Spacelabs qube Compact Monitor, Model 91390. The monitor… recallZ-1608-2015 · Spacelabs Healthcare Inc · initiated May 8, 2015

Class II Terminated All recalls posted in May 2015

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Spacelabs Healthcare Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The input power can fail even though the power source at the wall has not been interrupted. In addition, there is the potential for the qube Compact Monitor to fail to turn on. This is a retrospective recall related to a recall initiated in 2012 (Z-0403-2013).

FDA root cause category: Software design

What the company is doing

The firm sent the Urgent Medical Device Correction- Spacelabs Healthcare Inc.-qube Compact Monitor, Model 91390, Power Input PCBA Failure letter, dated 8 May 2015, via priority service mail to US consignees. This letter described the affected system, problem, actions to be taken by users and corrective actions conducted by Spacelabs. For additional information or technical assistance, please contact: Technical Support Spacelabs Healthcare, Inc. 35301 SE Center St. Snoqualmie, WA 98065 1-800-522-7025 and select 2 for Technical Support

Recall details

Recall number
Z-1608-2015
Recall class
Class II
Status
Terminated
Initiated by the company
May 8, 2015
Posted by the FDA
May 13, 2015
Recalling firm
Spacelabs Healthcare Inc
Quantity
5 units distributed in the US
Product code
MHX
FDA submissions
K120616
Distribution
distributed in North Carolina, Texas, and Wyoming

Product description

Spacelabs qube Compact Monitor, Model 91390. The monitor is functioning as either bedside or central monitors; passively displays data generated by Spacelabs Healthcare parameter modules, Flexports interfaces, and other SDLC based products in the form of waveform and numeric displays, trends and alarms.

Code information (models, lots, serial numbers)

Serial Numbers:  1390-000054,  1390-000064,  1390-000313,  1390-000047,  1390-000898.

Other devices under the same FDA submission

3 devices cleared under K120616 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.