R3(Tm) 22 Mm I.D., 58 Mm O.D. Constrained Liner, Ref… recallZ-1737-2015 · Smith & Nephew, Inc. · initiated May 6, 2015
Class II Terminated All recalls posted in June 2015
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Smith & Nephew, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The R3 Constrained outer liner locking groove diameter is 0.030 inches oversized and may not allow the locking ring to engage into the R3 shell.
FDA root cause category: Process change control
What the company is doing
Smith & Nephew sent an Urgent-Product Recall Notification letter dated May 6, 2015. to Sales offices by E-mail and Hospitals via Fed- Ex. The letter identified the affected product, problem and actions to be taken. Customers were asked to quarantine the product and return it. Distributors were asked to notified their customers of the recall and ask them to return the product.
Recall details
- Recall number
- Z-1737-2015
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- May 6, 2015
- Posted by the FDA
- Jun 9, 2015
- Recalling firm
- Smith & Nephew, Inc.
- Quantity
- 12 units
- Product code
- KWZ
- FDA submissions
- K083566
- Distribution
- Worldwide Distribution - US Nationwide in the states of CA, FL, IL, IN, MI, NE, NY, TN, TX, VA, WI and the countries of Australia, Switzerland, Chile, Canada, and Spain
Product description
R3(TM) 22 MM I.D., 58 MM O.D. CONSTRAINED LINER, REF 71339158, QTY: (1), STERILE EO. Product Usage: The R3 Constrained liner Acetabular System is a cemented or uncemented prosthesis intended to replace a hip joint. The Constrained liner is intended for primary or revision patients at high risk for hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability and for whom all other options to constrained acetabular components have been considered. The R3 Constrained liner is intended for single use only.
Code information (models, lots, serial numbers)
Lot numbers: 14GM08613, 14HM08037
Other devices under the same FDA submission
8 devices cleared under K083566 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- R3 71339152Smith & Nephew, Inc. · Prosthesis, hip, constrained, cemented or uncemented, metal/polymer14 reports
- R3 71339154Smith & Nephew, Inc. · Prosthesis, hip, constrained, cemented or uncemented, metal/polymer11 reports
- R3 71339156Smith & Nephew, Inc. · Prosthesis, hip, constrained, cemented or uncemented, metal/polymer11 reports
- R3 71339158Smith & Nephew, Inc. · Prosthesis, hip, constrained, cemented or uncemented, metal/polymer10 reports
- R3 71339160Smith & Nephew, Inc. · Prosthesis, hip, constrained, cemented or uncemented, metal/polymer5 reports
- R3 71339162Smith & Nephew, Inc. · Prosthesis, hip, constrained, cemented or uncemented, metal/polymer3 reports
- R3 71339164Smith & Nephew, Inc. · Prosthesis, hip, constrained, cemented or uncemented, metal/polymer1 report
- R3 71339166Smith & Nephew, Inc. · Prosthesis, hip, constrained, cemented or uncemented, metal/polymer1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 71330).
- Z-1735-2015 · TerminatedR3(TM) 22 MM I.D., 52 MM O.D. CONSTRAINED LINER, REF 71339152, QTY: (1), STERILE EO. Product Usage: The R3 Constrained liner Acetabular System is a cemented or uncemented prosthes…
- Z-1736-2015 · TerminatedR3(TM) 22 MM I.D., 56 MM O.D. CONSTRAINED LINER, REF 71339156, QTY: (1), STERILE EO. Product Usage: The R3 Constrained liner Acetabular System is a cemented or uncemented prosthes…
- Z-1738-2015 · TerminatedR3(TM) 28 MM I.D., 60 MM O.D. CONSTRAINED LINER, REF 71339160, QTY: (1), STERILE EO, Product Usage: The R3 Constrained liner Acetabular System is a cemented or uncemented prosthe…
- Z-1739-2015 · TerminatedR3(TM) 28 MM I.D., 62 MM O.D. CONSTRAINED LINER, REF 71339162, QTY: (1), STERILE EO. Product Usage: The R3 Constrained liner Acetabular System is a cemented or uncemented prosthes…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page