207438KRAD 00000016184HL4 207454CRAD1 00000073627HL2 207454RAD2 00000053617HL7 207662PRO 00000074527HL3 207664PROTEUS 00000077762HL3 207695PROXR 00000071635HL7 207794RAD2 00000039695HL2 207795CM3 00000059338HL4 207973PROTEUS 00000018756HL7 208227RM1 00000019410HL0 208227RM5 00000040540HL7 208265PROT 00000020071HL7 208359RM2 00000020723HL3 208373RAD2 00000023269HL4 208373RAD3 00000019231HL0 208381ED2 00000102868HL7 208402RAD1 00000042863HL1 208455RAD3 00000042194HL1 208455RAD4 00000035581HL8 208463RAD1 00000015875HL8 208463RAD2 00000029231HL8 208476CRAD 00000104065HL8 208524RM3 00000037591HL5 208524ROOM2 00000015942HL6 208529ER 00000038334HL9 208667KS 00000071922HL9 208678PROTEUS1 00000028682HL3 208746VX1A 00000608866HL1 208882IX1 00000030965HL8 208962SRAD 00000024470HL7 208983SRAD 00000043397HL9 209223PROTAM1 00000023266HL0 209385RAD1 00000024368HL3 209524GR5 00000022586HL2 209577RAD3 00000042688HL2 209754RAD1 00000029281HL3 209832RAD1 00000047229HL0 209835RAD2 00000020769HL6 209848OVRAD2 00000041213HL0 210292LRCR1 00000028979HL3 210297BMCR5 00000025613HL1 210297NERAD 00000028575HL9 210297SEBR4 00000114264HL5 210403MSPRO 00000105865HL0 210447ER 00000046134HL3 210575LCRAD 00000081382HL4 210590RAD1 00000029536HL0 210616MCTRAD2 00000029813HL3 210616R2 00000056738HL8 210617NWPRO 00000043175HL9 210617PRO 00000028677HL3 210646NEMR1 00000037574HL1 210646RAD2 00000016585HL2 210704SRER 00000102502HL2 210877PRO 00000042478HL8 210921PRO 00000044380HL4 210925KRM177 00000044382HL0 210925KRM185 00000071924HL5 210MIRAD1 00000021398HL3 212241RAD 00000023112HL6 212241RAD 00000021395HL9 212604PROT3 00000021893HL3 212604PROT3 00000019405HL0 213744PROTXR2 00000019596HL6 213744PROTXR2 00000064045HL8 214266IRVPRO 00000036576HL7 214266RAD1 00000051937HL1 214268RAD1 00000016400HL4 214320PTOMO 00000021323HL1 214345RAD1 00000071488HL1 214370BCCPRO1 00000090860HL8 214491RPG 00000059715HL3 214590UROPRO 00000081591HL0 215245PROFP 00000036599HL9 215441PROT1 00000024475HL6 215453PRO3 00000042133HL9 215481PROT1 00000052236HL7 215827PROT1 00000016819HL5 215952PROT 00000043861HL4 215955PROER 00000102504HL8 216445PROTEUS 00000104754HL7 216445PROTEUS 00000047515HL2 216778RAD1 00000024659HL5 217222QR6 00000036243HL4 251368ARAD 00000023400HL5 252209RCR3 00000036063HL6 252436RAD1 00000036096HL6 252436RAD2 00000035974HL5 252443NGER1 00000036550HL2 252443NGER2 00000044033HL9 252493RPCC 00000019877HL0 252522LRAD1 00000036281HL4 252637ECR2 00000032262HL8 252794BR2 00000100701HL2 252809MGRAD 00000031861HL8 252847EDR1 00000031899HL8 252847EDR2 00000037483HL5 252847PRE1 00000042285HL7 252847RAD8 00000042324HL4 252847RAD8 00000080062HL3 253841DINW1 00000033956HL4 253ECN2420RM4 00000036192HL3 253ECN3535RM2 00000065036HL6 253ECN47859RM5 00000036864HL7 254288DAR2 00000044518HL9 254288DAR4 00000021705HL9 254562RAD2 00000016600HL9 254897GRRAD 00000016735HL3 254897GRRAD 00000017809HL5 256235ORT1 00000064043HL3 256332RRM1 00000061620HL1 256356RAD1 00000053535HL1 256386RM4 00000082187HL6 256429CR2 00000073849HL2 256782PRO1 00000090522HL4 256CCP1 00000020716HL7 262334RAD1 00000060367HL0 262569PRO 00000079179HL8 262697PRO1 00000076419HL1 262697PRO2 00000016824HL5 262785RM1 00000047062HL5 262884PRO1 00000049034HL2 262896PRO1 00000063695HL1 269327PROTEUS 00000050597HL4 269381PROTEUSA 00000052225HL0 269381PROTEUSB 00000046049HL3 269463PROTEUS 00000023834HL5 269637PROTEUS 00000051939HL7 269657PROTEUS 00000051664HL1 269683PRO 00000025422HL7 269945PROTEUS 00000046338HL0 269968PROTEUS 00000015954HL1 270259R3 00000030413HL9 270444LR4 00000030408HL9 270444MNPOC 00000090521HL6 270483GPROT 00000045942HL0 270798AHC 00000052239HL1 270798ORTHO 00000059446HL5 270798ORTHO2 00000040539HL9 270798RAD 00000041616HL4 270798RAD3 00000047227HL4 270798RAD7 00000053615HL1 270956LR 00000055873HL4 270956LR 00000057040HL8 270965LR2 00000076042HL1 276666RM4 00000022892HL4 276964PRORAD1 00000046669HL8 281348PROTEUSER 00000017458HL1 281558PRO2 00000024908HL6 281588PROER 00000043329HL2 281897MHPRO 00000107532HL4 281ACERSPROT 00000080730HL5 281FCPROTEUS 00000107536HL5 281HHVPRO 00000072983HL0 281WHGPPRO 00000043070HL2 301609ER 00000025063HL9 301698ER2 00000024681HL9 301698RM3 00000060600HL4 301874UBRAD 00000029280HL5 302421PRO 00000036693HL0 3024307 00000052080HL9 302653RAD1 00000052228HL4 302674RAD1 00000054535HL0 302684RAD 00000042781HL5 302744ER 00000023177HL9 303320ER1 00000016402HL0 303450ED 00000045632HL7 303563XR 00000022157HL2 303730RM5 00000030414HL7 303778EDXR 00000073388HL1 304373R2 00000032394HL9 304598PROT1 00000033385HL6 304624PRT 00000031611HL7 304792PROT1 00000041735HL2 304792PROT2 00000069100HL6 304855PROT 00000041992HL5 304927R2 00000041992HL9 304927R2 00000023115HL9 305295R1 00000040960HL7 305325RM9 00000031150HL6 305682ER1 00000064048HL2 305FISHPRO 00000094364HL7 305MCHDORAL 00000089697HL7 305MRCOPRM1 00000016183HL6 307265GERAD 00000057042HL4 307578MOBRAD 00000038280HL4 307634XRA 00000015874HL1 309343CR2 00000046880HL1 309343CRAD1 00000028520HL5 309454BER 00000060101HL3 309454BX3A 00000064533HL3 309647GMGR1 00000031155HL5 309662JER 00000016406HL1 309944HHPROT 00000015945HL9 313359PROT 00000027950HL7 