Smith & Nephew BIRMINGHAM HIP (TM) RESURFACING DYSPLASIA… recallZ-2747-2015 · Smith & Nephew, Inc. · initiated Jun 3, 2015
Class II Terminated All recalls posted in September 2015
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Smith & Nephew, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The firm observed that the BIRMINGHAM HIP(TM) Resurfacing (BHR) System had revision rates which were higher than established benchmarks.
FDA root cause category: Device Design
What the company is doing
Smith & Nephew sent an notification letter dated June 3, 2015 to customers via E-mail and Fed- Ex. The letter identified the affected product, problem and actions to be taken. Customers were asked to quarantine the product and return it. Distributors were asked to notified their customers of the recall and ask them to return the product.
Recall details
- Recall number
- Z-2747-2015
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Jun 3, 2015
- Posted by the FDA
- Sep 10, 2015
- Recalling firm
- Smith & Nephew, Inc.
- Quantity
- 624 units
- Product code
- NXT
- FDA submissions
- P040033
- Distribution
- Worldwide Distribution - US Nationwide and countries of: Australia, Austria, Bahrain (UAE), Belgium, Canada, China, Denmark, Finland, Germany, Hong Kong, Italy, Japan, Malaysia, Mexico, Netherlands, New Zealand, Norway, Poland, Portugal, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand and United Arab Emirates (UAE).
Product description
Smith & Nephew BIRMINGHAM HIP (TM) RESURFACING DYSPLASIA CUP W/IMPACTOR, REF Numbers: 74120246 (46 MM), 74122248 (48 MM), 74120250 (50 MM), 74122252 (52 MM), 74120254 (54 MM) Product Usage: Hip joint prosthesis
Code information (models, lots, serial numbers)
all codes
Other devices under the same FDA submission
57 devices cleared under P040033 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- BIRMINGHAM HIP RESURFACING (BHR) 74120156Smith & Nephew, Inc. · Prosthesis, Hip, Semi-Constrained, Metal/Metal, Resurfacing3 deaths145 reports
- BIRMINGHAM HIP RESURFACING (BHR) 74120152Smith & Nephew, Inc. · Prosthesis, Hip, Semi-Constrained, Metal/Metal, Resurfacing139 reports
- BIRMINGHAM HIP RESURFACING (BHR) 74121146Smith & Nephew, Inc. · Prosthesis, Hip, Semi-Constrained, Metal/Metal, Resurfacing91 reports
- BIRMINGHAM HIP RESURFACING (BHR) 74121150Smith & Nephew, Inc. · Prosthesis, Hip, Semi-Constrained, Metal/Metal, Resurfacing1 death86 reports
- BIRMINGHAM HIP RESURFACING (BHR) 74122154Smith & Nephew, Inc. · Prosthesis, Hip, Semi-Constrained, Metal/Metal, Resurfacing75 reports
- BIRMINGHAM HIP RESURFACING (BHR) 74120154Smith & Nephew, Inc. · Prosthesis, Hip, Semi-Constrained, Metal/Metal, Resurfacing2 deaths70 reports
- BIRMINGHAM HIP RESURFACING (BHR) 74122158Smith & Nephew, Inc. · Prosthesis, Hip, Semi-Constrained, Metal/Metal, Resurfacing69 reports
- BIRMINGHAM HIP RESURFACING (BHR) 74120160Smith & Nephew, Inc. · Prosthesis, Hip, Semi-Constrained, Metal/Metal, Resurfacing62 reports
- BIRMINGHAM HIP RESURFACING (BHR) 74120158Smith & Nephew, Inc. · Prosthesis, Hip, Semi-Constrained, Metal/Metal, Resurfacing60 reports
- BIRMINGHAM HIP RESURFACING (BHR) 74120150Smith & Nephew, Inc. · Prosthesis, Hip, Semi-Constrained, Metal/Metal, Resurfacing58 reports
- BIRMINGHAM HIP RESURFACING (BHR) 74121142Smith & Nephew, Inc. · Prosthesis, Hip, Semi-Constrained, Metal/Metal, Resurfacing1 death57 reports
- BIRMINGHAM HIP RESURFACING (BHR) 74120148Smith & Nephew, Inc. · Prosthesis, Hip, Semi-Constrained, Metal/Metal, Resurfacing1 death56 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 71985).
- Z-2745-2015 · TerminatedSmith & Nephew BIRMINGHAM HIP (TM) RESURFACING FEMORAL HEAD, STERILE R, Rx only, REF Numbers: 74121138 (38 MM), 74123140 (40 MM), 74121142 (42 MM), 74123144 (44 MM), 74121146 (46 M…
- Z-2746-2015 · TerminatedSmith & Nephew BIRMINGHAM HIP (TM) RESURFACING ACETABULAR CUP W/IMPACTOR, REF Numbers: 74120144 (44 MM), 74120146 (46 MM), 74122146 (46 MM), 74122148 (48 MM), 74120148 (48 MM), 74…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page