This Datex-Ohmeda Aespire 7900 anesthesia gas machine delivers breathing gases during surgery. The 46 adverse event reports mostly describe the machine failing to deliver gas or pressure as expected, battery failures during use or service, and problems during pre-use checks. No deaths or injuries are reported. There are 3 open recalls, all from July 2024, over the risk that a breathing circuit limb could be mistakenly plugged into the auxiliary gas outlet.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
48
Deaths reported
0
Injuries reported
0
Recalls naming it
3

Summary

The FDA's MAUDE database contains 48 adverse event reports that list this exact device (DI 00840682102261), received between Feb 8, 2018 and Sep 29, 2026 — classified as 48 malfunctions.

The most frequently reported specific device problem is “Therapeutic or Diagnostic Output Failure” (9 reports).

3 FDA recall records name this device's identifier, 3 still open.

25 other recalls cover devices cleared under the same FDA submission (K050626, K092864, K122445, K172045); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 2 reportsJun 2024: 0 reportsJul 2024: 2 reportsAug 2024: 3 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 1 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 3 reportsJul 2025: 4 reportsAug 2025: 2 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 3 reportsDec 2025: 0 reportsJan 2026: 4 reportsFeb 2026: 0 reportsMar 2026: 5 reportsApr 2026: 1 reportsMay 2026: 1 reportsJun 2026: 1 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 3 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 12, 2024Open, Classified

    Potential issue that may arise if the limb of a patient breathing circuit is mistakenly connected to the Auxiliary Common Gas Outlet (ACGO) port.

    Z-0074-2025 · Root cause: Device Design · FDA record ↗

  2. Jul 12, 2024Open, Classified

    Potential issue that may arise if the limb of a patient breathing circuit is mistakenly connected to the Auxiliary Common Gas Outlet (ACGO) port.

    Z-0076-2025 · Root cause: Device Design · FDA record ↗

  3. Jul 12, 2024Open, Classified

    Potential issue that may arise if the limb of a patient breathing circuit is mistakenly connected to the Auxiliary Common Gas Outlet (ACGO) port.

    Z-0075-2025 · Root cause: Device Design · FDA record ↗

Most reported problems

Device problems

  • Therapeutic or Diagnostic Output Failure9
  • Failure to Run on Battery7
  • Mechanical Problem7
  • Insufficient Flow or Under Infusion6
  • Unexpected Therapeutic Results4
  • Electrical /Electronic Property Problem3
  • Gas/Air Leak3
  • Increase in Pressure3

Patient problems

  • No Clinical Signs, Symptoms or Conditions45
  • No Known Impact Or Consequence To Patient2
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 25 most recent of 48 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 29, 2026Therapeutic or Diagnostic Output Failure

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50928101 ↗
  • MalfunctionSep 25, 2026Mechanical Problem

    It was reported that there was a malfunction during a case resulting inloss of ventilation. There was no patient injury.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50900415 ↗
  • MalfunctionSep 8, 2026Insufficient Flow or Under Infusion

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50675151 ↗
  • MalfunctionJul 1, 2026No Flow

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN LOSS OF OXYGEN DURING SERVICE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25679677 ↗
  • MalfunctionJun 3, 2026Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN LOSS OF MECHANICAL VENTILATION DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25372994 ↗
  • MalfunctionMay 6, 2026Mechanical Problem

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN AN INABILITY TO SWITCH VENTILATION MODES AS EXPECTED DURING A CASE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25092535 ↗
  • MalfunctionApr 1, 2026Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN LOSS OF MECHANICAL VENTILATION DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24752943 ↗
  • MalfunctionMar 24, 2026Increase in Pressure

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN AN OVER-DELIVERY OF PRESSURE DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24679716 ↗
  • MalfunctionMar 20, 2026Increase in Pressure

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN OVER DELIVERY OF PRESSURE DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24650816 ↗
  • MalfunctionMar 20, 2026Failure to Run on Battery

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN A BATTERY FAILURE THAT COULD RESULT IN A LOSS OF MECHANICAL VENTILATION. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24649174 ↗
  • MalfunctionMar 12, 2026Failure to Run on Battery

    IT WAS REPORTED THAT THERE WAS A BATTERY FAILURE THAT COULD CAUSE LOSS OF MECHANICAL VENTILATION DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24580804 ↗
  • MalfunctionMar 11, 2026Electrical /Electronic Property Problem

    IT WAS REPORTED THAT THERE WAS AN ERROR RESULTING IN LOSS OF MECHANICAL VENTILATION DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24566421 ↗
  • MalfunctionJan 28, 2026Gas/Air Leak

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN A LEAK GREATER THAN 4.5 LPM. THERE WAS NO REPORT OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24199828 ↗
  • MalfunctionJan 27, 2026Failure to Run on Battery

    IT WAS REPORTED THAT THERE WAS A BATTERY FAILURE THAT COULD CAUSE LOSS OF MECHANICAL VENTILATION DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24186490 ↗
  • MalfunctionJan 16, 2026Electrical /Electronic Property Problem; Improper Flow or Infusion