314577CP1 00000027845HL7 314577CP2 00000019952HL1 314625WER 00000060462HL9 315255XRAY1 00000061886HL8 315255XRAY1 00000036424HL0 315332NR4 00000057437HL6 315332RM3 00000044645HL0 315338ER 00000016593HL6 315782RAD1 00000016590HL2 315782RAD2 00000088600HL2 315787TB1 00000017157HL9 316689WC3 00000017145HL4 316689WC4 00000018489HL5 316689WC4 00000089289HL3 316962ERPRO1 00000064817HL0 316962ORTHO 00000042862HL3 316962PRO 00000054415HL5 316962WCPRAD 00000034303HL8 317621PROT 00000020066HL7 317745HEPR5 00000019408HL4 317745HEPR6 00000021328HL0 317745HEPR7 00000040979HL7 318251PRO 00000029039HL5 318254PRO 00000030645HL6 318335ORM1 00000048112HL7 318456BRM3 00000048522HL7 318648RM2 00000029120HL3 319268PROT 00000025377HL3 319268SER 00000015897HL2 319376PROT 00000029656HL6 319863PROT 00000061263HL0 320243RAD1 00000017150HL4 320632RAD 00000062472HL6 320693RAD2 00000024203HL2 325670PRO2 00000029448HL8 325670PRO3 00000037156HL7 325696RAD1 00000037250HL8 325696RAD2 00000016734HL6 330332R4 00000040636HL3 330363RAD1 00000039369HL4 330363RAD2 00000043396HL1 330386ER 00000029128HL6 3304382 00000041371HL6 330543XRA 00000046532HL8 330707PX 00000040889HL8 330834RAD 00000023409HL6 332047 00000077527HL0 333IMM 00000046759HL7 333PRO 00000033384HL9 336372ARAD1 00000043809HL3 336719NR140 00000082321HL1 336846AMRAD 00000083632HL0 337289RAD 00000039583HL0 337457RAD1 00000034559HL5 337981ER1 00000028578HL3 337981R3 00000033498HL7 337981RAD3 00000077764HL9 33RM4 00000040538HL1 352253RAD1 00000096686HL1 352596PROM 00000040990HL4 352596XRA 00000019688HL1 352861RAD1 00000083056HL2 360346PROTEUS 00000042298HL0 360414OTHO 00000036323HL4 360493PROT 00000046764HL7 361275RM1 00000099137HL2 361293PROT 00000027095HL9 361572CER 00000032556HL3 361857DOR1 00000044409HL1 361888CCOPC 00000052226HL8 373ORTHO 00000023299HL1 386238FHR1 00000024469HL9 386238FHR3 00000028035HL4 386446RM1 00000028984HL3 386446RM2 00000035980HL2 386774RM1 00000034561HL1 386774RM2 00000016581HL1 401274ER1 00000057947HL4 401348WR1 00000022760HL3 401782SR2 00000029727HL5 401782SR3 00000057194HL3 402269RAD1 00000016818HL7 402481BX3 00000016186HL9 402481LGMX10 00000032263HL6 402643SMX1 00000020352HL1 402685RAD 00000073178HL6 402878PHSRAD 00000024428HL5 402925RAD 00000022507HL8 404616PRO1 00000062018HL7 404686CRAD5 00000056737HL0 404686MOT2 00000029226HL8 404778EP1 00000080938HL4 405262EFCIRAD 00000054414HL8 405271ETPRO 00000043639HL4 405271PRO 00000044311HL9 405271UPRO 00000017623HL0 405271UPX1 00000106222HL3 405307PRO 00000069327HL5 405359PBPRO 00000016405HL3 405372SP1 00000016404HL6 405372SP2 00000040891HL4 405527RAD 00000016813HL8 405736MR4 00000034664HL3 405927R1 00000102143HL5 405949ER1 00000018693HL2 406222LR2 00000036549HL4 406228FMRAD3 00000027945HL5 406238BCW1 00000071926HL0 406327PLX1 00000065368HL3 406363PROT 00000101781HL3 406363XRP 00000064535HL8 406543SX3 00000032297HL4 406547MVR1 00000091772HL4 406585PXR1 00000091104HL0 406585PXR2 00000031268HL6 406654PXR 00000023175HL3 406657BCH 00000062563HL2 406665BHR1 00000045432HL2 406728PRT 00000022161HL4 406782SJER 00000028570HL0 406782SX1 00000072530HL9 406787RRAD 00000042783HL1 406842RVXR 00000044310HL1 406863RAD 00000078008HL0 406873PXR 00000023375HL9 407296ERP 00000071318HL0 407296RM2 00000078169HL0 407296RM2 00000049930HL1 407321ERPRO1 00000032763HL5 407518PR1 00000049766HL9 408559RAD12 00000021194HL6 408559RAD2 00000042782HL3 408636PROTEUS1 00000028036HL2 408885EDA 00000025060HL5 408885EDB 00000052077HL5 408885GIL 00000064173HL8 408885MIL 00000056257HL9 408885MOOR 00000038406HL5 408885SVC 00000035372HL2 408885TUL 00000081134HL9 408885VHC1 00000052073HL4 408885VSC2 00000054167HL2 408885VSC3 00000086111HL2 408957ECF1 00000070006HL2 409212RAD1 00000050389HL6 409384RAD1 00000020072HL5 410392R3 00000015889HL9 4105462 00000029688HL9 410641ER 00000061621HL9 410740R1 00000064046HL6 410740R2 00000072984HL8 412232ER1 00000076507HL3 412232ER2 00000020867HL8 412367PROT1 00000111659HL9 412367XR1 00000068950HL5 412623PROT5 00000062099HL7 412623UCC 00000018114HL9 412641PROT1 00000042527HL2 412641R1 00000088302HL5 412647WDR1 00000031904HL6 412692PROT1 00000032706HL4 412692PROT2 00000105546HL6 412942WDR 00000019230HL2 414527PR1 00000029786HL1 414647RAD2 00000029496HL7 414761RAD1 00000100277HL3 414774PRO 00000042454HL9 417256XR2 00000096524HL4 417837XR1 00000026636HL1 417883SG23 00000076312HL8 419228PR 00000112817HL2 419238RAD 00000036577HL5 4193543 00000023369HL2 419383HCRAD 00000034844HL1 419394ER 00000069102HL2 419394RAD2 00000039584HL8 419399RAD 00000079710HL0 419538RAD 00000020957HL7 419578RAD 00000086112HL0 419578RAD3 00000050096HL7 419586PR 00000021704HL2 419592RAD2 00000015947HL5 419666RAD1 00000016739HL5 419692RAD 00000048797HL5 419738PR 00000034455HL6 419891ER 00000034922HL5 419891RM2 00000108377HL3 419946PR1 00000019879HL6 423493ER 00000025067HL0 423542RAD1 00000019690HL7 423778EE1 00000019772HL3 423844R1 00000069330HL9 423844R2 00000052234HL2 432837RAD1 00000039605HL1 434924HW2 00000071315HL6 434CCRAD 00000091295HL6 434SCHRM3 00000057196HL8 434SCHRM4 00000056812HL1 435207RM1 00000028574HL2 435251RM3 00000028977HL7 435251RM5 00000032663HL7 