Zimmer Natural Nail CM Long. Intramedullary Fixation Rod recallZ-0119-2016 · Zimmer Gmbh · initiated Sep 3, 2015

Class II Terminated All recalls posted in October 2015

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Zimmer Gmbh or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

In about 5% of tested samples, the outer tray was found to be compromised. Risks: A damaged outer tray detected during surgery may cause a delay in surgery to obtain another implant. If a compromised packaging is not detected prior or during a surgery by the user, it may lead to periprosthetic infection. This may result in revision or multi-stage revision to treat the infection.

FDA root cause category: Package design/selection

What the company is doing

On 9/3/2015, URGENT MEDICAL DEVICE RECALL notifications were sent via courier to the affected distributors and Hospital Risk Managers with instructions for locating and inspecting the affected devices and contacting Zimmer if damage is observed. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. If after reviewing this notification you have questions concerning this notification, please call 1-877-946-2761.

Recall details

Recall number
Z-0119-2016
Recall class
Class II
Status
Terminated
Initiated by the company
Sep 3, 2015
Posted by the FDA
Oct 14, 2015
Recalling firm
Zimmer Gmbh
Quantity
57,603 units
Product code
HSB
FDA submissions
K091566
Distribution
US Nationwide distribution.

Product description

Zimmer Natural Nail CM Long. Intramedullary Fixation Rod. The Zimmer Natural Nail System is intended for temporary fracture fixation and stabilization of the bone.

Code information (models, lots, serial numbers)

Show the full code information (8k characters)

