MEDPOR Surgical Implant Contoured Two Piece Chin 5 mm recallZ-1090-2015 · Stryker Craniomaxillofacial Division · initiated Dec 17, 2014
Class II Terminated All recalls posted in February 2015
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Stryker Craniomaxillofacial Division or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential for implant damage, implant cracking off/breaking intra-operatively.
FDA root cause category: Process control
What the company is doing
Stryker sent an Urgent Medical Device Recall Notice dated December 23, 2014, to all affected customers with a response form to be completed and returned upon receipt and review. The letter identified the affected product, potential hazards, risk mitigation and the actions to be taken. For questions contact Stryker CMF at 1-800-962-6558, or e-mail at [email protected].
Recall details
- Recall number
- Z-1090-2015
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Dec 17, 2014
- Posted by the FDA
- Feb 11, 2015
- Recalling firm
- Stryker Craniomaxillofacial Division
- Quantity
- 1051
- Product code
- FWP
- FDA submissions
- K922489
- Distribution
- Worldwide Distribution - US Nationwide in the states of PA, KY, VA, CT, UT, MD, AR, FL, TX, AL, DE, IL, IA, MI, CO, OH, GA, CA, DC, MO, LA, WI, ID, SC, NC, NY, MA, NH, CA, OK, MN, KY, IN, OR, TN and the countries of: Argentina, Australia, Canada, Chile, China, Netherlands, France, Spain, South Korea, Colombia, Switzerland, Singapore, and South Africa.
Product description
MEDPOR Surgical Implant Contoured Two Piece Chin 5 mm; Part Number 86001; Lot Numbers A1405048, A1404026 Product Usage: MEDPOR Implants in block, sheet and pre-formed shapes are intended for non-weight-bearing applications of craniofacial reconstruction/cosmetic surgery and repair of craniofacial trauma. MEDPOR Surgical Implants and MEDPOR BARRIER Implants are intended for the augmentation or restoration of bony contour in the craniofacial skeleton.
Code information (models, lots, serial numbers)
Part Number 86001; Lot Numbers A1405048, A1404026
Other devices under the same FDA submission
9 devices cleared under K922489 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- MEDPOR 82007Stryker Leibinger GmbH & Co. KG · PROSTHESIS, CHIN, INTERNAL4 reports
- MEDPOR 80010Stryker Leibinger GmbH & Co. KG · PROSTHESIS, CHIN, INTERNAL3 reports
- MEDPOR 8320Stryker Leibinger GmbH & Co. KG · PROSTHESIS, CHIN, INTERNAL1 report
- MEDPOR 9526Stryker Leibinger GmbH & Co. KG · PROSTHESIS, CHIN, INTERNAL1 report
- MEDPOR 9553Stryker Leibinger GmbH & Co. KG · PROSTHESIS, CHIN, INTERNAL1 report
- MEDPOR 9859Stryker Leibinger GmbH & Co. KG · PROSTHESIS, CHIN, INTERNAL1 report
- MEDPOR 80008Stryker Leibinger GmbH & Co. KG · PROSTHESIS, CHIN, INTERNAL1 report
- MEDPOR 80012Stryker Leibinger GmbH & Co. KG · PROSTHESIS, CHIN, INTERNAL1 report
- MEDPOR 86010Stryker Leibinger GmbH & Co. KG · PROSTHESIS, CHIN, INTERNAL1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 70166).
- Z-1093-2015 · TerminatedMEDPOR TITAN MAX Orbital Floor and Wall (OFW) MEDPOR-Titanium-Barrier (MTB) Right Product Usage: MEDPOR Implants in block, sheet and pre-formed shapes are intended for non-weight-…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page