435462XR 00000060063HL5 435628RROPRO1 00000024738HL7 435637RM1 00000029687HL1 435676RAD 00000046528HL6 435688RRPRO1 00000073387HL3 435716RM4 00000072706HL5 4357792RM2 00000023541HL6 435865PRO1 00000023297HL5 435865PRO2 00000088860HL2 435DCMCPRO1 00000088305HL8 440743PROTEUS 00000029725HL9 440827ER1 00000023836HL0 440988RAD1 00000062760HL4 443481OSM1 00000021703HL4 443849ER 00000031856HL8 443849ER1 00000025416HL9 443849RAD4 00000020591HL4 478633EC1 00000038033HL7 502259ER 00000039697HL8 502624R7 00000017153HL8 502726RAD1 00000015892HL3 502893ER 00000024426HL9 502897TOMO 00000031546HL5 503215PROTEUS 00000103418HL0 503399PRO 00000062196HL1 504371PRO 00000034454HL9 504391ER 00000020070HL9 504391R3 00000044037HL0 504447RAD1 00000041448HL2 504464ER 00000036910HL8 504842OUC 00000025177HL7 504853RAD1 00000055872HL6 504897OBED 00000031768HL5 505262PROT1 00000099908HL6 505426PROT 00000020864HL5 505522RAD1 00000020958HL5 505522RAD2 00000021895HL8 505538RAD1 00000024371HL7 505662PROT 00000033713HL9 505662PROT2 00000049768HL5 505727PROT101 00000037116HL1 505743RAD1 00000057440HL0 505893PROT 00000034458HL0 505DULCEXR 00000080623HL2 507385EDRAD1 00000047330HL6 507385NRRAD1 00000035373HL0 507385RAD 00000045364HL7 507385RAD1 00000020074HL1 507434RAD1 00000065133HL1 507526RAD1 00000062471HL8 507837RAD 00000059270HL9 507931RAD1 00000106551HL5 508457FR1 00000039585HL5 508460RAD1 00000041508HL3 508548FR2 00000081879HL9 508833SHRAD 00000044043HL8 508862R11 00000083055HL4 508FHCMRAD 00000105549HL0 508NOFALRAD 00000023108HL4 509525WX1P 00000068953HL9 509575YRX1 00000052081HL7 509633CX3 00000029953HL7 509682LX2 00000056042HL5 509754CX2 00000041260HL1 509758TRAD2 00000076134HL6 509786RAD1 00000041404HL5 509865TCX1 00000048360HL2 509962K2X 00000039057HL5 509SHRINPROT 00000024907HL8 510437PROT1 00000024912HL8 510437PROT2 00000017003HL5 510655SM11 00000019780HL6 510667PROTEUS 00000042197HL4 510797RM1 00000025176HL9 510797W2 00000029447HL0 510797W5 00000033554HL7 512397RAD2 00000033587HL7 512397RAD3 00000036425HL7 512447PRAD 00000082326HL0 512BASRAD 00000075818HL5 512BCRAD 00000074272HL6 512PFERRAD 00000026064HL6 513584PRO1 00000026635HL3 513584PRO2 00000032555HL5 513584RM9 00000019774HL9 513636EGXRAY 00000022324HL8 513636OPAPRO 00000040981HL3 513636PRO 00000028293HL9 513745PRM2 00000028097HL4 513745RM1 00000059449HL9 513BNOPRO1 00000059452HL3 513BNOPRO2 00000030913HL8 515222XRA 00000034540HL5 515222XRA2 00000060983HL4 515278IHPXRA 00000036551HL0 515382XRA 00000034429HL1 515643XRA 00000051938HL9 515832XRA 00000029537HL8 516520RM1 00000028372HL1 516632ER1 00000052235HL9 516632RM2 00000048521HL9 516632RM5 00000113338HL8 517334ER 00000035141HL1 517543PROTEUS 00000024580HL3 517663PROTEUS 00000032495HL4 518382RAD6 00000033386HL4 518382RAD8 00000016198HL4 518583RAD5 00000083633HL8 518697RAD1 00000101402HL6 518725RAD 00000028033HL9 518828RM1 00000035181HL7 520228PRO 00000029119HL5 520432RM2 00000039490HL8 520533PRO1 00000043497HL7 520533XR3 00000064170HL4 520STJOPRO 00000084007HL4 530225PROT 00000020956HL9 530662PRO 00000049127HL4 530662PRO2 00000063176HL2 530BAFPRO 00000016731HL2 540228RAD 00000019543HL8 540462RAD 00000048358HL6 540536ER1 00000049126HL6 540636RAD1 00000034306HL1 540776PR2 00000034304HL6 540776PR3 00000023171HL2 540862RAD1 00000023871HL7 540862RAD5 00000045706HL9 540953PRO 00000020950HL2 540981RX1 00000077288HL9 541276AOPROTEUS 00000038563HL3 541390RM1 00000038441HL2 541390RM2 00000045366HL2 541426WX1 00000097811HL4 541447CRM1 00000025561HL2 541451RM2 00000037746HL5 541677MOIRAD1 00000040780HL9 541768RM5PROT 00000047331HL4 559459XR1 00000048361HL0 559459XR2 00000083059HL6 561246RAD 00000019153HL6 561395RAD 00000030443HL6 561488RAD1 00000030407HL1 561488RAD3 00000042476HL2 561540RAD 00000022503HL7 561798ER1 00000029532HL9 561863RM6 00000088859HL4 561965RAD1 00000031442HL7 562493RM1 00000026063HL8 563589PROT 00000044515HL5 563652PROT 00000025563HL8 570286ER 00000029932HL1 570325PRO 00000046760HL5 570387PRO1 00000033376HL5 570523PRO1 00000060984HL2 570759RAD3 00000070211HL8 571231RAD1 00000083219HL6 571231RAD2 00000023404HL7 573486HR2 00000043528HL9 573582AUDRAD2 00000036574HL2 573756ER 00000065485HL5 573882RAD1 00000041505HL9 573884WXR 00000050387HL0 574535PRO 00000029284HL7 574537RAD 00000082190HL0 574546RAD1 00000028882HL9 574772PRO 00000087606HL0 575521RM1 00000085131HL1 575521RM5 00000030178HL8 580233ER1 00000018988HL6 580233PRO 00000041372HL4 580249SPRO 00000041258HL5 580249SPRO2 00000046530HL2 580323PRO 00000034661HL9 580335RAD1 00000019155HL1 580336PRO 00000029025HL4 580355RAD1 00000029286HL2 580355RAD2 00000029287HL0 580355RAD3 00000042456HL4 580548PRO 00000018499HL4 580822PRO 00000028573HL4 580938SPRO 00000079181HL4 585266RAD1 00000080942HL6 585266RAD1 00000058406HL0 585368RAD 00000022238HL0 585798RAD1 00000028684HL9 586263MO1 00000051942HL1 586366PRO 00000071316HL4 586573EDPROT 00000024582HL9 601200RM1 00000038332HL3 601288PROT1 00000019147HL8 601364RM3 00000015941HL8 601426ERRAD 00000066383HL1 601443RAD2 00000066269HL2 601587PROT 00000024436HL8 601672R1 00000038036HL0 601740PROT1 00000025373HL2 601764RAD1 00000020350HL5 601824ER1 