Skip to content

NA S-8NS-G1Recalls & reported problems

Synthes GmbH · 3 mm Fluted Matchstick. Use with QD8-G1, SHORT-G1, QD8, QD8-S, SHORT Attachments. · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00845384012542), received between Jun 8, 2021 and Aug 18, 2021 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Separation Problem” (4 reports).

39 other recalls cover devices cleared under the same FDA submission (K113476); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Separation Problem4
  • Dull, Blunt2
  • Material Fragmentation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 18, 2021Material Fragmentation; Separation Problem; Material Fragmentation; Separation Problem

    IT WAS REPORTED BY (B)(6) THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE CUTTER DEVICE WAS ABLE TO BE REMOVED FROM THE ATTACHMENT DEVICE. IT WAS FURTHER OBSERVED THAT THE ATTACHMENT AND CUTTER WERE IN BAD CONDITION, THE FLUTED MATCHSTICK HEAD HAD BROKEN OFF, AND THE SHAFT WAS CORRODED. IT WAS NOTED IN THE SERVICE ORDER THAT THE ATTACHMENT DEVICE WAS RECEIVED WITH A CUTTER DEVICE STUCK IN IT AND COULD NOT BE RELEASED. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12339033 ↗
  • MalfunctionJul 16, 2021Dull, Blunt; Separation Problem; Dull, Blunt; Separation Problem

    IT WAS REPORTED BY (B)(6) THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE CUTTER DEVICE FAILED VISUAL INSPECTION. IT WAS FURTHER OBSERVED THAT THE CUTTER SHOWED SIGNS OF CORROSION. IT WAS DETERMINED THAT THE CUTTING TOOL COULD BE RELEASED FROM ATTACHMENT AND HANDPIECE. IT WAS DETERMINED THAT THE DEVICE SHOWED SIGNS OF USED BLADES BUT HEAVY CORROSION ON ITS SHAFT, WHICH WAS CAUSED BY REPROCESSING THE PARTS WHILE ASSEMBLED. IT WAS NOTED IN THE SERVICE ORDER THAT THE ATTACHMENT DEVICE AND CUTTER DEVICE WERE STUCK IN THE MOTOR DEVICE AND COULD NOT BE RELEASED. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12182289 ↗
  • MalfunctionJul 6, 2021Separation Problem; Separation Problem

    IT WAS REPORTED FROM (B)(6) THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE CUTTER DEVICE WAS STUCK IN THE MOTOR DEVICE, THE LOCKING MECHANISM WAS BLOCKED, THE CUTTER DEVICE HAD TRACES OF A FOREIGN SUBSTANCE ON IT AND SHOWED SIGNS OF USAGE . IT WAS NOTED IN THE SERVICE ORDER THAT THE CUTTER DEVICE WAS STUCK IN THE ATTACHMENT DEVICE AND WAS DIFFICULT TO DISENGAGE. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12122527 ↗
  • MalfunctionJun 8, 2021Dull, Blunt; Separation Problem; Dull, Blunt; Separation Problem

    IT WAS REPORTED FROM (B)(6) THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE CUTTER DEVICE SHOWED NORMAL SIGNS OF WEAR ON THE CUTTING EDGES, AND THE SHAFT SHOWED SIGNS OF REPROCESSING (FOREIGN SUBSTANCE) INSIDE THE ASSEMBLY. IT WAS NOTED IN THE SERVICE ORDER THAT THE ATTACHMENT DEVICE WAS STUCK ON THE HANDPIECE DEVICE WITH A CUTTER DEVICE STUCK/JAMMED INSIDE AND COULD NOT BE RELEASED. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION, OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11960094 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K113476. One submission often covers many models, so check the product and lot details in each record.

  • Z-2837-2016 · Aug 26, 2016 · Terminated
    G1 Dissection Tool - 7mm FBALL L; For cutting and shaping bone including spine and cranium.
  • Z-2840-2016 · Aug 26, 2016 · Terminated
    G1 Dissection Tool - 6MM CRS DBALL, 2mm Shaft EXP, M; For cutting and shaping bone including spine and cranium.
  • Z-2839-2016 · Aug 26, 2016 · Terminated
    G1 Dissection Tool - 3mm FLUTED BALL; For cutting and shaping bone including spine and cranium.
  • Z-2838-2016 · Aug 26, 2016 · Terminated
    G1 Dissection Tool - 3mm CRS DBALL, 2mm SHAFT EXP, M; For cutting and shaping bone including spine and cranium.
  • Z-2309-2015 · Apr 13, 2015 · Terminated
    ANSPACH - 5MM Fluted Ball, 2MM Shaft Exposure; REF S-5B-2-G1 Cutting shaping bone including spine and cranium.
  • Z-2296-2015 · Apr 13, 2015 · Terminated
    ANSPACH - 5MM x 11.3MM Fluted Barrel; REF M-5BA-A-G1 Cutting shaping bone including spine and cranium.
  • Z-2282-2015 · Apr 13, 2015 · Terminated
    ANSPACH - Extra Coarse Diamond Drum, 6.8MM x 10.6MM REF L-68DRMDX-G1 Cutting shaping bone including spine and cranium.
  • Z-2305-2015 · Apr 13, 2015 · Terminated
    ANSPACH -3MM Coarse Diamond Ball,72MM Shaft Exposure; REF S-3SDC-22-G1. Cutting shaping bone including spine and cranium.
  • Z-2288-2015 · Apr 13, 2015 · Terminated
    ANSPACH - 8MM x 14MM Fluted Drum, Side Cut Only; REF L-8DRC-G1. Cutting shaping bone including spine and cranium.
  • Z-2284-2015 · Apr 13, 2015 · Terminated
    ANSPACH - Fluted Acorn, 7MM REF L-7AC-G1 Cutting shaping bone including spine and cranium.
  • Z-2310-2015 · Apr 13, 2015 · Terminated
    ANSPACH - 5MM Diamond Ball, 2MM Shaft Exposure; REF S-5D-2-G1 Cutting shaping bone including spine and cranium.
  • Z-2300-2015 · Apr 13, 2015 · Terminated
    ANSPACH - 7MM Extra Coarse Diamond Ball; REF M-7DX-G1 Cutting shaping bone including spine and cranium.
  • Z-2294-2015 · Apr 13, 2015 · Terminated
    ANSPACH - 4MM x 11 MM Fluted Barrel, 14 Flutes; REF M-4BA-A-G1. Cutting shaping bone including spine and cranium.
  • Z-2306-2015 · Apr 13, 2015 · Terminated
    ANSPACH -4MM Carbide Fluted Ball, Extends 27MM; REF S-4BL-C-G1. Cutting shaping bone including spine and cranium.
  • Z-2289-2015 · Apr 13, 2015 · Terminated
    ANSPACH - Extra Coarse Diamond Ball; 8MM REF L-8DX-G1 Cutting shaping bone including spine and cranium.

…and 24 more.

Other DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) devices

All DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page