GE Healthcare, Aespire 7900, Aespire View Anesthesia… recallZ-1376-2015 · GE Healthcare · initiated Feb 27, 2015
Class II Terminated All recalls posted in April 2015
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from GE Healthcare or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Specific Aespire 7900 Anesthesia Machines have an increased likelihood of failure of the CPU circuit board. This could result in a blank display or loss of mechanical ventilation.
FDA root cause category: Process control
What the company is doing
Consignees were sent a GE Healthcare "Urgent Medical Device Correction" letter GEHC Ref# 34064, dated February 27, 2015. The letter was addressed to Chief of Anesthesia, Health Care Administrator / Risk Manager and Director of Biomedical / Clinical Engineering. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information. The letter stated that customers can continue to use the affected device in Manual Ventilation mode, and contact their local GE Healthcare Service Representative. GE Healthcare will correct all affected devices at no cost to the customers. A Service Representative will contact customers to arrange for the correction. For questions or concerns contact GE Healthcare Service at 1-800-437-1171, or your local Service Representative.
Recall details
- Recall number
- Z-1376-2015
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Feb 27, 2015
- Posted by the FDA
- Apr 2, 2015
- Recalling firm
- GE Healthcare
- Quantity
- 46 (10 units US, 36 units OUS)
- Product code
- BSZ
- FDA submissions
- K122445
- Distribution
- Worldwide Distribution -- US, including the states of NC and OK; and the countries of CHINA, INDIA, JAPAN, MEXICO, and VENEZUELA.
Product description
GE Healthcare, Aespire 7900, Aespire View Anesthesia Machines. Model numbers 1009-9012-000. Intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients
Code information (models, lots, serial numbers)
Mfg Lot or Serial # ANCU00179 ANCU00114 ANCU00115 ANCU00116 ANCU00117 ANCU00118 ANCU00119 ANCU00120 ANCU00121 ANCU00178 ANCU00105 ANCU00123 ANCU00124 ANCU00125 ANCU00129 ANCU00130 ANCU00126 ANCU00108 ANCU00113 ANCU00112 ANCU00193 ANCU00174 ANCU00137 ANCU00175 ANCU00134 ANCU00135 ANCU00136 ANCU00150 ANCU00151 ANCU00138 ANCU00152 ANCU00140 ANCU00172 ANCU00139 ANCU00153 ANCU00141 ANCU00142 ANCU00143 ANCU00149 ANCU00147 ANCU00148 ANCU00146 ANCU00145 ANCU00101 ANCU00102 ANCU00103
Other devices under the same FDA submission
1 device cleared under K122445 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Aespire 7900Datex-Ohmeda Inc · Gas-machine, anesthesia3 recalls48 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page