00000067054HL7 601849PROT1 00000088066HL6 601883PRAD 00000081385HL7 601883PRAD1 00000031444HL3 601894RAD1 00000079510HL4 601928PXR 00000020687HL0 601947GCRAD 00000022321HL4 602239ER1 00000021508HL7 602344ER12 00000063514HL4 602344PRORM4 00000037378HL7 602344RM2 00000042198HL2 602588RM8 00000025560HL4 602824RM4 00000036451HL3 602867RM3 00000040502HL7 603292NRAD 00000045109HL6 603335FER 00000055344HL6 603335WMMCRAD 00000020595HL5 603432P3 00000037406HL6 603577R10 00000033832HL7 603577R4 00000032397HL2 603577RT5 00000027092HL6 603752PROT 00000024373HL3 603778RAD4 00000094527HL9 603924PROT 00000059713HL8 605271PROTEUS 00000023534HL1 605322PROTEUS 00000053536HL9 605322PROTEUS2 00000068785HL5 605322PROTEUSCC 00000015883HL2 605322RM10 00000043745HL9 605345RAD 00000050593HL3 605347RAD 00000101404HL2 605353PROTEUS 00000035144HL5 605353RAD 00000041405HL2 605385RAD 00000043020HL7 605432RAD 00000111897HL5 605455RAD 00000087141HL8 605472PROTEUS 00000041450HL8 605578RAD 00000053318HL2 605644RAD 00000064167HL0 605673RAD 00000015869HL1 605685RAD 00000039522HL8 605697RAD 00000046879HL3 605725RAD 00000034456HL4 605854RAD 00000043527HL1 605925PROTEUS 00000073180HL2 605964RAD 00000016988HL8 606325PROT 00000066384HL9 606666PRO 00000113340HL4 606759PRO2 00000076949HL7 606759PROT 00000041642HL0 606833PROT2 00000049770HL1 606833PROT3 00000064529HL1 606878RM3 00000041187HL6 606878XR2 00000088064HL1 606920PROT 00000048921HL1 607535PRO 00000016823HL7 607796RAD1 00000017457HL3 607796RAD2 00000098074HL8 608323PRO 00000051588HL2 608324CLPR2 00000048215HL8 608324PR3 00000065487HL1 608339RAD 00000034560HL3 608342PRT 00000066677HL6 608355PROT 00000033954HL9 608364PROT 00000046529HL4 608524PRT 00000029026HL2 608745CLRAD2 00000028875HL3 608745RAD1 00000063802HL3 608775PR10 00000028518HL9 608775PRE 00000027093HL4 608775PRT 00000023599HL4 608775RM9 00000070885HL9 608776PROT 00000018105HL7 608782PRO 00000096523HL6 608785PRO 00000016193HL5 608873RAD 00000017628HL9 609584ER1 00000017638HL8 609584ER2 00000021702HL6 609584PRO 00000016989HL6 610237ERPRO 00000015865HL9 610237PRO 00000019951HL3 610278PRO 00000041374HL0 610284PRO2 00000021698HL6 612798RAD 00000036259HL0 614257ER 00000037004HL9 614293DIAG1 00000028379HL6 614293ERB 00000068952HL1 614293GPRO 00000064044HL1 614293HC 00000032679HL3 614293RPRO 00000019225HL2 614446CX3 00000029882HL8 614566ER3 00000024429HL3 614566PTB 00000066823HL6 614688PROT 00000029027HL0 614839PROT 00000016820HL3 614861PROT 00000033831HL9 615284ERPROT 00000108820HL2 615341NGPRO 00000083396HL2 615342CRAD1 00000018379HL8 615342PROT1 00000089700HL9 615342RAD 00000068444HL9 615435WRAD4 00000028253HL3 615768SCMCCR01 00000028297HL0 615768SCMCCRER 00000025558HL8 616392PROTEUS3 00000040673HL6 616394HMI 00000041263HL5 616394PROTEUS2 00000021162HL3 616659PROTEUS 00000059575HL1 61763HL9 00000032661HL1 617726RS1Y3 00000029533HL7 617983PR2 00000018383HL0 618395R1 00000023629HL9 618395WR2 00000026065HL3 618465SAR3 00000034711HL2 618549CCR 00000029933HL9 618635RAD 00000085827HL4 618798GWER 00000029955HL2 619644PRO1 00000047228HL2 619644PRO2 00000050099HL1 619644PRO3 00000019883HL8 620223MRAD 00000016987HL0 620272JEDI 00000021836HL2 620272RM3 00000054660HL6 620340ORTHO 00000059847HL4 620356PRO 00000028381HL2 620672PRO 00000098079HL7 620886PRO 00000041990HL3 623433XR1 00000021031HL0 623445PROT 00000038259HL8 630579PROT 00000041173HL6 630652PROT2 00000018108HL1 630833ER2 00000023737HL0 630933PROT1 00000030274HL5 630933PROT2 00000023109HL2 631351ER 00000015891HL5 631476ER 00000020582HL3 631723PRO 00000018110HL7 631726PRO 00000047007HL0 631726PRO1 00000020722HL5 636239JR1 00000029959HL4 636326FRM1 00000029159HL1 636625WRM3 00000052072HL6 641444RAD 00000023371HL8 651232RAD3 00000020684HL7 651982R2 00000075269HL1 660223MRAD2 00000047518HL6 660259PROT 00000038282HL0 660562SPRO 00000015876HL6 660646PRO 00000016394HL9 660886FR3 00000049130HL8 660886FXR2 00000031438HL5 661945HDH2 00000028101HL4 661KRRAD 00000021699HL4 662289RAD 00000100282HL3 662378PRO 00000038210HL1 662615R1 00000070719HL0 662647PRO 00000047066HL6 678583XR 00000052992HL5 678604RM3 00000092361HL5 678741PRO 00000027949HL7 678945PROT 00000105205HL9 701256RAD 00000038283HL8 701567RAD 00000091770HL8 701665RAD 00000029957HL8 702346XR1 00000044030HL5 702614XR1 00000042528HL0 702614XR2 00000016399HL8 702731ERXR 00000015895HL6 702731RAD2 00000015959HL0 702731RAD4 00000039523HL6 703478PROT 00000030186HL1 703478PROT1 00000028292HL1 703689RM06 00000028100HL6 704484MOC 00000055250HL5 704873DPRO 00000049937HL6 706282SR2 00000050600HL6 706282SR2 00000045790HL3 706335BR1 00000044741HL7 706396COM 00000043328HL4 706468XR 00000019406HL8 706494DER 00000031272HL8 706517MMP 00000044742HL5 706544CC1 00000065366HL7 706544HC 00000047371HL0 706544MA3 00000039733HL1 706544MA4 00000044489HL3 706544MA5 00000050392HL0 706544WC2 00000025169HL4 706596PER 00000044065HL4 706632PRO 00000095423HL0 706655PRO 00000071149HL9 706678RAD 00000016822HL9 706754HR1 00000052783HL8 706787R3 00000023735HL4 707429NBER1 00000022469HL1 707429NBGV1 00000047069HL0 707464RAD1 00000039058HL3 707554ERRAD 00000049129HL0 707554PROT1 00000027943HL0 708388PROT 