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN LOSS OF MECHANICAL VENTILATION. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24101352 ↗
  • MalfunctionJan 7, 2026Insufficient Flow or Under Infusion

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN A POTENTIAL IMPACT TO THE DELIVERY OF O2 OR FRESH GAS FLOW DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23988805 ↗
  • MalfunctionNov 26, 2025Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN LOSS OF MECHANICAL VENTILATION. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23659082 ↗
  • MalfunctionNov 21, 2025Insufficient Flow or Under Infusion; Insufficient Flow or Under Infusion

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN INSUFFICIENT OXYGEN OR FRESH GAS FLOW DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23621004 ↗
  • MalfunctionNov 4, 2025Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN LOSS OF MECHANICAL VENTILATION. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23468076 ↗
  • MalfunctionSep 19, 2025Failure to Run on Battery; Failure to Run on Battery

    IT WAS REPORTED THAT THERE WAS A BATTERY FAILURE THAT COULD CAUSE LOSS OF MECHANICAL VENTILATION. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23095818 ↗
  • MalfunctionAug 29, 2025Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN LOSS OF MECHANICAL VENTILATION. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22928375 ↗
  • MalfunctionAug 1, 2025Increase in Pressure; Increase in Pressure

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN OVER DELIVERY OF PRESSURE. THERE WAS NO PATIENT INVOLVEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22687817 ↗
  • MalfunctionJul 21, 2025Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THAT THERE WAS AN ERROR RESULTING IN LOSS OF MECHANICAL VENTILATION DURING A CASE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22555248 ↗
  • MalfunctionJul 16, 2025Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN LOSS OF MECHANICAL VENTILATION DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22521808 ↗
  • MalfunctionJul 8, 2025Insufficient Flow or Under Infusion; Insufficient Flow or Under Infusion

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN A LEAK THAT COULD CAUSE LOSS OF MECHANICAL VENTILATION DURING A CASE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22434686 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K050626, K092864, K122445, K172045. One submission often covers many models, so check the product and lot details in each record.

  • Z-3189-2026 · Jul 30, 2026 · Open, Classified
    Aisys, Datex-Ohmeda anesthesia system, Part Number 1011-9050-000
  • Z-3192-2026 · Jul 30, 2026 · Open, Classified
    Et Control, Software option for anesthesia gas machine to achieve and maintain targeted end tidal oxygen and anesthetic agents, Field replaceable unit (FRU) part numbers: 1011-8000
  • Z-3191-2026 · Jul 30, 2026 · Open, Classified
    Aisys CS2 with Et Control, Datex-Ohmeda anesthesia system, Part Number 1011-9000-000
  • Z-3190-2026 · Jul 30, 2026 · Open, Classified
    Aisys CS2, Datex-Ohmeda anesthesia system, Part Number 1011-9055-000
  • Z-0084-2025 · Jul 12, 2024 · Open, Classified
    Auxiliary Common Gas Outlet (ACGO) conversion/install kits sold under product names CONV KIT SCGO TO ACGO and ACGO ASSY FIELD INSTL. These kits are not affected specifically, howev
  • Z-0081-2025 · Jul 12, 2024 · Open, Classified
    Aisys. This anesthesia gas machine is intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals trained
  • Z-0082-2025 · Jul 12, 2024 · Open, Classified
    Aisys CS2. This anesthesia gas machine is intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals tra
  • Z-0078-2025 · Jul 12, 2024 · Open, Classified
    Avance. This anesthesia gas machine is intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals traine
  • Z-0080-2025 · Jul 12, 2024 · Open, Classified
    Avance CS2 and Avance CS2 Pro. These anesthesia gas machines are intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by me
  • Z-0077-2025 · Jul 12, 2024 · Open, Classified
    Aespire View. This anesthesia gas machine is intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals
  • Z-1389-2022 · Jun 1, 2022 · Terminated
    GE Healthcare Aisys CS2 Anesthesia System, Part number 1011-9050-000; and Avance CS2 and Avance CS2 Pro Anesthesia System, Part number 1009-9050-000.
  • Z-1427-2022 · May 27, 2022 · Open, Classified
    GE Healthcare Avance CS2, Avance CS2 Pro Anesthesia Systems, Part #1009-9050-000; and Aisys CS2 Anesthesia Systems, Part # 1011-9050-000.
  • Z-1285-2022 · Apr 25, 2022 · Open, Classified
    Back up batteries in Avance CS2 and Avance CS2 Pro anesthesia devices manufactured on or after April 1, 2019
  • Z-1286-2022 · Apr 25, 2022 · Open, Classified
    Field replacement back up batteries distributed on or after April 1, 2019 for Avance CS2, Avance CS2 Pro, Avance, Amingo, Aespire View anesthesia devices
  • Z-2595-2021 · Aug 2, 2021 · Open, Classified
    Flow sensors kept as stand-alone user replaceable spare parts for the GEHC anethesia machines: 2089610-001 FLOW SENSOR, LEGACY VAR ORF BCG (blue, cleanable) 2089610-001-S FLOW

…and 10 more.

Other Gas-machine, anesthesia devices

All Gas-machine, anesthesia devices →