ANSPACH - Fluted Drum, Side Cut Only recallZ-2285-2015 · The Anspach Effort, Inc. · initiated Apr 13, 2015
Class II Terminated All recalls posted in July 2015
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from The Anspach Effort, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
G1 Dissection Tools are not compatible with all attachments listed on the labeling; G1 cutting tools are labeled with a Specified Shaft Exposure which may not be consistent when used with each attachment on the product label. Shaft exposure may be greater.
FDA root cause category: Device Design
What the company is doing
Anspach sent an Urgent Notice: Medical Device Recall (Correction) letter of Anspach G1 Dissection Tools dated May 6, 2015, to all affected customers. Customers were not required to return their inventory of GI Dissection tools. Customers were instructed to complete, sign, and return the attached reply form confirming receipt of the letter by fax or email. Customers were also instructed to forward the letter as appropriate immediately. Customers with questions should contact the Complaint Handling Unit Manager at 561-494-3673 or contact their Anspach Sales Representative. For questions regarding this recall call 561-627-1080.
Recall details
- Recall number
- Z-2285-2015
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Apr 13, 2015
- Posted by the FDA
- Jul 29, 2015
- Recalling firm
- The Anspach Effort, Inc.
- Quantity
- 115 tools.
- Product code
- HBE
- FDA submissions
- K113476
- Distribution
- Worldwide Distribution - US including AZ, CA, CO, FL, GA, IL, KY, LA, MN, MO, NC, NJ, NY, OH, OR, PA, TX, and Internationally to Canada and China.
Product description
ANSPACH - Fluted Drum, Side Cut Only; 7MM x 14MM REF L-7DRC-G1 Cutting shaping bone including spine and cranium.
Code information (models, lots, serial numbers)
REF L-7DRC-G1 ; ALL lot numbers
Other devices under the same FDA submission
47 devices cleared under K113476 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- NA S-1510TD-G1Synthes GmbH · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)104 reports
- NA P-CRN-G1Synthes GmbH · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)15 reports
- NA A-CRN-G1Synthes GmbH · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)11 reports
- NA S-1R-G1Synthes GmbH · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)10 reports
- NA SHD-6R-G1Synthes GmbH · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)8 reports
- NA SHD-9AC-G1Synthes GmbH · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)6 reports
- NA M-4B-G1Synthes GmbH · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)4 reports
- NA M-8NS-G1Synthes GmbH · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)4 reports
- NA S-8NS-G1Synthes GmbH · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)4 reports
- NA S-5AC-G1Synthes GmbH · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)3 reports
- NA L-4B-G1Synthes GmbH · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)2 reports
- NA S-4DC-G1Synthes GmbH · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)2 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 71212).
- Z-2280-2015 · TerminatedANSPACH - Fluted Drum, Side Cut only 10MM x 14MM , REF L-10DRC-G1 Cutting shaping bone including spine and cranium.
- Z-2281-2015 · TerminatedANSPACH - Fluted Drum, Side Cut only 5MM x 14MM , REF L-5DRC-G1. Cutting shaping bone including spine and cranium.
- Z-2282-2015 · TerminatedANSPACH - Extra Coarse Diamond Drum, 6.8MM x 10.6MM REF L-68DRMDX-G1 Cutting shaping bone including spine and cranium.
- Z-2283-2015 · TerminatedANSPACH - Fluted Drum, Side Cut only, 6MM x 14MM REF L-6DRC-G1 Cutting shaping bone including spine and cranium.
- Z-2284-2015 · TerminatedANSPACH - Fluted Acorn, 7MM REF L-7AC-G1 Cutting shaping bone including spine and cranium.
- Z-2286-2015 · TerminatedANSPACH - Fluted Barrel; 8MM x 12.5MM REF L-8BA-G1 Cutting shaping bone including spine and cranium.
- Z-2287-2015 · TerminatedANSPACH - Fluted Barrel; 8MM x 12.5MM REF L-8BA-G1 Cutting shaping bone including spine and cranium.
- Z-2288-2015 · TerminatedANSPACH - 8MM x 14MM Fluted Drum, Side Cut Only; REF L-8DRC-G1. Cutting shaping bone including spine and cranium.