All Lots. Item number and name: 47-2493-481-10 ZNN CMN Nail 10mmX48cm 125 L 47-2493-442-10 ZNN CMN Nail 10mmX44cm 130 R 47-2493-480-10 ZNN CMN Nail 10mmX48cm 125 R 47-2493-445-10 ZNN CMN Nail 10mmX44cm 135 L 47-2493-483-10 ZNN CMN Nail 10mmX48cm 130 L 47-2493-444-10 ZNN CMN Nail 10mmX44cm 135 R 47-2493-482-10 ZNN CMN Nail 10mmX48cm 130 R 47-2493-441-11 ZNN CMN Nail 11.5mmX44cm 125 L 47-2493-485-10 ZNN CMN Nail 10mmX48cm 135 L 47-2493-440-11 ZNN CMN Nail 11.5mmX44cm 125 R 47-2493-484-10 ZNN CMN Nail 10mmX48cm 135 R 47-2493-443-11 ZNN CMN Nail 11.5mmX44cm 130 L 47-2493-481-11 ZNN CMN Nail 11.5mmX48cm 125 L 47-2493-442-11 ZNN CMN Nail 11.5mmX44cm 130 R 47-2493-480-11 ZNN CMN Nail 11.5mmX48cm 125 R 47-2493-445-11 ZNN CMN Nail 11.5mmX44cm 135 L 47-2493-483-11 ZNN CMN Nail 11.5mmX48cm 130 L 47-2493-444-11 ZNN CMN Nail 11.5mmX44cm 135 R 47-2493-482-11 ZNN CMN Nail 11.5mmX48cm 130 R 47-2493-441-13 ZNN CMN Nail 13mmX44cm 125 L 47-2493-485-11 ZNN CMN Nail 11.5mmX48cm 135 L 47-2493-440-13 ZNN CMN Nail 13mmX44cm 125 R 47-2493-484-11 ZNN CMN Nail 11.5mmX48cm 135 R 47-2493-443-13 ZNN CMN Nail 13mmX44cm 130 L 47-2493-481-13 ZNN CMN Nail 13mmX48cm 125 L 47-2493-442-13 ZNN CMN Nail 13mmX44cm 130 R 47-2493-480-13 ZNN CMN Nail 13mmX48cm 125 R 47-2493-445-13 ZNN CMN Nail 13mmX44cm 135 L 47-2493-483-13 ZNN CMN Nail 13mmX48cm 130 L 47-2493-444-13 ZNN CMN Nail 13mmX44cm 135 R 47-2493-482-13 ZNN CMN Nail 13mmX48cm 130 R 47-2493-421-10 ZNN CMN Nail 10mmX42cm 125 L 47-2493-485-13 ZNN CMN Nail 13mmX48cm 135 L 47-2493-420-10 ZNN CMN Nail 10mmX42cm 125 R 47-2493-484-13 ZNN CMN Nail 13mmX48cm 135 R 47-2493-423-10 ZNN CMN Nail 10mmX42cm 130 L 47-2493-463-13 ZNN CMN Nail 13mmX46cm 130 L 47-2493-422-10 ZNN CMN Nail 10mmX42cm 130 R 47-2493-461-10 ZNN CMN Nail 10mmX46cm 125 L 47-2493-425-10 ZNN CMN Nail 10mmX42cm 135 L 47-2493-460-10 ZNN CMN Nail 10mmX46cm 125 R 47-2493-424-10 ZNN CMN Nail 10mmX42cm 135 R 47-2493-463-10 ZNN CMN Nail 10mmX46cm 130 L 47-2493-421-11 ZNN CMN Nail 11.5mmX42cm 125 L 47-2493-462-10 ZNN CMN Nail 10mmX46cm 130 R 47-2493-420-11 ZNN CMN Nail 11.5mmX42cm 125 R 47-2493-465-10 ZNN CMN Nail 10mmX46cm 135 L 47-2493-423-11 ZNN CMN Nail 11.5mmX42cm 130 L 47-2493-464-10 ZNN CMN Nail 10mmX46cm 135 R 47-2493-422-11 ZNN CMN Nail 11.5mmX42cm 130 R 47-2493-461-11 ZNN CMN Nail 11.5mmX46cm 125 L 47-2493-425-11 ZNN CMN Nail 11.5mmX42cm 135 L 47-2493-460-11 ZNN CMN Nail 11.5mmX46cm 125 R 47-2493-424-11 ZNN CMN Nail 11.5mmX42cm 135 R 47-2493-463-11 ZNN CMN Nail 11.5mmX46cm 130 L 47-2493-421-13 ZNN CMN Nail 13mmX42cm 125 L 47-2493-462-11 ZNN CMN Nail 11.5mmX46cm 130 R 47-2493-420-13 ZNN CMN Nail 13mmX42cm 125 R 47-2493-465-11 ZNN CMN Nail 11.5mmX46cm 135 L 47-2493-423-13 ZNN CMN Nail 13mmX42cm 130 L 47-2493-464-11 