AMSCO C and AMSCO 400 Steam Sterilizers, STERIS Corporation recallZ-1561-2015 · Steris Corporation · initiated Feb 19, 2015
Class II Terminated All recalls posted in April 2015
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Steris Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
STERIS has identified that the control board software in select AMSCO 400 and AMSCO C units will interrupt and cancel a processing cycle should the selected sterilizer shutdown time coincide with a processing cycle. This may result in a procedure delay.
FDA root cause category: Software design (manufacturing process)
What the company is doing
On 3/9/2015, the firm sent Urgent Field Correction Notices, to their customers. The letter identified the affected product and gave a description of the problem. The letter states that the control software is being revised to ensure the utility shutdown feature will not interrupt an ongoing processing cycle. STERIS Field Representatives will contact customers to schedule installation of the new control board on affected units. Customers may direct questions regarding the issue to STERIS Customer Service at 1-800-548-4873; Marie LaFrance, Product Manager, at 440-392-7648, or their local sales representative.
Recall details
- Recall number
- Z-1561-2015
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Feb 19, 2015
- Posted by the FDA
- Apr 30, 2015
- Recalling firm
- Steris Corporation
- Quantity
- 701 units
- Product code
- FLE
- FDA submissions
- K111223, K112055
- Distribution
- Worldwide Distribution -- US, including the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, I, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, & WY; and, the countries of Afghanistan, Canada, Japan, Kuwait, Mexico, Republic of Korea, & United Arab Emirates.
Product description
AMSCO C and AMSCO 400 Steam Sterilizers, STERIS Corporation. AMSCO C Small Sterilizer, AMSCO 400 Small Sterilizer, AMSCO 400 Medium Sterilizer. Designed for sterilization of heat and moisture-stable materials used in healthcare facilities.
Code information (models, lots, serial numbers)
AMSCO 400 Units: Serial #'s: 030051501 - 033641440; AMSCO C Units: Serial #'s: R020051501 - R023641403.
Other devices under the same FDA submission
27 devices cleared under K111223, K112055 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- AMSCO AMSCO 400 Steam SterilizerSteris Mexico, S. de R.L. de C.V. · Sterilizer, steam11 reports
- AMSCO AMSCO 400 Steam SterilizerSteris Mexico, S. de R.L. de C.V. · Sterilizer, steam9 reports
- AMSCO AMSCO 400 Steam SterilizerSteris Mexico, S. de R.L. de C.V. · Sterilizer, steam7 reports
- AMSCO 16CSteris Mexico, S. de R.L. de C.V. · Sterilizer, steam6 reports
- AMSCO AMSCO 400 Steam SterilizerSteris Mexico, S. de R.L. de C.V. · Sterilizer, steam6 reports
- AMSCO AMSCO 400 Steam SterilizerSteris Mexico, S. de R.L. de C.V. · Sterilizer, steam5 reports
- AMSCO AMSCO 400 Steam SterilizerSteris Mexico, S. de R.L. de C.V. · Sterilizer, steam3 reports
- AMSCO AMSCO 400 Steam SterilizerSteris Mexico, S. de R.L. de C.V. · Sterilizer, steam3 reports
- AMSCO AMSCO 400 Steam SterilizerSteris Mexico, S. de R.L. de C.V. · Sterilizer, steam3 reports
- AMSCO 20CSteris Mexico, S. de R.L. de C.V. · Sterilizer, steam2 reports
- AMSCO AMSCO 400 Steam SterilizerSteris Mexico, S. de R.L. de C.V. · Sterilizer, steam2 reports
- AMSCO AMSCO 400 Steam SterilizerSteris Mexico, S. de R.L. de C.V. · Sterilizer, steam2 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page