Skip to content

Fabius MRI Anesthesia Machine recallZ-2271-2015 · Draeger Medical, Inc. · initiated Jun 22, 2015

Class II Terminated All recalls posted in July 2015

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Draeger Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

the Fabius MRI or parts of the system were attracted by the magnetic field of the MRI. If the Fabius MRI is positioned too close to the MRI, its parts can be loosened as a result of the magnetic force of the MRI.

FDA root cause category: Labeling False and Misleading

What the company is doing

Draeger Medical sent an Urgent Medical Device Recall Letter, dated June 2015, and Instructions for Use Supplement (one supplement for each Fabius MRI Anesthesia Machine at the facility) to end users. The letter identified the affected product, problem and actions to be taken. For questions regarding the recall letter contact Michael Kelhart at 1-800-543-5047 (press 1 at the prompt, then 2, then 32349). For questions regarding the operation and/or servicing of the Drger Fabius MRI anesthesia machine contact DrgerService Technical Support at 1-800-543-5047 (press 4 at the prompt).

Recall details

Recall number
Z-2271-2015
Recall class
Class II
Status
Terminated
Initiated by the company
Jun 22, 2015
Posted by the FDA
Jul 27, 2015
Recalling firm
Draeger Medical, Inc.
Quantity
373
Product code
BSZ
FDA submissions
K072884
Distribution
US Nationwide Distribution

Product description

Fabius MRI Anesthesia Machine; Fabius MRI is an inhalation anesthesia machine for use in MRI environments in operating, induction and recovery rooms.

Code information (models, lots, serial numbers)

Catalog Number(s): 8607300

Other devices under the same FDA submission

1 device cleared under K072884 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page