Volcano s5/s5i/CORE Intravascular Ultrasound Systems… recallZ-2134-2015 · Volcano Corporation · initiated Jun 22, 2015
Class II Terminated All recalls posted in July 2015
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Volcano Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
A software issue was discovered where an inaccurate FFR/iFR value could be calculated under certain circumstances.
FDA root cause category: Software design
What the company is doing
Urgent Medical Device Correction letters were sent on June 22, 2015 to all affected sites via FED Ex.
Recall details
- Recall number
- Z-2134-2015
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Jun 22, 2015
- Posted by the FDA
- Jul 20, 2015
- Recalling firm
- Volcano Corporation
- Quantity
- A total of 4007 devices, all models
- Product code
- IYO
- FDA submissions
- K123898
- Distribution
- Worldwide Distribution-US (nationwide) and regions in Europe and the countries of Japan, Canada, South Africa, Taiwan, Thailand, Singapore, Philippines, Australia, New Zealand, El Salvador, Malaysia, Mexico, Lebanon, Kuwait, South Korea, Kenya, Jordan, Iran, India, Israel, Hong Kong, Egypt, Algeria, Dominican Republic, Costa Rica, Chile, Brazil, Bolivia, and Argentina
Product description
Volcano s5/s5i/CORE Intravascular Ultrasound Systems software versions 3.3 and 3.4 (with and without iFR Scout Technology) and v.3.4 software kits; Model s5i Radiology: The Volcano s5/s5i and CORE Series Intravascular Imaging and Pressure System is used for the qualitative and quantitative evaluation of vascular morphology in the coronary arteries and vessels of the peripheral vasculature
Code information (models, lots, serial numbers)
Model s5i. Part number: 806300-020.
Other devices under the same FDA submission
2 devices cleared under K123898 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- CORE Mobile System Medical Equipment 400-0100.07Philips Image Guided Therapy Corporation · System, Imaging, Pulsed Echo, Ultrasonic3 reports
- CORE Mobile System Medical Equipment 400-0100.01Philips Image Guided Therapy Corporation · System, Imaging, Pulsed Echo, Ultrasonic1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 71654).
- Z-2131-2015 · TerminatedVolcano s5/s5i/CORE Intravascular Ultrasound Systems with software versions 3.3 and 3.4 (with and without iFR Scout Technology) and v.3.4 software kits; Model s5. Radiology: The …
- Z-2132-2015 · TerminatedVolcano s5/s5i/CORE Intravascular Ultrasound Systems software versions 3.3 and 3.4 (with and without iFR Scout Technology) and v.3.4 software kits; Model s5x. Radiology: The Vol…
- Z-2133-2015 · TerminatedVolcano s5/s5i/CORE Intravascular Ultrasound Systems software versions 3.3 and 3.4 (with and without iFR Scout Technology) and v.3.4 software kits; Model CORE Mobile. Radiology:…
- Z-2135-2015 · TerminatedVolcano s5/s5i/CORE Intravascular Ultrasound Systems software versions 3.3 and 3.4 (with and without iFR Scout Technology) and v.3.4 software kits; Model s5ix/s5iz Radiology: T…
- Z-2136-2015 · TerminatedVolcano s5/s5i/CORE Intravascular Ultrasound Systems software versions 3.3 and 3.4 (with and without iFR Scout Technology) and v.3.4 software kits; Model CORE Radiology: The Vo…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page