ZNN CMN Nail 11.5mmX46cm 135 R 47-2493-422-13 ZNN CMN Nail 13mmX42cm 130 R 47-2493-461-13 ZNN CMN Nail 13mmX46cm 125 L 47-2493-425-13 ZNN CMN Nail 13mmX42cm 135 L 47-2493-460-13 ZNN CMN Nail 13mmX46cm 125 R 47-2493-424-13 ZNN CMN Nail 13mmX42cm 135 R 47-2493-462-13 ZNN CMN Nail 13mmX46cm 130 R 47-2493-401-10 ZNN CMN Nail 10mmX40cm 125 L 47-2493-465-13 ZNN CMN Nail 13mmX46cm 135 L 47-2493-400-10 ZNN CMN Nail 10mmX40cm 125 R 47-2493-464-13 ZNN CMN Nail 13mmX46cm 135 R 47-2493-403-10 ZNN CMN Nail 10mmX40cm 130 L 47-2493-441-10 ZNN CMN Nail 10mmX44cm 125 L 47-2493-402-10 ZNN CMN Nail 10mmX40cm 130 R 47-2493-440-10 ZNN CMN Nail 10mmX44cm 125 R 47-2493-405-10 ZNN CMN Nail 10mmX40cm 135 L 47-2493-443-10 ZNN CMN Nail 10mmX44cm 130 L 47-2493-404-10 ZNN CMN Nail 10mmX40cm 135 R 47-2493-401-11 ZNN CMN Nail 11.5mmX40cm 125 L 47-2493-340-10 ZNN CMN Nail 10mmX34cm 125 R 47-2493-400-11 ZNN CMN Nail 11.5mmX40cm 125 R 47-2493-343-10 ZNN CMN Nail 10mmX34cm 130 L 47-2493-403-11 ZNN CMN Nail 11.5mmX40cm 130 L 47-2493-342-10 ZNN CMN Nail 10mmX34cm 130 R 47-2493-402-11 ZNN CMN Nail 11.5mmX40cm 130 R 47-2493-341-11 ZNN CMN Nail 11.5mmX34cm 125 L 47-2493-405-11 ZNN CMN Nail 11.5mmX40cm 135 L 47-2493-340-11 ZNN CMN Nail 11.5mmX34cm 125 R 47-2493-404-11 ZNN CMN Nail 11.5mmX40cm 135 R 47-2493-343-11 ZNN CMN Nail 11.5mmX34cm 130 L 47-2493-401-13 ZNN CMN Nail 13mmX40cm 125 L 47-2493-342-11 ZNN CMN Nail 11.5mmX34cm 130 R 47-2493-400-13 ZNN CMN Nail 13mmX40cm 125 R 47-2493-341-13 ZNN CMN Nail 13mmX34cm 125 L 47-2493-403-13 ZNN CMN Nail 13mmX40cm 130 L 47-2493-340-13 ZNN CMN Nail 13mmX34cm 125 R 47-2493-402-13 ZNN CMN Nail 13mmX40cm 130 R 47-2493-343-13 ZNN CMN Nail 13mmX34cm 130 L 47-2493-405-13 ZNN CMN Nail 13mmX40cm 135 L 47-2493-342-13 ZNN CMN Nail 13mmX34cm 130 R 47-2493-404-13 ZNN CMN Nail 13mmX40cm 135 R 47-2493-321-10 ZNN CMN Nail 10mmX32cm 125 L 47-2493-381-10 ZNN CMN Nail 10mmX38cm 125 L 47-2493-320-10 ZNN CMN Nail 10mmX32cm 125 R 47-2493-380-10 ZNN CMN Nail 10mmX38cm 125 R 47-2493-323-10 ZNN CMN Nail 10mmX32cm 130 L 47-2493-383-10 ZNN CMN Nail 10mmX38cm 130 L 47-2493-322-10 ZNN CMN Nail 10mmX32cm 130 R 47-2493-382-10 ZNN CMN Nail 10mmX38cm 130 R 47-2493-321-11 ZNN CMN Nail 11.5mmX32cm 125 L 47-2493-381-11 ZNN CMN Nail 11.5mmX38cm 125 L 47-2493-320-11 ZNN CMN Nail 11.5mmX32cm 125 R 47-2493-380-11 ZNN CMN Nail 11.5mmX38cm 125 R 47-2493-323-11 ZNN CMN Nail 11.5mmX32cm 130 L 47-2493-383-11 ZNN CMN Nail 11.5mmX38cm 130 L 47-2493-322-11 ZNN CMN Nail 11.5mmX32cm 130 R 47-2493-382-11 ZNN CMN Nail 11.5mmX38cm 130 R 47-2493-321-13 ZNN CMN Nail 13mmX32cm 125 L 47-2493-381-13 ZNN CMN Nail 13mmX38cm 125 L 47-2493-320-13 ZNN CMN Nail 13mmX32cm 125 R 47-2493-380-13 ZNN CMN Nail 13mmX38cm 125 R 47-2493-323-13 ZNN CMN Nail 