Medtronic EOPA Arterial Cannula, Model # 77620. Lot:… recallZ-2420-2015 · Medtronic Perfusion Systems · initiated Jul 24, 2015
Class II Terminated All recalls posted in August 2015
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Perfusion Systems or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
EOPA Arterial Cannula Devices in this lot were shipped without the guidewire.
FDA root cause category: Packaging process control
What the company is doing
On July 24, 2015 Medtronic sales representatives began verbally notifying the customers of the issue, requesting quarantine and return of un-used, affected product Consignees were sent a Medtronic "Urgent Medical device Recall" letter dated August 2015. The letter was addressed to Risk Manger and Health Care Professional. The letter described the problem and the product involved in the recall. Advised consignees to quarantine and remove the product from their inventory and return it to Medtronic. Requested consignees to complete and return the Customer Confirmation Certificate via email to RS.CFQFCA@Medtronic com or fax to Medtronic 651-367-0612 to the attention of Customer Focused Quality. For questions contact your Medtronic Sales Representative or Lifeline Technical Services at 1-877-526-7890.
Recall details
- Recall number
- Z-2420-2015
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Jul 24, 2015
- Posted by the FDA
- Aug 19, 2015
- Recalling firm
- Medtronic Perfusion Systems
- Quantity
- 221 (111 US, 110 OUS)
- Product code
- DWF
- FDA submissions
- K031037
- Distribution
- Worldwide distribution. US nationwide including: FL, GA, IL, MI, NY, NC, PA, WV; Germany, Hong Kong, and Italy.
Product description
Medtronic EOPA Arterial Cannula, Model # 77620. Lot: 2015060544 intended to deliver oxygenated blood to the heart during cardiopulmonary bypass procedures for up to 6 hours in duration.
Code information (models, lots, serial numbers)
Lot number 2015060544;
Other devices under the same FDA submission
28 devices cleared under K031037 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- EOPA™ 77422Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls19 reports
- EOPA™ 77622Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls4 reports
- EOPA™ 77420Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass1 recall14 reports
- EOPA™ 77522Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass1 recall10 reports
- EOPA™ 77524Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass1 recall5 reports
- EOPA™ 77418Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass1 recall5 reports
- EOPA™ 77522Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass1 recall1 report
- EOPA™ 77722Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass1 recall0 reports
- EOPA™ 77424Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass17 reports
- EOPA™ 77620Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass10 reports
- EOPA™ 77422Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass8 reports
- EOPA™ 77624Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass7 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page