- Z-2289-2015 · TerminatedANSPACH - Extra Coarse Diamond Ball; 8MM REF L-8DX-G1 Cutting shaping bone including spine and cranium.
- Z-2290-2015 · TerminatedANSPACH - Fluted Ball; 9MM REF L-9B-G1 Cutting shaping bone including spine and cranium.
- Z-2291-2015 · TerminatedANSPACH - Fluted Drum, Side Cut Only; 9MM x 14 MM REF L-9DRC-G1 Cutting shaping bone including spine and cranium.
- Z-2292-2015 · TerminatedANSPACH - 3MM Course Diamond Ball, 2 MM Shaft Exposure; REF M-3DC-2-G1 Cutting shaping bone including spine and cranium.
- Z-2293-2015 · TerminatedANSPACH - 3MM Fluted Ball, 2 MM Shaft Exposure; REF M-3SB-2-G1. Cutting shaping bone including spine and cranium.
- Z-2294-2015 · TerminatedANSPACH - 4MM x 11 MM Fluted Barrel, 14 Flutes; REF M-4BA-A-G1. Cutting shaping bone including spine and cranium.
- Z-2295-2015 · TerminatedANSPACH - 5MM Fluted Ball, 2 mm Shaft Exposure; REF M-5B-2-G1. Cutting shaping bone including spine and cranium.
- Z-2296-2015 · TerminatedANSPACH - 5MM x 11.3MM Fluted Barrel; REF M-5BA-A-G1 Cutting shaping bone including spine and cranium.
- Z-2297-2015 · TerminatedANSPACH - 6MM x 11.3MM Fluted Barrel; REF M-6BA-A-G1 Cutting shaping bone including spine and cranium.
- Z-2298-2015 · TerminatedANSPACH - 6MM Coarse Diamond Ball, 2 mm Shaft Exposure; REF M-6DC-2-G1 Cutting shaping bone including spine and cranium.
- Z-2299-2015 · TerminatedANSPACH - 7MM Fluted Acorn; REF M-7AC-G1 Cutting shaping bone including spine and cranium.
- Z-2300-2015 · TerminatedANSPACH - 7MM Extra Coarse Diamond Ball; REF M-7DX-G1 Cutting shaping bone including spine and cranium.
- Z-2301-2015 · TerminatedANSPACH - 8MM x 12.5MM Fluted Barrel; REF M-8BA-G1 Cutting shaping bone including spine and cranium.
- Z-2302-2015 · TerminatedANSPACH - 4MM Fluted Ball, Extends 27MM; REF M-4B-27-G1. Cutting shaping bone including spine and cranium.
- Z-2303-2015 · TerminatedANSPACH - 5MM Fluted Ball, Extends 27MM; REF M-5B-27-G1. Cutting shaping bone including spine and cranium.
- Z-2304-2015 · TerminatedANSPACH -1.5MM x 7.5MM Cylindrical Carbide Drum, Extends 25MM; REF S-15DRMC-L-G1. Cutting shaping bone including spine and cranium.
- Z-2305-2015 · TerminatedANSPACH -3MM Coarse Diamond Ball,72MM Shaft Exposure; REF S-3SDC-22-G1. Cutting shaping bone including spine and cranium.
- Z-2306-2015 · TerminatedANSPACH -4MM Carbide Fluted Ball, Extends 27MM; REF S-4BL-C-G1. Cutting shaping bone including spine and cranium.
- Z-2307-2015 · TerminatedANSPACH -4MM Coarse Diamond Ball, 22MM Shaft Exposure; REF S-4DC-22-G1 Cutting shaping bone including spine and cranium.
- Z-2308-2015 · TerminatedANSPACH - 4MM Diamond Ball, Extends 27MM; REF S-4D-L-G1 all lot numbers Cutting shaping bone including spine and cranium.
- Z-2309-2015 · TerminatedANSPACH - 5MM Fluted Ball, 2MM Shaft Exposure; REF S-5B-2-G1 Cutting shaping bone including spine and cranium.
- Z-2310-2015 · TerminatedANSPACH - 5MM Diamond Ball, 2MM Shaft Exposure; REF S-5D-2-G1 Cutting shaping bone including spine and cranium.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page