13mmX32cm 130 L 47-2493-383-13 ZNN CMN Nail 13mmX38cm 130 L 47-2493-322-13 ZNN CMN Nail 13mmX32cm 130 R 47-2493-382-13 ZNN CMN Nail 13mmX38cm 130 R 47-2493-301-10 ZNN CMN Nail 10mmX30cm 125 L 47-2493-361-10 ZNN CMN Nail 10mmX36cm 125 L 47-2493-300-10 ZNN CMN Nail 10mmX30cm 125 R 47-2493-360-10 ZNN CMN Nail 10mmX36cm 125 R 47-2493-303-10 ZNN CMN Nail 10mmX30cm 130 L 47-2493-363-10 ZNN CMN Nail 10mmX36cm 130 L 47-2493-302-10 ZNN CMN Nail 10mmX30cm 130 R 47-2493-362-10 ZNN CMN Nail 10mmX36cm 130 R 47-2493-301-11 ZNN CMN Nail 11.5mmX30cm 125 L 47-2493-361-11 ZNN CMN Nail 11.5mmX36cm 125 L 47-2493-300-11 ZNN CMN Nail 11.5mmX30cm 125 R 47-2493-360-11 ZNN CMN Nail 11.5mmX36cm 125 R 47-2493-303-11 ZNN CMN Nail 11.5mmX30cm 130 L 47-2493-363-11 ZNN CMN Nail 11.5mmX36cm 130 L 47-2493-302-11 ZNN CMN Nail 11.5mmX30cm 130 R 47-2493-362-11 ZNN CMN Nail 11.5mmX36cm 130 R 47-2493-301-13 ZNN CMN Nail 13mmX30cm 125 L 47-2493-361-13 ZNN CMN Nail 13mmX36cm 125 L 47-2493-300-13 ZNN CMN Nail 13mmX30cm 125 R 47-2493-360-13 ZNN CMN Nail 13mmX36cm 125 R 47-2493-303-13 ZNN CMN Nail 13mmX30cm 130 L 47-2493-363-13 ZNN CMN Nail 13mmX36cm 130 L 47-2493-302-13 ZNN CMN Nail 13mmX30cm 130 R 47-2493-362-13 ZNN CMN Nail 13mmX36cm 130 R 47-2493-305-10 ZNN CMN Nail 10mmX30cm 135 L 47-2493-341-10 ZNN CMN Nail 10mmX34cm 125 L 47-2493-305-11 ZNN CMN Nail 11.5mmX30cm 135 L 47-2493-325-10 ZNN CMN Nail 10mmX32cm 135 L 47-2493-305-13 ZNN CMN Nail 13mmX30cm 135 L 47-2493-325-11 ZNN CMN Nail 11.5mmX32cm 135 L 47-2493-365-13 ZNN CMN Nail 13mmX36cm 135 L 47-2493-325-13 ZNN CMN Nail 13mmX32cm 135 L 47-2493-344-10 ZNN CMN Nail 10mmX34cm 135 R 47-2493-304-10 ZNN CMN Nail 10mmX30cm 135 R 47-2493-344-11 ZNN CMN Nail 11.5mmX34cm 135 R 47-2493-304-11 ZNN CMN Nail 11.5mmX30cm 135 R 47-2493-344-13 ZNN CMN Nail 13mmX34cm 135 R 47-2493-304-13 ZNN CMN Nail 13mmX30cm 135 R 47-2493-364-10 ZNN CMN Nail 10mmX36cm 135 R 47-2493-324-10 ZNN CMN Nail 10mmX32cm 135 R 47-2493-364-11 ZNN CMN Nail 11.5mmX36cm 135 R 47-2493-324-11 ZNN CMN Nail 11.5mmX32cm 135 R 47-2493-364-13 ZNN CMN Nail 13mmX36cm 135 R 47-2493-324-13 ZNN CMN Nail 13mmX32cm 135 R 47-2493-385-10 ZNN CMN Nail 10mmX38cm 135 L 47-2493-345-10 ZNN CMN Nail 10mmX34cm 135 L 47-2493-385-11 ZNN CMN Nail 11.5mmX38cm 135 L 47-2493-345-11 ZNN CMN Nail 11.5mmX34cm 135 L 47-2493-385-13 ZNN CMN Nail 13mmX38cm 135 L 47-2493-345-13 ZNN CMN Nail 13mmX34cm 135 L 47-2493-384-10 ZNN CMN Nail 10mmX38cm 135 R 47-2493-365-10 ZNN CMN Nail 10mmX36cm 135 L 47-2493-384-11 ZNN CMN Nail 11.5mmX38cm 135 R 47-2493-365-11 ZNN CMN Nail 11.5mmX36cm 135 L 47-2493-384-13 ZNN CMN Nail 13mmX38cm 135 R

Other devices under the same FDA submission

10 devices cleared under K091566 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.