00000039123HL5 708425XR5 00000034948HL0 7087471635 00000035093HL4 7087471635 00000035143HL7 7087471639 00000024649HL6 708756JR3 00000048214HL1 708756PRO3 00000043588HL3 708783PRO1 00000043744HL2 708783PRO2 00000022465HL9 708783RM5 00000016582HL9 708783RM7 00000022591HL2 708783RM9 00000030278HL6 712279PROTEUS 00000107529HL0 712324RAD 00000029726HL7 712423BRAD1 00000045431HL4 712722PROTEUS 00000030007HL9 713285PPPRAD1 00000043019HL9 713359PROT 00000044041HL2 713383MHPROD 00000043172HL6 713778PRO 00000025556HL2 713873B15 00000028376HL2 713873B18 00000029087HL4 713873BPRO1 00000031683HL6 713873QMRAD1 00000101782HL1 714537RM3 00000065592HL8 715243PRO 00000035152HL8 715268RAD 00000082191HL8 715356PR 00000020209HL3 715425R1 00000052229HL9 715635RAD 00000037820HL8 715726RAD 00000022162HL2 715732RAD1 00000036423HL2 715743PR1 00000023602HL6 716675RAD1 00000050243HL5 717225PROT 00000019776HL4 717245R6 00000024202HL4 717290R1 00000053317HL4 717299RAD1 00000054661HL4 717624R1 00000032886HL4 717637ER1 00000073626HL4 717637R1A 00000043333HL4 717637RAD1 00000036862HL1 717653R1 00000040635HL5 717684R1 00000041256HL9 717684R1 00000095424HL8 717733GDSR1 00000093306HL9 717733GR1 00000021389HL2 717733HP 00000094650HL9 812265PROT 00000031686HL9 812265TOMO 00000073390HL7 812268PROT 00000085129HL5 812385RAD 00000023264HL5 812450RT11 00000033953HL1 812482ER2 00000034670HL0 812482TOMO 00000039122HL7 812663PROT2 00000041451HL6 812683RAD 00000022464HL2 812738RAD1 00000107533HL2 812847PROT 00000036694HL8 812858MRAD 00000102503HL0 812932ORTHO 00000086878HL6 812934PROT1 00000067371HL5 813558ERPRO 00000018752HL6 813571XPRO 00000016814HL6 813634RAD 00000082188HL4 813763BRPCER1 00000042901HL9 813873ERPRO 00000040837HL7 813971PPRO 00000018987HL8 814226R1 00000018301HL2 814274R1 00000064532HL5 814450RAD 00000042475HL4 814467R1 00000028572HL6 814533ER1 00000020351HL3 814533PROT1 00000035151HL0 814533R8 00000025615HL6 814643R4 00000017160HL3 814827R 00000020069HL1 814838XR 00000019950HL5 815235RAD3 00000016401HL2 815243RAD 00000038208HL5 815478RAD 00000036241HL8 815664MX1 00000054530HL1 815673PROT1 00000023903HL8 815741RM2 00000019158HL5 815741RM5 00000044407HL5 816276PROT1 00000045283HL9 816276PROT2 00000025419HL3 817255NHTOMO 00000018808HL6 817347MPTOMO 00000078549HL3 817347PRODIG 00000048255HL4 817408PRO1 00000053541HL9 817473PRO 00000060395HL1 817599PREOP 00000061518HL7 817599PREOP 00000077761HL5 817639ALLPRO 00000035576HL8 818502PROT1 00000103414HL9 818563DR 00000025417HL7 818676PROT1 00000028678HL1 818884ER 00000042326HL9 828213AI118 00000042328HL5 828213AI119 00000041643HL8 828213RCR1 00000041615HL6 828213RCR2 00000029539HL4 828315FRC 00000033427HL6 828681PRER1 00000045794HL5 828697LPR1 00000034241HL0 828754CER 00000050603HL0 830298PRO 00000050872HL1 830298PRO 00000075815HL1 830426MRH2 00000087778HL7 830608DIGPRO 00000018695HL7 831424RAD1 00000078009HL8 84016HL5 00000045156HL7 843236CFR1 00000052227HL6 843355RAD1 00000052071HL8 843549RAD2 00000040638HL9 845279P4 00000016591HL0 845534ER 00000048623HL3 845938RM3 00000033961HL4 847316PROER1 00000023263HL7 847432ER 00000028579HL1 847471RAD 00000025124HL9 850325RAD1 00000015957HL4 850434BER 00000024552HL2 850494WER2 00000023374HL2 850584RAD1 00000041582HL8 850626SR1 00000032182HL8 850747GC2 00000018098HL4 850863WR3 00000018099HL2 850863WR3 00000063513HL6 850875RAD1 00000031859HL2 850995PMC1 00000073389HL9 856246PROT2 00000029124HL8 859572ER1 00000029124HL5 859572ER2 00000023174HL6 859572RM1 00000022508HL6 859962RM1 00000022472HL5 859962RM2 00000028178HL2 859967PROT 00000029625HL1 859986XR1 00000022919HL5 860364RM4 00000029535HL2 860444XER2 00000031441HL9 860444XER2 00000032557HL1 860446EDX 00000032302HL2 860585RM1 00000035975HL2 863298RM1 00000082843HL4 863314PROT 00000019946HL3 863763RM3 00000087138HL4 209895RAD1 00000024745HL2 217333LXR2 00000096683HL8 218935PRO 00000080729HL7 781762RM1 00000034451HL5 303399RAD11 00000023033HL4 336768OSR1 00000023539HL0 205387PRO 00000047450HL2 209578ER1 00000107535HL7 713236PRAD1 00000029820HL8 803407R1 00000052075HL9 912287R608 00000032265HL1 434924KR1 00000028502HL3 602231XR1 00000042282HL4 918756XR1 00000002068HL3 479484PROT 00000020681HL3 479484PROT 00000045707HL7 864255XRA1 00000056810HL5 864400XRA1 00000018496HL0 856488PROT 00000017812HL9 856582PROT 00000020948HL6 251275RAD 00000080063HL1 251275RAD2 00000019233HL6 423744RAD3 00000043496HL9 423745ER 00000041756HL8 518525RAD1 00000074624HL8 AE1059RX01 00000016589HL4 0910072078 00000021506HL1 0910072103 00000021502HL0 0910072104 00000031275HL1 0910072122 00000042530HL6 0910072144 00000042616HL3 0910072146 00000042784HL9 0910072146 00000051665HL8 0910072171 00000076508HL1 0910072216 00000090520HL8 0910072230 00000091292HL3 0910072231 00000021393HL4 0910073062 00000024662HL9 0910073066 00000045406HL6 0910073081 00000046047HL7 0910073082 00000081588HL6 0910073111 00000061397HL6 0910074024 00000091102HL4 0910074035 00000082325HL2 0910075021 00000082324HL5 0910075022 00000076420HL9 0910076047 00000094362HL1 83089838806 00000096687HL9 83091661241 00000100279HL9 83093535827 00000024740HL3 0910073065 00000021396HL7 0910072090 