Siemens RAPIDPoint 500 Measurement Cartridge Lactate (400)… recallZ-2814-2015 · Siemens Healthcare Diagnostics Inc · initiated Aug 24, 2015
Class II Terminated All recalls posted in September 2015
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Healthcare Diagnostics Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
RAPIDPoint 500 measurement cartridge serial numbers 2517102517 through 2519000012, negative sodium bias on both QC and patient results
FDA root cause category: Process control
What the company is doing
Siemens Healthcare sent an Urgent Field Safety Notice letter dated August 2015 to all affected customers. The notice identified the affected product, problem and actions to be taken. Customers were instructed to complete and return the Field Correction Effectiveness Check Form attached. For questions contact your Siemens Customer Care Center or ther local Siemens technical support representative.
Recall details
- Recall number
- Z-2814-2015
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Aug 24, 2015
- Posted by the FDA
- Sep 24, 2015
- Recalling firm
- Siemens Healthcare Diagnostics Inc
- Quantity
- 1213 cartridges
- Product code
- KHP
- FDA submissions
- K113216
- Distribution
- Worldwide Distribution - US Nationwide and the countries of Canada, Albania, Austria, Australia, Bosnia Herzeg., Bangladesh, Belgium, Brazil, Switzerland, Chile, China, Colombia, Czech Republic, Germany, Denmark, Denmark, Estonia, Spain, Finland, France, Great Britain, Georgia, Greece, Croatia, Hungary, Ireland, India, Italy, Lithuania, Luxembourg, Latvia, Macedonia, Malaysia, Netherlands, Norway, New Zealand, Poland, Portugal, Paraguay, Romania, Sweden, Singapore, Slovenia, Slovakia, Turkey, Taiwan, South Africa, Unit.Arab Emirates
Product description
Siemens RAPIDPoint 500 Measurement Cartridge Lactate (400) tests Part Number: 10491448 Product Usage: The RAPIDPoint 500 system is designed for professional use in a point-of-care or laboratory environment. This system tests blood gases, electrolytes, metabolites, total hemoglobin, and hemoglobin derivatives in arterial, venous, and capillary whole blood samples. The following parameters are tested: pH, pC02, p02, Na+, K+, Ca++, Cl-, glucose, lactate, tHb, F02Hb, FCHb, FMetHb, FHHb, nBili
Code information (models, lots, serial numbers)
Serial Numbers: 2517102517 through 2519000012. Cartridges within this serial number range that have a green dot on the box and/or the cartridge are not affected
Other devices under the same FDA submission
1 device cleared under K113216 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- RAPIDPoint® 500 System 10492730Siemens Healthcare Diagnostics Inc. · ACID, LACTIC, ENZYMATIC METHOD1 recall1 death32 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 72044).
- Z-2813-2015 · TerminatedSiemens RAPIDPoint 500 Measurement Cartridge Lactate (750 tests) Part Number: 10491449 Product Usage: The RAPIDPoint 500 system is designed for professional use in a point-of-care…
- Z-2815-2015 · TerminatedSiemens RAPIDPoint 500 Measurement Cartridge Lactate (250) tests Part Number: 10491447 Product Usage: The RAPIDPoint 500 system is designed for professional use in a point-of-care…
- Z-2816-2015 · TerminatedSiemens RAPIDPoint 500 Measurement Cartridge Lactate (100) tests Part Number: 10844813 Product Usage: The RAPIDPoint 500 system is designed for professional use in a point-of-care…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page