Smith & Nephew BIRMINGHAM HIP (TM) RESURFACING FEMORAL… recallZ-2745-2015 · Smith & Nephew, Inc. · initiated Jun 3, 2015

Class II Terminated All recalls posted in September 2015

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Smith & Nephew, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The firm observed that the BIRMINGHAM HIP(TM) Resurfacing (BHR) System had revision rates which were higher than established benchmarks.

FDA root cause category: Device Design

What the company is doing

Smith & Nephew sent an notification letter dated June 3, 2015 to customers via E-mail and Fed- Ex. The letter identified the affected product, problem and actions to be taken. Customers were asked to quarantine the product and return it. Distributors were asked to notified their customers of the recall and ask them to return the product.

Recall details

Recall number
Z-2745-2015
Recall class
Class II
Status
Terminated
Initiated by the company
Jun 3, 2015
Posted by the FDA
Sep 10, 2015
Recalling firm
Smith & Nephew, Inc.
Quantity
5,987 units
Product code
NXT
FDA submissions
P040033
Distribution
Worldwide Distribution - US Nationwide and countries of: Australia, Austria, Bahrain (UAE), Belgium, Canada, China, Denmark, Finland, Germany, Hong Kong, Italy, Japan, Malaysia, Mexico, Netherlands, New Zealand, Norway, Poland, Portugal, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand and United Arab Emirates (UAE).

Product description

Smith & Nephew BIRMINGHAM HIP (TM) RESURFACING FEMORAL HEAD, STERILE R, Rx only, REF Numbers: 74121138 (38 MM), 74123140 (40 MM), 74121142 (42 MM), 74123144 (44 MM), 74121146 (46 MM). Product Usage: Hip joint prosthesis

Code information (models, lots, serial numbers)

all codes

Other devices under the same FDA submission

57 devices cleared under P040033 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 71985).