Skip to content

Volcano s5/s5x/CORE Mobile Intravascular Ultrasound Imaging… recallZ-2043-2015 · Volcano Corp · initiated Sep 23, 2013

Class II Terminated All recalls posted in July 2015

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Volcano Corp or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A limitation in interoperability between the Philips Xper Flex Cardia Physiomonitoring System and Volcano's sS/Si, Core, and ComboMap systems. This limitation may impact the calculation of Fraction Flow Reserve (FFR) within the Volcano System when the system is connected to the Xper Flex Cardia system via its aortic pressure auxiliary output.

FDA root cause category: Nonconforming Material/Component

What the company is doing

Volcano and Philips sent an Advisory Notice letter dated September 23, 2013 to all impacted customers. The letter identified the affected product, problem and actions to be taken. For question contact your Volcano and Philips representative.

Recall details

Recall number
Z-2043-2015
Recall class
Class II
Status
Terminated
Initiated by the company
Sep 23, 2013
Posted by the FDA
Jul 8, 2015
Recalling firm
Volcano Corp
Quantity
227 total units affected by the issue - all models
Product code
IYO
FDA submissions
K123898
Distribution
Worldwide Distribution - US Nationwide and the countries of Belgium, United Kingdom, Denmark, Netherlands, Sweden, Iceland, France, Saudi Arabia, Australia, Philippines and Thailand.

Product description

Volcano s5/s5x/CORE Mobile Intravascular Ultrasound Imaging and Pressure System; Product/Part numbers: s5 - 804200-002 and 804200-014; s5x - 807300-001; CORE MOBILE - 400-0100.01. Intravascular Imaging and Pressure System for evaluation of blood flow, perfusion and vascular structures.

Code information (models, lots, serial numbers)

s5 Part number 804200-002 and 804200-014; s5x Part number 807300-001; CORE MOBILE Part Number 400-0100.01. All software versions

Other devices under the same FDA submission

2 devices cleared under K123898 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 71404).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page