Artis Q/ Q zen systems recallZ-1445-2015 · Siemens Medical Solutions USA, Inc · initiated Mar 19, 2015
Class II Terminated All recalls posted in April 2015
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
potential issue with the Artis Q/ Q zen systems where it is possible that an electrical connection in the equipment cabinet has not been installed correctly. The potentially defective connection in the equipment cabinet will be inspected and, if necessary, corrected.
FDA root cause category: Employee error
What the company is doing
A customer notification, dated March 19, 2015, was distributed to all affected customer via AX018/15/S. Additionally, Siemens has initiated an inspection of Artis Q/ Q zen systems via Update Instruction AX017/15/S.
Recall details
- Recall number
- Z-1445-2015
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Mar 19, 2015
- Posted by the FDA
- Apr 14, 2015
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Quantity
- 139
- Product code
- OWB
- FDA submissions
- K123529
- Distribution
- Nationwide
Product description
Artis Q/ Q zen systems; Artis Q/ Q zen systems are dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed include cardiac angiography, neuro angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures.
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Model Number of device:10848280, 10848281, 10848282, 10848283, 10848353, 10848354, 10848355 with serial numbers: 109033 123026 121061 121089 121094 109080 109074 121006 121013 121040 109119 111005 109016 121062 109118 111002 121082 121011 121039 123008 109066 109067 109068 121053 121054 117001 117023 123003 123011 103030 109084 121014 121051 109011 109096 121004 117021 121018 121036 123013 123014 117029 117006 121012 121049 103013 109102 121078 109015 109009 121019 109057 121010 117034 109108 103008 109103 121081 109038 109040 109034 117020 105016 109029 121096 111003 123025 121015 109064 117018 117030 109058 121016 109050 109076 117026 123022 109007 103001 111000 109071 109072 109075 121020 109091 109090 109041 109039 109060 121047 123023 123024 117024 121088 109088 121027 103028 109092 109032 121009 121056 121057 109021 121034 121008 109048 109020 109036 109043 123027 109101 109104 109105 109107 121058 121085 109047 117040 109031 109027 109028 105017 109010 109093 121071 109069 117014 123010 117009 121023 123015 123016 121079 121041 109044 121037 121046 105007 121072
Other devices under the same FDA submission
7 devices cleared under K123529 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Artis zeego 10848283Siemens Healthcare GmbH · Interventional fluoroscopic x-ray system5 recalls1 death7 reports
- Artis Q ceiling 10848281Siemens Healthcare GmbH · System, x-ray, angiographic4 recalls1 death49 reports
- Artis Q biplane 10848282Siemens Healthcare GmbH · System, x-ray, fluoroscopic, image-intensified4 recalls1 death46 reports
- Artis Q floor 10848280Siemens Healthcare GmbH · System, x-ray, angiographic4 recalls18 reports
- Artis Q.zen ceiling 10848354Siemens Healthcare GmbH · System, x-ray, angiographic4 recalls10 reports
- Artis Q.zen floor 10848353Siemens Healthcare GmbH · System, x-ray, angiographic4 recalls5 reports
- Artis Q.zen biplane 10848355Siemens Healthcare GmbH · Interventional fluoroscopic x-ray system4 recalls5 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page