00000063689HL4 BA4106RX01 00000063693HL6 BA4147RX01 00000063800HL7 BA4149RX01 00000063691HL0 BA4151RX02 00000063517HL7 BA4153RX01 00000025123HL1 90700 00000025119HL9 2543552 00000024906HL0 85179 00000022509HL4 85182 00000025068HL8 85186 00000024903HL7 85194 00000025065HL4 85211 00000042385HL5 BG4549RX02 00000042689HL0 BG4549RX03 00000062474HL2 0847070048 00000062476HL7 0847070075 00000072708HL1 306338RAD1 00000048359HL4 306873RAD2 00000051824HL1 709832RAD 00000036190HL7 506422XRA 00000045938HL8 506623PROTEUS 00000040400HL4 204239RAD 00000031440HL1 204765XRA 00000020170HL7 204836RAD1 00000035918HL2 204945XRA 00000031151HL4 250314XRA 00000029880HL2 250365XRA 00000030182HL0 250395XRA 00000024661HL1 250412RAD1 00000029787HL9 250428XRA 00000035026HL4 250491XRA 00000030801HL5 250545XRA 00000018992HL8 250632XRA 00000036242HL6 780945XRA 00000024468HL1 807343BRAD1 00000036261HL6 807343BRAD2 00000020073HL3 807345PROTEUS 00000032266HL9 807626RAD3 00000036189HL9 902427XRA 00000042200HL6 902428ARAD1 00000046335HL6 902597RAD1 00000020344HL8 902625PROTEUS 00000035368HL0 902765XRA 00000041646HL1 902794PROTEUS 00000091294HL9 902794XR1 00000072366HL8 902862RAD1 00000072364HL3 902862RAD2 00000020077HL4 905336RAD1 00000020171HL5 905682RAD 00000018807HL8 905732RAD1 00000016181HL0 905845RAD 00000017461HL5 905845RAD2 00000020357HL0 905895XRA 00000033500HL0 905895XRA2 1-89097903605 905895XRA2 00000065034HL1 306778RAD7 00000028680HL7 780585RAD2 00000039207HL6 250782XRA 00000030962HL5 250945XRA 00000036193HL1 250945XRA 00000038034HL5 250949XRA 00000039751HL3 250956RAD1 00000030957HL5 306332RAD1 00000016584HL5 306352RAD1 00000042195HL8 306359RAD1 00000054815HL6 306435RAD 00000055251HL3 306628RAD2 00000048258HL8 306752RAD1 00000031793HL3 306763RAD1 00000020212HL7 306775RAD1 00000037345HL6 306778RAD4 00000042329HL3 306778RAD5 00000045281HL3 306778RAD6 00000048363HL6 306862RAD1 00000031153HL0 306865RAD1 00000046136HL8 403212RAD1 00000021841HL2 403227RAD1 00000041994HL5 403259XRA 00000022915HL3 403343RAD2 00000021507HL9 403343RM1 00000020862HL9 403556RAD1 00000033876HL4 403777GRAD1 00000023597HL8 403777HRAD1 00000025378HL1 403782RAD1 00000022159HL8 403783RAD1 00000040959HL9 416348XRA 00000089901HL3 418589XR1 00000036132HL9 418677XRA 00000018384HL8 418682PROTEUS 00000031277HL7 418844PROTEUS 00000051826HL6 418844PROTEUS2 00000036282HL2 450358XRA 00000017460HL7 506632RM11B 00000057993HL8 514398RAD1 00000078550HL1 514495XR1 00000083638HL7 514495XR2 00000083668HL7 514495XR2 00000040589HL4 514697PROTEUS 00000040672HL8 519254PRO1 00000040587HL8 519254PRO2 00000017139HL7 519396RAD 00000033662HL8 519426XRA 00000034452HL3 519485XRA 00000003958ML8 519661RAD1 00000032493HL9 519735PRO 00000034024HL0 519842XRA 00000028878HL7 519843XRA 00000015884HL0 604278XRA 00000023594HL5 604775XRA 00000063803HL1 604775XRA2 00000041174HL4 604869XRA 00000017820HL2 604875CXRA 00000031445HL0 613392XRA 00000036244HL2 613541XRA 00000067609HL8 613548BRAD1 00000036262HL4 613687XRA 00000019407HL6 613933RAD 00000016194HL3 705286RAD 00000036191HL5 705424XRA 00000016195HL0 705856RAD 00000016811HL2 709884PROTEUS 00000022918HL7 819864PROTEUS 00000064042HL5 604775XRA3 00000059775HL7 613351RAD1 00000059772HL4 613536RAD1 00000045615HL2 HSRXR1 00000046045HL1 HSRXR2 00000096685HL3 XR24609M30 00000081876HL5 XR98881HL6 00000041133HL0 082407100301 00000045209HL4 082407160217 00000034242HL8 082407030307 00000022923HL7 082407100129 00000032893HL0 082407250034 00000025171HL0 082407010123 00000028634HL4 082407020293A 00000023603HL4 082407040103 00000028038HL8 082407040129 00000028031HL3 082407040130 00000028250HL9 082407040169 00000029534HL5 082407040209 00000023898HL0 082407050043 00000026061HL2 082407050046 00000028039HL6 082407050055 00000029334HL0 082407060018 00000051667HL4 082407060050 00000019403HL5 082407070037 00000024741HL1 082407070048 00000035239HL3 082407070077 00000040504HL3 082407070085 00000054659HL8 082407070095 00000064699HL2 082407070120 00000031857HL6 082407090069 00000066962HL2 082407090136 00000043174HL2 082407140124 00000024206HL5 082407160084 00000038947HL8 082407160198 00000031858HL4 082407180033 00000034161HL0 082407180038 00000025173HL6 082407200025 00000019882HL0 082407270078 00000025174HL4 082407290024 00000030280HL2 082407020303 00000029285HL4 082407030105 00000048520HL1 082407030329 00000058227HL0 082407030339 00000020219HL2 082407040074 00000022556HL5 082407040092 00000031024HL3 082407040216 00000032892HL2 082407040233 00000046668HL0 082407040367 00000047451HL0 082407040377 00000047885HL9 082407040378 00000029629HL3 082407070068 00000055685HL2 082407070103 00000031609HL1 082407080015 00000018302HL0 082407090029 00000028635HL1 082407100182 00000028633HL6 082407100197 00000049457HL5 082407100370 00000039732HL3 082407110020 00000030272HL9 082407120103 00000040989HL6 082407140120 00000029204HL5 082407150021 00000028176HL6 082407180023 00000040768HL4 082407180040 00000032883HL1 082407190034 00000032890HL6 082407190037 00000052991HL7 082407290040 00000057434HL3 082407290043 00000041100HL9 2608914PROTEUS 00000082525HL7 XR81884HL9 00000024733HL8 73244PROTEUS 00000037005HL6 2625645PROT 