Helical Blade Inserter for Trochanteric Fixation… recallZ-2091-2015 · Synthes (USA) Products LLC · initiated Jun 10, 2015
Class II Terminated All recalls posted in July 2015
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Synthes (USA) Products LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The height of one of the three guiding pins of the Helical Blade Inserter for the TFNA is oversized. The Helical Blade Inserter is therefore unable to pass through the Blade/Screw Guide Sleeve (part 03.037.017).
FDA root cause category: Process control
What the company is doing
A recall notification letter, dated June 10, 2015, was sent to sales consultants and Synthes distribution sites requesting return of affected devices.
Recall details
- Recall number
- Z-2091-2015
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Jun 10, 2015
- Posted by the FDA
- Jul 16, 2015
- Recalling firm
- Synthes (USA) Products LLC
- Quantity
- 32
- Product code
- HSB
- FDA submissions
- K131548
- Distribution
- Nationwide Distribution.
Product description
Helical Blade Inserter for Trochanteric Fixation Nail-ADVANCED (TFNA); intended to secure the head element to the head element inserter to prevent disengagement during head element insertion. Manual orthopedic surgical instrument.
Code information (models, lots, serial numbers)
Part Number Lot Numbers 03.037.024 T102759
Other devices under the same FDA submission
172 devices cleared under K131548 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- NA 03.037.010Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES168 reports
- NA 03.037.030Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES124 reports
- NA 03.037.024Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES120 reports
- NA 03.037.025Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES119 reports
- NA 03.037.013Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES117 reports
- NA 03.037.011Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES106 reports
- TFN-ADVANCE 04.037.242SSynthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall26 reports
- NA 03.037.012Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES75 reports
- TFN-ADVANCE 04.037.144SSynthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall6 reports
- TFN-ADVANCE 04.037.212SSynthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall5 reports
- NA 03.037.032Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES54 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page