Why it was recalled
The guide sleeve has the potential to jam in the Aiming Arms, which may delay disassembly and removal of instruments.
FDA root cause category: Component design/selection
What the company is doing
DePuy Synthes sent an Urgent Notice: Medical Device Recall letter, dated March 2, 2015, to affected customers. The letter identified the affected product, problem and actions to be taken. The letter instructed customers to review inventory for the affected parts and following the actions provided in the letter for removal and replacement of these devices. For questions call 610-719-5450.
Recall details
- Recall number
- Z-1352-2015
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Feb 24, 2015
- Posted by the FDA
- Mar 30, 2015
- Recalling firm
- Synthes, Inc.
- Quantity
- 131
- Product code
- HSB
- FDA submissions
- K131548
- Distribution
- US Nationwide Distribution - NY, TX, MD, OR, MS, ND, CA, AZ, MI, IL, PA, TN, AR, SD, SC, FL, VA, OK, and GA
Product description
Synthes 130 Degree Aiming Arm and 125 Degree Aiming Arm; The 130 Degree Aiming Arm and 125 Degree Aiming Arm are a part of the TFN-ADVANCED (TFNA) System. The TFN-ADVANCED Proximal Femoral Nailing System is intended for treatment of fractures in adults and adolescents (12-21) in which the growth plates have fused.
Code information (models, lots, serial numbers)
Part Numbers: 03.037.013, 03.037.014, Lot Numbers: 8903848, 9124017, 903851, 9124019, 8918566, 9124022, 9124012, 9124023, 9124014, 9124026, 9124016, 9124027, 8918563
Other devices under the same FDA submission
172 devices cleared under K131548 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- NA 03.037.010Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES168 reports
- NA 03.037.030Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES124 reports
- NA 03.037.024Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES120 reports
- NA 03.037.025Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES119 reports
- NA 03.037.013Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES117 reports
- NA 03.037.011Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES106 reports
- TFN-ADVANCE 04.037.242SSynthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall26 reports
- NA 03.037.012Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES75 reports
- TFN-ADVANCE 04.037.144SSynthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall6 reports
- TFN-ADVANCE 04.037.212SSynthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall5 reports
- NA 03.037.032Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES54 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page