00000070588HL9 2906074PROT 00000043176HL7 CS1304RX01 00000084212HL0 CS1337RX01 00000036600HL5 600033RX04 00000043495HL1 CS1042RX01 00000016407HL9 CS1156RX01 00000018382HL2 CS1156RX02 00000018602HL3 CS1175RX01 00000018601HL5 CS1175RX02 00000043810HL1 CS1305RX01 00000052565HL9 7245PROTEUS 00000064321HL3 EG1039RX05 00000064322HL1 EG1039RX06 00000017156HL1 FI1112RX01 00000104757HL0 M4145030 00000015877HL4 M4163744 00000017154HL6 M4205408 00000035365HL6 M4207611 00000030903HL9 M4448013 00000023265HL2 M5484815 00000025375HL7 M9624229 00000031159HL7 HC3746XR02 00000095425HL5 HC7638XR01 00000041315HL3 HROS01XR04 00000019685HL7 PC8109XR01 00000101037HL0 PX5326XR02 00000016342HL8 H2014RAD02 00000019151HL0 H2022RAD08 00000048918HL7 H2097RAD06 00000058606HL5 H2165RAD02 00000044488HL5 H2188RAD04 00000058724HL6 H2199RAD02 00000058723HL8 H2212RAD01 00000058604HL0 H2213RAD01 00000058605HL7 H2214RAD01 00000058607HL3 H2216RAD01 00000085825HL8 XR85143HL6 00000081137HL2 XR84557HL8 00000081381HL6 XR84563HL6 00000061398HL4 2836332XRAY 00000072090HL4 2933763XR 00000075816HL9 2948102XR 00000016733HL8 0828070089 00000020594HL8 0828070092 00000022759HL5 0828070092 00000104752HL1 HU1008RX33 00000088857HL8 0835070198 00000099911HL0 0835070204 00000104064HL1 0835070215 00000032301HL4 IE1017RX02 00000044745HL8 IE1019RX03 00000021900HL6 IE1047RX01 00000050388HL8 IE1135RX01 00000057597HL7 IE1138RX01 00000020767HL0 IS1002RX03 00000023628HL1 A5413816 00000018107HL3 A5699201 00000018504HL1 B5356001 00000016196HL8 XU0009 00000016981HL3 XU0010 00000016740HL3 XU0011 00000018604HL9 XU0013 00000018616HL3 XU0014 00000019536HL2 XU0015 00000058130HL6 XU0017 00000060272HL2 XU0018 00000069563HL5 XU0019 00000021029HL4 KA1022RX01 00000090523HL2 KW1009RX03 00000102500HL6 LV4009RX01 00000030416HL2 LT4001RX02 00000100470HL4 0834070213 00000104277HL9 0834070227 00000021390HL0 72164PROTEUS 00000023739HL6 78797PROHM 00000020680HL5 8320106PROTEUS 00000027948HL9 21250PROTEUS 00000023630HL7 65593PRO2HM 00000032398HL0 662PROTEUS 00000031795HL8 TLB0211 00000016398HL0 0920079046 00000045284HL7 0920079084 00000045367HL0 0920079085 00000067372HL3 0920079126 00000074883HL0 0920079133 00000083057HL0 0920079145 00000045110HL4 NO1108RX03 00000055249HL7 PK1300RX05 00000079707HL6 PK1354RX03 00000084772HL3 0843070185 00000023738HL8 PL1005RX03 00000029232HL6 PL1043RX02 00000040430HL1 PL1107RX02 00000030081HL4 PL1228RX01 00000030008HL7 PL1555RX01 00000099136HL4 PL1639RX01 00000046430HL5 PL1671RX01 00000052240HL9 PL1696RX01 00000038333HL1 PL3101RX01 00000023111HL8 PF1006RX02 00000030728HL0 PER08733 00000056260HL3 PER100200 00000067236HL0 PER100300 00000055065HL7 PER100400 00000074015HL9 PER12310 00000071637HL3 PER18123 00000070886HL7 PER18127 00000028876HL1 PER18604 00000018380HL6 PER18606 00000021188HL8 PER18606 00000056261HL1 PER30002 00000041212HL2 PER39502 00000055875HL9 PER59501 00000055348HL7 PER99500 00000055345HL3 PER99600 00000019683HL2 PPR01322 00000030725HL6 PPR15708 00000021903HL0 PPR64101 00000043161JL0 PPR00172 00000041316HL1 600130RX05 00000041318HL7 600130RX07 00000029883HL6 RU1029RX01 00000055681HL1 RU1067RX02 00000017629HL7 RU1111RX07 00000029123HL7 RU1116RX01 00000040586HL0 RU1283RX05 00000018102HL4 RU2166RX01 00000027846HL5 RU2238RX01 00000110793HL7 RU2250RX02 00000020214HL3 RU2252RX01 00000054934HL5 RU2441RX02 00000034710HL4 RU2526RX01 00000032762HL7 RU2589RX01 00000032664HL5 RU2706RX01 00000040781HL7 RU2714RX01 00000039205HL0 RU2725RX01 00000064325HL4 RU3421RX01 00000070590HL5 RU3665RX02 00000069441HL4 RU3725RX02 00000069443HL0 RU3815RX01 00000075648HL6 RU7172RX01 00000090524HL0 RU7884RX01 00000021500HL4 SA1039RX06 00000033722HL0 SA1010RX03 00000033233HL8 SA1071RX02 00000023107HL6 SA1208RX01 00000024645HL4 SA1208RX03 00000022387HL5 SA1251RX01 00000022396HL6 SA1254RX01 00000024650HL4 SA1258RX01 00000024583HL7 SA1276RX01 00000029122HL9 SA1304RX01 00000029446HL2 SA1347RX01 00000032560HL5 SA1361RX01 00000036454HL7 SA1387RX02 00000046255HL6 SA2116RX01 00000024648HL8 SA1014RX09 00000033425HL0 SA1129RX02 00000019402HL7 SA1158RX02 00000019234HL4 SA1231RX04 00000018609HL8 SA1231RX05 00000060270HL6 ZA2255RX01 00000069562HL7 ZA1030RX04 00000069844HL9 ZA1030RX05 00000070884HL2 ZA1030RX06 00000070718HL2 ZA1030RX07 00000070716HL6 ZA1030RX08 00000020775HL3 ZA2122RX01 00000027094HL2 0850070692 00000018986HL0 0850070477 00000020770HL4 0850070485 00000023169HL6 0850070637 00000030005HL3 0850070707 00000042129HL7 0850070792 00000015890HL7 DE647A06 00000019954HL7 DP226703 00000058127HL2 DP226711 00000052989HL1 DPB08419 00000028981HL9 DPB17711 00000094523HL8 EPP05800 00000019162HL7 FE394804 00000043638HL6 FEG85406 00000042381HL4 FP970408 00000104066HL6 FPE60907 00000055350HL3 FPG07402 00000094655HL8 FPG10903 00000087603HL7 FPG23602 00000021327HL2 FPG30519 00000109669HL2 FPG66604 00000028032HL1 GEQ12201 00000045616HL0 GPX02109 00000037484HL3 ME203017 00000044032HL1 ME203022 00000041824HL4 ME242402 00000041759HL2 ME246210 00000042131HL3 ME246211 00000042158HL6 ME250406 00000041825HL1 ME256647 00000021697HL8 ME612427 00000022163HL0 ME667518 00000027852HL3 ME667520 00000058407HL8 ME707211 00000042077HL8 MEX00808 00000038281HL2 MP228411 00000016350HL1 MP491301 00000023631HL5 MP697607 00000046332HL3 MP926412 00000097812HL2 MPX16800 00000042457HL2 MPX28402 00000028685HL6 MPX39902 00000029021HL3 MPX39903 00000037409HL0 MPX77008 00000037482HL7 MPX77009 00000036863HL9 MPX82224 00000025566HL1 NE219630 00000040377HL4 NE435014 00000019775HL6 NE655316 00000056043HL3 NPV79304 00000056091HL2 NPV79305 00000019771HL5 UE276008 00000024366HL7 UE535A75 00000023902HL0 UE535A76 00000028781HL3 UE743A01 00000017828HL5 YP231724 00000092088HL4 YP600801 00000022385HL9 Z008RX01 00000022391HL7 Z009RX01 00000022920HL3 547110XR02 00000021326HL4 475030XR03 00000025069HL6 0853070274 00000022761HL1 0853070266 00000022755HL3 0853070267 00000023833HL7 0853070272 00000023179HL5 0853070285 00000032491HL3 0853070302 00000036324HL2 0853070305 00000037715HL0 0853070315 00000037714HL3 0853070316 00000043587HL5 0853070319 00000043743HL4 0853070326 00000059450HL7 0853070345 00000111467HL7 0853070357 00000034088HL5 0853070307 00000035238HL5 0853070309 00000057598HL5 0856070193 00000089903HL9 0856070257 00000040982HL1 0856070177 00000094361HL3 0856070248 00000034092HL7 UK1092RX02 00000019880HL4 UK5026RX01 00000037889HL3 10509RAD01 00000058727HL9 10599RAD01 00000020717HL5 77119RAD06 00000031796HL6 77119RAD07 00000019687HL3 00006RAD01 00000036611HL2 00019RAD01 00000023535HL8 00053RAD13 00000023740HL4 00053RAD14 00000034026HL5 00053RAD17 00000034430HL9 00068RAD07 00000015894HL9 00082RAD07 00000020585HL6 00082RAD09 00000050590HL9 00091RAD09 00000052079HL1 00091RAD10 00000052566HL7 00091RAD11 00000064818HL8 00094RAD04 00000024585HL2 00095RAD09 00000041649HL5 00100RAD02 00000063696HL9 00100RAD03 00000063511HL0 00100RAD04 00000018612HL2 00116RAD03 00000031789HL1 00129RAD09 00000019785HL5 00131RAD06 00000032267HL7 00131RAD07 00000031903HL8 00140RAD02 00000080059HL9 00141RAD05 00000059714HL6 00169RAD09 00000023406HL2 00233RAD10 00000023403HL9 00233RAD11 00000017452HL4 00243RAD09 00000017626HL3 00243RAD10 00000017625HL5 00243RAD11 00000017620HL6 00243RAD12 00000019409HL2 00253RAD12 00000094200HL3 00254RAD01 00000025126HL4 00263RAD15 00000025555HL4 00282RAD02 00000036188HL1 00303RAD02 00000033497HL9 00309RAD01 00000053414HL9 00313RAD07 00000053613HL6 00313RAD08 00000019684HL0 00329RAD01 00000100281HL5 00341RAD04 00000035301HL1 00373RAD04 00000037817HL4 00373RAD06 00000028251HL7 00378RAD03 00000019566HL9 00442RAD03 00000031271HL0 00488RAD07 00000019400HL1 00508RAD03 00000015886HL5 00510RAD08 00000016349HL3 00510RAD09 00000029815HL8 00521RAD03 00000031684HL4 00521RAD04 00000049935HL0 00554RAD01 00000050098HL3 00572RAD02 00000060271HL4 00572RAD04 00000099138HL0 00617RAD08 00000017148HL8 00680RAD02 00000031443HL5 00695RAD02 00000047425HL4 00701RAD05 00000047449HL4 00701RAD06 00000060158HL3 00709RAD02 00000023262HL9 00755RAD04 00000028291HL3 00755RAD05 00000063804HL9 00762RAD02 00000031608HL3 00784RAD02 00000032164HL6 00784RAD03 00000031687HL7 00785RAD08 00000017149HL6 00793RAD02 00000054813HL1 00824RAD03 00000050095HL9 00824RAD04 00000019154HL4 00833RAD02 00000046131HL9 00959RAD02 00000024466HL5 00965RAD02 00000030181HL2 01007RAD01 00000017459HL9 01203RAD02 00000031027HL6 01205RAD03 00000019686HL5 01255RAD02 00000047517HL8 05228RAD02 00000019689HL9 05452RAD03 00000039489HL0 05452RAD05 00000045943HL8 05523RAD03 00000063181HL2 05528RAD03 00000022320HL6 10134RAD02 00000094524HL6 10183RAD02 00000030912HL0 10251RAD01 00000051935HL5 10307RAD01 00000017816HL0 10362RAD01 00000018111HL5 10362RAD02 00000017143HL9 10390RAD01 00000019540HL4 10396RAD02 00000022066HL5 10428RAD01 00000021163HL1 10433RAD03 00000023743HL8 10448RAD02 00000023627HL3 10448RAD03 00000023741HL2 10448RAD04 00000030802HL3 10459RAD01 00000030803HL1 10459RAD02 00000029056HL9 10460RAD01 00000033721HL2 10469RAD01 00000033720HL4 10477RAD01 00000034160HL2 10490RAD01 00000037230HL0 10493RAD01 00000036095HL8 10494RAD01 00000035972HL9 10494RAD02 00000036793HL8 10498RAD02 00000040962HL3 10516RAD01 00000040836HL9 10516RAD02 00000041134HL8 10518RAD01 00000041262HL7 10519RAD02 00000046432HL1 10548RAD01 00000051827HL4 10563RAD01 00000054814HL9 10575RAD01 00000054933HL7 10575RAD02 00000055682HL9 10591RAD01 00000061065HL9 10605RAD01 00000060160HL9 10606RAD01 00000060065HL0 10609RAD01 00000060159HL1 10609RAD01 00000064528HL3 10625RAD01 00000081138HL0 10670RAD01 00000070370HL2 IE1502RX02 00000101403HL4 IE1564RX03 00000094653HL3 XR104248HL0 00000036979HL3 2571797PROTEUS 00000055680HL3 3882756515XRAND 00000035535HL4 AN-2561355-XR 00000042617HL1 BD-2629649-XR 00000042618HL9 BD-2629649-XR 00000056598HL6 TT-2790634-XR 00000043960HL4 GU-2647968-XR 00000049767HL7 SU-2714528-XR 00000046044HL4 3882672727XRCTR 00000049035HL9 3882706094XRAND 00000059337HL6 3882756170XRCTR 00000056258HL7 3882756601XROCC 00000028498HL4 38842510XRMET 00000022587HL0 0859070084 00000022752HL0 0859070084
Other devices under the same FDA submission
1 device cleared under K993090 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Proteus XRaGE Hualun Medical Systems Co., Ltd · System, x-ray, stationary2 recalls0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page